manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Kobold
  6. •
  7. Transducer
  8. •
  9. Kobold KP46 Series User manual

Kobold KP46 Series User manual

Series KP46
Explosion-Proof
Pressure Transducers
FM Rev. 05/10/13
KP46_manual_5-13
•User's Responsibility for Safety: KOBOLD manufactures a
wide range of process sensors and technologies. While each
of these technologies are designed to operate in a wide
variety of applications, it is the user's responsibility to select a
technology that is appropriate for the application, to install it
properly, to perform tests of the installed system, and to
maintain all components. The failure to do so could result in
property damage or serious injury.
•Wiring and Electrical: Because this is an electrically
operated device, only properly trained personnel should install
and maintain this product. Be sure that the power supplied to
the unit is appropriate for the electronics version supplied.
Electrical wiring of the sensor MUST be performed in
accordance with all applicable national, state and local codes
AND in accordance with the most recent version of the
National Electric Code ANSI/NFPA 70 (NEC) Articles 500
thorough 505.
•Make a Fail-Safe System: Design a fail-safe system
that accommodates the possibility of sensor, relay or
power failure. In critical applications, KOBOLD
recommends the use of redundant backup systems and
alarms in addition to the primary system.
•Hazardous Location Approval: CSA approved to
standards CSA 30 and U.L. 1203, Explosion-Proof for
Class 1, Div 1 Groups A, B, C, D and Div. 2 Groups A,
B, C, D for use in Hazardous Areas
• - ANSI/ISA 12.27.01-2003 Single
Seal Approved
• - CE EN61326 as per CECV,
Canadian Electrical Code
Variance
Specifications:
Available Ranges: 0-50 through 0-20000 PSIG
Over Range Protection: 2X full scale
Burst Pressure: 5X full scale
Cycle Life: >100 million cycles
Process Connection: ¼” MNPT (to 10K PSIG),
½” MNPT (15K & 20K PSIG)
Wetted Materials: 316 L Stainless Steel
others available upon request
Corrosion Resistance: Complies with NACE MR0175
Linearity: ±0.25% BFSL
Zero Offset: <±1% of full scale
Span Tolerance: <±2% of full scale
Media Temp Range: -55 to +250°F
Ambient Temp Range: -40-185°F
Compensated Range: 30 to 130°F
Temp. Drift: <±1.5% of full scale for zero
and span over the compensated range
Shock Resistance:100G, 11mSec.
Vibration Resistance: 10G peak, 20-2000Hz
Stability (1 year): ±0.25% of full scale
Electrical Specifications:
Output: 4-20 mA, 2-wire
Input Power Supply: 10-28 VDC
Max. Loop Load: 800 Ohms
Elec. Connection: ½” NPT Conduit with 6 Ft. 18 AWG wire leads
Elec. Protection: NEMA 4X / IP 66
Precautions
KOBOLD Instruments Inc.
1801 Parkway View Drive
Pittsburgh, PA 15205
PH: 412-788-2830
FAX: 412-788-4890
www.koboldusa.com
Wiring Connections:
CAUTION: BE SURE THAT POWER IS DISCONNECTED
BEFORE PROCEEDING!
4-20 mA, 2-wire Output
Red: +V Supply (10-28VDC)
Black: + Signal Output (4-20mA)
Green: Case/Equipment Ground
Installation:
1. Mount the transmitter so that the sensor port will not
become clogged with sediment or debris that could
hamper the sensing ability.
2. Although the transmitter is highly vibration resistant, it is
always good practice to mount the device appropriately to
minimize any potentially damaging shock and vibration.
3. Apply PTFE tape, or other appropriate thread sealant to
the NPT pressure fitting thread before installation. Tighten
the device using the 7/8” Hex next to the pressure
connection. DO NOT USE A PIPE WRENCH ON THE
TRANSMITTER HOUSING!
KOBOLD Instruments, Inc.
Standard Electrical Connections For Model KPD and KP46 Cable
2/09/18/13 www.koboldusa.com
Cable / Hookup
Wire Connections Standard Cable Conduit with Cable Explosionproof Conduit with Cable
Electrical Connection option ‘”A, B, C, D” “L, M, N, P” “T, U, W”
Gauge Wire 24 AWG 24 AWG 18 AWG
Available on KOBOLD Models: KPD KPD KP46
Cable / Hookup
Wire Connections Standard Cable Conduit with Cable Explosionproof Conduit with Cable
Output Type Green Black Red White Green Black Red White Green Black Red White
mV & 4-Wire Voltage +S Ground +V -S +S Ground +V -S N/A N/A N/A N/A
3 Wire Voltage N/C Ground +V S N/C Ground +V S Case Ground +V S
4-20 mA N/C Ground +V N/C N/C Ground +V N/C Case Ground +V N/C
S= Output Signal
V= Voltage Supply
N/C= Not Connected
Case= Case Ground
mV Voltage 4-20mA
0.5-4.5V ratiometric 1-5V, 1-6V, 0-5V 0-10V, 1-10V
Supply Voltage 5VDC typ., 10VDC max. 5VDC ± 0.01V 10-28VDC 15-28VDC 10-28VDC
Output Load >1 MΩ >10 KΩ >10 KΩ >10 KΩ 0-800Ω
Operating Temperature -40 to 105ºC -40 to 85ºC -40 to 85ºC -40 to 85ºC -40 to 85ºC

Other Kobold Transducer manuals

Kobold MM Series User manual

Kobold

Kobold MM Series User manual

Kobold NRF-2 User manual

Kobold

Kobold NRF-2 User manual

Kobold DOR User manual

Kobold

Kobold DOR User manual

Popular Transducer manuals by other brands

Tecsis P3230 instruction manual

Tecsis

Tecsis P3230 instruction manual

Zimmer PRO -V manual

Zimmer

Zimmer PRO -V manual

Campbell ST350 instruction manual

Campbell

Campbell ST350 instruction manual

Humminbird transducer Assembly & operating instructions

Humminbird

Humminbird transducer Assembly & operating instructions

Humminbird TROLLING MOTOR TRANSDUCER installation guide

Humminbird

Humminbird TROLLING MOTOR TRANSDUCER installation guide

MCS MCS-***B Series Application note

MCS

MCS MCS-***B Series Application note

Mercoid PBLTX Series Installation and operating instructions

Mercoid

Mercoid PBLTX Series Installation and operating instructions

Airmar DT800 Owner's guide & installation instructions

Airmar

Airmar DT800 Owner's guide & installation instructions

Fishman SBT-C installation guide

Fishman

Fishman SBT-C installation guide

turck IMX12-FI01-2SF-2I Series quick start guide

turck

turck IMX12-FI01-2SF-2I Series quick start guide

Simrad ES333-7CD -  REV A datasheet

Simrad

Simrad ES333-7CD - REV A datasheet

Humminbird APEX Series installation guide

Humminbird

Humminbird APEX Series installation guide

HBM C9B Mounting instructions

HBM

HBM C9B Mounting instructions

Balluff BTL7-A/E1 0-M-H-SA262-KA Series Condensed guide

Balluff

Balluff BTL7-A/E1 0-M-H-SA262-KA Series Condensed guide

GHM GRMU 2000 MP operating manual

GHM

GHM GRMU 2000 MP operating manual

LEI accessories HS-NS Installation and operation instructions

LEI accessories

LEI accessories HS-NS Installation and operation instructions

M-system PV instruction manual

M-system

M-system PV instruction manual

JDC Instruments SKYWATCH WWS Instructions for use

JDC Instruments

JDC Instruments SKYWATCH WWS Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.