Ingos COR 400 User manual

INGOS Ltd. COR400
CIRCULATION
Producer:
INGOS Ltd., Division of Laboratory Instruments
Supplier a service:
INGOS Ltd.
K Nouzovu 2090; 143 00 Prague 4
; Czech Republic
Tel.: + 420 225 983 400
+ 420 225 983 409
Fax: + 420 244 403 051
e-mail:instruments@ingos.cz
www:ingos.cz
17.7.2019
CIRCULATION
CHILLER
COR 400
User manual
INGOS Ltd., Division of Laboratory Instruments
INGOS Ltd.
; Czech Republic
Strana 1
COR 400

INGOS Ltd. COR400 17.7.2019
Strana 2
1. INTRODUCTION
The Chiller is designed to cool the cooler rotary vacuum evaporator (it can also be used for
two rotary evaporators together). The chiller can be used for other device which need cooling. The
control module controls temperature in the cooling tanks. The chiller is designed as a single
functional instrument.
4
3
5
7
6
2
8
1
1. Switch 4. Output cooling medium 7. Grip
2. Thermostat 5. Input cooling medium 8. Socket or switch
3. Circulation pump 6. Overflow circulation pump
Fig.1 Chiller
Warning :
Using devices other than as described in the instructions can cause security breaches!
2. DESCRIPTION AND SPECIFICATION
2.1 Description
The instrument consists of a box in which is thermostatic tank, as a source of cold a modern
compressor cooling unit Lindr is used that uses the input electric energy for direct transfer to
chilling. Cooling loop in a thermostated tank is made of stainless steel. Cooled thermostated tank is
equipped with a circulation pump. Temperature control and cooler run is through digital thermostat.
Used materials meet the highest requirements for hygiene standards and guarantee a long
working life of this cooler.
Two grips located on the sides of the cooler facilitate handling.

INGOS Ltd. COR400 17.7.2019
Strana 3
2.2 Specification
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1/4 HP
Performance 22°C =>7°C . . . . . . . . . . . . . . . . . . . . . . . . 80 l/hour
Minimum temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 1°C for water
Minimum temperature . . . . . . . . . . . . . . . . . . . . . . . . . . -20°C for antifreezing mix
Regulation accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . .±1°C
Volume of cooling tank . . . . . . . . . . . . . . . . . . . . . . . . 20 l
Displacement pump . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 m
Pump flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 l/hour
Cooling gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R 290
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230V, 50Hz
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max. 391 VA
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 kg
Dimensions (w x d x h) . . . . . . . . . . . . . . . . . . . . . . . . . 530 x 330 x 474 mm
The instrument is designed for an ambient temperature 15 - 35 °C; with humidity up to 80%
without acid and caustic vapors.
3. COMMISSIONING
3.1 Unpac ing
Carefully unpack the instrument from the shipping container. Check the exterior of the
instrument, suffer during the transport from visible damage. Should any damage occur during the
transport, contact the shipping company.
According to the delivery note check all items. In case of discrepancy carefully inspect all
packing material. If any item is missing, contact the manufacturer or your supplier.
Place the instrument on the table and familiarize yourself with the distribution and function of
the controls and connections. Before operating read the instructions of this manual thoroughly.
Place the chiller so that the heat generated by the chiller may be sufficiently ventilated,
must not be placed near radiant heat or direct sunlight. The chiller should be placed on a
horizontal surface. Do not place objects the chiller that would prevent air circulation.
The chiller can in no case be put on its side even during transport!
3.1.1 What you will need
To put into operation, you will need the following items:
1. Tubing with insulation and couplings for connection to a rotary vacuum evaporator cooler.
2. 20 - 25 liters of antifreeze with a freezing point of about 5°C lower than the lowest temperature
set on the thermostat. When setting the lowest limit of temperature of 1°C and above, you can be
used water.
Fill the plastic tank with liquid to the overflow.

INGOS Ltd. COR400 17.7.2019
Strana 4
3.1.2 Connecting elements, instrument assembly
The power switch is located on the panel (green with glow lamp). Furthermore, on the panel
is switch displacement pump (red with glow lamp) and a temperature controller.
Cooled liquid output is in labeled outlet. Returning liquid input is in labeled input. Using the
supplied hoses and couplings assemble the cooling circuit and check the connection hose. Hoses
must be straight cut-away in the event that it can not be inserted into the coupling, it is necessary
to moisten the end.
3.2 Temperature controller
Display along with a button keyboard provide user comfort for direct control.
The function of the individual keys is as follows:
Button to turn on/ turn of temperature control.
Set Button to enter the setup mode.
Setting button.
Setting button.
Setting parameters
1. Enter the setup mode.
Press the SET button for more than 5 seconds, the LED turns on lights up and displays the set
temperature.
2. Setting the temperature.
Use the settings button to set the desired temperature.
3. Leave the settings mode.
Press the SET button more than 5 seconds, or in 30 seconds the thermostat (when there is no
button operation) will return to normal operating mode.

INGOS Ltd. COR400 17.7.2019
Strana 5
Entry to the mode menu
1. Entry to management menu
Press the SET button and setting button for more than 5 seconds, the LED lights and the
display shows the code F0.
2. Setting parameters
Setting buttons to select the code and press SET button.
3. Change settings
Setting buttons to set the parameters of the selected code.
4. To return to the parameter setting
Press the button SET.
5. Leave the setting mode
Press the SET button more than 5 seconds, or in 30 seconds the thermostat (when there is
no button operation) will return to normal operating mode.
Setting item Range setting parameters Factory setting code
Temperature difference 1°C ~ 16°C 1°C F0
Compressor delay 0 ~ 9 minute 0 minute F1
Lower limit temperature up to -40°C -20°C F2
Upper limit temperature up to +70°C 20°C F3
Selectable mode 1:cooling; 2:heating; 3:alarm 1 F4
Temperature calibration -5°C ~ +5°C 0 F5
3.3 Switching on the instrument
The procedure to turn on the instrument is as follows:
1. Plug the instrument into the mains with corresponding voltage.
2. Check hose connections.
3. Turn on the power switch.
4. Set the temperature regulator.
5. Turn on the circulating pump.

INGOS Ltd. COR400 17.7.2019
Strana 6
4. ROUTINE MAINTENANCE AND SERVICE
At the latest after six months of operation, replace the contents of the bath. Periodically check
the cleanliness and filling of bath. During sanitizing the water temperature must not be higher than
40°C. Cleaning of all external surfaces of the instrument should be done with a soft cloth. You can
use common cleansers. Do not leave the instrument on a surface contaminated from corrosive
substances. This could cause damage of paint on the instrument. The surface of the instrument clean
with a soft cloth.
Cleaning the condenser perform at least 1x per month by sweeping, or air pressure, or CO
2
.
Fan cover by pulling upward pull of metal trails. Unscrew the fastening screws from the cooler on
the condenser side. Loose cover first pull up upwards 3 cm, than pull to the side. Loosen the other
side of the cover so that you can pull the whole cover of metal trails. Now you can clean the
condenser.
Warning: When cleaning, unplug the instrument!
Ejecting the fan Putting the fan in place Unscrewing the cover
Removing the cover Condenser ready for cleaning
5. ACCESSORIES
2 pcs Set of hoses and couplings for connection RVO 400

INGOS Ltd. COR400 17.7.2019
Strana 7
6. WARRANTY, OPERATING CONDITIONS
For the instrument the manufacturer provides a warranty of two years from the delivery to the
customer. The instrument may be used only in the manner described in this manual. The producer is
not liable for damages resulting for failure to comply with the conditions set in this manual.
The instrument is designed for an ambient temperature 15 - 35 °C; with humidity up to 80%
without acid and caustic vapors. The inlet cord plugs into a socket circuit featuring protection 16 A.
All warranty and post warranty repairs are to be performed by the manufacturer, or its
authorized body.
Warning!
For using water as coolant, set in mode F2 lower limit of the temperature at 1°C and above.
7. WASTE DISPOSAL
At the end of life discard the instrument according to valid regulations on waste or return to
the vendor or producer for liquidation.
The instrument contains a cooling gas R 290.
The instrument is destroyed as electronic waste.
8. CONTENT
1.
INTRODUCTION ......................................................................................................................2
2.
DESCRIPTION AND SPECIFICATION ..................................................................................2
2.1
Description..............................................................................................................................2
2.2
Specification............................................................................................................................3
3.
COMMISSIONING....................................................................................................................3
3.1
Unpacking ...............................................................................................................................3
3.1.1
What you will need..........................................................................................................3
3.1.2
Connecting elements, instrument assembly.....................................................................4
3.2
Temperature controller............................................................................................................4
3.3
Switching on the instrument....................................................................................................5
4.
ROUTINE MAINTENANCE AND SERVICE .........................................................................6
5.
ACCESSORIES..........................................................................................................................6
6.
WARRANTY, OPERATING CONDITIONS ...........................................................................7
7.
WASTE DISPOSAL...................................................................................................................7
8.
CONTENT..................................................................................................................................7
Table of contents
Popular Chiller manuals by other brands

Daikin
Daikin EWAD TB Installation, operation and maintenance manual

Lindab
Lindab PLAFOND Series installation instructions

Unical
Unical HP_OWER ONE 70R User's and installer's manual

teca
teca RLC-1402HC Product information packet

MTA
MTA Aries AS 201 FC Operating and instruction manual

Daikin
Daikin EWYD-4ZB Installation, operation and maintenance manual