  9. HEMCO AFA 500 MK 3 Instruction manual

HEMCO AFA 500 MK 3 Instruction manual

Installation & Calibration
Phone (816) 796-2900
Fax. (816) 796-3333
Email [email protected]
HEMCO Corporation
711 S. Powell Road
Independence MO, 64056
MM-TEL-AIRFLOW-500.07.14
Visit www.HEMCOcorp.com Call Toll Free: (800) 779-4362
Lab Planning Solutions
HEMCO
R
The monitor or sensor must be positioned
where it can ‘see’ the room pressure of the
laboratory. The back connection spigot of the
monitor or sensor will accept the 25mm (1”)
hose which should be connected to the inner
chamber of the fume hood. The ideal position
for the hose connection is 100mm (4”) back
from the sash and 100mm (4”) higher than the
normal sash opening height.
AFA500 or Remote Airflow Sensor Installation Diagram
Monitor Panel Dimensions Monitor Panel Cutout Dimensions
Fume hood
inner skin Fume hood
outer skin
Normal Sash
working height
TEL Airflow Monitor Model AFA 500 MK 3
Installation, Operation, Maintenance Manual
Phone (816) 796-2900
Fax. (816) 796-3333
Email [email protected]
HEMCO Corporation
711 S. Powell Road
Independence MO, 64056
Visit www.HEMCOcorp.com Call Toll Free: (800) 779-4362
Lab Planning Solutions
HEMCO
R
2
Model # AFA 500 MK3 Airflow Monitor
STARTUP
The AFA500 MK3 must be field-calibrated once the room air supply and exhaust is balanced.
When the unit is powered up, the following sequence of events occurs:
1. The alarm performs a self-test of its functions, LEDs and audible alarm (approximately 2
seconds) and then initiates a delay timer of 30 seconds to allow the airflow sensor to stabilize.
2. During the 30-second stabilizing period, all alarms and relay outputs are deactivated and
the red and green LEDs remain on.
3. At the end of the delay, the unit will do one of two things:
a. If the monitor has been calibrated, the unit enters normal operating mode (solid green light
for safe velocity, red light and audible alarm if low velocity).
b. If the unit has not been calibrated, the red and green LEDs will flash, the audible alarm
will be muted.
CALIBRATION Single Point (default)
1. Determine the required low air alarm point, then position the sash and use a calibrated
instrument so that the face velocity is equal to the required alarm point.
2. Press and hold the Enter button for 5 seconds to enter Calibration Mode. This is indicated by
both red and green LEDs flashing with the audible alarm beeping.
3. To initiate calibration, press and hold the Enter and Set buttons at the same time. The unit
will sample the airflow for 5 seconds, during which time the green LED goes off and the red
LED flashes. The audible alarm continues to sound during the air sampling.
4. If calibration is successful, the monitor will give a two-tone beep at the end of the air sample,
and then automatically enter run mode.
5. If the Enter or Set button is released during the air sampling period, or if the airflow is fluctuating
too much, the alarm will emit a lower-frequency buzzing for a short period and then re-enter
calibration mode. If this occurs, press the Enter and Set buttons again to repeat the airflow
sampling.
6. When the calibration is complete, lower the sash to the normal operating height and the Green
LED should be illuminated, indicating that the airflow is greater than the calibrated alarm point.
If, during normal operation, the airflow drops below the alarm point, the unit will go into alarm
condition (red LED flashing, audible horn beeping). Push the enter button to temporarily mute
the horn.
7. The horn can be permanently disabled by pressing and holding the Set button for 10 seconds.
The horn will sound 3 times to indicate that it has been disabled. In Safe mode the Green LED
will flash if the horn is disabled or be solid to indicate that the horn is enabled. To enable the
horn press and hold the Set button for 10 seconds, the horn will sound 3 times to indicate that
it has been enabled.
3
Installation & Calibration
Horn Red LED Green LED Operation
Beep* 1 On On Power up 30 second time period to allow sensor stabilization
Off Flashing Flashing Requires Calibration - follow calibration procedure
Off Off On Air SAFE - audible alarm enabled
Off Off Flashing Air SAFE - audible alarm permanently disabled
On On Off Air Fail
Off On Flashing Air Fail - Setback mode (remote horn mute activated)
On Flashing Off Sash High alarm
Operation

Popular Measuring Instrument manuals by other brands

Sentry 200 owner's manual

Sentry

Sentry 200 owner's manual

Endress+Hauser Gammapilot M FMG50 Brief operating instructions

Endress+Hauser

Endress+Hauser Gammapilot M FMG50 Brief operating instructions

GTC FF310 instruction manual

GTC

GTC FF310 instruction manual

Keysight N9923A Service guide

Keysight

Keysight N9923A Service guide

XFtp SAT LITE Quick reference guide

XFtp

XFtp SAT LITE Quick reference guide

Apera Instruments YD300 instruction manual

Apera Instruments

Apera Instruments YD300 instruction manual

AEMC AmpFlex 300A user manual

AEMC

AEMC AmpFlex 300A user manual

Valeport MIDAS ECM operating manual

Valeport

Valeport MIDAS ECM operating manual

BD Sensors DM10 Mounting instructions

BD Sensors

BD Sensors DM10 Mounting instructions

Omega Engineering DPG280 Series user guide

Omega Engineering

Omega Engineering DPG280 Series user guide

Gsertel ATSC 3.0 user manual

Gsertel

Gsertel ATSC 3.0 user manual

Rover HD TAB 9 user guide

Rover

Rover HD TAB 9 user guide

MRC SP-82AM manual

MRC

MRC SP-82AM manual

Bender NGRM700 quick start

Bender

Bender NGRM700 quick start

GasLab SAN-20 user manual

GasLab

GasLab SAN-20 user manual

Saturn South SS9005 user manual

Saturn South

Saturn South SS9005 user manual

NEO TOOLS 75-100 instruction manual

NEO TOOLS

NEO TOOLS 75-100 instruction manual

UGT Tensio 140 operating manual

UGT

UGT Tensio 140 operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.