
3
INTRODUCTION
The ICAM e-series is available in several versions, depending on the number of sample tube inlets and build options fitted. All types, except
the FT1 and Pico, incorporate a rotary valve enabling individual sample pipes to be selected. Under normal conditions, the FireTracer will
monitor all tubes simultaneously. Once smoke has been detected, it will go into Trace mode, sampling each pipe in turn until the source
has been located. The FT1 has an inlet manifold which combines samples from up to four inlet pipes and the Pico uses a single inlet which
samples from a T-piece giving two inlet pipes. All models, except the Pico, have flow monitoring on each inlet.
Two fundamentally different methods of aspiration are used. The Pico, FT1, FT4 and FT6 use an internal fan and depend on relatively high
flow rate sampling through wide-bore rigid sample tubes at near ambient pressure. These are referred to as low pressure systems. The
FT8 and FT15 have an internal vacuum pump capable of drawing air samples down small bore tubes. These are referred to as high
pressure systems. Where installations need to be unobtrusive, small bore systems will often be preferred. Wide bore tubes generally have
holes drilled at intervals within the protected area. Small bore tubes normally sample at their open end points.
The FIRETRACER 4/6/8/15 has an optional reference channel using a dedicated sample tube. Monitoring the fresh air to the protected
area, the system automatically adjusts measurements for changes in the ambient level.
Other features can be added as options to the basic system, including a high intensity warning beacon, ethernet and Web capability, and
environmental features such as temperature, humidity, flood and security. High pressure systems may also incorporate monitoring of up to
four gases.
E-Series Model Pico FT1 FT4 FT6 FT8 FT15
Number of Sample Tubes 2 4 4 6 8 15
Individual Sampling (rotary valve) No No Yes Yes Yes Yes
Individual Flow Monitoring No Yes Yes Yes Yes Yes
Flow Monitoring System Thermal Pressure
Sample tube inlet diameter 25mm O.D 21.4mm or ½”BSP 6mm O.D/4mm ID
Aspirator system Fan Pump
Area Coverage 850m22000m22000m22000m22000m22000m2
Available with Beacon Option No Yes Yes Yes Yes Yes
Available with Ethernet Option Yes Yes Yes Yes Yes Yes
Available with Gas Detection Option Yes No No No Yes Yes
Available with Environmental Option Yes Yes Yes Yes Yes Yes
Principle of Operation
The smoke detector itself has a laser beam directed across an optical chamber, through which the air sample flows. A photo-detector built
into the optical chamber measures the amount of light scattering. A clean air sample will cause very little scattering. As the smoke density
of the sample increases, the amount of light directed onto the detector will also increase. The light signal is processed to become a direct
measurement of obscuration.
Samples from all inlets (sectors) are combined as an overall sample, filtered, and fed to the smoke detector. Systems incorporating a rotary
valve have an Overall position which draws air equally from all sectors. The FT1 has an inlet manifold which performs the same function. If
smoke is detected, one or more Alarm relays, preset to activate at incremental smoke densities will signal the host panel after a preset
time delay. There are four stages of Alarm: ALERT, ACTION, FIRE1 and FIRE2. Systems with a rotary valve (FT4 and above) have an
additional preset trigger level, TRACE, usually set below the ALERT level. When TRACE is activated, the rotary valve will sequentially scan
the sectors, in order to determine the source of the event. While in TRACE mode, the default levels for ALERT, ACTION, FIRE1 and FIRE2
are the same as used for OVERALL monitoring. If required, different levels for ALERT, ACTION, FIRE1 and FIRE2 may be set for each
sector in the SECTOR ALARMS menu. The OVERALL Alarm levels can also be changed as can all delay times. Alarm states are shown on
the display panel, and an audible warning is given. An optional Alarm beacon can also be fitted.
Flow Monitoring
The control system monitors for blockages or disconnection of tubes on an individual sector basis. High pressure systems must have equal
pipe lengths for all sectors. Low pressure systems normalise flow readings to nominally 100% for each sector; a calibration sequence to set
this up must be performed at the installation stage. The unit has default high and low limits and associated delay times, but these may be
changed if necessary in the CONFIGURE menu.