11
Setup Procedure
Gladiator Acoustic Switch Series
1. Mount the unit in its actual position
(see mounting - pages 7-8)
2. Check where the actual level or target is
relative to the sensors.
Make sure that the material or target is not blocking
the path between the transducers.
4. Turn the power on
The display will turn on and the fail-safe relay will
switch. The display will scroll through the following
messages: HAWK, Amp SerialNo, Type, Amp Soft
Ver, Device ID, SensorSerial, SensorModel, Sens
SoftVer, Sensor Addrs, Gladiator System Amp.
The unit will then go into operational mode
displaying ‘Switch’ with a % value. This % value
represents the changing amount of signal
loss between the transducers.
If you see an error code, see ‘troubleshooting’.
5. Simple “1-minute” Setup - Follow the flow chart
!
QuickSet
Cal Mounting
Relay Action
Unlock
0
App Type
CAL
CAL
CAL
CAL
CAL
CAL
CAL
CAL
CAL
CAL
RUN
Switch Point
CAL
On Delay Adj
OffDelay Adj
CAL
CAL
Select the Switch point
• The output relay will switch at the entered % value.
• The default value of 75% will be suitable for detecting most media.
• For detection of lighter / fine products select a lower % value and vice versa.
• For heavy materials almost any setting will work a higher % setting may be required. When
the level or target falls below the sensors the relay will switch back on.
Select the Switch Time Delay
Set the delay time for the relay switching on and off.
Mounting Calibration
• Do not proceed with this step unless the material or target position is well beneath the line
between the sender and receiver.
• Select ‘Yes’ to start the mounting calibration.
• ‘Wait’ will be displayed during the calibration for up to 30 seconds.
• Unit is now able to cancel the influence of the mounting.
• The % reading on the back lit display has been zeroed with the existing process conditions
and the measurement history log has been cleared.
• Always CAL MOUNT after changing application type or moving the transducers.
Select the required relay action
The Relay action can be defaulted to either ‘ON’ or ‘OFF’ and switch ‘ON’ or ‘OFF’ in response
to an instrument failure.
• Failsafe High - Relay is energised during normal operation and de-energised during blocked
conditions or unit failure.
• Failsafe Low - Relay is de-energised during normal operation and energised during blocked
conditions or unit failure.
Choose Application Type
• Alignment - For transducer alignment at long range (boom protection applications)
• Blocked Chutes - Configures the unit for blocked chute applications.
• Boom Protection - For anti collision / machinery detection.
• Switch - Allows selection of Sensitivity% for standard switch application.
Note: If prompted to ‘re-calc app params’ this means the unit is already set to that app type -
the unit is asking if you wish to re-configure the settings.