Hoffrichter AquaTREND uni User manual

User´s Manual
Quality makes the Difference
Humidifier for Therapeutic Air Flow
AquaTREND uni

2 Info
SERIAL NUMBER
HOFFRICHTER GmbH delivers all its systems with a
serial number in order to ensure traceability. The serial
number is indicated on the type label on the bottom
of the system.
Please always specify the serial number in inquiries
and complaints.
CONFORMITY
The device complies with the requirements of the Direc-
tive 93/42/EEC.

Table of Contents 3
TABLE OF CONTENTS
Legend of Symbols ................................................ 4
Safety Instructions ................................................. 5
Technical Description ............................................. 8
Operation of the Device........................................ 10
Cleaning .............................................................. 13
Troubleshooting ................................................... 15
Disposal .............................................................. 16
Humidifier ......................................................... 16
Packaging......................................................... 16
Technical Data ..................................................... 17

4 Legend of Symbols
LEGEND OF SYMBOLS
Observe the accompanying documenta-
tion
Applied part type BF
Electric safety class II (with protective
insulation)
CE Declaration of Conformity
Manufacturer
Observe the User´s Manual
Caution! Hot!
Do not dispose of the device with nor-
mal domestic waste. For information on
the proper disposal of the device, please
contact your responsible customer service.

Safety Instructions 5
SAFETY INSTRUCTIONS
Before using your AquaTREND uni for the first time,
please thoroughly read these User´s Manual. Keep
these User´s Manual at a safe place for immediate
consultation, if necessary.
For the period stipulated by law, HOFFRICHTER GmbH
warrants that the product is free from any defects pro-
vided it is used as intended.
HOFFRICHTER GmbH is not liable for any damage and
consequential damage nor for any effects on the safety,
reliability and performance of the device, if:
•
manipulations, modifications, expansions, adjust-
ments, repairs and maintenance measures are
carried out by persons not authorized by us to do so
• use is made of accessories and spare parts which
come from other manufacturers and have not been
released by us for use with the point
•
the device is used in a way that is different from
that described in the User´s Manual
• the hygiene and cleaning instructions contained in
these User´s Manual fail to be observed

6 Safety Instructions
In order to prevent potential hazards, please fol-
low the instructions below:
•
Use the humidifier in combination with a respiration
therapy device of the TREND II series only and
observe all instructions for use given for that device.
•
Use accessories only which have been recom-
mended and offered by your specialized dealer.
•
Do not operate the humidifier if it is damaged or
not working properly. If that is the case, please
immediately notify your specialized dealer.
•
Place the device on a firm and flat surface, avoid
an inclined position of the device during operation.
Never transport or tilt the respiration therapy device
if the humidifier is attached and filled. This might
cause water to enter the therapy device and thus
damage it.
•
Do not expose the therapy device to direct sunlight
or set it up near heaters or similar if the humidifier
is attached and filled.
• Change the water everyday to prevent the spread
of bacteria and germs.

Intended Use 7
INTENDED USE
AquaTREND uni is used for humidifying the therapy air
flow which is supplied to the patient.
Use the humidifier only together with TREND II devices.
The water used for humidifying must not contain any
additives such as medication, salt, aroma oil or any
other substances.

8 Technical Description
TECHNICAL DESCRIPTION
The device is a warm air humidifier which can be mounted
on devices of the respiration therapy line TREND II.
It is intended to humidify the therapy air flow with mois-
ture to prevent the mucous membranes of the upper
respiratory tract from becoming dry.
The heater output can be adjusted in five levels thus
allowing the device to be adjusted to the individual
patient’s requirements.
When turning off the therapy device or taking off the
humidifier the device saves the heating level that was
set last.

Device Description 9
DEVICE DESCRIPTION
1 Air outlet (connection for
therapy tube)
2 Cap
3 Air guide
4 Air inlet
5 Contact pins
6 Reservoir with heating
surface

10 Operation of the Device
OPERATION OF THE DEVICE
Rinse the humidifier thoroughly with clean fresh water
prior to its first use.
Do this as described in the chapter “Cleaning” from
page 13.
When filling the humidifier it must be disconnected from
the therapy device.
Open the humidifier by turning the cap in anti-clock-
wise direction and removing it from the water reservoir
(see illustration on page 13).
Fill the humidifier with cold distilled water for medical
applications or boiled cold water with a low lime in con-
tact. The water must be cold and contain no additives.
Fill in the water only up to the specified maximum fill-
ing level.
Close the water reservoir again by placing the cap onto
the water reservoir with slightly pressure and turning it
in clockwise direction until it stops.
While carrying the filled humidifier please keep hold
of the reservoir.

Operation of the Device 11
Never tilt or turn over the humidifier after filling it. When
connecting the humidifier with your respiration therapy
device please observe the following instructions:
•
Place the contact pins and air inlet of the humidifier
on the therapy device without twisting or canting
them.
•The release button has to snap in place audibly.

12 Operation of the Device
The heating power can be set to five levels at the ther-
apy device.
For starting operation of the humidifier push the heater
key on the TREND II device. The display now shows
the currently set heating level and the heater LED lights.
After starting the operation of the device with the control
key the heating level will be continuously be displayed.
If you interrupt or end the therapy with the control key
the heater LED will flash and the humidifier is set to
standby.
For changing the heating level press down the heater
key for approx. 1 second until the heater menu is dis-
played. By pressing the key the heating level can be
selected. You can now set the heating level with the
help of the key and (higher level = higher tem-
perature). Confirm by pressing .
CAUTION!
Operating the blower when the humidifier is filled
but the therapy tube is disconnected may cause
water droplets to be swept along in the air flow.
Therefore the therapy tube should always be firmly
in place before switching on the humidifier.
.!00EH
=< L ?@!/ L =
=< L ?@!/ L =
B@$
4%/!*1
Q!/%*#MMMMMMMMMMMMMM =

Cleaning 13
CLEANING
Please carefully follow the hygiene and cleaning instruc-
tions in order to preserve your health and to avoid dis-
turbances and failures.
Except from cleaning it regularly, the device requires
no further maintenance.
The regular, daily cleaning prevents the built-up of lim-
escale which is difficult to remove.
Press the release button, separate the humidifier from
the therapy device, and take the humidifier to a sink.
Open the humidifier by turning the cap anti-clockwise
and removing it from the water reservoirs (1).
ATTENTION!
Never use sharp or pointed objects for cleaning
the humidifier.
Empty the water reservoir and rinse it with running
water(2). Remove any limescale by wiping it with a
soft cloth (3).
If tough limescale has built up we recommend using
normal limescale remover or filling the water reservoir
with a thin vinegar solution and let it soak for a while.
Then empty the reservoir, wipe off any residue limes-
cale with a soft cloth and then rinse the reservoir thor-
1
2
3
3

14 Cleaning
oughly under running water.
Before cleaning the cap remove the air guide as shown
(4). It is sufficient to rinse the device daily with run-
ning water.
It is important to let the humidifier dry completely after
cleaning it. Do not refill it before it has not dried out
completely. You can also clean the water reservoir and
the cap in the dishwasher at a temperature of up to
60 °C. However, we recommend thoroughly rinsing it
with running water to avoid deposition of food leftovers
from the dishwasher.
A chemical or heat disinfection of the humidifier by the
user is not required and is not recommended.
AquaTREND uni is intended for use by only one patient.
Before using it for another patient it has to be profes-
sionally sanitized. Please take the humidifier to your
specialized dealer to have it sanitized.
4

Troubleshooting 15
TROUBLESHOOTING
Problem Remedy
Water in the therapy tube The heating has been set too high, select a lower heating
level.
Run the tube under the blanket.
No or insufficient humidifi-
cation
Select a higher heating level.
Selected heating level is not
displayed
Check if the contact pins have been inserted into the res-
piration therapy device all the way and if the release but-
ton has firmly snapped into place. The respiration therapy
device has to be switched on.
Water is escaping from the
air outlet
Check if the water reservoir has been filled beyond the
maximum filling level.

16 Disposal
DISPOSAL
HUMIDIFIER
For the proper disposal of the device please contact
your specialized dealer.
PACKAGING
Your specialized dealer will take back the packaging,
however, you can also separate the packaging accord-
ing to material and dispose of it in your normal domes-
tic waste.

Technical Data 17
TECHNICAL DATA
Dimensions (W x D x H) 110 x 150 x 125 mm
Mass without water approx. 350 g
Operational voltage 24 V DC
Electrical power 0…20 W
Heating level 1 4 W
Heating level 2 8 W
Heating level 3 12 W
Heating level 4 16 W
Heating level 5 20 W
Relative humidity flow level 1 level 5
4 hPa 16 l/min > 75 % 100%
10 hPa 26 l/min > 65 % > 90%
20 hPa 36 l/min > 50 % > 70%
Values measured at 22 °C and 55 % rel. air humidity
Dispensary of the humidifier 4 hPa : 16 l/min : 13 mg/l
(Heat level 5) 20 hPa : 36 l/min : 10 mg/l
Values measured at 24 °C and 35 % rel. air humidity
NOTE:
The humidification ratio can vary at different ambient
conditions.

18 Technical Data
Max. capacity approx. 300 ml
Max. water temperature 43 °C
Classification acc. to EN 60601-1 Safety class II, device type BF
Operating temperature + 5 °C ... + 35 °C
Storage temperature - 40 °C ... + 70 °C
Rel. air humidity 15 ... 95 %
Operating altitude (altitude above MSL) - 400 m ... 3500 m (1060 hPa ... 700 hPa)
Air Outlet 22 mm, conus (acc. to ISO 5356-1)
CE mark according to EC Directive 93/42/EEC,
device in compliance with MDD, Class IIa
Technical specifications are subject to technical modifications.