KERN PLS User manual

GB KERN PLS/PLJ/ALS/ALJ
Version 1.9 03/2008
Operating Instructions
Electronic precision and analytical balance
Contents
1Technical Data_________________________________________________ 37
2Declaration of conformity _______________________________________ 41
3Fundamental information (general)________________________________ 43
3.1 Intended Use ____________________________________________________ 43
3.2 Inappropriate use ________________________________________________ 43
3.3 Guarantee ______________________________________________________ 43
3.4 Monitoring the test substances_____________________________________ 44
4Fundamental safety information __________________________________ 44
4.1 Observe the information in the operating instructions__________________ 44
5Transport and storage __________________________________________ 44
5.1 Acceptance check _______________________________________________ 44
5.2 Packaging ______________________________________________________ 44
6Unpacking, installation and commissioning ________________________ 45
6.1 Place of installation, place of use ___________________________________ 45
6.2 Unpacking ______________________________________________________ 45
6.2.1 Installation __________________________________________________________ 46
6.2.2 List of items supplied __________________________________________________ 47
6.3 Mains supply____________________________________________________ 47
6.4 Connection of peripheral equipment ________________________________ 47
6.5 Initial start-up ___________________________________________________ 47
6.5.1 Power display ________________________________________________________ 47
6.5.2 Stability display_______________________________________________________ 47
6.5.3 Balance zero-display __________________________________________________ 48
6.5.4 Auto-zero-function ____________________________________________________ 48
6.6 Adjustment _____________________________________________________ 48
6.6.1 Adjusting with internal weight (only PLJ/ALJ) ________________________________ 49
6.7 Adjusting with external weight (only PLS/ALS)________________________ 49
6.8 Verification _____________________________________________________ 50
6.9 Under floor weighing _____________________________________________ 51
PLS/PLJ/ALS/ALJ-BA-e-0819 3
5

7Operation_____________________________________________________ 52
7.1 Operating elements ______________________________________________ 52
7.1.1 Rear illuminated Display________________________________________________ 52
7.1.2 Keyboard: ___________________________________________________________ 53
7.2 Weighing _______________________________________________________ 54
7.2.1 Simple weighing ______________________________________________________ 54
7.2.2 Weighing Units _______________________________________________________ 54
7.3 Taring__________________________________________________________ 56
7.4 Adding of display values (only PLJ)_________________________________ 56
7.5 Parts counting __________________________________________________ 58
7.6 Percentage determiation __________________________________________ 59
7.6.1 Determination of a reference weight by weighing_____________________________ 59
7.6.2 Determination of a reference weight by numeric entry _________________________ 60
8Functions_____________________________________________________ 61
8.1 Weighing functions ______________________________________________ 61
8.2 General functions ________________________________________________ 62
8.3 Parameter for the serial interface ___________________________________ 63
9Data output RS 232 C ___________________________________________ 64
9.1 Technical Data __________________________________________________ 64
9.2 Pin allocation of the balance exit plug (front view) _____________________ 64
9.3 Interface cables _________________________________________________ 64
9.4 Description of the data transfer (data format) _________________________ 65
9.4.1 Output when pressing the PRINT-button ___________________________________ 65
9.4.2 Remote commands ___________________________________________________ 65
9.4.3 Output format ________________________________________________________ 66
10 Maintenance, upkeep, disposal _________________________________ 67
10.1 Cleaning _______________________________________________________ 67
10.2 Maintenance, upkeep _____________________________________________ 67
10.3 Disposal________________________________________________________ 67
11 Troubleshooting _____________________________________________ 68
PLS/PLJ/ALS/ALJ-BA-e-0819
36

1 Technical Data
KERN ALS 120-4 ALS 220-4 ALJ 120-4
Readability (d) 0.1 mg 0.1 mg 0.1 mg
Weighing range (Max) 120 g 220 g 120 g
Taring range (subtractive) 120 g 220 g 120 g
Reproducibility 0.2 mg 0.2 mg 0.2 mg
Linearity ± 0.2 mg ± 0.2 mg ± 0.2 mg
Minimal piece weight for
Parts counting > 0.5 mg > 0.5 mg > 0.5 mg
Recommended adjusting weight,
not included (class) 100 g (E2) 200 g (E2) internal
Warming up time (operating
temperature) 8 h
Verifiable no
Reference piece numbers for
Parts counting 10, 20, 50, freely selectable
Weighing Units mg, g, ct, gn, mom, oz, dwt
Stabilization time (typically) 4 sec.
Permissible ambient temperature + 15° C .... +30° C
Humidity of air max. 80 % (not condensing)
Under floor weighing device Loop for under floor weighing, serial
Housing (W x D x H) mm 206 x 312 x 260
Weighing plate mm 85
Weight kg (net) 6,3
PLS/PLJ/ALS/ALJ-BA-e-0819 37

KERN ALJ 220-4 ALJ 160-4M ALJ 220-4M
Readability (d) 0,1 mg 0,1 mg 0,1 mg
Weighing range (Max) 220 g 160 g 220 g
Minimum load (Min) - 10 mg 10 mg
Taring range (subtractive) 220 g 160 g 220 g
Reproducibility 0,2 mg 0,2 mg 0,2 mg
Linearity ± 0,2 mg ± 0,2 mg ± 0,2 mg
Minimal piece weight for
Parts counting > 0,5 mg > 0,5 mg > 0,5 mg
Adjustment weight internal internal internal
Verifiable no yes yes
Verification value (e) - 1 mg 1 mg
Accuracy category - I I
Warming up time (operating
temperature) 8 h
Reference piece numbers for parts
counting 10, 20, 50, freely selectable
Weighing Units mg, g, ct, gn,
mom, oz, dwt mg, g, ct
Stabilization time (typically) 4 sec.
Permissible ambient temperature + 18° C .... + 30° C
Humidity of air max. 80 % (not condensing)
Under floor weighing device Loop for under floor weighing, serial
Weighing plate mm 206 x 312 x 260
Housing (W x D x H) mm 85
Weight kg (net) 6,3
PLS/PLJ/ALS/ALJ-BA-e-0819
38

KERN PLS 360-3 PLS 510-3 PLS2100-2 PLS 4000-2
Readability (d) 0.001 g 0.001 g 0.01 g 0.01 g
Weighing range (Max) 360 g 510 g 2,100 g 4,000 g
Taring range (subtractive) 360 g 510 g 2,100 g 4,000 g
Reproducibility 0.002 g 0.002 g 0.02 g 0.02 g
Linearity ± 0.002 g ± 0.002 g ± 0.02 g ± 0.02 g
Minimal piece weight for
Parts counting > 0.005 g > 0.005 g > 0.05 g > 0.05 g
Recommended adjusting
weight, not included
(class) 200 g (F1) 500 g (E2) 2000 g (F1) 2000 g (F1)
Verifiable no no no no
Warming up time
(operating temperature) 2 h 4 h 2 h 4 h
Reference piece numbers
for parts counting 10, 20, 50, freely selectable
Weighing Units mg, g, ct, gn, mom, oz, dwt
Stabilization time
(typically) 4 sec.
Permissible ambient
temperature + 15° C .... +30° C
Humidity of air max. 80 % (not condensing)
Under floor weighing
device Loop for under floor weighing, serial
Weighing plate mm 128 x 128 165 x 165
Housing (W x D x H) mm 206 x 312 x 160
(with draft shield) 206 x 312 x 98
(without draft shield)
Weight kg (net) 4,6 5,4
PLS/PLJ/ALS/ALJ-BA-e-0819 39

KERN PLJ 360-3M PLJ 510-3M PLJ 2100-2M PLJ 4000-2M PLJ 6100-2
Readability (d) 0.001 g 0.001 g 0.01 g 0.01 g 0.01 g
Weighing range (Max) 360 g 510 g 2,100 g 4,000 g 6,100 g
Minimum load (Min) 0,02 g 0,02 g 0,5 g 0,5 g -
Taring range (subtractive) 360 g 510 g 2,100 g 4,000 g 6,100 g
Reproducibility 0.002 g 0.002 g 0.02 g 0.02 g 0.01 g
Linearity ± 0.002 g ± 0.002 g ± 0.02 g ± 0.02 g ± 0.03 g
Minimal piece weight for
Parts counting > 0.05 g > 0.005 g > 0.05 g > 0.05 g > 0.05 g
Adjustment weight internal internal internal internal internal
Verifiable yes yes yes yes no
Verification value (e) 0.01 g 0.01 g 0.1 g 0.1 g -
Accuracy category II II II II -
Warming up time (operating
temperature) 2 h 4 h 2 h 4 h 4 h
Reference piece numbers for
parts counting 10, 20, 50, freely selectable
Weighing Units mg, g, ct, mg, g, ct, gn,
mom, oz, dwt
Stabilization time (typically) 4 sec.
Permissible ambient temperature + 15° C .... +30° C
Humidity of air max. 80 % (not condensing)
Under floor weighing device Loop for under floor weighing, serial
Weighing plate mm 128 x 128 165 x 165
Housing (W x D x H) mm 206 x 312 x 160
(with draft shield) 206 x 312 x 98
(without draft shield)
Weight kg (net) 4,6 5,4
PLS/PLJ/ALS/ALJ-BA-e-0819
40

2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Español Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de
acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
Electronic Balance: KERN ALS, ALJ
KERN PLS, PLJ
Mark applied EU Directive Standards
89/336EEC EMC EN 61000-4-2 :1999
EN 61000-4-3 :1996
EN 61000-4-4 : 1999
EN 61000-4-5 : 1998
EN 61000-4-6 : 1999
EN 61000-4-11 : 1997
EN 55022 :2000
Date: 22.12.2005 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149
PLS/PLJ/ALS/ALJ-BA-e-0819 41

KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Konformitätserklärungen
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
This declaration is only valid with the certificate of conformity by a notified body.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer
benannten Stelle.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Cette déclaration est valide seulement avec un certificat de conformité d’un
organisme notifié.
Español Manifestamos en la presente que el producto al que se refiere esta declaración está de
acuerdo con las normas siguientes.
Esta declaración solo será válida acompañada del certificado de conformidad de
conformidad de la parte nominal.
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità
della parte nominale.
Model: KERN ALJ/PLJ
EU Directive Standards EC-type-approval
certificate no.
Issued by Model
T6656 NMI PLJ 360-3M
PLJ 510-3M
PLJ 2100-2M
PLJ 4000-2M
90/384/EEC EN 45501
TCM 128/06-4438 CMI ALJ 160-4M
ALJ 220-4M
Date: 15.9.2006 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149
PLS/PLJ/ALS/ALJ-BA-e-0819
42

3 Fundamental information (general)
3.1 Intended Use
The balance you have acquired serves to determine the weighing value of the
material to be weighed. It is intended to be used as a “non-automatic“ balance, i.e.
the material to be weighed is manually and carefully placed in the centre of the
weighing plate. The weighing value can be read off after a stable weighing value has
been obtained.
3.2 Inappropriate use
Do not use the balance for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly
draining fluids from a container on the balance.)
Do not leave a permanent load on the weighing plate. This can damage the
measuring system.
Be sure to avoid impact shock and overloading the balance in excess of the
prescribed maximum load rating (max.), minus any possible tare weight that is
already present. This could cause damage to the balance.
Never operate the balance in hazardous locations. The series design is not
explosion-proof.
Structural alterations may not be made to the balance. This can lead to incorrect
weighing results, faults concerning safety regulations as well as to destruction of the
balance.
The balance may only be used in compliance with the described guidelines. Varying
areas of application/planned use must be approved by KERN in writing.
3.3 Guarantee
The guarantee is not valid at the following:
•non-observation of our guidelines in the operating instructions
•use outside the described applications
•alteration to or opening of the device
•mechanical damage and damage caused by media, liquids, natural wear and tear
•inappropriate erection or electric installation
•overloading of the measuring equipment
PLS/PLJ/ALS/ALJ-BA-e-0819 43

3.4 Monitoring the test substances
The metrology features of the balance and any possible available adjusting weight
must be checked at regular intervals within the scope of quality assurance. For this
purpose, the answerable user must define a suitable interval as well as the nature
and scope of this check. Information is available on KERN’s home page (www.kern-
sohn.com) with regard to the monitoring of balance test substances and the test
weights required for this. Test weights and balances can be adjusted quickly and at a
reasonable price in KERN’s accredited DKD calibration laboratory (return to national
normal).
4 Fundamental safety information
4.1 Observe the information in the operating instructions
Please read the operating instructions carefully before erecting and commissioning,
even if you already have experience with KERN balances.
4.2 Staff training
The device may only be operated and looked after by trained members of staff.
5 Transport and storage
5.1 Acceptance check
Please check the packaging immediately upon delivery and the device during
unpacking for any visible signs of external damage.
5.2 Packaging
Please retain all parts of the original packaging in case it should be necessary to
return items at any time.
Only the original packaging should be used for return consignments.
Before despatch, disconnect all attached cables and loose/movable parts.
Apply any intended transport security devices. Secure all parts e.g. weighing plate,
mains power supply etc. against slipping and damage.
PLS/PLJ/ALS/ALJ-BA-e-0819
44

6 Unpacking, installation and commissioning
6.1 Place of installation, place of use
The balance is constructed in such a way that reliable weighing results can be
achieved under normal application conditions.
By selecting the correct location for your balance, you will be able to work quickly and
precisely.
Therefore please observe the following at the place of installation:
•Place the balance on a firm, level surface;
•Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
•Protect the balance against direct draughts due to open windows and doors;
•Avoid jarring during weighing;
•Protect the balance against high humidity, vapours and dust;
•Do not expose the device to extreme dampness for longer periods of time.
Inadmissible bedewing (condensation of air moisture on the device) can occur
if a cold device is taken into a significantly warmer environment. In this case,
please acclimatise the device for approx. 2 hours at room temperature after it
has been disconnected from the mains.
•Avoid static charging of items to be weighed, or weighing container.
Major display deviations (incorrect weighing results) are possible if electromagnetic
fields occur as well as due to static charging, currents and instable power supply. It is
then necessary to change the location.
6.2 Unpacking
Carefully remove the balance from its packaging, remove the plastic wrapping and
position the balance in its intended working location.
PLS/PLJ/ALS/ALJ-BA-e-0819 45

6.2.1 Installation
Use the foot screws to level the balance until the air bubble in the bubble level is in
the prescribed circle.
Analysing balances (ALS/ALJ):
Precision balances (PLS/PLJ):
Models d = 10 mg Models d = 1 mg
PLS/PLJ/ALS/ALJ-BA-e-0819
46

6.2.2 List of items supplied
Standard accessories:
•Balance
•Weighing plate
•Mains device (Power adaptor)
•Operating Instructions
•Draft shield (only for models up to d = 1mg)
•Protective working cover
6.3 Mains supply
Electric power supply is by means of the external mains supply circuit. The printed
voltage level must comply with the local voltage.
Only use original KERN mains supply circuits. The use of other makes is subject to
approval by Kern.
6.4 Connection of peripheral equipment
The balance must be disconnected from the mains before connecting or
disconnecting additional equipment (printer, PC) to or from the data interface.
Only use KERN accessories and peripheral equipment with your balance. These
have been ideally coordinated to your balance.
6.5 Initial start-up
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1). During this
warming up time, the balance must be connected to the power supply (mains,
accumulator or battery).
For verified balances series PLJ…-3M there is warming up time, during which the
display indicates [burn - in].
The accuracy of the balance depends on the local acceleration of the fall.
Instructions in chapter Adjustment must be observed.
6.5.1 Power display
If this character [ : : ] is visible, the balance it supplied with power via the mains
power supply. Actuating the ON/OFF –button will switch the balance into weighing
mode.
This way, the power display in the display overview is no longer visible.
6.5.2 Stability display
If the display shows the stability display [ ] the balance is in a stable status. If the
status is instable the [ ] display disappears (see also chapter 8.2.1). Stable
environment conditions may be reached by using a draft shield.
PLS/PLJ/ALS/ALJ-BA-e-0819 47

6.5.3 Balance zero-display
If the balance does not show exactly zero although the pan scale is unloaded, press
the TARE-button and the balance will be reset to zero [ a].
6.5.4 Auto-zero-function
In the menu you can turn on or off the auto-zero-function. To do so, proceed as
follows:
ÖTurn on balance using the ON/OFF-button
ÖPress the Tare-button; the display shows horizontal lines
ÖWhile this display shows, press the CAL-button, until AUtO is displayed
ÖPress the PRINT-button; the display now shows the current menu item
blinking
ÖUse the F-button to choose between the following settings:
AUtO 0: Auto-zero-function activated
(Display [ /a])
AUtO 1: Auto-zero-function deactivated
ÖUse the PRINT-button to confirm the selected setting
6.6 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. It is
also recommendable to adjust the balance periodically during weighing operation in
order to obtain exact measured values.
PLS/PLJ/ALS/ALJ-BA-e-0819
48

6.6.1 Adjusting with internal weight (only PLJ/ALJ)
The installed adjusting weight allows checking and resetting the weighing accuracy at
any time.
The automatic adjustment function is always active.
You can also start adjustment manually by pressing the Cal-button.
Automatic adjustment is started when the balance is
•cut off from power supply,
•when changing temperature
•after a time interval
Temperature / time controlled adjustment is taking place:
5 minutes before the start of automatic adjustment, this will be announced by a "º C“,
(change of temperature) or a “►“ (after a certain time interval ends) symbol on the
display. The user must complete his/her weighing process within this time. After 5
minutes [ CAL 30 ] appears on the display. This starts a “count down” of 30 seconds
[ CAL 30 ] →[ CAL 0]. During these 30 seconds it is possible to cancel the
adjustment by pressing the TARE key. This makes the balance return to weighing
mode in order to e. g. complete an unfinished measurement.
6.6.2 Adjusting with external weight (only PLS/ALS)
Carry out adjustment with the recommended adjusting weight (see Chapter 1
„Technical Data“).
Procedure for adjustment:
Observe stable environment conditions. A warming-up time of ca. 2 hours for
stabilisation is necessary.
ÖUse the ON/OFF-button to turn on the balance.
ÖPress CAL-button; the display shows NO CAL.
ÖWait until LOAD is displayed, then place adjusting weight (see Chapter
1 „Technical Data“) carefully in the centre of the weighing plate.
ÖThe display shows CAL, adjustment is started.
ÖThe display shows UNLOAD, adjustment is completed.
ÖRemove adjusting weight, the balance will automatically return to
weighing mode.
In case of an adjustment error or wrong adjusting weight CAL Err is displayed;
repeat adjustment process
Keep test weight with the balance. Daily checking of the weighing accuracy is
recommended for quality-relevant applications.
PLS/PLJ/ALS/ALJ-BA-e-0819 49

6.7 Verification
General:
According to the EU guideline 90/384/EEC balances must be verified officially if they
are to be used as follows (legally regulated area):
a) For commercial transactions if the price of goods is determined by weighing
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory
c) For official purposes
d) For the production of finished packages
In case of doubt, please contact your local office of weights and measures.
Verification information
An EU qualification approval is available for those balances marked as appropriate
for verification in the technical data. In the event that the balance is applied in an
area subject to verification as described above, it must be officially verified and re-
verified at regular intervals.
Re-verification of a balance is carried out in compliance with the respective legal
provisions of the states. The term of verification validity for balances in Germany, for
example, is normally 2 years.
The legal provisions of the country of use are to be observed.
After verification the balance is sealed at the marked positions.
Verification of the balance is invalid without the seal.
Position of the seal:
Seal
Balances with obligatory verification must be taken out of service if:
- The weighing result of the balance is outside the maximum limits of
operating errors. Therefore load balance is regular intervals with known test
weight (ca. 1/3 of the max. load) and compare with displayed value.
- The date for subsequent or periodical verification has passed.
PLS/PLJ/ALS/ALJ-BA-e-0819
50

6.8 Under floor weighing
Using under floor weighing allows weighing of objects that because of their size of
shape cannot be placed on the pan scale.
Proceed as follows:
•Turn off balance.
•Open the closing lid at the bottom of the balance.
•Carefully and completely hook in the hook for under floor weighing.
•Place balance over an opening.
•Suspend the item to be weighed from the hook and carry out weighing.
Fig. 1: Setting up balance for under floor weighing
CAUTION
•Ensure that all suspended objects are stable enough to hold the desired items
to be weighed securely (danger of breaking).
•Never suspend loads that exceed the stated maximum load (Max)
(danger of breaking)
Always ensure that underneath the load there are no living beings or objects that
might be damaged.
!
NOTE
When the under floor weighing the opening in the bottom of the balance must be
closed (protection against dust).
!
PLS/PLJ/ALS/ALJ-BA-e-0819 51

7 Operation
7.1 Operating elements
7.1.1 Rear illuminated Display
Contrasting display which can also be read in the dark.
To choose the backlight mode, please follow the below instruction for the setting:
Operation: Display:
ÖUse the ON/OFF-button to turn on the balance
0.00
g
a
ÖPress Tare-button
iiiiiiii
ÖWhile this display is on, press the CAL-button.
The first function “BL“ appears.
BL
ÖPress the PRINT-key; display shows actual value for
this function
BL 0
ÖUse the F-key to choose between the following
settings:
0 Backlight is turn off
1 Backlight is turn on
BL 0
Ú
BL 1
ÖPress key PRINT to confirm setting
PLS/PLJ/ALS/ALJ-BA-e-0819
52

7.1.2 Keyboard:
Key Function
ON/OFF •Turn on/off
PRINT
•Printout of the weighed value on an external
appliance
(printer or PC)
•Save settings
CAL •Adjusting function (manual)
F
•Function key
•Switching between weighing units
TARE/→0←
•Taring
•Set weight display at zero
PLS/PLJ/ALS/ALJ-BA-e-0819 53

7.2 Weighing
7.2.1 Simple weighing
Operation: Display:
ÖUse the ON/OFF-button to turn on the balance
ÖAs soon as the weight display shows „0.000" your
balance is ready for weighing 0,0000
ÖPlace items to be weighed on balance, the weighed
value is displayed. 19.6879 g
ÖPressing the F-button you can switch to another
weighing unit, e.g. ct (see Chapter 7.2.2) 98.4380 ct
ÖTo turn off the balance press the ON/OFF-button
7.2.2 Weighing Units
Switching option for a weighing unit by multiple pressing of the F-button:
Weighing unit g
↕
Display with three decimal digits
↕
Weighing unit mg ↔ct ↔GN ↔mo ↔oz ↔dwt
Display Conversion factor
1 g =
Gram g 1.
Ounce oz 0.035273962
Grain GN 15.43235835
Pennyweight dwt 0.643014931
Momme mo 0.2667
Carat ct 5
Different foreign weighing units are integrated in the various balance models.
For details see Chapter 1 „Technical Data“.
PLS/PLJ/ALS/ALJ-BA-e-0819
54
This manual suits for next models
18
Table of contents
Other KERN Scale manuals

KERN
KERN OEX 6K-3HM User manual

KERN
KERN RIB Series User manual

KERN
KERN PLJ-N User manual

KERN
KERN EOB B Series User manual

KERN
KERN FIS User manual

KERN
KERN SFB 10K1HIP User manual

KERN
KERN KBP V20 User manual

KERN
KERN EMS 300-3 User manual

KERN
KERN MCB 300K100M User manual

KERN
KERN PLJ-GM Series User manual