SUG7000 issue 1 03.13
Expiratory limb heated
wire connector
Airway temperature
probe
Chamber
temperature
probe
Mains lead
Heated
wire
adaptor
Dual temperature
probe connector
Inspiratory limb heated
wire connector
Mount Humidifier Heater Base. Mount the heater base using an appropriate bracket;
ensure the top of the humidifier chamber is placed lower than patient airway entry position.
Mount humidifier chamber onto humidifier base. Ensure chamber clicks into place.
Connect patient circuit to chamber. See example below.
Add water to humidifier chamber. Follow chamber manufacturers guidelines.
Heated Wire. Ensuring that the white arrows on both connectors are aligned, connect the
appropriate heated wire electrical adaptor to the Black/Yellow Push-Fit connector on heater
base. Connect the cloverleaf connector to the inspiratory heated wire limb; if using a Dual
Heated Circuit, connect oval connector to the expiratory heated limb connector.
Dual Temperature Probe. Ensuring that the white lines on both connectors are aligned,
connect the dual temperature probe to the Green/Silver Push-Fit connector on heater base.
Then insert the temperature sensors in to their appropriate ports on the inspiratory limb, as
shown in diagram below.
Connect power cord to AC wall socket
Turn on gas supply equipment (Ventilator). Connect the ventilator to the gas supply
and turn on and carry out pressure (tightness) testing applicable to that ventilators protocol.
Expiratory limb
Water Bag Spike
Exp
Insp
Inspiratory limb
Important: This guide is not intended to replace the Instructions for Use. It is the users
responsibility to ensure the device is operated correctly and in accordance with the Instruc-
tions for Use.