IDEAL INDUSTRIES 61-724 User manual

IDEAL INDUSTRIES, INC.
TECHNICAL MANUAL
MODEL: 61-724
The Service Information provides the following information:
•Precautions and safety information
• Specifications
• Performance test procedure
• Calibration and calibration adjustment procedure
• Basic maintenance (replacing the battery)
Form number: TM61724
Revision: 2. Date: September 2006
Form number TM61724 Rev 2 Sept 2006

TABLE OF CONTENTS
Title Page
Introduction 1
Precautions and Safety Information 1
Safety Information 1
Specifications 2
General Specification 2
Voltage Specifications 2
Current Specifications 2
Resistance Specifications 2
Frequency Specification 2
Diode Specifications 2
Continuity Specifications 2
Performance Verification 3
Calibration 4
Clamp Jaw Alignment Figures 5
Replacing the Battery 5
Form number TM61724 Rev 2 Sept 2006

Page 1
Introduction Warning
To avoid shock or injury, do not perform the verification tests or calibration
procedures described in this manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel only.
Caution
The 61-724 contains parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
For additional information about IDEAL INDUSTRIES, INC. and its products,
and services, visit IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION.
These statements identify conditions or practices that could result in damage to the equipment or other
property.
WARNING.
These statements identify conditions or practices that could result in personal injury or loss of life.
Specific precautions
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without the covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
Form number TM61724 Rev 2 Sept 2006

Page 2
General specifications
Characteristics Description
Display 3½ Digit LCD display
Display Count 2000 count, maximum reading 1999
Overrange Indication “OL” is displayed
Sampling Rate 2.5 time/second
Operating Environment:
Relative Humidity 0°C to 50°C (32°F to 122°F)
0 ~ 75% RH
Storage Environment: -20°C to 60°C (-4°F to 140°F) at <80% relative humidity
Power source: 9V Battery (NEDA 1604)
Battery Live: 150 hours typical (alkaline)
Low Battery Indicator: symbol indicates low battery voltage
Dimensions 9.7” H X 3.7” W X 1.8” D
247mm H X 94mm W X 46mm D
Maximum Cable Size ACA 1.8” (46mm), DCA 1.9” (48mm)
Weight: Approximately 13.4 oz. or 430g including battery
RANGES and ACCURACY SPECIFICATION
Accuracy: Accuracy specifications at 23°C ±5°C (73.4°F ±9°F) less than 75% RH.
Temperature Coefficient: 0.1 times the applicable accuracy specification per °C
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F)
Electrical Specification: Accuracy are ±(reading plus number of digits) at 23°C ±5°C <75% RH
Function Setting Ranges Accuracy
AC Voltage 200.0/600V at 50-60 Hz 1.2% ±4digits
AC Voltage 200.0/600V at 40-500 Hz 2.0% ±4 digits
DC Voltage 600V 0.5% ±1 digit
AC Current 200.0/1000A at 50-60 Hz 1.75% ±4 digit
AC Current 200.0/1000A at 40- 500 Hz 3.50% ±5 digit
DC Current 200/1200A 1.5% ±5 digits
Frequency 2K ~ 20KHz (Auto-ranging only) 0.1% ±3 digits
Resistance 2.000K/200.0K 1.2% ±1 digit
Diode Check 1mA ±0.6mA 6.0% ±3 digits
Continuity Approximately < 30Ωon 2KΩ,
<100Ωon Diode, Continuity 6.0% ±3 digits
AC Converter: True RMS Responding, RMS Calibrated to Sine Wave
Overload Protection: AC and DC Voltage: 600 VDC or VAC rms for no more than one minute.
Resistance, Diode check, Continuity.: 600 VDC or VAC rms
Current: Not to exceed 1200 amp DC or 1000 amp AC for
no more than one minute.
Form number TM61724 Rev 2 Sept 2006

Page 3
PERFORMANCE VERIFICAITONS
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning
correctly. If the meter does not conform to any of the listed limits the calibration procedure must be
performed.
Performance Verification Preparation
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization
should be reached after 30 minutes.
2. Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5V
DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test.
3. Input the values listed in Table 1
Table 1 Performance Verification
Function
Setting/Range
Input
Low Limit
High Limit
ACV 200
ACV 200 190V AC @ 50Hz
190V AC @ 500Hz 187.3
185.8 192.7
194.2
ACV 600
ACV 600 500V AC @ 60Hz
500V AC @ 500Hz 490
486 510
514
DCV 600 500V DC 496 503
ACA 200A 100A @ 50Hz 97.9 102.1
ACA 200A 100A @ 500Hz 96.0 104.0
ACA 1000A 1000A @ 50Hz 979 1021
DCA 200A
DCA 1000A 100A
1000A 98.0
980 102.0
1020
Hz
Hz 2.0 KHz @ 80V rms
20 KHz @ 80V rms 1.995
19.95 2.005
20.05
Ω2K 1.000K .987 1.013
Ω200K 100.0K 98.7 101.3
Diode Test .500V DC .467 .533
Continuity Test 20 Ω, Beep on
40 Ω, Beep off.
Form number TM61724 Rev 2 Sept 2006

Page 4
CALIBRATION
Calibration Preparation
1. Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization
should be reached after 30 minutes.
2. Disconnect the test leads and turn the range switch to “OFF”.
3. Remove the screws holding the battery cover and one at the jaw.
4. Remove the case bottom using care not to damage the battery connector and leads
to the continuity beeper. (Beeper is attached to the bottom case cover.)
5. Using a calibrated meter ensure the battery measures a minimum of 7.5V DC.
If the battery measures under 7.5V DC, replace the battery.
Calibration Procedure
It is recommended that all IDEAL meters undergo the following calibration procedure on
an annual basis.
The class of calibrator or equipment should have an accuracy that exceeds, by an expectable ratio the
accuracy of this instrument.
Volts DC Calibration
1. Set the function / range to 600V DC
2. Connect the calibrator to the Vand COM inputs on the meter.
3. Output 100V DC.
Adjust R43 until unit display reads 100V
De-energize source and remove test leads
DCA Zero Calibration:
1. Remove the Jumper
2. Set the function / range to the 200A DC
3. Without any signal applied (keeping Jaw away from other devices)
adjust R54 for a display reading of 0.0 ±1 digit
4. Replace the Jumper, then press the “ZERO” button
adjust R506 for a display reading of 0.0 ±1 digit
DCA Balance Calibration
1. Set the function / range to the 200A DC.
2. Clamp the unit to a 100A DC source.
2.a Move the jaw to the top of the clamp {see figure 1a} and note reading
2.b Move the jaw to the bottom of the clamp {see figure 1b} and note reading
2.c Repeating steps 2a and 2b, Adjust R581 until the difference of measured readings
in step 2a and 2b is within 5 digits
DCA Calibration
1 Set the function / range to 200A DC
2. Clamp the unit to a 100A DC Source
Note: Use care to keep the Jaw (CT) of the unit properly centered to the Source
3. Adjust R62 until the display reads 100.0 ±2 digit
Calibration of the 61-724 is complete.
Remove all leads from the calibrator and equipment.
Return unit to proper operating condition.
Form number TM61724 Rev 2 Sept 2006

Page 5
Placement of source for DC Balance Calibration
Battery Replacement (refer to Figure 2)
1. Disconnect the test leads from any circuit under test and turn off meter.
2. Use a Philips head screwdriver to remove the two screws on battery cover
3. Remove battery from compartment and unsnap the battery connector.
4. Install new 9V battery (NEDA #1604). An alkaline type is recommended.
5. Install new battery into compartment using care not to pinch or bind battery leads
6. Reinstall battery cover.
Figure 2
Form number TM61724 Rev 2 Sept 2006
Table of contents
Other IDEAL INDUSTRIES Measuring Instrument manuals

IDEAL INDUSTRIES
IDEAL INDUSTRIES 61-721 User manual

IDEAL INDUSTRIES
IDEAL INDUSTRIES 61-773 User manual

IDEAL INDUSTRIES
IDEAL INDUSTRIES 61-795 User manual

IDEAL INDUSTRIES
IDEAL INDUSTRIES SignalTEK II User manual

IDEAL INDUSTRIES
IDEAL INDUSTRIES 806 User manual

IDEAL INDUSTRIES
IDEAL INDUSTRIES 61-340 User manual

IDEAL INDUSTRIES
IDEAL INDUSTRIES 62-160 User manual
Popular Measuring Instrument manuals by other brands

Gossen MetraWatt
Gossen MetraWatt MAVOLUX 5032C USB operating instructions

ESD SYSTEMS
ESD SYSTEMS SPl-20686 Operation installation and maintenance

Syco
Syco SWR quick start guide

Argus
Argus AA300 user manual

Mirion Technologies
Mirion Technologies RDS-31 S user manual

Elcometer
Elcometer 456 3 operating instructions