IDS IDS-iSYS User manual

IS-310400 IDS-iSYS
(Multi-Discipline Automated System)
User Manual
Software version V 14
For further assistance,
please contact IDS Technical Support:
Tel: 0191 519 6153
Fax: 0191 519 0760
Email: [email protected]
IDS France , 42, rue Stéphane Mazeau, 21320 POUILLY-EN-AUXOIS, FRANCE
IDS Ltd 10 Didcot Way, Boldon Business Park,
Boldon, Tyne and Wear, NE35 9PD
Tel: 0191 519 0660

This instruction manual is to be used in conjunction with the
in vitro diagnostic analyser and must be read before
installing or using the analyser.
in vitro diagnostic analyser must only be used by
personnel trained by approved IDS staff.
The purpose of the user manual is to explain:
•The way the analyser works.
•How to use it in routine working practice.
•The preventive maintenance required.
Revision M1 of the User Manual was produced on the 21st December 2015, for the software
version:
V 14.04
The information contained in this document is applicable to any subsequent software version identified as «V
14.XX». Changes with the decimals of the version identifier are used to account for minor software
enhancements, concerning neither the features nor the use of the system.
The manual comprises the following sections and appendices:
Section 1: !Operating Principle
Section 2: !User Interface Software
Section 3: !Use !
Section 4: !Messages
Section 5:!Maintenance
Section 6:!Problems And Corrective Action
Section 7:!System Configuration
Appendix I:!Waste Disposal
Appendix II: !Decontaminating The Analyser
Appendix III: !Disposal Of The Analyser
Appendix IV: !IDS-iSYS Cuvettes
Appendix V: !Sample barcode symbology managed by the system
Appendix VI: !Analyser long stoppage period
Attached document: Protocol for connection, revision N1
!If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
If the recommendations contained in this manual are not respected, the level of performance
offered by the analyser may be impaired and the results generated may be incorrect.
Preface
IDS-iSYS User Manual - Revision M1"P 1
Software version V 14

New in the manual Revision M1
Section 1
•Modification in Characteristics, pages 7 to 9.
•Modification in 1-1-6-Precautions for use, page 15: use of the sample type «Other» in immunoassays.
Section 2
•Modification in 2-3-1-FILE Menu, page 25: adding of a new feature accessible from this menu.
•Modification in 2-3-4-MAINTENANCE Menu, page 28: adding of a new automatic maintenance accessible
from this menu
•Modification in 2-3-5-SETUP Menu, page 30: adding of a new feature accessible from this menu.
Section 3
•Modification in 3-3-1-Installation of internal ancillary reagents, page 47 and in 3-3-2-Installation of
external ancillary reagents, page 48: adding of automatic priming of ancillary reagents.
•Modification in 3-4-System Performance Checks, pages 42 and following: programming the automatic
qualification profile now in section 7.
•Adding of 3-5-1-Programming calibrations, page 55: new display options in the calibrations/controls
window.
•Modification in 3-7-Programming Samples, pages 69 and following: use of the sample type «Other» in
immunoassays.
•Adding of 3-8-6-Performing reflex tests, page 76.
•Adding of 3-8-7-Performing dilutions, page 77: post dilutions carried out on user request.
•Modification in 3-10-1-Result of a calibration, page 84: the validation criteria «calibrator CV in RLU» is
modified.
•Modification in 3-10-3-Sample results, page 87: change in the reporting of results outside the measuring
range; adding of reflex test launch.
•Modification in 3-10-4-Work list results, pages 87 and 88: adding of reflex test launch.
•Modification in 3-11-Messages associated with results, pages 89 and following: adding of the messages
RFX and AIN; modification of messages OMR- and OMR+.
•Modification in 3-12-Results Storage, pages 93 and 94: storage of incomplete profiles and removal of
stored profiles
•Modification in 3-13-1-Cumulative analysis, pages 95 and following: modification in Levey-Jennings
access.
•Modification in 3-14-1-Emptying the solid waste, page 101: use of disposable container.
Section 4
•Modification in 4-3-Messages associated with results, pages 108 and following: adding of the messages
RFX and AIN; modification of messages OMR- and OMR+.
Section 5
•Adding of 5-4-13-2-Cleaning the probe, page 145.
Section 7
•Adding of Section 7: System Configuration, pages 151 and following.
Appendices
•Modification in A-6- Analyser long stoppage period, pages A-8 and following.
Preface
IDS-iSYS User Manual - Revision M1"P 2
Software version V 14

New features in the software version V 14.04
•The system allows reflex testing: tests can be defined with a Supervisor level of access and can be carried
out either automatically or on request.
•Ancillary solutions IDS-iSYS TRIGGER A & B, IDS-iSYS WASH S and IDS-iSYS DSORB are now
automatically primed when installed in the system.
•Incomplete patient profiles can now be stored together with the completed profiles. In this case, the
analysis requests that have not been performed are automatically removed.
•The results calculated outside the measuring range are now interpreted in relation to the limits of the
measuring range.
•The number of available tests is now indicated on Immunology cartridges common to several analytes (eg
CCS cartridge). The minimum and the maximum numbers of available tests are displayed next to the
icon.
•The pubertal development stage (Tanner stage) can now be defined in the patient identification tab. This
data is transferred using the connection protocol «ASTM Compatible V3».
•The qualification profile tests are no longer displayed in the calibrations/controls window.
•New display options are available in the calibrations/controls window.
•The criteria for automatic validation of Immunoassay calibrations are modified: a specific limit can now be
applied to the CV calculated on the RLU of each calibrator. This CV may also be excluded from the
calibration validation criteria.
•A new criteria is now applied to Immunoassays calibrated without master curve: it allows verifying the
difference between the target and the calculated value of each calibrator.
•The validity of an expired calibration can be extended one day.
•CV calculated with raw data of each control is now displayed on the first tab of calibration for
immunoassays measured by spectrophotometer.
•An identifier entered for a calibrator or control is recorded by the system. When this ID is used again, the
lot data are automatically associated.
•In the CONSUMABLES menu, the filling of the solid waste is now displayed proportional to a maximal
capacity of 400 cuvettes.
•Post-dilutions can be performed either automatically or on request.
•An automatic maintenance, requested by infectious disease assays, is automatically performed when the
system is placed in standby mode.
•In the reference tables of analytical configuration, it is no longer mandatory to define an interval that does
not generate a message.
•The connection protocol « ASTM Compatible V3» is added:
•transfer the identifier of the operator who validates the profile.
•transfer the Tanner stage.
•The system can now be connected to the centralised computer via an Ethernet link.
•Access to the Levey-Jennings menu is simplified.
•The function for updating the analytical configuration is now integrated to the user’s interface.
•Biochemistry assay setup is modified: modification in step configuration (compatible with previous versions
without modifying the analytical process).
Issues in software package V14.04
•Using Westgard Rules, when a calibration has failed, a normal WE! Error message is attached to the results. When the
calibration is then restarted and been successful, normally the WE! error message should disappear. That is not the
case in this software version, the flag is still wrongly displayed on control results but not on patient profile results
•Using Westgard Rules, in the case of a calibration with two levels of controls violating a rule (WE! Error message), only
the first control level is displayed on the Westgard graph and in the table of values.
•When a calibration curve is displayed, only the control level 1 is shown on the curve. Other control levels must be
selected each time to be displayed.
•An error message displays each time a Levey-Jennings graph is displayed for a control that does not have a target
value defined. This error message must be accepted prior to select the appropriate graphic representation (mean/
deviation).
•When restoring a high volume database, the option «Stored results» for a partial restoring cannot be used. Full
restoring or the other options for partial restoring can be used.
Preface
IDS-iSYS User Manual - Revision M1"P 3
Software version V 14

Issues in software package V 14.04 (continued)
•In the Storage menu displayed in standard mode, when the transfer to the LIS of a patient profile is requested while
one of multiple replicate results is selected, this selected result is not sent with the profile results. The relevant profile
must be selected to ensure sending of the complete profile.
•In certain cases, assays schedule to a first cartridge but not complete, may remain in error (red tick) despite a second
cartridge of the same lot being on-board. The first cartridge may also not be fully used. In this case, the run cycle must
be stopped and the relevant assays must be deselected (uncheck the box) prior to be selected again.
•Rounding at 0.1 mAbs may be different for the same result.
•When two analytes use the same cartridge, the number of available tests can be incorrect. In this case, all assays
requested for the two analytes will be carried out.
•The blue contour is applied only to removable racks in the sample compartment.
•A countdown before use may not resume when the cartridge is removed less than 5 minutes. In this case, a new
countdown begins.
•The status of a faulty calibration is not updated when excluding one of the calibrator replicates during a run. Patient
results and controls are calculated. The calibration status will be updated once the run cycle is stopped.
Preface
IDS-iSYS User Manual - Revision M1"P 4
Software version V 14

List of symbols used on the analyser
Manufacturer.
In vitro diagnostic medical device.
Consult the instructions for use.
Caution recommended: see Safety Precautions.
Risk of biological contamination.
Risk of crushing injury.
Electrical and electronic waste.
Dispose of in accordance with current country-
specific laws.
High temperature.
Risk of hand injury.
Preface
IDS-iSYS User Manual - Revision M1"P 5
Software version V 14


Section 1 Operating principle 1
1-1- Field of Use!2
1-2- General Overview!3
1-3- Characteristics!7
1-4- Installation!10
1-4-1- Environment!10
1-4-2- Electricity supply!11
1-4-3- Connections!11
1-5- Operating principles!12
1-5-1- Luminescence measurements!12
1-5-2- Absorbancy measurements!12
1-5-3- Immunoassays!13
1-5-4- Biochemistry assays!14
1-6- Precautions for Use!15
1-6-1- General precautions!15
1-6-2- Special precautions!16
1-6-3- Safety precautions!16
Section 2 User Interface Software 17
2-1- Structure of the Software!18
2-2- Main Screen!19
2-3- Menus!25
2-3-1- FILE Menu!25
2-3-2- SESSIONS menu!26
2-3-3- DATA menu!27
2-3-4- MAINTENANCE menu!28
2-3-4- MAINTENANCE menu (continued)!29
2-3-5- SETUP menu!30
2-3-6- MANAGEMENT OF LOTS menu!31
2-3-7- HELP menu!32
2-4- Functions keys!33
Table of Contents
IDS-iSYS User Manual - Revision M1 !Table of Contents i
Software version V 14

Section 3 Use 34
Information on the lid opening!35
3-1- Start up!37
3-1-1- Initial Start up!37
3-1-2- Start up from standby mode!37
3-2- Installation of Reagents!38
3-2-1- Colour codes associated with reagent positions in the reagent compartment!43
3-2-2- Information displayed with reagents!43
3-2-3- Management of on board reagents!44
3-3- Installation of Ancillary Reagents!46
3-3-1- Installation of internal ancillary reagents!47
3-3-2- Installation of external ancillary reagents!48
3-3-3- Installation of IDS-iSYS cuvettes!50
3-4- System Performance Checks (Immunoassay only)!52
3-4-1- Programming the qualification profile!52
3-4-2- Management of results!53
3-5- Programming Calibrations and Controls!55
3-5-1- Programming calibrations!55
3-5-2- Programming Quality Controls!58
3-6- Loading of Calibrators, Controls or Samples in the Sample Compartment!60
3-6-1- Removable trays!61
3-6-2- Fixed tray!64
3-6-3- Configure a position without barcode!65
3-6-4- End of loading!66
3-6-5- Colour codes associated with sample positions!67
3-6-6- Colour codes associated with removable trays!67
3-6-7- Colour codes associated with position contour!68
3-7- Programming Samples!69
3-8- Assays!70
3-8-1- Performing assays!70
3-8-2- Adding samples during an assay!71
3-8-3- Adding or replacing a reagent during an assay!72
3-8-4- Releasing an alarm during an assay!73
3-8-5- Unloading samples during an assay!75
3-8-6- Performing reflex tests!76
3-8-7- Performing dilutions!77
3-9- Adding an Emergency (STAT) Sample!79
3-10- Results!80
3-10-1- Result of a calibration!81
3-10-2- Result of controls!85
3-10-3- Sample results!86
3-10-4- Work list results!87
3-11- Messages Associated with Results!89
3-12- Results Storage!93
3-13- Quality Control Management!95
3-13-1- Cumulative analysis!95
3-13-2- Westgard rules!98
3-14- Switching The Analyser Off! 100
3-14-1- Emptying the solid waste!101
3-14-2- Emptying the liquid waste!102
3-15- Switching The Analyser OffCompletely!103
Table of Contents
IDS-iSYS User Manual - Revision M1 !Table of Contents ii
Software version V 14

Section 4 Messages 104
4-1- Messages Associated with Calibrations!105
4-2- Messages Associated with Calibration Controls!107
4-3- Messages Associated with Results!108
4-4- Warning Messages!112
4-5- Error Messages!113
Section 5 Maintenance 114
5-1- Daily Maintenance!115
5-1-1- General Maintenance!116
5-1-1-1- Checking reagent drawer and Plexiglas®#116
5-1-1-2- Checking sample drawer and Plexiglas®#116
5-1-1-3- Cleaning needle exterior!117
5-1-1-4- Decontamination of the probe!117
5-2- Weekly Maintenance!118
5-2-1- General Maintenance!119
5-2-1-1- Cleaning the reagent compartment and Plexiglas®#119
5-2-1-2- Cleaning the sample tray and Plexiglas®#119
5-2-1-3- Cleaning the rinsing well!120
5-2-1-4- Checking dilutors and IDS-iSYS System Liquid pumps!120
5-2-2- Immunoassay Maintenance!121
5-2-2-1- Rinsing of the IDS-iSYS Triggers tubing!121
5-3- Monthly Maintenance!122
5-3-1- General Maintenance!122
5-3-1-1- Cleaning the IDS-iSYS System Liquid pump shafts!122
5-3-1-2- Cleaning the liquid waste pump shaft!122
5-3-1-3- Cleaning the IDS-iSYS D-Sorb pump and level sensor shafts!123
5-3-1-4- Cleaning the liquid waste level sensor!123
5-3-1-5- Checking lamp references!124
5-3-1-6- Switch off the instrument!124
5-3-2- Immunoassay Maintenance!124
5-3-2-1- Cleaning the IDS-iSYS Wash Solution pump and level sensor shafts!124
5-3-2-2- Rinsing of the AP Substrate tubing!125
5-4- Analyser Interventions!126
5-4-1- Replacement of lamp!126
5-4-2- Replacement of probe!127
5-4-3- Replacement of primary fuses!128
5-4-4- Replacement of secondary fuses!129
5-4-5- Replacement of IDS-iSYS Wash Solution pump (Immunoassay)!130
5-4-6- Replacement of IDS-iSYS System Liquid pump!130
5-4-7- Replacement of liquid waste pump!130
5-4-8- Replacement of IDS-iSYS D-Sorb pump!131
5-4-9- Removal of the on-board IDS-iSYS Cuvettes cube!131
5-4-10- Adjustment of probe reference position!132
5-4-10-1- Adjustment procedure for analyser without lid locking system!133
5-4-10-2- Adjustment procedure for analyser with lid locking system!136
5-4-11- Unclogging the sampling probe!141
5-4-12- Intervention in System Configuration menu!143
Table of Contents
IDS-iSYS User Manual - Revision M1 !Table of Contents iii
Software version V 14

Section 5 Maintenance (continued)
5-4-13- Repeat an automatic maintenance!145
5-4-13-1- Washer needle cleaning!145
5-4-13-2- Cleaning the probe!145
5-5- Analyser Cleaning!146
5-5-1- Decontaminating the containers of solid and liquid waste!147
Section 6 Problems & Corrective Action 148
6-1 Resolving Errors in Cartridge Check System (CCS)!149
Section 7 System Configuration 151
7-1- Programming the automatic qualification profile!152
7-2- Creating the reflex tests!153
7-2-1- Creating a new rule!154
7-2-2- Launch option of reflex tests!156
7-2-3- Modifying an existing rule!157
7-2-4- Importing/exporting rules!157
7-2-5- Removing a rule!157
7-3- Configuring the print settings!158
7-4- Options of validating and transferring the results!159
7-5- User management!160
7-6- Updating the analytical configuration!162
Table of Contents
IDS-iSYS User Manual - Revision M1 !Table of Contents iv
Software version V 14

Appendices A 1
A-1 Waste Disposal!A 2
A-2 Decontaminating the Analyser!A 3
Cleaning/Decontaminating Declaration!A 4
A-3 Disposal of the Analyser!A 5
A-4 IDS-iSYS Cuvettes!A 6
A4-1 List of symbols used on the IDS-iSYS Cuvettes cube!A 6
A4-2 Storage of the IDS-iSYS Cuvette cubes"A 6
A-5 Sample barcode symbology managed by the system!A 7
A-6- Analyser long stoppage period!A 8
A6-1-Less than14 days"A 8
A6-2-Between 15 and 30 days"A 10
A6-3-Analyser start up after a long stoppage period"A 13
Table of Contents
IDS-iSYS User Manual - Revision M1 !Table of Contents v
Software version V 14

SECTION1:
Operatingprinciple

Operating principle 1
1-1- Field of Use!2
1-2- General Overview!3
1-3- Characteristics!7
1-4- Installation!10
1-4-1- Environment!10
1-4-2- Electricity supply!11
1-4-3- Connections!11
1-5- Operating principles!12
1-5-1- Luminescence measurements!12
1-5-2- Absorbancy measurements!12
1-5-3- Immunoassays!13
1-5-4- Biochemistry assays!14
1-6- Precautions for Use!15
1-6-1- General precautions!15
1-6-2- Special precautions!16
1-6-3- Safety precautions!16
Section 1
IDS-iSYS User Manual - Revision M1! ! ! ! ! 1
Software version V 14

1-1- Field of Use
is an in vitro diagnostic analyser. It enables Immunoassay and Biochemistry assays to be
carried out on a single analytical platform:
•Immunoassay.
•Bone and Growth.
•Infectious Diseases.
•Hypertension.
•Autoimmunity.
•Biochemistry.
•Substrates.
•Enzymes.
•Electrolytes.
•Specific Proteins.
The analyser is intended for professional use and must only be used by trained personnel
working in compliance with the precautions for use set out in this manual (see Section 1-6-1, page 15) and
good laboratory practice (GLP).
Operating principle
1-1- Field of Use
IDS-iSYS User Manual - Revision M1"Operating principle 1-1 2
Software version V 14

1-2- General Overview
is an in vitro diagnostic analyser. It enables assays using different systems of detection to be
carried out on a single analytical platform: luminescence and spectrophotometry, applied to the different fields
of clinical biology, Immunoassays and Biochemistry.
The analyser can support Immunoassays based on the principle of chemiluminescence as well as assays using
an enzymatic detection.
The analyser enables the complete automation for Immunoassays and Biochemistry tests.
Individual assays are carried out in disposable cuvettes which are automatically loaded onto a carousel.
Asynchronous management allows each cuvette to be processed individually and transferred to the relevant
reaction modules positioned around the carousel.
The cuvette is able to ensure compatibility in both fields. The measurements specific to each principle are
carried out directly in the reaction cuvette:
•Luminescence measurements carried out in the luminometer.
•Absorbency measurements carried out continuously by the spectrophotometer.
Designed for continuous loading, the analyser works on a sample-by-sample basis.
Operating principle
1-2- General Overview
IDS-iSYS User Manual - Revision M1"Operating principle 1-2 3
Software version V 14

With all modules fitted the analyser comprises the following key components:
7
2
4
1 5
8
3
6
1. A refrigerated reagent compartment, consisting of 15 rails containing the Immunoassay or Biochemistry
racks as well as a specialised rack for chronometric reagents.
2. A compartment for samples , calibrators and controls with, depending on the analyser configuration, a
fixed sample tray with 64 positions , or a rack containing 6 removable trays each with 20 positions.
3. A pipetting arm that pipettes both reagents and samples.
4. A thermo-regulated carousel set at 37°C with 90 positions for disposable cuvettes. Incorporated into the
carousel is the spectrophotometer, used for measuring absorbancy from certain Immunoassay,
Biochemistry and Turbidimetry reactions.
5. A sedimentation module for magnetic particles.
6. An automatic cuvette loader, holding 960 cuvettes at a time (pre-formed as a cube).
7. A luminometer measuring luminescence in Immunoassay reactions (Immunoassay version).
8. Four washers for washing magnetic particles.
The barcode reader integrated into the reagent compartment identifies reagent cartridges supplied by IDS.
A second barcode reader, located on the front of the analyser identifies ancillary reagents.
Samples, calibrators and controls are identified or by the reader located on the front for analysers with a fixed
sample tray, either by a reader integrated into the sample compartment for analysers with removable sample
trays.
Operating principle
1-2- General Overview (continued)
IDS-iSYS User Manual - Revision M1"Operating principle 1-2 4
Software version V 14

For the fixed-tray configuration, each position is equipped with a detection sensor. Samples are identified by
the barcode reader (on the front) prior to loading the sample, or manually programmed after sample loading.
For the removable-tray configuration, the tray and each of its positions are identified when loading by the
integrated barcode reader.
IDS
Site de POUILLY
Specifications:
Rack Sample Module
Revision: 0
Page : 8/
20
Embedded software has to be extended with direct LED commande order, a GUI detection method (based on question
intervalle time out) and a possibility to grab all known tests and set LEDs.
4.2 Mechanical views of mock up:
Configuration with removable sample trays
The reagent cartridges are stored in the refrigerated reagent compartment between 12 and 15°C while the
analyser is operating and between either 8 and 10°C or 12 and 15°C in standby mode (depending on the
analyser configuration).
Depending on the configuration of the analyser, up to 15 Immunoassay parameters can be programmed
simultaneously or up to 36 Biochemistry parameters or a combination of both, up to the maximum number of
positions available.
For Immunoassays, the analyser only uses reagents supplied by IDS and its partners. For Biochemistry, it is
possible to use reagents available from suppliers other than IDS, however these reagents will not be managed
using the barcode system, and the information required for traceability must be manually programmed by the
operator.
When Biochemistry regents available from third-party suppliers are used, IDS takes no responsibility for the
validity of results obtained. It is the responsibility of the operator to ensure any non-IDS reagents are optimised
and validated for use on the analyser.
Operating principle
1-2- General Overview (continued)
IDS-iSYS User Manual - Revision M1"Operating principle 1-2 5
Software version V 14

Samples and reagents are aspirated using a probe equipped to detect liquid level using capacitance. The
sample and reagents are pipetted in accordance with the validated parameters for each assay and transferred
into a cuvette where the reaction takes place. Between each sample, the probe is rinsed internally and
externally in order to eliminate any risk of contamination.
Immunoassays:
The cuvettes are transferred to the washer module where magnets hold the solid phase (magnetic particles)
whilst washing takes place.
When the reaction is over, depending on the type of the assay, the cuvette is either transferred to the
luminometer for chemiluminescence assays or kept in the carousel for enzymatic assays.
For chemiluminescence assays, trigger reagents are added to produce luminescence.
For enzymatic assays, substrate is added and the photometric changes are measured continuously by the
spectrophotometer.
When measurement is complete, cuvettes are automatically ejected into a re-usable solid waste tray.
Biochemistry assays:
Photometric changes taking place in the reaction are measured continuously by the spectrophotometer.
The analyser is connected to a computer via an Ethernet link.
The software is able to program the analyser workload and carry out the following functions:
•Management of lots of reagents, calibrators and controls.
•Quality controls (Levey-Jennings and Westgard).
•Management of ancillary reagents.
•Operator traceability.
•Transfer of results to a centralised computer system.
•Printing, storing and traceability of results.
Operating principle
1-2- General Overview (continued)
IDS-iSYS User Manual - Revision M1"Operating principle 1-2 6
Software version V 14

1-3- Characteristics
Analyser
System
•Multidiscipline Immunoassay and Biochemistry analyser.
•Continuous loading.
•Configuration:
•Immunoassay (4 washers) + Biochemistry.
Analyser Physical
Characteristics
•Analyser dimensions:!L 105 cm x H 70 cm x W 75 cm.
•Overall dimensions: !L 124 cm x H 110 cm x W 71 cm.
•Weight: ! ! ! 103 kg.
•Basis weight: 130 kg/m2
•Sound level:! ! 55 dB (average in run cycle); up to 66 dB
•Emitted heat: up to 530 W (1808 BTU/h or 16,5 cal/s)
•Computer System: !L 60 cm x H 40 cm x W 50 cm.
Throughput
•Immunoassay! ! Up to 120 tests/hour (Assay dependent)*.
•Biochemistry! ! 240 tests/hour.
* According to configuration.
Number of tests on board
•15 Immunoassay.
•Up to 50 Biochemistry.
Immunoassay detection
•Chemiluminescence (luminometer).
•Enzymatic assays (spectrophotometer)
Biochemistry detection"
•Colorimetric and enzymatic assays.
•Turbidimetric assays.
Samples
•Serum, plasma or urine.
•Primary tubes 5 mL (13 x 75 mm), 7 mL (13 x 100 mm).
10 mL (16 x 75 mm).
•Secondary tubes (13 x 75 mm).
•Cups (ref. IS-CSC105 and ref. IS-CSC115).
•Depending on the configuration, positions for samples, calibrators and
controls:
•Fixed tray : 64 positions
•Removable trays : 20 positions per tray. Up to 6 trays on board.
Possibility of special tray for calibrator and control vials (16 positions
for 2ml vials, diameter 18 mm).
•Liquid level detection by capacitance.
•Clot detection.
•Dilutions and automatic pre-treatments.
Barcode readers
•Reader on front face for barcode identification of cuvettes and ancillary
reagents. Used for sample identification in fixed tray configuration.
•Integrated reader for reagents barcode
•For configuration with removable trays, reader integrated into the
sample compartment.
Operating principle
1-3- Characteristics
IDS-iSYS User Manual - Revision M1"Operating principle 1-3 7
Software version V 14
Table of contents
Popular Measuring Instrument manuals by other brands

HOP
HOP Smart Spot-Noise installation guide

Milwaukee
Milwaukee 2205-40 Original instructions

International Light Technologies
International Light Technologies CureRight ILT800 Quick Start Software Manual

Performance Tool
Performance Tool W89719 owner's manual

PASCO
PASCO OS-8516A instruction sheet

E-Mon
E-Mon Din-Mon D5 installation instructions