Thank you for purchasing KYOWA’s product Handy Data Logger SME
series.
Read this Instruction Manual carefully in order to make full use of the high
performance capabilities of the product.
Do not use the product in methods other than described in this Manual.
Copyright © KYOWA ELECTRONIC INSTRUMENTS CO., LTD. All rights reserved.
This Instruction Manual may not be copied or reproduced, in whole or part, without consent of KYOWA.
Generally, company names and trade names described in this Instruction Manual are trademarks or
registered trademarks of the companies.
The contents of the Instruction Manual are subjected to change without prior notice.

CONTENTS
STANDARD ACCESSORIES ........................................................................................................................ 1
OPTIONS......................................................................................................................................................... 1
SAFETY PRECAUTIONS ............................................................................................................................. 1
PRIOR TO USE........................................................................................................................................... 1
SAFETY SYMBOLS................................................................................................................................... 1
NOTATIONS USED IN INSTRUCTION MANUAL................................................................................... 3
INFORMATIONAL NOTES...................................................................................................................... 3
QUOTATIONS............................................................................................................................................. 3
IMPORTANT PRECAUTIONS WHEN USING SME SERIES................................................................. 3
1. OUTLINE OF SME SERIES ..................................................................................................................... 4
1-1 OUTLINE OF SME SERIES............................................................................................................... 4
1-2 SYSTEM CONFIGURATION............................................................................................................. 5
2. PARTS AND FUNCTIONS ........................................................................................................................ 6
2-1 DESCRIPTION OF OPERATION PANEL........................................................................................ 6
2-2 DESCRIPTION OF DISPLAY ............................................................................................................ 7
3. HOW TO USE............................................................................................................................................ 10
3-1 BEFORE USING................................................................................................................................. 10
3-2 INSERTING DRY CELL ................................................................................................................... 10
3-3 INSERTINGAND REMOVING SD CARD......................................................................................11
3-4 POWER ON/OFF................................................................................................................................ 12
3-5 INPUT SETTING................................................................................................................................ 13
3-6 SENSOR CONNECTION .................................................................................................................. 14
3-7 SAVING DATA.................................................................................................................................... 16
4. MEASUREMENT WITH NTB SERIES CONNECTED...................................................................... 18
5. USEFUL FUNCTIONS............................................................................................................................. 20
5-1 OPERATION MODE OF SME SERIES .......................................................................................... 20
5-2 RESISTANCE VALUE CHECKING................................................................................................ 22
5-3 INSULATION CHECKING............................................................................................................... 23
5-4 AUTO POWER OFF FUNCTION .................................................................................................... 24
5-5 RELATIVE VALUE MEASUREMENT........................................................................................... 26
5-6 CALC. CAL COEF. (FOR DISPLAYING PHYSICAL QUANTITY)........................................... 28
5-7 READING TEDS INFORMATION .................................................................................................. 30
5-8 INTERVAL MEASUREMENT.......................................................................................................... 32
5-9 FILE OPERATION............................................................................................................................. 34
6. MENU LIST............................................................................................................................................... 39
7. SPECIFICATIONS ................................................................................................................................... 40
HANDY DATA LOGGER: SME-30A/31A ............................................................................................. 40
HANDY DATA LOGGER: SME-100A/101A ......................................................................................... 42
8. OUTSIDE DRAWING .............................................................................................................................. 44

STANDARD ACCESSORIES
The Handy Data Logger SME series (hereinafter referred to as the SME series) comes with the following
accessories. After unpacking, check to be sure that they are included.
SME-30A/31A SME-100A/101A
CD-R (Instruction Manual: This file) 1 1
Input Cable 1 1
Size AA Alkaline Dry Cell 2 2
Communication Cable N-102 (Length: Approx. 1 m) 1
―
Warranty and Test Data Sheet 1 each 1 each
Shoulder Strap 1 1
Wrist Strap 1 1
OPTIONS
AC Adapter (for 100 VAC, usable with the SME-31A and 101A) : DR-523E
NTB-100A series : Refer to the NTB catalog.
SAFETY PRECAUTIONS
The SME series is designed in accordance with “7. SPECIFICATIONS”.
Do not use the SME series in an environment exceeding the specifications. Or, failure of the SME series
may result.
PRIOR TO USE
For safe use of the SME series, do not forget to read the <Safety Precautions> prior to use.
Kyowa Electronic Instruments Co., Ltd. assumes no liability for any damages resulting from user’s failure
to comply with the safety precautions.
SAFETY SYMBOLS
For safety operation of the SME series, the following symbol marks are used in the Instruction Manual.
Improper operation of the system may result in death or severe injury of the
WARNING
1
operator.
Improper operation of the system may result in injury of the operator and
CAUTION physical damage of the system.

WARNING
●Do not short-circuit positive (+) and negative (-) terminals of the dry cell.
Or, it may produce fever or ignite due to a rush current.
●Do not heat or throw into fire.
Or, it may produce fever, ignite or explode.
●Correctly insert dry cells.
●If liquid leaking from the dry cell gets into your eyes, do not rub your eyes.
Wash them well with clean water and consult with a doctor.
●If any liquid from the dry cell comes in contact with your skin, wash quickly with water.
●Do not recharge the alkaline dry cells since the AA size alkaline dry cells are primary dry cells.
●When using the AC adapter with the SME series compatible to the AC adapter, check the power
voltage of the outlet before connecting.
●Avoid using in environment with inflammable gas, etc.
To prevent the risk of fire hazard or explosion, do not use the SME series in environment with
inflammable gas, vapor or dust.
●If any trouble occurs.
To prevent a fire hazard, if smoke is emitted from the SME series, immediately stop using the SME
series.
When using the AC adapter, disconnect the AC adapter from the receptacle.
●Avoid measuring object applied with high voltage.
CAUTION
●Use the SME series in the specified operating temperature range -10 to 50ºC.
Use at temperatures exceeding the specified range may lower the performance and cause trouble.
If use under direct sunlight or in a cold place is inevitable, prepare a sunscreen or take proper measures
to keep it warm.
●Use the SME series in the specified operating humidity range 20 to 85% RH.
Use in a humid place exceeding the specified range or where it is exposed to splashing water may lower
the performance and cause trouble.
●Immediately stop the SME series if operating environment changes abruptly.
Leave the SME series as it is until it becomes adaptable to the environment.
Abrupt change in ambient temperature due to transportation, etc. may cause dew condensation result to
deteriorate its performance and cause trouble.
●Do not use the SME series in strong electromagnetic field.
Use the SME series in a magnetic field environment where the PC may be used.
If the SME series is used near devices generating strong magnetic field such as a telemetry system,
microwave oven, electric furnace, etc., its performance may be lowered and erroneous operation and
trouble may result.
●Do not pull cords and cables.
Lay cords and cables with a certain allowance so that unreasonable force is applied to the connections.
Pulling or applying unreasonable force may cause trouble or interrupt the measurement.
2

NOTATIONS USED IN INSTRUCTION MANUAL
INFORMATIONAL NOTES
Certain notations are used as necessary to attract your attention to information that requires special care
when handing the SME series, and to information provided for reference purposes.
Examples of Notations
3
QUOTATIONS
Operation keys are enclosed as follows.
Example: ENTER key
Names described on the panel surface are expressed in angled bracket (< >). (Excluding menu)
Example: <ON>
IMPORTANT PRECAUTIONS WHEN USING SME SERIES
Essential precautions required when handling the SME series.
NOTE
Reference items required when handling the SME series.
MEMO
NOTE
1. Do not take off the batteries during saving the data.
If the power is forcibly turned OFF during saving the data, recording data may lead to
invalid.
2. Do not use solvent such as thinner for cleaning.
To clean the case, dip a soft cloth to a neutral detergent and twist it tightly.
If the solvent such as thinner is used, the case or panel may melt or be discolored.
3. Do not press on the LCD screen.
Do not puncture the LCD screen by an object with a fine head such as ballpoint pen.
Or, it may cause trouble.
4. Do not insert foreign matters to a SD card slot. Or, it may cause trouble.
5. Do not insert foreign matters or water to a dry cell place.
6. The SME series is not waterproof. After the water is attached, sufficiently clean with a
dry and clean cloth and dry the SME series before using.
7. When not using the SME series for a long time, remove dry cells from the SME series.

1. OUTLINE OF SME SERIES
1-1 OUTLINE OF SME SERIES
The SME series is an indicator that is used with a strain gage or strain gage transducer connected.
Capable of measuring strain easily with the accessory input cable since the SME series incorporates a
bridge box.
Power is supplied by AA size alkaline dry cells (2).
Capable of saving the measured data in a commercially available SD card.
The SME-100A/101A can be used for checking the operation of an optional Network Terminal Box
NTB-100A (hereinafter referred to as the NTB). Furthermore, the SME-100A/101A can also measure up to
33-CH by connecting the multiple NTB units.
An optional AC adapter can be used with the SME-31A/101A.
4
[SME series]
Model Measurement with the NTB Usage of AC adapter
SME-30A × ×
SME-31A × ○
SME-100A ○×
SME-101A ○○
Note: ○: Available
×: Not available
SD card of SDHC type is not available.
NOTE

1-2 SYSTEM CONFIGURATION
1) SME-30A/31A
5
*1 Only the SME-31A can use the AC adapter.
*2 For measuring a strain gage, use the accessory input cable.
2) SME-100A/101A
*1 Only the SME-101A can use the AC adapter.
*2 For measuring a strain gage, use the accessory input cable.
SD card
(
o
p
tion
)
Input:
Strain Gage *2
Strain Gage Transducer
Civil Engineering Transducer
with temperature measuring
function
Power supply: Size AA dry cells (2)
AC adapter *1 (option)
NTB-100 series (option)
Power supply: Size AA dry cells (2)
Input:
Strain Gage *2
Strain Gage Transducer
Civil Engineering Transducer
with temperature measuring
function
SD card
(
o
p
tional
)
AC adapter *1 (option)

2. PARTSAND FUNCTIONS
2-1 DESCRIPTION OF OPERATION PANEL
6
1) Dry cell cove
r
For mounting and dismounting, slide the cover with pressing.
Be sure not to dismount the cover carelessly.
2) Input connecto
r
Directly connect a sensor with a NDIS connector or connect the input cable
for measurement.
3) POWE
R
key POWER ON/OF
F
.
For turning ON the power, keep pressing for about 3 seconds.
For turning OFF the power, press for about 1 second.
4) INPUT key Displays input modes.
Used for selecting the target mode applicable to the sensor.
5) MENU key Displays a detailed menu.
Used for coefficient calculation, interval measurement, etc.
6) TEDS key Displays a menu related to TEDS.
Used for reading TEDS information.
7) SAVE key For saving the set value.
8) Curso
r
key Used for moving a cursor or switching a menu.
9) DEL key Used for deleting a file or default setting.
10) ENTE
R
key Used for determining a setting content.
11) ESC key Get back to the previous window.
12) SD card slot cove
r
Open the cover to find a SD card slot and DC jack.
13) DC jack Connect the optional AC adapter.
(For the SME-31A/101A only.)
14) SD card slo
t
Correctly insert the SD card.
(Insert the SD card with the slanted corner set to the top-righ
t
.)
15) NTB connection and maintenance connecto
r
Connec
t
the NTB with the communication cable.
(For the SME-100A/101A only.)
15)
1)
11)
10)
8)
4) 9)
12)
2) 3) 5) 7)
6) 14)
13)

2-2 DESCRIPTION OF DISPLAY
1) Monitor window
7
(1) Time Setting time of the SME series is displayed in due order from day,
month, year, and time.
Month is displayed as follows.
JAN: January FEB: February MAR: March
APR: April MAY: May JUN: June
JUL: July AUG: Augus
t
SEP: Septembe
r
OCT: Octobe
r
NOV: Novembe
r
DEC: Decembe
r
Displays(2) CH
t
he selected CH No.
When the SME-30A/31A and SME-100A/101A is selected, CH001 is
displayed.
When a CH of the NTB is selected, Box No. and CH of the target NTB
is displayed.
[Example]
When the Box No. of the NTB is 01 and 3CH is selected, CH013 is
displayed.
(3) Input mode Displays a setting input mode.
(4) Relative value measuremen
t
When the Relative Value Measurement (Rel.) is set to ON, <Rel.> is
displayed.
(5) Measuring value Displays a measuring value.
When the Calculation Calibration Coefficient (Calc. Cal Coef.) is set to
ON, decimal point position or unit can be set.
(6) Temperature
T
e
m
perature of the connecting civil engineering
t
ransducer with
temperature measuring function is displayed.
(7) Dry cell remaining capacity Dry cell remaining capacity in 3 levels.
(8) Interval measurement During an interval measurement, is displayed.
2) Dry cell remaining capacity
Dry cell remaining capacity Alkaline Dry Cell Nickel-hydrogen Dry Cell
Full Full
There is sufficient power to
use the device.
New dry cells are required for just
in case.
Dry cells may need to be
replaced.
Replace the dry cells.
Exhaust in a very short
period of time.
Replace the dry cells.
Nearly exhaust dry cells.
Replace the dry cells.
I
(2)
(1)
(6) (7)
(5)
(4)
(8)
(3)

3) Basic method for displaying window
This section describes how to display the target window. Blinking
(1) How to describe the selecting item.
If the target item or character is selected on the SME series
window, it blinks.
(<4G/T> in the right figure that is displayed with an outline
character means blinking.) Blinking
(2) Selection window
Window for selecting the target from several items
[Example] Selection window of input mode
Select a target item with the cursor key and push the
ENTER key for setting.
(3) Edit window
Window for selecting and inputting characters and numbers
[Example] File name input window Blinking
Select target characters and numbers with the cursor key
and push the ENTER key for setting.
8

9
(White page)

3. HOW TO USE
3-1 BEFORE USING
[How to install a strap]
Install the shoulder strap or wrist strap as follows.
Also, be sure to connect the small strap (dry cell cover strap) from the top left loop and the second from
top left loop to prevent loosing the dry cell cover.
Dry cell cover strap
10
With the wrist strap
With the shoulder strap
3-2 INSERTING DRY CELL
1) Slide a dry cell cover while pushing in direction of arrows
for removing.
2) Correctly insert dry cells.
3) Slide a dry cell cover while pushing in direction of arrows
for closing.
It is recommended to seal the dry cell cover so as not to open it carelessly.
MEMO
Use the size AA alkaline dry cells.

3-3 INSERTINGAND REMOVING SD CARD
1) For inserting the SD card
(1) Open a SD card slot cover to the arrow direction.
(2) Check the direction of the SD card.
Thoroughly insert the SD card until <click> sound is
heard.
(3) Close the SD card slot cover.
11
2) For removing the SD card
(1) Open a SD card slot cover to the arrow direction.
(2) Push the SD card until <click> sound is heard.
Remove your finger from the SD card, it will pop up.
(3) Pull the card straight up to remove.
(4) Close the SD card slot cover.
NOTE
Before inserting or removing the SD card, be sure to turn OFF the SME series or get back to the
monitor window.
Correctly insert the SD card. Or, the SME series may cause trouble.
The SD card of the SDHC type is not available.
It is recommended to format the SD card before using.
Format the SD card in FAT16 format.
NOTE
Do not remove the dry cells or SD card while accessing the SD card.
Or, the SD card may be broken.

3-4 POWER ON/OFF
1) POWER ON
12
Push POWER key approx. 3 seconds to turn ON the
(1)
power.
(2) Model appears for a moment.
(3) Monitor window appears.
2) POWER OFF
Push POWER key approx. 1 second to turn OFF the
(1)
power.
If the POWER key is pushed too long for turning OFF, the power is turned ON again. Release
your finger from the POWER key after the display disappears.
MEMO

3-5 INPUT SETTING
Select the target input mode that fits to the target sensor.
This section describes how to connect the civil engineering transducer with the temperature measuring
function.
For connecting the civil engineering transducer with the temperature measuring function, set input
mode to 4G/T.
Push INPUT key to display mode selection window.
1)
Blinking
2) Select the target mode with the cursor key.
Blinking
Push ENTER key for setting.
3)
4) The selected mode appears on the monitor window.
5) Connect the sensor.
13

3-6 SENSOR CONNECTION
For the strain gage transducer with a connector plug (NDIS standard) applied to the cable tip, directly
connect the transducer to the SME series connector. For the strain gage or transducer without the connector
plug, use the accessory input cable for connection. Wirings with the input connector or input cable are
described as follow.
1) Input connector
Measuring Target Connection Method Remarks
Strain gage A: +BV
Strain gage transducer B: -SIG
Civil engineering transducer with temperature C: -BV
14
measuring function
Quarter bridge method
120,240,350 ohm
Full bridge method
120 to 1000 ohm
D: +SIG
E: Shield
F: TEDS+
G: TEDS-
TEDS compatible
2) Input cable
Bridge Circuit
: Strain gage
: Built-in fixed
Measuring Method resistor
A: Active gage
E: Bridge exciter
e: Bridge output
Connection Method
A: Active gage
CH mode
No measurement.
SKIP
Measuring data is space.
Quarter bridge method
(2-wire system connection)
Gage resistance
120, 240, 350 ohm
1G120
1G240
1G350
Red (+BV)
White (-SIG)
Black (-BV)
Green (+SIG)
ACrocodile clip
(Shield)
Quarter bridge method
(3-wire system connection)
Gage resistance
120, 240, 350 ohm
1G120
1G240
1G350
Red (+BV)
White (-SIG)
Black (-BV)
Green (+SIG)
ACrocodile clip
(Shield)

Bridge Circuit
: Strain gage
: Built-in fixed
15
Measuring Method resistor
A: Active gage
E: Bridge exciter
e: Bridge output
Connection Method
A: Active gage
CH mode
Half bridge method
(Active-active method)
Gage resistance
120 to 1000 ohm
2G
Red (+BV)
A1 White (-SIG)
A2 Black (-BV)
Green (+SIG)
Crocodile clip
(Shield)
Full bridge method
(Strain gage transducer)
Gage resistance
120 to 1000 ohm
4G
4GI
Red (+BV)
AWhite (-SIG)
A
ABlack (-BV)
AGreen (+SIG)
Crocodile clip
(Shield)

3-7 SAVING DATA
Push SAVE key.
1)
16
<Saving…> appears on the monitor window and the
2)
measured data is recorded to the SD card.
NOTE
While displaying the <Saving…> on the monitor window, the SME series is accessing to the SD
card.
Do not remove the batteries or SD card while accessing the SD card.
Or, the SD card may be broken.
MEMO
<Write Protect> appears if the SD card is locked. In addition, the data cannot be saved.
•
Unlock the SD card before operation.
Push ESC key to clear the <Write Protect>.
<No Disk> appears if the operation related to the SD card is conducted with the SD card
•
not inserted.
Insert the SD card into the slot.
Push ESC key or insert the unlocked SD card to clear the <No Disk.>.
<Disk Err> appears if error (other than read and write errors) occurred in the SD card.
•
Push ESC key or solve the problem to clear the <Disk Err.>
[Example] When the capacity of the SD card for recording the measured data is short
<Disk Read Err> appears if the read error occurred in the SD card for some reason.
•
Push ESC key or solve the problem to clear the <Disk Read Err>.
<Disk Write Err> appears if the write error occurred in the SD card for some reason.
•
Push ESC key or solve the problem to clear the <Disk Write Err>.

17
(White page)

4. MEASUREMENT WITH NTB SERIES CONNECTED
The SME-100A/101A can check operation of the optional NTB or save the measured value through the
dedicated communication cable.
18
*1 Only the SME-101A can use the AC adapter.
*2 For measuring a strain gage, use the accessory input cable.
NTB-100 series (option)
SD card
(
o
p
tion
)
Input:
Strain Gage*2
Strain Gage Transducer
Civil Engineering
Transducer with the
temperature measuring
function
Power supply: Size AA dry cells (2)
AC ada
p
ter *1
(
o
p
tion
)
This manual suits for next models
4
Table of contents