HTI BioChem FC-360 User manual

HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
High Technology, Inc.
109 Production Road, Walpole MA 02081 USA
Automatic Chemistry Analyzer
HTI BioChem FC-360
Operator’s Manual
Document Number
:
OM-E-FC-360-30
Revision Level
:
Rev. 3
Effective Date
:
8/15/2016


High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL I
Contents
INTENDED USE ........................................................................................................................................................1
1TRANSPORTATION, STORAGE AND UNPACKING......................................................................................4
TRANSPORTATION AND STORAGE.......................................................................................................4
UNPACKING INSTRUCTIONS..................................................................................................................5
2INSTALLATION.................................................................................................................................................7
ELECTRICAL REQUIREMENTS...............................................................................................................7
DIMENSIONS.............................................................................................................................................7
ENVIRONMENTAL REQUIREMENTS......................................................................................................8
MOVING THE INSTRUMENT....................................................................................................................8
PC MINIMUM REQUIREMENTS...............................................................................................................8
INSTRUMENT PLACEMENT AND PREPARATION.................................................................................8
WIRES AND TUBING CONNECTION.......................................................................................................9
Washing solution..............................................................................................................................11
2.7.1.1 Washing solution for probe and cuvettes .....................................................................................................................11
2.7.1.2 Washing solution for probe and cuvettes when running turbidimetric methods.........................................................12
INSTALLATION........................................................................................................................................12
PC configuration...............................................................................................................................12
2.8.1.1 Install analyzer software. .............................................................................................................................................. 12
Analyzer setup..................................................................................................................................13
3ANALYZER TECHNICAL SPECIFICATIONS ................................................................................................14
GENERAL SPECIFICATIONS.................................................................................................................15
SAMPLE/REAGENT TRAY......................................................................................................................15
CUVETTES/REACTION TRAY................................................................................................................15
DISPENSING ARM AND HYDRAULIC SYSTEM....................................................................................16
CUVETTE WASHER................................................................................................................................16
PHOTOMETER........................................................................................................................................16
ISE MODULE (OPTIONAL) .....................................................................................................................16
SOFTWARE.............................................................................................................................................17
POWER REQUIREMENT........................................................................................................................17
DESCRIPTION OF ELECTRONIC CIRCUITS........................................................................................17
TYPES OF SAMPLES .............................................................................................................................18
TYPES OF REACTIONS .........................................................................................................................18
TYPES OF METHODS ............................................................................................................................18
4GENERAL FUNCTIONS..................................................................................................................................18
MAIN PARTS OR MODULES..................................................................................................................18
Sample/Reagent Tray ......................................................................................................................19
Dispensing Arm and Probe ..............................................................................................................20
Cuvette/Reaction Tray .....................................................................................................................20
Cuvette Washer................................................................................................................................22
Hydraulic System .............................................................................................................................23
Optical System.................................................................................................................................24
ISE Module (Optional)......................................................................................................................25
DESCRIPTION OF THE OPERATING CYCLE.......................................................................................25
Reaction Preparation Cycle .............................................................................................................26
Incubation and Reading Cycle .........................................................................................................26
Cuvette Washing Cycle....................................................................................................................26
ISE processing cycle (Optional).......................................................................................................26

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
II HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
5BASIC OPERATIONS .....................................................................................................................................27
INITIAL CONFIGURATION OF THE PROGRAM....................................................................................27
DAILY OPERATION.................................................................................................................................27
PROGRAMMED MAINTENANCE ...........................................................................................................28
6SOFTWARE TABS..........................................................................................................................................29
MAIN WINDOW OF THE SOFTWARE....................................................................................................29
OPERATIONS TAB..................................................................................................................................30
REAGENTS TRAYS TAB ........................................................................................................................33
Reagent position ..............................................................................................................................33
PAT RESULTS TAB.................................................................................................................................35
Patient results search.......................................................................................................................37
Advanced search..............................................................................................................................40
CAL RESULTS TAB.................................................................................................................................44
CON RESULTS TAB................................................................................................................................45
SAMPLE POSITIONS TAB......................................................................................................................47
7OPERATIONS MENU......................................................................................................................................48
ADD SAMPLES AND METHODS............................................................................................................48
Add Sample......................................................................................................................................49
7.1.1.1 Add Patient...................................................................................................................................50
7.1.1.2 Add Calibrator ..............................................................................................................................52
7.1.1.3 Add Control ..................................................................................................................................52
Add Methods....................................................................................................................................53
MODIFY SAMPLE....................................................................................................................................54
DELETE SAMPLES AND METHODS......................................................................................................55
Delete Sample..................................................................................................................................56
Delete Selected Method...................................................................................................................57
PENDING: SAVE, LOAD AND IMPORT PENDING REQUESTS ...........................................................58
Save Pending...................................................................................................................................59
Load Pending...................................................................................................................................59
Import from CSV...............................................................................................................................60
SAMPLES TO TRAY................................................................................................................................60
SAMPLES TO TRAY (BARCODE) (OPTIONAL).....................................................................................63
WASH CUVETTES ..................................................................................................................................67
START......................................................................................................................................................68
STOP........................................................................................................................................................69
PAUSE .....................................................................................................................................................69
PRINT RESULTS.....................................................................................................................................71
REPEAT RESULTS .................................................................................................................................74
RELOAD SAMPLES ................................................................................................................................75
EXPORT RESULTS.................................................................................................................................75
8METHODS MENU............................................................................................................................................76
SETTINGS: METHODS CONFIGURATION............................................................................................76
General Tab .....................................................................................................................................77
Factor Tab........................................................................................................................................79
Reference Values Tab......................................................................................................................82
Specials Tab.....................................................................................................................................83
8.1.4.1 End Point Reagent Blank.............................................................................................................83
8.1.4.2 End Point Sample Blank ..............................................................................................................85
8.1.4.3 Kinetics.........................................................................................................................................87
8.1.4.4 Fixed Time Kinetics......................................................................................................................89
Advanced tab ...................................................................................................................................91
ISE METHODS SETTINGS (OPTIONAL)................................................................................................94

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL III
General Tab (ISE)............................................................................................................................94
Factor tab (ISE)................................................................................................................................96
Special Tab (ISE).............................................................................................................................97
Advanced tab (ISE)..........................................................................................................................98
REAGENTS IN USE AND PROFILES...................................................................................................100
SAVING METHODS...............................................................................................................................101
LOADING METHODS............................................................................................................................102
SAVING PROFILES...............................................................................................................................103
LOADING PROFILES ............................................................................................................................103
9CALIBRATORS MENU .................................................................................................................................104
SETTINGS .............................................................................................................................................104
SAVING CALIBRATORS FILES............................................................................................................106
LOADING CALIBRATORS FILES .........................................................................................................106
10 CONTROLS MENU ...................................................................................................................................106
SETTINGS .............................................................................................................................................107
SAVING CONTROLS FILES..................................................................................................................108
LOADING CONTROLS FILES...............................................................................................................108
11 STATISTICS MENU...................................................................................................................................108
LEVY-JENNINGS OF PATIENTS..........................................................................................................108
LEVY-JENNINGS OF CALIBRATORS..................................................................................................110
LEVY-JENNINGS OF CONTROLS........................................................................................................111
QUALITY CONTROL .............................................................................................................................112
12 MAINTENANCE MENU.............................................................................................................................115
SETTINGS .............................................................................................................................................116
Serial Port Tab ...............................................................................................................................116
Files Tab.........................................................................................................................................116
Database Tab.................................................................................................................................118
Ages Tab........................................................................................................................................120
Reports Tab....................................................................................................................................121
Memory Tab...................................................................................................................................123
Language/ Style Tab......................................................................................................................124
Access Control Tab........................................................................................................................125
CUVETTES STATUS.............................................................................................................................125
CONTAINERS STATUS ........................................................................................................................127
PROGRAMMED WASHING ..................................................................................................................127
TIP CLEANING ......................................................................................................................................128
COMMUNICATIONS..............................................................................................................................129
Commands Console.......................................................................................................................129
Serial Port Log................................................................................................................................130
VALIDATIONS........................................................................................................................................131
Stray Light......................................................................................................................................132
Photometer Precision.....................................................................................................................133
Photometric Accuracy ....................................................................................................................134
Photometric Linearity .....................................................................................................................135
6.
Select Maintenance Validations Photometric Linearity Cuvette Number OK ..............135
Diluter Precision.............................................................................................................................137
Precision test P150 and P250........................................................................................................139
12.7.6.1 P150 Method..............................................................................................................................141
12.7.6.2 P250 Method..............................................................................................................................143
INSTRUMENT........................................................................................................................................145
Calibration......................................................................................................................................145

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
IV HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
12.8.1.1 Photometer calibration ...............................................................................................................145
12.8.1.2 Cuvettes calibration....................................................................................................................146
Diluter purge...................................................................................................................................146
Washer purge.................................................................................................................................147
Initialize ..........................................................................................................................................147
ISE (Optional).................................................................................................................................147
12.8.5.1 Start up.......................................................................................................................................148
12.8.5.2 Calibrant A purge .......................................................................................................................149
12.8.5.3 Calibrant B purge .......................................................................................................................149
12.8.5.4 Calibration ..................................................................................................................................149
12.8.5.5 Clean..........................................................................................................................................150
12.8.5.6 Pumps Calibration......................................................................................................................150
12.8.5.7 Bubble Sensor Calibration .........................................................................................................151
12.8.5.8 Electrode replacement ...............................................................................................................151
12.8.5.9 Reagent Kit Replacement ..........................................................................................................152
12.8.5.10 Reconnect ..............................................................................................................................153
13 ADVANCED OPERATIONS......................................................................................................................153
CONFIGURATION OF THE PRINTING FORMATS..............................................................................153
FILE EXCHANGE...................................................................................................................................156
IMPORTING SAMPLES FROM AN ADMINISTRATION PROGRAM ...................................................156
EXPORTING RESULTS TO THE ADMINISTRATION PROGRAM ......................................................157
DETERMINATION OF PRECISION OF AN ANALYTIC METHOD.......................................................161
DIRECT READING OF THE SOLUTION ABSORBANCE.....................................................................161
14 ISE MODULE: DESCRIPTION AND OPERATION (OPTIONAL) ............................................................163
ELECTRODES.......................................................................................................................................164
FLUID MANAGEMENT..........................................................................................................................165
SAMPLE TYPE ......................................................................................................................................165
Sample Handling and Collection....................................................................................................166
14.3.1.1 Serum.........................................................................................................................................166
14.3.1.2 Plasma .......................................................................................................................................166
14.3.1.3 Urine Samples............................................................................................................................166
14.3.1.4 Matrix Effects..............................................................................................................................166
Reagent Kit Electronic Key ............................................................................................................167
SAMPLE ASPIRATING AND DISPENSING..........................................................................................167
CALIBRATION CYCLE ..........................................................................................................................167
CLEAN CYCLE ......................................................................................................................................169
PURGE A CYCLE..................................................................................................................................169
PURGE B CYCLE..................................................................................................................................169
SIP CYCLE.............................................................................................................................................169
OFF CYCLE (UNDER ELECTRODE REPLACEMENT COMMAND) ...............................................169
PUMP CALIBRATION CYCLE...........................................................................................................170
BUBBLE CAL CYCLE........................................................................................................................170
MAINTENANCE.................................................................................................................................170
Before starting daily operation ...................................................................................................170
Daily Maintenance......................................................................................................................170
Shutdown Procedure..................................................................................................................171
ISE Module re-activation............................................................................................................171
ISE TROUBLESHOOTING ................................................................................................................171

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL V
15 MAINTENANCE PROGRAM.....................................................................................................................175
DAILY MAINTENANCE..........................................................................................................................175
Before starting daily operation .......................................................................................................175
Maintenance Instrument Diluter Purge .............................................................................................175
During the daily operation ..............................................................................................................176
Upon finishing with the daily operation...........................................................................................176
WEEKLY MAINTENANCE.....................................................................................................................176
MONTHLY MAINTENANCE ..................................................................................................................176
External washing of cuvettes .........................................................................................................176
General cleaning of the instrument................................................................................................177
Back-up of files in use....................................................................................................................177
ISE MODULE MAINTENANCE PROGRAM (OPTIONAL) ....................................................................178
Recommended Component Replacement Schedule (low volume user).......................................178
Recommended Replacement Schedule (high volume user) .........................................................178
MAINTENANCE ACCORDING TO PROGRAMMED ALARMS ............................................................179
Maintenance Levels according to cycles of use counters..............................................................179
Maintenance Level 1......................................................................................................................179
15.5.2.1 Disinfecting the reagent tray ......................................................................................................179
15.5.2.2 Disinfecting the tubing and cleaning of the containers...............................................................180
15.5.2.3 General cleaning of the instrument............................................................................................180
15.5.2.4 Axes lubrication..........................................................................................................................180
15.5.2.5 Cleaning of the washer circuit filter ............................................................................................181
15.5.2.6 Replacement of the Teflon seals in the piston pump.................................................................181
15.5.2.7 Flushing Pump Membrane replacement....................................................................................181
15.5.2.8 Diluter Liquid Diaphragm Pump Membrane replacement..........................................................181
Maintenance Level 2......................................................................................................................181
15.5.3.1 Disinfecting the reagent tray ......................................................................................................181
15.5.3.2 Disinfecting the tubing and cleaning of the containers...............................................................181
15.5.3.3 General cleaning of the instrument............................................................................................181
15.5.3.4 Axes lubrication..........................................................................................................................181
15.5.3.5 Cleaning of the washer circuit filter ............................................................................................181
15.5.3.6 Replacement of the Teflon seals in the piston pump.................................................................181
15.5.3.7 Flushing Pump Membrane Replacement...................................................................................181
15.5.3.8 Diluter Liquid Diaphragm Pump Membrane Replacement ........................................................182
15.5.3.9 Cuvettes replacement ................................................................................................................182
AS NEEDED MAINTENANCE...............................................................................................................182
Database Initialization (every 10.000 tests)...................................................................................182
Lamp replacement..........................................................................................................................183
Cuvettes Replacement (see item 15.5.3.8)....................................................................................185
15.6.3.1 Interferential filters replacement.................................................................................................185
Sample / Reagent probe replacement ...........................................................................................185
Replacing the fuses........................................................................................................................189
Special Wash .................................................................................................................................189
Replacing O-rings of piston pump..................................................................................................189
Replacing filters of washing system...............................................................................................189
Electrode replacement (For analyzers with ISE Module installed) ................................................190
15.6.9.1 Replacing ISE peristaltic pumps tubing (For analyzers with ISE Module installed)...................191

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
VI HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
16 PREPARING SOLUTIONS........................................................................................................................191
WASH SOLUTION.................................................................................................................................191
Washing solution for probe and cuvettes.......................................................................................191
Washing solution for probe and cuvettes when running turbidimetric methods ............................192
TIP CLEANING SOLUTION...................................................................................................................192
SPECIAL WASHING SOLUTION FOR MAINTENANCE ACCORDING TO NEED..............................192
SPECIAL WASHING SOLUTION FOR TURBIDIMETRIC TECHNIQUES WITH LATEX.....................192
DISINFECTING SOLUTION ..................................................................................................................192
17 TROUBLESHOOTING...............................................................................................................................192
INTRODUCTION....................................................................................................................................192
CHEMICAL PROBLEMS........................................................................................................................194
High results ....................................................................................................................................195
Low Results....................................................................................................................................196
Erratic Results................................................................................................................................196
Only one affected sample for all of the methods............................................................................197
Only one affected method for all of the samples............................................................................198
INSTRUMENTAL PROBLEMS..............................................................................................................199
Connections and supplies..............................................................................................................199
Diluter.............................................................................................................................................200
Washer...........................................................................................................................................200
Sample/reagent probe....................................................................................................................200
Reaction Cuvettes Tray..................................................................................................................201
Sample/Reagent Tray ....................................................................................................................201
Software .........................................................................................................................................201
HARDWARE MESSAGES.....................................................................................................................202

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 1 of 222
INTENDED USE
BioChem FC-360 is a photometric analyzer for clinical chemistry, it is meant to perform automated analytic
procedures for “in vitro” diagnosis.
Terms of use
1.
The instrument must be installed by trained personnel.
2.
This instrument is Class 1, Type b, IPX0.
3.
The instrument must be used by trained personnel
4.
Operation room requirements: Room temperature 20-25ºC; Humidity: 20- 85% (no condensation).
5.
The instrument must be connected to a supply line according to current national regulations.
6.
This instrument must not be used for purposes other than the ones it has been built for.
7.
After switching on the instrument, wait 10 minutes before beginning analysis.
8.
After powering off, wait at least 20 seconds before restarting.
9.
If line variations are higher than 10%, a ferroresonant voltage regulator or a 1500 VA uninterruptible power
supply (UPS) is recommended.
10.
Read the entire User’s Manual before using the instrument
11.
The laboratory must count on a special residues collection service, to handle wastes.
12.
Do not discard the analyzer, neither its accessories, along with domestic residues. Check the local terms for
its correct elimination. Is the user’s responsibility to deliver the analyzer in the indicated pickup point for
recycling of electric and electronic devices, otherwise get in contact with HTI or its authorized agent to
proceed to its removal in a safe and ecological way.
13.
Use only expendables and spare parts provided by HTI or its authorized agent.
14.
If any failure occurs, contact your local authorized technical service dept., or contact HTI.
Precautions
Electrical Risks
As with any electronic equipment, electric shock is always a potential threat. Extreme caution should be used
when working around the instrument to avoid contact with any electrical wire or components. DO NOT attempt to
work in any electronic compartment while the power is ON.
Mechanical Risks
Keep lids closed when the instrument is working.
Do not touch mechanical arms or other moving parts while the instrument is working.
Do not wear rings, bracelets, necklaces, etc. as they may get caught in operating mechanical systems.
Do not try to change or touch sample tubes, reagents, etc. when the instrument is running to prevent breakage
and/or an accident when the Reagent Sample Tray moves.
Make sure that the instrument is paused to add reagents, samples, etc.

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 2 of 217 HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Chemical risks
Always wear protective apparel when operating the analyzer (gloves, safety glasses, etc.). Follow specific
recommendations on the bottles of each reagent or washing solution.
Read safety information of materials provided by manufacturers to be aware of possible danger and to learn how
to prevent it.
Biological risks
Rings and long nails can easily break gloves and increase exposure to biological risks. Always treat samples and
reagents as potentially infectious.
Waste and waste deposits must be treated as toxic and biologically hazardous. Handle them and eliminate them
in accordance with routine laboratory protocols.
Follow in force standards given by local sanitary authority.
Eye risks
Do not look straight at barcode detector light beams.
Do not look straight at the light source of spectrophotometers since they can emit ultraviolet rays.
Warnings about the instrument and laboratory practices
The use of calibrators is recommended; refer to reagent manufacturer instructions for specific calibration
instructions.
Symbols used in the manual and the instrument
Warning
Ground Connection
Manufacturing Date
Consult instructions for use
Biological risk
Fuse
This symbol denotes the Catalog number

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 3 of 222
The symbols for “SERIAL NUMBER”, The serial number shall be after
or below the symbol, adjacent to it
The symbol indicates the manufacturer and its address, after which are
shown its name and address
This symbol denotes to Consult instructions for use
This symbol denotes that this product conforms to the directive 98/79/EC
(IVD-directive)
The symbol indicates EU representatives of the manufacturer and their
addresses, after which are shown their names and addresses
The symbol means the product is in vitro diagnostic medical device.
Symbols used on the box
This means that instrument should handle with care in the course of transportation, so as not to
damage it.
The symbol means that the environment of instruments must be damp-proof in the course of
transport, and instrument must be kept in a dry environment.
The symbol indicates temperature range of the analyzers during storage and transportation
Warranty Terms
HTI warrants the instrument for a period of 12 months as of the date the instrument leaves the factory. This
warranty is only effective as long as the pertinent warranty card together with the installation record with all the
data is duly completed and sent to us by certified mail.
Warning: This warranty only covers defective materials or manufacturing defects, according to
examination carried out in the factory and shall be limited to the replacement of defective materials.
Accessories which are not provided by HTI, such as: wash solutions, standards, controls, reagents and
consumables are not covered by the warranty. Furthermore, costs arising from mishandling the
instrument and/or damage to the instrument are not covered by this warranty.
This warranty shall not be valid if personnel not authorized by HTI attempt to repair the instrument.
WARRANTY SERVICE: Without charge at HTI. Otherwise, transport and traveling expenses of a technician to
the place where the instrument is to be repaired shall be paid.

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 4 of 217 HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
HTI RESERVES THE RIGHT TO MODIFY THE INSTRUMENTS WITHOUT AFFECTING THEIR OPERATING
PARAMETERS.
NOTE: The following elements are not covered by warranty terms: halogen lamps, fuses, cuvettes,
reagents bottles, tubes and tubing.
1 TRANSPORTATION, STORAGE AND UNPACKING
Transportation and storage
Do not pile up more than four wooden boxes
Keep dry, protect from humidity and rain
Fragile. Handle with care
Store at 0°C –50°C
Move using forklift

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 5 of 222
Unpacking instructions
Warning: Unpacking shall be carried out by trained personnel. Carefully unpack the instrument to
prevent damage.
Labels indicating this operation are found on each side of the packaging box.
Figure 1-1
Remove the wooden box lid by unscrewing the eight screws located in the front, back and sides.
Figure 1-2

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 6 of 217 HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
NOTE: Instrument remains fastened to the bottom.
Remove the instrument from the bottom of the box as shown in Figure 1-2 and as explained below. On the
upper part of the instrument you will find an Allen wrench Nº 6, which will be used to remove the screws
joining the bottom to the instrument.
Figure 1-3
Unscrew the four screws located on the sides.
Figure 1-4
Screws are placed this way:

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 7 of 222
Figure 1-5
Once the screws have been removed and the instrument is released, the protective Styrofoam is removed
and the instrument can be placed on the working surface.
2 INSTALLATION
Installation must be carried out by trained personnel.
Electrical requirements
A standard 100 Volts 60 Hz or 220 Volts 50 Hz socket for 400 VA consumption. The supply line has a third
terminal to ensure proper ground connection. If line variations are higher than 10%, a ferroresonant
voltage regulator, or a 1500 VA uninterruptible power supply (UPS sine wave output) is recommended.
Dimensions & Weight
Width: 800 mm, Height: 464 mm, Depth: 647 mm
Weight: 48kg

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 8 of 217 HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
Environmental requirements
Operating temperature: 20°C ~25°C
Operating Humidity: 20~85% (no condensation)
Storage Temperature: 0°C~50°C
Storage Humidity: ≤80%
Moving the instrument
Avoid hitting the instrument as well as subjecting it to any kind of vibration while moving it.
PC minimum requirements
2 GHz Pentium IV
RAM Memory, 512 MB
Independent video card
40 GB Hard Drive
CD Rom
CD Reader
Windows XP
Printer (optional)
The PC must comply with safety regulations IEC 60950 and have a certified source.
Instrument placement and preparation
Warning: Avoid hitting the instrument as well as subjecting it to any kind of vibration while moving
it.
The instrument must be placed on a counter in a room free from dust and corrosive vapors to ensure its
proper operation.
The instrument must not be subject to vibrations or sudden changes of temperature.
Leave a 20 centimeter separation between the equipment and the wall to ensure proper ventilation. Liquid
containers must be under the counter. Avoid collapsing and/or curving of waste exit tubing.
Avoid any centrifuge, lifts or x-ray supply lines or any other kind of noise-generating equipment in the
supply line.
Avoid direct sun light or illumination on the instrument.
The supply line has a third ground conductor for protection purposes.

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 9 of 222
Connect the instrument to the computer and its peripherals before connecting to the supply line. If the
above requirements are not met, the quality of results may be affected.
Warning: Make sure that ground connection in the line meets the standard requirements for its
operating power. Not performing ground connection poses significant safety risk for the operator
and may damage one or several parts of the equipment or the computer.
Warning: Waste solution tubing must be placed correctly so that it drains by gravity. They must
not be curved and/or collapsed.
Warning: The instrument must be connected to a PC which meets the requirements previously
mentioned.
Warning: The instrument has been tested with a tensioactive solution which is added to the
distilled water. Not using or altering the product or its dilution will affect the operation of the
instrument.
Wires and Tubing Connection
Figure 2-1
1. Connect the interlock to 100~220V, 50~60Hz supply.
If line variations are higher than 10%, a ferroresonant voltage regulator or a 1500 VA uninterruptible
power supply (UPS sine wave output) is recommended.
2. Connect the RS-232 wire to the PC before switching on the instrument.
For analyzers with ISE Module installed continue with step (3). Otherwise go to step (8)
3. Connect the ISE module connector to the reagent Kit.
4. Remove all electrodes from sealed bags.
5. Remove tape from K+ and Li+ electrode.

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 10 of 217 HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
If necessary, soak the reference electrode in warm water until the lumen of the electrode has been
cleared of salt build-up.
6. Depress the compression tray to install electrodes in position (as shown in Figure 2-2)
Figure 2-2
7. Press the lock stick and move down the ISE module to put in place
Warning: Do not turn off the analyzer by pressing general button unless maintenance cycle has
been done and electrodes removed.
NOTE: ISE module must be power on all time that the electrodes are in it. Purge cycles are
performed by the ISE module automatically every 30 minutes.
8. Connect tubing to the appropriate containers.

High Technology, Inc.
109 Production Road,
Walpole MA 02081 USA
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared by
J.S. Bolio
Approved by
S. Titov
Effective Date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 11 of 222
Figure 2-3
Warning: The tubing with the biggest diameter (waste funnel) must be kept straight downwards
without curves to allow free drain of fluids.
Washing solution
Warning: Containers must be placed under the analyzer.
Warning: Container must not be placed on the same analyzer level.
2.7.1.1 Washing solution for probe and cuvettes
First wash concentrate solution has to be prepared from Triton X-100 Solution. Then, use wash
concentrate solution to prepare washing solution for probe and cuvettes.
Wash Concentrate: dilute 1 part v/v of Triton X-100 in 9 parts of distilled water, Ex: 100 ml Triton X-
100 in 900 ml of distilled water.
The Triton X-100 is extremely viscous. It is suggested to pour the desired quantity into a graduated
cylinder and gradually add the water. Heat the distilled water to 70 –80 ºC to help dissolve the Triton
X-100.
Washing solution for probe and cuvettes: dilute 7 ml of wash concentrate in 1 liter of distilled water.
Ex: 140 ml of wash concentrate per 20 liters of distilled water.

High Technology, Inc.
109 Production Road,
Walpole MA USA 02081
Document Number
OM-E-FC-360-30
Revision Level
Rev. 3
Prepared By
J.S. Bolio
Approved By
S. Titov
Effective date
8/15/2016
Operator’s Manual: HTI BioChem FC-360
Page 12 of 217 HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL
2.7.1.2 Washing solution for probe and cuvettes when running turbidimetric methods
In those cases where latex reagents are used (most turbidimetric methods), is highly
recommended to use the following washing solutions:
Washing solution for probe: use only DI water
Washing solution for cuvettes: dilute 3 ml of wash concentrate in 1 liter of distilled water. Ex: 70 ml
of wash concentrate per 20 liters of distilled water.
Warning: not following the instructions given above will affect the results.
Installation
PC configuration
Open Control Panel Display Settings into Screen Resolution field set 1024x768 pixels
Control Panel Date and Time Internet Time unselect Automatically synchronize with an
Internet time server
Uninstall antivirus
Uninstall screensaver
2.8.1.1 Install analyzer software.
1.
Open “Computer”.
2.
Find the device where the software installer is.
3.
Double Click on the analyzer software installer executable.
4.
Press Next to continue with the installation or Cancel to exit.
5.
Choose between “Everyone and Just Me” option (see NOTE).
6.
Press Next to continue with the installation or Cancel to exit. In this step the file folder path can be
changed.
7.
Press Next to start installation process.
8.
Press Close to close the window.
NOTE: “Everyone” option means that the software will be installed in all the user accounts
existing at the moment of installation. “Just Me” option means that the software will be
installed in the account that the installer is being executed (if a new account is created then
the analyzer software must be installed in the new account). If “Just Me” option is chosen the
other users cannot use analyzer software, unless they install the analyzer software too, but
they have to change the path of all the files when the software is being installed. Maintenance
Settings
Files tab: all path must be changed for a different location in this user account.
9.
Open analyzer software from desktop.
Table of contents
Other HTI Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

ETI
ETI DEC-1 quick start guide

Vega
Vega VEGAPULS 68 operating instructions

Chauvin Arnoux
Chauvin Arnoux C.A 6681 LOCAT-N user manual

HEIDENHAIN
HEIDENHAIN MSE 1201 installation instructions

Klein Tools
Klein Tools CL2200 instruction manual

Photon Systems Instruments
Photon Systems Instruments FluorPen FP 110 Manual and user guide