KETEK SiPM User manual

Silicon Photomulplier
Evaluaon Kit
Quick Start Guide
SiPM
Eval Kit
www.ketek.net
info@ketek.net
Hofer Str. 3
81737 Munich
Germany
phone +49 89 673 467 70
fax +49 89 673 467 77
KETEK GmbH
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

Revision history:
Rev. 1.0, August 2016 - Inial release
This quick start guide covers the KETEK SiPM Evaluaon Kit only.
For other products please refer to www.ketek.net.
Table of Contents
1 Introducon..........................................................................................................................3
2 Setup Example......................................................................................................................3
3 Single Photon Measurements ..............................................................................................4
3.1 Dark Count Spectrum....................................................................................................................... 4
3.2 Single Photon Spectrum .................................................................................................................. 4
3.3 Using the Full Dynamic Range.......................................................................................................... 4
4 Measurements with Scinllator ...........................................................................................5
4.1 Energy Spectrum.............................................................................................................................. 5
4.2 Coincidence Time Resoluon........................................................................................................... 6
5 Appendix ..............................................................................................................................7
5.1 Variant for Opcal Bench Mount ..................................................................................................... 7
5.2 SiPM Bias Source.............................................................................................................................. 7
SiPM
Eval Kit
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

08/2016 Rev. 1.0 Page 3
SiPM
Eval Kit
1 Introducon
The KETEK SiPM Evaluaon Kit allows an easy operaon and evaluaon of any KETEK SiPM. It can be used
for a wide range of applicaons which require e.g. single photon counng or measurements with scinlla-
tors. Two of the Evaluaon Kits can be mounted face to face for coincidence measurements.
The SiPM Evaluaon Kit is equipped with a preamplier and set up modularly so that the PCBs with the
presoldered SiPMs can be swapped to a dierent SiPM model.
For the operaon of the Evaluaon Kit only a +12 V DC power supply, a bias source and an oscilloscope are
required.
For full technical details of the SiPMs and the Evaluaon Kit, please refer to the KETEK SiPM Datasheet and
the KETEK SiPM Evaluaon Kit Datasheet.
Addionally, KETEK provides a SiPM Bias Source which can be directly connected to the Evaluaon Kit.
Please refer to the KETEK SiPM Bias Source Datasheet for further informaon.
2 Setup Example
• The signal and bias connectors are SMA type, the preamplier is equipped with solder pins. An over-
view is shown in gure 1. Signal lines need to be be terminated with 50 Ω.
• Preamplier power: +12V DC, 150 mA
• Bias: posive with max. + 40 V, typical current limit 2 mA
• Signal: connected to the preamplier
• Note: Signal corresponds to 90% of the SiPM signal
• Monitor: connected to an oscilloscope (50 Ω DC)
• Note: Monitor corresponds to 10% of the SiPM signal
• Amplied Signal: connected to an oscilloscope (50 Ω DC)
• Since the SiPM is a highly sensive photodetector, it must be operated under dark condions.
• Aer biasing the SiPM with e.g. 4 V above the breakdown voltage, dark counts should be visible at the
Amplied Signal path, using e.g. a mebase of 100 ns/div and a vercal resoluon of 20 mV/div. The
Monitor is used for higher photon uxes for which the preamplier saturates (e.g. when coupling a
bright scinllator).
• Note: The procedure may vary depending on the used readout electronics. For the examples shown here, a
digital oscilloscope is used.
Fig. 1 SiPM Evaluaon Kit Connecon Scheme
Amplied
Signal
+12 V DC GND
Preamplier PCB with SiPM
Bias
Monitor
Signal
Mounng Holes for
Coincidence Setup
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

SiPM Evaluaon Kit Quick Start Guide
SiPM
Eval Kit
3 Single Photon Measurements
3.1 Dark Count Spectrum
• At the oscilloscope, set the trigger to the amplied SiPM signal at 0.5 pe amplitude (1 pe is the smallest
occurring pulse height, corresponding to a photoelectron pe).
• Note: This measurement has to be done under dark condions without illuminang the SiPM.
• Measure the area respecvely the charge of the signal at the posion of the trigger point in e.g. a 25 ns
wide gate. Best results are obtained by integrang the whole posive pulse area.
• Histogramming of the integrated charge is the resulng dark count spectrum.
• Note: E.g. crosstalk probability can be extracted from the dark count spectrum.
3.2 Single Photon Spectrum
• For this measurement, a pulsed light source is needed. This can be e.g. a pulsed LED or a pulsed laser.
Connect the electrical trigger output of the pulser to the oscilloscope and set the trigger to it.
• Note: In case a synchronized electrical trigger output is not provided by the pulser, a second SiPM Evaluaon
Kit can be used to generate the trigger signal. This second SiPM should be fully illuminated by the same opcal
pulse e.g. with an opcal beam splier. The other path for the single photon measurement can be aenuate
e.g. with neutral density lters.
• Measure the area respecvely the charge of the SiPM signal in e.g. a 25 ns wide gate. The histogram of
the measured charge is the single photon spectrum (cf. g. 2).
• Note: E.g. the relave gain, breakdown voltage and photo detecon eciency can be extracted.
3.3 Using the Full Dynamic Range
• The amplied SiPM signal output and the Monitor output can be connected and used simultaneously.
Typically for low light levels down to single photons, the measurement is done with the amplied out-
put. For higher light levels, the preamplier will saturate at 0.5 V signal amplitude. Aer this point, the
Monitor output is measured. This allows to exploit the the full dynamic range of the SiPM.
Fig. 2 Example of a Single Photon Spectrum
Measured with PM3350 at 4 V overvoltage
SiPM Signal
Histogram
Laser Trigger SiPM Signal
20 mV/divCounts [a.u.]
5 ns/div
Integrated Charge [nVs]
End Integraon Gate
Start Integraon Gate
pedestal
1 pe
2 pe
3 pe
4 pe
5 pe
6 pe
7 pe
8 pe
9 pe
10 pe
11 pe
12 pe
13 pe
14 pe
15 pe
16 pe
17 pe
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

08/2016 Rev. 1.0 Page 5
SiPM
Eval Kit
4 Measurements with Scinllator
• With bright scinllators typically the charge released by the SiPM exceeds the dynamic range of the
preamplier. In this case, the Monitor output is used. The charge released by the SiPM is a measure
for the detected number of scinllang photons and thus a measure for the deposited energy in the
scinllator.
• Note: Alternavely the preamplier can be disconnected to connect the Signal output directly to the oscillo-
scope (50 Ω DC). This is an opon in case ming is not measured.
• For simultaneous measurements of ming and energy, the Amplied Signal is used for ming and the
Monitor output for the energy measurement.
• Note: Even though the preamplier may be in saturaon, ming can sll be measured at the steep rising edge
of the Amplied Signal close to the baseline.
• Usually the monitor output is used to measure the charge released by the SiPM corresponding to the
deposited energy in the scinllator.
• Evaluaon Kits can be ordered either with unmounted or premounted LYSO crystals on the SiPMs:
• For PM11 (1.2 x 1.2 mm² acve area) LYSO with 1.2 x 1.2 x 4.0 mm³
• For PM33 (3.0 x 3.0 mm² acve area) LYSO with 3.0 x 3.0 x 5.0 mm³
• For PM66 (6.0 x 6.0 mm² acve area) LYSO with 5.0 x 5.0 x 5.0 mm³
4.1 Energy Spectrum
• In analogy to secon 3.1, at the oscilloscope the trigger has to be set to the SiPM signal above the base-
line and the pulse area as a measure for the energy is histogrammed.
• Figure 3 shows an example energy spectrum of a ²²Na source measured with the Monitor output. The
used scinllator is LYSO with 3 x 3 x 5 mm³.
Fig. 3 Example of an Energy Spectrum
Measured with PM3350 at 4 V overvoltage, ²²Na source and LYSO 3.0 x 3.0 x 5.0 mm³
0
50
100
150
200
250
300
350
0 250 500 750 1000 1250
Energy [keV]
Counts [a.u.]
Energy Resoluon
11.9% FWHM at 511 keV
(corrected for saturaon)
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

SiPM Evaluaon Kit Quick Start Guide
SiPM
Eval Kit
4.2 Coincidence Time Resoluon
• Two Evaluaon Kits are mounted facing each other
• Note: Two M6 mounng bolts are supplied for alignment when ordering two Evaluaon Kits, they are screwed
to one of the Evaluaon Kits, the second Evaluaon Kit is slid on the mounng bolts
• This example uses two PM3350, each equipped with 3.0 x 3.0 x 5.0 mm³ LYSO with a ²²Na source placed
exactly in between the SiPMs
• Note: ²²Na has two decay branches. It emits two 511 keV annihilaon γ rays in opposing direcons, originang
from a β+ decay. The second decay emits a single 1.27 MeV γ ray. Of interest here are the two 511 keV γ rays for
the coincidence measurement.
• Monitor outputs are connected to a coincidence logic, e.g. realized by the oscilloscope
• Lower threshold of the coincidence logic denes the lower energy cut-o. The thresholds for both
Monitor outputs are set to the minimum between Compton edge and 511 keV photopeak to lter
for the energy respecvely coincident events.
• The Coincidence Timing can then be measured by the me dierence of both Amplied Signals. The
output of the coincidence logic triggers the measurement of the mestamps of both Amplied Signals.
• Note: Best minig is achieved with a ming measurement close to the baseline. Typically with LYSO the op-
mum is around 2 - 3 pe.
• The me stamps of both Amplied Signals can be measured e.g. with a leading edge discriminator,
a constant fracon discriminator or using an oscilloscope.
• The me dierence between both me stamps is then histogrammed as shown in gure 4. The Coinci-
dence Time Resoluon is the FWHM of the Gaussian distribuon.
Fig. 4 Example of a CTR Histogram
Measured with two PM3350 at 4 V overvoltage, ²²Na source and LYSO 3.0 x 3.0 x 5.0 mm³
0
500
1000
1500
2000
2500
-1 -0.5 0 0.5 1
Time Dierence [ns]
Counts [a.u.]
CTR
180 ps FWHM
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk

08/2016 Rev. 1.0 Page 7
SiPM
Eval Kit
5 Appendix
5.1 Variant for Opcal Bench Mount
The SiPM Evaluaon Kit as also available in a dierent form factor for opcal bench mount including the
preamplier (cf. g. 5). The PCB with the SiPM is premounted to a Thorlabs® CP02/M Frame (SM1-Threaded
30 mm Cage Plate, 0.35” Thick, 2 Retaining Rings, M4 Tap). The SiPM is centered on the opcal axis. The
cables are MCX to SMA and are supplied together with the Evaluaon Kit.
5.2 SiPM Bias Source
As an addon to both Evaluaon Kits, a SiPM Bias Source is available (cf. g. 5). It oers an adjustable bias
voltage from 20 to 40 V with either posive or negave polarity. Current limit can be set to either 2 or 20
mA. For full technical specicaons please refer to the KETEK SiPM Bias Source Datasheet.
Fig. 5 SiPM Evaluaon Kit for Opcal Bench Mount with SiPM Bias Source
Preamplier
SiPM Bias Source SiPM premounted in optomechanic frame
www.aptechnologies.co.uk
A
P
Technologies Limited The Coach House Watery Lane Bath BA2 1RL
T: +44 (0) 1225 780400 F: +44 (0) 8701 266449 E: info@aptechnologies.co.uk
Table of contents