Operating manual
Platform balance
KERN OEX
Version 1.1
10/2014
GB
Doc.No.OEX-BA-e-1411

GB
KERN OEX
Version 1.1 10/2014
Operating manual
Platform balance
Contents
1Pay attention to the instructions in the Operation Manual............................. 3
2Intended use for OEX balances......................................................................... 4
3Design of OEX balances with one weighing cell............................................. 4
5ATEX marking – meaning of icon ..................................................................... 7
6Installation instruction for balance................................................................... 9
7Changing the installation site for the balance............................................... 13
8Cleaning and maintenance.............................................................................. 14
9Technical examination..................................................................................... 15
2 Doc.No.OEX-BA-e-1411

1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with
KERN balances.
All language versions contain a non-binding
translation.
The original German is binding.
Conformity with these operating instructions is a precautionary measure for the safe
operation of IEX balances in ex-protected spaces.
Employees should familiarize themselves with the device’s components described in
the operating instructions:
•Display KEX-TM
•Power unit IEX-A01-U, IEX-A01-E
These operating instructions should be freely available for consultation to the
employees.
The operating instructions contain the icons shown below:
•Continuously prescribed actions
•Actions prescribed under certain circumstances
•Marking essential paragraphs dealing with explosion protection
These operating instructions may be supplemented by additional warnings and safety
information provided by KERN.
All employees must be informed about new supplements added to these operating
instructions.
Doc.No.OEX-BA-e-1411 3

2 Intended use for OEX balances
Load in ex-protected spaces. It is possible to tare the balance over the entire working
range, thus allowing determination of the net weight.
The design of the balance (construction of non-electrical part of the platform balance
and material used) is in conformity with Directive 94/9/EE, applicable in countries of
the European Union.
•Due to their design, the balances of the IEX series can be used with
a non-
electrical part on the platform and the material used in zones 1
and 2 where
ogasses, mixtures of gasses, vapour types IIA, IIB or IIC
oor explosive groups can develop,
o
as well as for temperature classes T1, T2, T3 or T4.
3 Design of OEX balances with one weighing cell
4 The construction of balances with one weighing cell of the IEX series is based on
the modular design principle and comprise:
•Display KEX-TM
•Power unit IEX-A01 by KERN, positioned outside the hazardous zone
•Weighing cell executed as ex-model
•Weighing platform, non-electric, mechanical balance component.
The display unit of the KEX-TM series is connected by a cable to the integrated
dynamometer (a weighing cell) of the platform. The platform balance comprises a
weighing pan made of stainless steel providing sufficient protection and allowing
easy cleaning. The balances of the OEX series have a stainless steel weighing cell
providing advanced sturdiness against damaging influences in the ambient
conditions.
The cable used for connecting the weighing cell to the display unit is protected by a
rod or approx. 3 m long tube.
The display unit is supplied with energy by the KERN mains adapter IEX-A01
positioned outside the hazardous area. The mains adapter is plugged into the mains
power supply of 230V.
The housing of the display unit has a marked terminal for the PE conductor
(electrostatic discharge). The housing is connected electrically to the platform
balance by a steel tube.
4 Doc.No.OEX-BA-e-1411

The balances of the OEX series are equipped with a KEX-TM display unit.
Safe area
Ex-protected area
1 load cell balance
Doc.No.OEX-BA-e-1411 5

Models with tripod:
Balance model
A1
A2
B
C1
C2
D
E
OEX 6K-3HM
306
246
103±3
268
208
~535
~485
OEX 10K-3HM
306
246
103±3
268
208
~535
~485
OEX 30K-2HM
410
410
98±2
370
370
~617
~613
OEX 60K-2HM
500
500
150±5
400
400
~790
~710
OEX 100K-1HM
500
500
150±5
400
400
~790
~710
Balances of the OEX series - dimensions
Models without tripod:
Balance model
A1
A2
B
C1
C2
OEX 6K-3M
306
246
103±3
268
208
OEX 10K-3M
306
246
103±3
268
208
OEX 30K-2M
410
410
98±2
370
370
OEX 60K-2M
500
500
150±5
400
400
OEX 100K-1M
500
500
150±5
400
400
Balances of the OEX series - dimensions
6 Doc.No.OEX-BA-e-1411

5 ATEX marking – meaning of icon
•The balances of the OEX series meet the requirements of group II,
class 2 for devices in as per Directive 94/9/EU and are marked as -
II 2 G IIC T4
.
Icon declaration as per Directive 94/9/EG:
The balances of the OEX series are intended for indoor and outdoor use. However,
they should not be exposed to aggressive chemical vapours, critical temperatures
and high humidity. The platform balance construction is made of stainless steel.
Category for group II
devices:
1
– The device
guarantees a very
high safety level,
- for operation in
areas 0,1,2
2
– The device
guarantees a very
high safety level,
- for operation in
areas 1,2
3
– The device
guarantees a normal
safety level,
- for operation in
Area 2
Explosive atmosphere:
G
– Caused by gas, steam
and mist.
D
– Caused by dust
Icon for explosion
protection used
Construction
type:
o
– Oil protection
p
– Gas
protection
with
excessive
pressure
q
– Sand
protection
d
– Flameproof
protection
e
– Reinforced
construction
ia
– Intrinsically
safe
construction
for operation
in areas
0,1,2,
ib
– Intrinsically
safe
construction
for operation
in areas 1,2,
Electrical
equipment
–
compatible with
one or few types
of explosion
protection
constructions
Gas explosivity group -
Examples:
-
IIA: Propane (T1)
Benzene (T3)
Butane (T2)
Ethyl alcohol (T2)
-
IIB: Ethylene (T2)
-
IIC: Acetylene (T2)
Hydrogen (T1)
,
Temperature class – max
surface temperature of device
parts exposed to explosive
mixtures:
T1: 4500C
T2: 3000C
T3: 2000C
T4: 1350C
T5: 1000C
T6: 850C
Temperature class
of gas and vapour
spontaneous ignition:
T1: > 4500C
T2: > 3000C
T3: > 2000C
T4: > 1350C
T5: > 1000C
T6: > 850C
Safety level
Gas atmosphere:
-
Ga
-
Gb
-
Gc
Dust atmosphere:
-
Da
-
Db
-
Dc
Doc.No.OEX-BA-e-1411 7

•Ambient temperature during operation -10°C to +40°C.
•The balance must be protected against mechanical shocks and
vibrations.
•The OEX balance must be installed and operated by expert staff.
•The balance should be serviced from time to time.
•Prevent electrostatic charge. Ensure that the earth conductor
(equipotential) is connected to the KEX-TM display unit at all times.
•Do not disconnect PE conductor.
The use of equipotential bonding is
only admissible where the development of
explosive gasses can be
completely ruled out.
•
In the case of damage to the balance, unplug the plug, disconnect
the supply cable and take the device to a safe area for interim
storage so as to prevent any possible use of the balance before
repair.
•The display unit KEX-TM is designed intrinsically safe and is in
conformity with the following standards: PN-EN 60079-0 and PN-EN
60079-11 certified by KDB 14ATEX0110 certificate.
•The display unit is powered by mains adapter IEX-A01 made by
KERN and positioned outside the hazardous and is in conformity
with the standards: PN-EN 60079-0 and PN-EN 60079-11, certificate
KDB 14ATEX0109.
•The platform balance is used with weighing cells PC6, PC22, PC42,
PC60 and PCB supplied by Flintec which are certified for safe use in
hazardous areas - certificate BSV 09 ATEX E 086X.
•All non-electrical parts of the platform balance have been
manufactured in accordance with standard PN-EN 13463-1.
•All repair and maintenance work during which the housing of the
display unit needs to be opened should be carried out by KERN
customer services. Replace subassemblies with genuine parts. The
balance must not be opened by unauthorised persons as this will
result in loss of all rights to compensation under warranty.
Any installation or repair work deviating from those described in this manual will
inevitably result in loss of warranty.
8 Doc.No.OEX-BA-e-1411

6 Installation instruction for balance
•Prepare the installation site for the OEX balance in a professional manner.
•Ensure that the installation site is dry, level and firm.
•The permissible operating temperate range is from -10°C to +40°C
•The floor base for the installation of the platform balance must be firm enough
for supporting the maximum weight allowed for the load plus the platform
balance.
•Balances of the OEX series without tripod have a display unit that can be
connected to the wall via a cable with the help of a holder.
•The display unit is designed for attachment to a table or for installation on a
wall with the help of a holder.
oUnpack the balance outside the hazardous area.
oRemove all transport locking devices after unpacking:
oAttach tripod to platform.
oFeed excess cable into the tripod.
oLift the weighing pan and remove the transport locking device.
Weighing cell conductor
Transport Securing
Tripod
Washer
Nut
Doc.No.OEX-BA-e-1411 9

CAUTION!
•To prevent electrostatic charge, connect the balance to earth.
•The contact slots for the wires are marked with the icon „ “.
•Prior to transport, prepare a PE conductor for electrostatic connection to earth of
the display unit.
•The PE conductor should end in a loop that is screwed to the display unit with the
help of nuts (M4).
•To ensure that the loop is fixed permanently and in contact with the display unit,
use a lock washer.
•The PE conductor should be of yellow and green colour and provide a sectional
surface of 4mm².
The diagram below shows the balance’s connection to earth:
Place the platform balance and the display unit on the intended installation site in the
ex-area and connect it to earth.
Transport Securing
Weighing pan
Safe area
Ex-protected area
10 Doc.No.OEX-BA-e-1411

CAUTION!
•The assembly of the display unit, installation of the platform
balance and the connection to earth must be carried out within
a time frame that rules out the risk of developing an explosive
atmosphere.
•Place the platform balance in a level, stable place as far away as possible from
heat sources and drafts.
•Level the balance with the help of the adjustable feet and a spirit level.
•Check, whether the adjustable feet stand on the balance base.
•Ensure that the weight is evenly distributed among the feet.
•Check once again that the installation site for the balance is level. Exact flatness
is established when the air bubble of the spirit level is in the centre of the spirit
level ring.
•If further levelling of the platform balance is required, turn the adjustable feet of
the balance in order to adjust the height.
•Check the earth contact springs between the platform balance and the weighing
pan. This ensures that the electrostatic charge is discharged from the weighing
pan.
Doc.No.OEX-BA-e-1411 11

•Put the weighing pan on the platform balance.
•Position the mains adapter outside the hazardous area in which an explosive
atmosphere may be expected to be present (that is in a safe place).
•Connect the connecting cable (max. 20 m) to the mains terminal marked DC IN
on the display unit.
1. Mains connection
2. Connection cable
3. Marking PE conductor
4. Shielded load cell cable
5. Sticker power connection
•Prior to initial start-up, make sure that there is no risk of explosion –
that no explosive atmosphere is present. If it is obvious that the
balance is not working correctly, disconnect the balance from the
power supply, secure it and contact KERN customer services.
12 Doc.No.OEX-BA-e-1411

•Connect the adapter to a power supply of 230V AC.
•For prolonged periods of inactivity disconnect the balance from the power supply.
•Press the ON/OFF button and hold for approx. 1 second.
•After turning on the balance will carry out a self-test.
•Afterwards the balance will change to zero display, thus indicating readiness for
operation.
•The following icons will appear:
Zero indicator
Stability display
kg Weighing mode / weighing unit
7 Changing the installation site for the balance
•If the balance is to be taken to another installation site carry out all
the precautionary measures described in chapter 6 of these
operating instructions.
Doc.No.OEX-BA-e-1411 13

8 Cleaning and maintenance
•Unplug the mains plug of power unit IEX-A01.
•Check, whether the electrostatic earth connection is plugged in correctly and
make sure it is in good working order.
•Clean the balance and the housing, using a slightly moistened cloth.
Simply apply any commercially available cleaning agent available for
domestic use.
•Do not use any agents for cleaningthat may scratch the surface
mechanically. Among other things it is covered in a conductive coat that is
important for safe operation.
•The adjustable feet of the platform balance and the shock absorbers
are made of rubber. When they are inside the hazard area, you will
have to use a slightly moistened cloth for cleaning. Moistness
protects against electrostatic charges.
•Make sure that you do not damage the earth contact springs during
cleaning. They serve as a retainer of electrostatic charge on the
weighing pan. They should be approx. 5 mm above the shock
absorbers.
•Do not clean in the presence of explosive gasses.
•The use of concentrated acids, alkaline solutions or alcohols for
cleaning purposes is prohibited.
•The same applies to the use of compressed air and high pressure
cleaning equipment.
14 Doc.No.OEX-BA-e-1411

9 Technical examination
•Technical examination by expert staff is required every three
months.
The examination should include the following:
•Condition of connection to earth for balance display unit, including:
oFastening of display housing and electrical resistance - max. 100Ω,
oFit a flexible steel tube for the protection of the conductor leading to the
platform balance and the display unit (regular balances with “cabled”
display).
•The quality of the connection between the platform and the display unit (flexible
steel tube connecting both subassemblies).
•Damage and loss of a connection is inacceptable.
•The electrical connection between the housing and the platform – the maximum
resistance between the weighing pan (put on) should be less than100Ω.
•Stability and levelling of the balance.
Doc.No.OEX-BA-e-1411 15
This manual suits for next models
9
Table of contents
Other KERN Scale manuals

KERN
KERN MBS User manual

KERN
KERN PES 620-3M User manual

KERN
KERN ABT series User manual

KERN
KERN ADB-A Series User manual

KERN
KERN CKE 6k0.02 User manual

KERN
KERN MPC 250K100NM User manual

KERN
KERN FKB 6K0.02 User manual

KERN
KERN GAB-P Series User manual

KERN
KERN RIB Series User manual

KERN
KERN PLS User manual
Popular Scale manuals by other brands

Ideal Life
Ideal Life Body-Manager BWM 0001 instruction manual

Steinberg Systems
Steinberg Systems SBS-LW-300E user manual

Tanita
Tanita RD-901 innerScan Dual instruction manual

HiWeigh
HiWeigh NSS user manual

Salter
Salter 1015 Instructions and guarantee

Health O Meter
Health O Meter 3001KL-AM User instructions