IBIOM Instruments Ltd TRIAD User manual

OPERATION MANUAL-TRIAD™
IBIOM Instruments Ltd, 670 Joseph-Louis-Mathieu Street, Sherbrooke, Quebec, J1R 0X3
TABLE OF CONTENTS
1. CONTACT US.................................................................................................................... 3
2. INTENDED USE ................................................................................................................ 4
3. SYMBOL IDENTIFICATION .............................................................................................. 4
4. WARNINGS....................................................................................................................... 5
5. LABEL............................................................................................................................... 7
6. OPERATION INSTRUCTIONS .......................................................................................... 8
6.1 INSTALLATION .................................................................................................................. 8
6.2 CENTRAL AND DIRECTION LOCK........................................................................................10
6.3 CHAIR FUNCTIONS...........................................................................................................11
6.4 ARMREST .......................................................................................................................12
6.5 EMERGENCY POSITION -ELECTRICAL................................................................................15
6.6 SIDE TABLE.....................................................................................................................16
6.7 RETRACTABLE SIDE TABLE...............................................................................................17
6.8 RECHARGEABLE BATTERY................................................................................................18
6.9 EMERGENCY FOOT SWITCH..............................................................................................19
7. CLEANING INSTRUCTIONS............................................................................................20
7.1 RECOMMENDED CLEANING PRODUCTS..............................................................................20
7.2 CLEANING PROCEDURE ...................................................................................................21
8. SPECIFICATIONS............................................................................................................22
8.1 PRODUCT SPECIFICATIONS ..............................................................................................22
8.2 ELECTRICAL SPECIFICATIONS...........................................................................................23
8.3 ENVIRONMENTAL CONDITIONS..........................................................................................23
8.4 CERTIFICATION ...............................................................................................................24
9. LIMITED LIFETIME WARRANTY.....................................................................................25
10. WASTE DISPOSAL..........................................................................................................26
11. MAINTENANCE................................................................................................................27
12. TROUBLESHOOTING......................................................................................................28
13. SPARE PARTS.................................................................................................................29

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 3 of 30 Revision: January 2021
1. CONTACT US
Before contacting us, make sure to have the following information ready:
➢7-digit serial number (example: 1234567).
➢Model number or name (example: TRIAD).
This information is available on label located on base of the chair.
Phone
1-877-563-5777
Fax
450-445-9837
Email
Web
www.ibiom.com
Address
IBIOM Instruments Ltd
670 Joseph-Louis-Mathieu Street
Sherbrooke, Quebec
J1R 0X3
Canada

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 4 of 30 Revision: January 2021
2. INTENDED USE
TRIAD chair is AC powered treatment chair that can be adjusted to different position. It is used to
support the patient during medical examination or procedure. It is intended to be operated by
healthcare professional in a medical environment including hospitals, clinics and doctor’s office.
3. SYMBOL IDENTIFICATION
The following symbols and their definitions are used throughout this manual.
Symbol
Definition
WARNING. Alerts the reader about a situation, which if not avoided, could result
in death or serious injury. It may also describe potential serious adverse reactions
and safety hazards.
Warning, electrical shock hazard.
Operator must read manual before using this equipment.
Operator must read manual before using this equipment.
Class II or double insulated electrical equipment.
Electrical safety, type B applied part.
No sitting.
No stepping.
This device contains materials that are potentially hazardous to the environment.
In accordance with directive on wasteelectrical and electronic equipment(WEEE),
TRIAD electrical system and accessories should not be disposed as unsorted
municipal waste. Consultyour instructional policies and local regulations regarding
disposal.

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 5 of 30 Revision: January 2021
4. WARNINGS
MANDATORY. Read, understand and keep this manual before starting to use
IBIOM’s products. This manual must be available for everyone using IBIOM’s
products.
WARNING. TRIAD chair is designed for indoor use on flat surfaces.
WARNING. Always lock wheels before the patient is placed in the chair.
WARNING. Never use braking system to stop a moving chair.
WARNING. Do not use TRIAD chair for any purpose other than the intended use.
WARNING. Do not modify this equipment without IBIOM's authorization.
WARNING. Do not use in oxygen-rich environment.
WARNING. Do not use in MRI environment.
WARNING. Do not use in wet environment.
WARNING. Do not use for CPR or X-ray procedures or with defibrillator.
WARNING. Never exceed safe working load of 450lb or 205kg on TRIAD
treatment chair. Safe working load includes patient’s weight and accessories.
WARNING. Never exceed safe working load of 100lb or 45kg on TRIAD’s
armrests. Safe working load includes patient’s weight and accessories (optional
retractable side table for example).
WARNING. Neverexceed safe working loadof 25lb or 11kg on TRIAD’s side table.
WARNING. Never exceed safe working load of 10lb or 4.5kg on TRIAD’s
retractable side table.
WARNING. Never exceed safe working load of 100lb or 45kg on TRIAD’s footrest.
Safe working load includes patient’s weight and accessories (optional exercise
bike for example).
WARNING. Do not sit on backrest.
WARNING. Do not sit on leg rest.
WARNING. Do not sit on armrests.
WARNING. Do not leave patient unattended.
WARNING. Do not use TRIAD chair as a transport stretcher to move the patient
from one room or floor to another.
WARNING. To ensure proper operation of the chair, maintenance must be done
as described in this manual.
WARNING. Read the warnings in the various sections of this manual.
WARNING. Do not roll on power cable or hand control cable.
WARNING. Keep all cables out of reach of children.
WARNING. Keep all cables away from moving parts.

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 6 of 30 Revision: January 2021
WARNING. Do not move chair by pulling hand control, power cable or any other
cable.
WARNING. Never use the chair if the power cable or any other cables are
damaged.
WARNING. Keep the power cable out of heat sources.
WARNING. Always ensure there is sufficient clearance around the chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
WARNING. Never exceed duty cycle of 2/18 (2 minutes ON and 18 minutes OFF).
WARNING. Medical staff should inform the patient on how to adjust the chair.
WARNING. If patient is not able to adjust the chair, the hand control should be
placed out of the patient's reach.
WARNING. Avoid placing hands andother body parts near moving parts when the
chair is adjusted.
WARNING. Always lower and close the armrests once the patient is in place on
the chair.
WARNING. Always ensure that the rotation of the armrest is locked in the closed
position once the patient is in place on the chair.
WARNING. Always lower chair to its lowest position before the patient is seated
in the chair.
WARNING. The patient must take position on the chair from the side of the chair
only.
WARNING. Always ensure that side table mechanism is tighten solidly before the
patient is placed in the chair.
WARNING. Take appropriate precautions to avoid injury when handling
accessories (side table, arms, etc.).
WARNING. Defective or damaged batteries may leak acid and adequate
precautions must be taken during handling and transportation.
WARNING. Do not open the battery case as damage to the cell or circuitry may
develop excessive heat.

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 7 of 30 Revision: January 2021
5. LABEL
The product identification label is located on the product base. This image and the information
contained herein is for informational purposes only.
Symbol identification
Manufacturer’s name and address
Product code or name
7-digit serial number
Product’s weight
Date of manufacture (YYYY-MM-DD)
Maximum safe working load (maximum patient weight)
Made in Canada
QR code pointing to IBIOM’s website
Barcode including date of manufacture and serial number
Type B applied part
Double electrical insulation
Obligation to read the operation manual
Required voltage (120V), alternating current
Frequency (60Hz), max amperage (3.9A)
Maximum activation time followed by minimum deactivation time

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 8 of 30 Revision: January 2021
6. OPERATION INSTRUCTIONS
6.1 Installation
WARNING. Always ensure there is sufficient clearance around the chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
WARNING. TRIAD chair is designed for indoor use on flat surfaces.
WARNING. Always lock wheels before the patient is placed in the chair.
WARNING. Never use braking system to stop a moving chair.
WARNING. Do not leave patient unattended.
WARNING. Do not use in oxygen-rich environment.
WARNING. Do not use in MRI environment.
WARNING. Do not use in wet environment.
WARNING. Do not use for CPR or X-ray procedures or with defibrillator.
WARNING. Do not roll on power cable or hand control cable.
WARNING. Keep all cables out of reach of children.
WARNING. Keep all cables away from moving parts.
WARNING. Do not move chair by pulling hand control, power cable or any other
cable.
WARNING. Never use the chair if the power cable or any other cables are
damaged.
WARNING. Keep the power cable out of heat sources.
WARNING. Never exceed duty cycle of 2/18 (2 minutes ON and 18 minutes OFF).
TRIAD chair is delivered assembled and ready for use. The chair must be unpacked and
inspected upon receipt to ensure that there has been no damage during transport.
1. Once unpacked and put in desired location, install all accessories (arms, side table).

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 9 of 30 Revision: January 2021
2. Position TRIAD chair in the desired position. IMPORTANT: Make sure you have enough
space around the chair to avoid collisions between its surroundings in ALL POSITIONS.
3. Once in place, lock all wheels.
4. Remove any remaining packaging from the product. Example: Protective plastic on the
hand control cable.
5. Connect the power cord.
6. Perform an operation test to ensure that everything is working properly.

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 10 of 30 Revision: January 2021
6.2 Central and direction lock
WARNING. Always lock wheels before the patient is placed in the chair.
WARNING. Never use braking system to stop a moving chair.
WARNING. Do not move chair by pulling hand control, power cable or any other
cable.
WARNING. Always ensure there is sufficient clearance around the chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
Central lock mechanism can be engaged with pedal located on both sides of treatment chair.
Pedal needs to be triggered by foot to one of the 3 positions:
•Red: all 4 casters are locked;
•Neutral: all casters unlocked and free to roll;
•Green: 1 caster at foot end locks in direction and 3 others remain free.
Direction
(green)
Central lock
(red)

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 11 of 30 Revision: January 2021
6.3 Chair functions
WARNING. Always ensure there is sufficient clearance around the chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
WARNING. Always lock wheels before the patient is placed in the chair.
WARNING. Always lower chair to its lowest position before the patient is seated
in the chair.
WARNING. Do not leave patient unattended.
WARNING. Medical staff should inform the patient on how to adjust the chair.
WARNING. If patient is not able to adjust the chair, the hand control should be
placed out of the patient's reach.
WARNING. Avoid placing hands andother body parts near moving parts when the
chair is adjusted.
WARNING. Never use the chair if the power cable or any other cables are
damaged.
Hand control device
Button
Function
A
Raise backrest
B
Lower backrest
C
Raise chair
D
Lower chair
E
Raise leg section
F
Lower leg section
G
Raise footrest
H
Lower footrest
I
Simultaneous movements: raise backrest and lower leg section.
J
Simultaneous movements: lower backrest and raise leg section.
A
B
C
D
E
F
G
H
I
J

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 12 of 30 Revision: January 2021
6.4 Armrest
WARNING. Never exceed safe working load of 100lb or 45kg on TRIAD’s
armrests. Safe working load includes patient’s weight and accessories (optional
retractable side table for example).
WARNING. Never exceed safe working load of 10lb or 4.5kg on TRIAD’s
retractable side table.
WARNING. Do not sit on armrests.
WARNING. Always lower and close the armrests once the patient is in place on
the chair.
WARNING. Always ensure that the rotation of the armrest is locked in the closed
position once the patient is in place on the chair.
WARNING. Do not leave patient unattended.
WARNING. The patient must take position on the chair from the side of the chair
only.
WARNING. Medical staff should inform the patient on how to adjust the chair.
WARNING. Take appropriate precautions to avoid injury when handling
accessories (side table, arms, etc.).
Removable armrest
1. Remove pin by holding metal ring. IMPORTANT : DO NOT pull lanyard holding pin to the
chair.
2. Remove armrest from tube.
1
2

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 13 of 30 Revision: January 2021
Swivel armrest
WARNING. Always ensure that the rotation of the armrest is locked in the closed
position once the patient is in place on the chair.
1. Push and hold lever.
2. Rotate armrest to desired position and release lever.
1

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 14 of 30 Revision: January 2021
Lifting armrest
WARNING. Always lower and close the armrests once the patient is in place on
the chair.
Each armrest can be lifted. Lifting may be locked with combination of some options.
1. Hold armrest underneath.
2. Raise armrest until it stops to its raised position. No locking system is required.
2
1

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 15 of 30 Revision: January 2021
6.5 Emergency position - electrical
WARNING. Do not use TRIAD chair as a transport stretcher to move the patient
from one room or floor to another.
WARNING. Always ensure there is sufficient clearance around the chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
To activate emergency position, use corresponding buttons on hand control device. This function
moves simultaneously backrest and leg section.

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 16 of 30 Revision: January 2021
6.6 Side table
WARNING. Neverexceed safe working loadof 25lb or 11kg on TRIAD’s side table.
WARNING. Always ensure that side table mechanism is tighten solidly before the
patient is placed in the chair.
WARNING. Take appropriate precautions to avoid injury when handling
accessories (arms and side table).
1. Slide mounting block on either left or right insert located underneath mattress.
2. Tighten knob until shelf has no loose.
3. To remove, loosen knob and remove whole assembly.
1
2

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 17 of 30 Revision: January 2021
6.7 Retractable side table
WARNING. Never exceed safe working load of 10lb or 4.5kg on TRIAD’s
retractable side table.
1. Slightly lift side table.
2. Fold side table gently.
1
2

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 18 of 30 Revision: January 2021
6.8 Rechargeable battery
WARNING. Never exceed duty cycle of 2/18 (2 minutes ON and 18 minutes OFF).
WARNING. Defective or damaged batteries may leak acid and adequate
precautions must be taken during handling and transportation.
WARNING. Do not open the battery case as damage to the cell or circuitry may
develop excessive heat.
•CAUTION. NEVER discharge completely battery or it may be damaged and reduce its
lifetime.
•Always allow battery to settle to room temperature before use.
•Do not exceed duty cycle 2/18 as it will shorten the life and reduce performance of the
battery.
•After every day of use, make sure to charge battery for 8 hours. Plug power outlet to the
wall to charge.
•If battery is not used every day, charge as needed.
•It is possible to use battery while it is plugged to wall outlet without any risk of damage.
•When battery level is low and needs to be charged, it will emit audible signal.

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 19 of 30 Revision: January 2021
6.9 Emergency foot switch
WARNING. Always ensure there is sufficient clearance around the chair before
activating the various movements to avoid collisions with the environment,
equipment and people.
WARNING. Avoid placing hands andother body parts near moving parts when the
chair is adjusted.
Emergency foot switches are located on both side of the chair.
1. To activate the emergency position, press the pedal displaying the negative symbol. This
function allows the backrest to be lowered and the leg section to be raised simultaneously
to place the patient in the lying position (Trendelenburg).
2. To straighten the backrest and lower the leg section simultaneously, press the pedal
displaying the positive symbol.
1
2

OPERATION MANUAL-TRIAD™
© Copyright IBIOM Instruments ltd 2021 Page 20 of 30 Revision: January 2021
7. CLEANING INSTRUCTIONS
WARNING. Always clean chair between patient to avoid contamination.
WARNING. Clean chair immediately after spillage of bodily or other fluids.
WARNING. Protect mattress cover from sharp and rough objects to avoid
damaging it.
WARNING. Always inspect mattress to ensure its integrity. A damaged mattress
can absorb contaminated liquids. In case of tear or wear, contact IBIOM to
immediately replace the damaged mattress.
WARNING. Do not wash with pressure or steam.
WARNING. It is the user's responsibility to wear adequate protective equipment to
avoid contamination from bodily fluids or other dirt on the chair.
WARNING. Any damage to the mattresses caused by the use of a product other
than those mentioned in this section or at a different concentration is not covered
by the warranty.
WARNING. Do not use cleaning products containing other products or in
concentration other than those specified in this section.
WARNING. Do not use abrasive cleaning products.
WARNING. Avoid placing hands and other body parts near moving parts when the
chair is adjusted.
WARNING. It is the user's responsibility to ensure that these instructions are in
accordance with the institution's internal infection control procedures.
7.1 Recommended cleaning products
-Disinfectant containing sodium hypochlorite (bleaching product) in a maximum concentration
of 0.5%. Always rinse with clean water and dry with a clean cloth.
-Disinfectant containing hydrogen peroxide diluted in a maximum concentration of 0.5%.
Always rinse with clean water and dry with a clean cloth.
-Disinfectant containing alcohol diluted in a maximum concentration of 22%. Always rinse with
clean water and dry with a clean cloth.
-It is recommended to use “ready-to-use” cleaning products.
Notes:
•Always test the cleaning product in an inconspicuous place to check the effect of this
product on the cleaned surface.
•Always follow manufacturer’s cleaning products instructions.
•Clean spill as soon as possible to prevent penetration and becoming a permanent stain.
Table of contents
Other IBIOM Instruments Ltd Indoor Furnishing manuals
Popular Indoor Furnishing manuals by other brands

Forte
Forte MTBT131 Assembling Instruction

TUHOME Furniture
TUHOME Furniture PERU FOUR DRAWER DRESSER Assembly instructions

Better Homes and Gardens
Better Homes and Gardens Canton BHS336158813011 manual

fantastic furniture
fantastic furniture Torkay Table D6 manual

Cheyenne
Cheyenne CPFV1130 Assembly instructions

Probuilder
Probuilder 52461 instruction manual

Tennsco
Tennsco ESPC-1236 Assembly Instructions/Parts Manual

Hillsdale Furniture
Hillsdale Furniture 4236-883 quick start guide

Germania
Germania Writing Desk 4036 Assembly instruction

ROOMS TO GO
ROOMS TO GO MAAH RU70137HM manual

PARAMOUNT BED
PARAMOUNT BED PE-6001 instruction manual

Next
Next Bronx 508137 Assembly instructions