
8
If there is a different O
2
concentration value, press the KAL-signed key and accomplish an
automatic calibration. After the calibration the display should show 20.8 - 21.0%, usually
20.9% .
2.Check the battery status following way. Disconnect the DC power adaptor and the
display lighting will disappear. If the oximeter continues to operate , everything is all right. In
case that the battery is totally discharged, the oximeter will stop to operate and the display
will not show anything. If the battery is very low but not totally discharged, display will show
another inscription:"Vymen bateriu" . If the battery is borderline charged, there will be an
exclamation right after the "Kyslik" inscription. The exclamation means, that you will have to
change the battery in a close future.
Plug in DC adaptor into the power source
3.Place the oxygene sensor into the environment where do you want to measure O
2
concentration.
4.Set the upper and lower alarm limits following way:
Press at the same time keys: ENTER and arrow downwards /-1/ for about 2 seconds. The
cursor will start to twinkle on the first number of lower alarm limit. Pressing keys + 1, or - 1,
will the first number of the alarm limit increase or decrease. Reaching the wanted value of the
first number press ENTER .Then cursor will move to the second number of the lower alarm
limit and you can repeat the procedure. Repeated pressing of ENTER will stop your editation
of lower alarm limit. If you do not want to change the alarm limit, press ENTER twice.
Upper alarm limit can be changed by concomittant pressing of ENTER and arrow upwards
/+1/keys. The setting procedure is the same as in lower alarm limit setting.
!!! It is impossible to set lower alarm limit below 18% . In case you will set the lower
alarm limit below 18%, after finishing the editation, it will be automatically setted on
18%.
The difference between lower and upper alarm limits is controled by the processor and it is
impossible to set the difference lower than 6% of O2. If your upper limit will be 40% and you
will try to set the lower one on 39%, upper limit will be automatically changed for 45%, that
means +6%.
5.Setting the oximeter in accordance with the proceeding described above , it is prepared to
monitor oxygene concentration.
7.4
Switch off the oximeter
1.Switch off the DC adaptor by pulling it out from the power
2.Switch the oximeter off / VYP position /
3.Pull out the oxygene sensor from the place of oxygene concentration measurement and
place it on free air.
4.If necessary, make a surface disinfection and sterilisation of theoxygene sensor
connection /seebelow/