TESTO 176 User manual

testo 17
6 · Data loggers
Instruction manual
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

1 Contents
31
1Contents
1 Contents .................................................................................................31
2 Safety and the environment..................................................................32
2.1. About this document......................................................................32
2.2. Ensure safety.................................................................................33
2.3. Protecting the environment............................................................33
3 Specifications ........................................................................................34
3.1. Use ................................................................................................34
3.2. Technical data ...............................................................................35
4 First steps ..............................................................................................43
4.1. Unlock the data logger...................................................................43
4.2. Connecting the data logger to PC..................................................43
5 Display and control elements...............................................................44
5.1. Display...........................................................................................44
5.2. LED ...............................................................................................47
5.3. Key functions.................................................................................48
6 Using the product..................................................................................48
6.1. Connecting a sensor......................................................................48
6.2. Programming data logger ..............................................................49
6.3. Menu overview ..............................................................................50
6.4. Mounting the wall bracket..............................................................53
6.5. Securing the data logger ...............................................................53
6.6. Reading out data ...........................................................................53
7 Maintaining the product........................................................................55
7.1. Replacing the battery.....................................................................55
7.2. Cleaning the instrument.................................................................56
8 Tips and assistance...............................................................................57
8.1. Questions and answers .................................................................57
8.2. Accessories and spare parts .........................................................58
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

2 Safety and the environment
32
2Safety and the environment
2.1. About this document
Use
> Please read this documentation through carefully and
familiarize yourself with the product before putting it to use. Pay
particular attention to the safety instructions and warning advice
in order to prevent injuries and damage to the products.
> Keep this document to hand so that you can refer to it when
necessary.
> Hand this documentation on to any subsequent users of the
product.
Symbols and writing standards
Representation
Explanation
Warning advice, risk level according to the
signal word:
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to
the equipment may occur.
> Implement the specified precautionary
measures.
Note: Basic or further information.
1. ...
2. ...
Action: more steps, the sequence must be
followed.
> ... Action: a step or an optional step.
- ... Result of an action.
Menu
Elements of the instrument, the instrument
display or the program interface.
[OK] Control keys of the instrument or buttons of
the program interface.
... | ... Functions/paths within a menu.
“...” Example entries
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

2 Safety and the environment
33
2.2. Ensure safety
> Only operate the product properly, for its intended purpose and
within the parameters specified in the technical data. Do not
use any force.
> Never use the instrument to measure on or near live parts.
> Before each measurement check that the connections are
correctly closed with blanking plugs or that appropriate sensors
have been correctly plugged in. The protection class in the
technical data specified for the corresponding instrument may
otherwise not be reached.
> testo 176 T3 testo 176 T4: The maximum permissible difference
in potential between the sensor inputs is 50 V. Take this into
account when using surface sensors with non-isolated
thermocouple.
> After the final measurement, allow probes and probe shafts to
cool down sufficiently in order to avoid burns from the hot
sensor tip or the probe shaft.
> Temperatures given on probes/sensors relate only to the
measuring range of the sensors. Do not expose handles and
feed lines to any temperatures in excess of 70 °C unless they
are expressly permitted for higher temperatures.
> Carry out only the maintenance and repair work on this
instrument that is described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Testo.
> After the final measurement, allow probes and probe shafts to
cool down sufficiently in order to avoid burns from the hot
sensor tip or the probe shaft.
> Do not use the device in a polluted environment (heavily dusty,
oil, foreign matter, volatile chemicals).
2.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate
collection for electric and electronic devices (observe local
regulations) or return the product to Testo for disposal.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
34
3Specifications
3.1. Use
The data loggers testo 176 serves the purpose of saving and
reading out individual readings and measurement series.
Readings are measured and saved with testo 176 and transferred
to the PC via USB-cable or SD-card, where they can be read out
and evaluated with the help of the testo Comfort Software. With the
software the data loggers can also be individually programmed.
Typical applications
testo 176 T1 is optimally suitable for temperature measurements
under extreme conditions, like e.g. on ships or in power plants,
because its metal housing provides excellent protection against
mechanical influences.
Due to the possibility of connecting external, high-precision Pt100
sensors the testo 176 T2 is optimally suitable for temperature
measurements, e.g. in food applications or in the laboratory.
With its external sensor connections (thermocouples type T, type K
and type J) and its robust metal housing the testo 176 T3 is able to
perform parallel temperature measurements on up to four
measuring locations under extreme conditions.
The external sensor connections (thermocouples type T, type K
and type J) make the testo 176 T4 and excellent choice for different
applications, such as e.g. simultaneous feed and return
temperature measurement in underfloor heating at up to four
measuring locations.
The testo 176 H1 is optimally suitable for parallel temperature and
humidity monitoring, e.g. in warehouses.
With its external sensor connections and its robust metal housing
the testo 176 H2 is able to execute temperature and humidity
measurements in parallel mode under extreme conditions.
The testo 176 P1 is able to measure pressure, temperature and
humidity at the same time and is thus optimally suitable for the
documentation of ambient conditions, e.g. in laboratories.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
35
3.2. Technical data
testo 176 T1 (0572 1761)
Feature Values
Measurement
parameter
Temperature (°C/°F)
Sensor type Pt100 class A internal
Measurement range
-35 to +70 °C
Instrument
accuracy
± 0.4 °C (-35 bis +70 °C) ± 1 digit
Resolution 0.01 °C
Operating
temperature
-35 ... +70 °C
Storage
temperature
-40 ... +85 °C
Battery type 1x lithium (TL-5903)
Life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP68 1m
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight approx. 410 g
Measuring cycle 1s – 24h (freely selectable, for online
measurement 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
Standards
2014/30/EC, EN 128304
Pos: 20 /TD/Leistungsbeschreibung/Technische Daten/testo 176 T2 @
6\mod_1278341523355_79.docx @ 63361 @ 5 @ 1
4Please note that, according to EN 12830, this instrument must be regularly checked and
calibrated as specified in EN 13486 (recommendation: every year) Contact us for more
information.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
36
testo 176 T2 (0572 1762)
Feature
Values
Measurement
parameter
Temperature (°C/°F)
Sensor type 2 x Pt100 class A external
Measurement range
-100 to +400 °C
Instrument
accuracy
± 0.2 °C (-100 bis +200 °C) ± 1 digit
± 0.3 °C (+200.1 bis +400 °C) ± 1 digit
Resolution 0.01 °C
Operating
temperature
-35 ... +70 °C
Storage
temperature
-40 ... +85 °C
Battery type 1x lithium (TL-5903)
Life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP65
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight approx. 220 g
Measuring cycle 1s – 24h (freely selectable, for online
measurement 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
Standards
2014/30/EC, EN 128305
Pos: 21 /TD/L ei st ung s bes chr eib ung /T ec hni sc h e Dat en/t es t o 17 6 T 3 @ 6\mod_1278341475870_79.docx@ 63287@ 5 @ 1
5Please note that, according to EN 12830, this instrument must be regularly checked and
calibrated as specified in EN 13486 (recommendation: every year) Contact us for more
information.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
37
testo 176 T3 (0572 1763)
Feature
Values
Measurement
parameter
Temperature (°C/°F)
Sensor type 4 thermocouples (type T, K, J) external
Measurement range
-100 to +750 °C (type J)
-200 to +1000 °C (type K)
-200 to +400 °C (type T)
Instrument
accuracy
± 1 % of the measurement value (-200
to -100.1 °C) ± 1 digit
± 0.3 °C (-100 bis +70 °C) ± 1 digit
± 0.5 % of the measurement value (+70.1 to
+1000 °C) ± 1 digit
Resolution 0.1 °C
Operating
temperature
-20 ... +70 °C
Storage
temperature
-40 ... +85 °C
Battery type 1x lithium (TL-5903)
Life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP65
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight approx. 430 g
Measuring cycle 1s – 24h (freely selectable, for online
measurement 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
EC Directive
2014/30/EC
Pos: 22 /TD/L ei st ung s bes chr eib ung /T ec hni sc h e Dat en/t es t o 17 6 T 4 @ 6\mod_1278341547926_79.docx@ 63393@ 5 @ 1
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
38
testo 176 T4 (0572 1764)
Feature
Values
Measurement
parameter
Temperature (°C/°F)
Sensor type 4 thermocouples (type T, K, J) external
Measurement range
-100 to +750 °C (type J)
-200 to +1000 °C (type K)
-200 to +400 °C (type T)
Instrument
accuracy
± 1 % of the measurement value (-200
to -100.1 °C) ± 1 digit
± 0.3 °C (-100 bis +70 °C) ± 1 digit
± 0.5 % of the measurement value (+70.1 to
+1000 °C) ± 1 digit
Resolution 0.1 °C
Operating
temperature
-20 ... +70 °C
Storage
temperature
-40 ... +85 °C
Battery type 1x lithium (TL-5903)
Life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP65
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight approx. 230 g
Measuring cycle 1s – 24h (freely selectable, for online
measurement 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
EC Directive
2014/30/EC
Pos: 23 /TD/L ei st ung s bes chr eib ung /T ec hni sc h e Dat en/t es t o 17 6 H 1 @ 6\mod_1278341570278_79.docx@ 63425@ 5 @ 1
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
39
testo 176 H1 (0572 1765)
Feature
Values
Measurement
parameter
Temperature (°C/°F) and humidity (%rF,
%RH, °Ctd, g/m³, WB)
Sensor type 2 NTC temperature sensors external or
2 capacitive humidity sensors external
Measurement range
0 to 100 %rF (non-dewing)
-20 to +70 °C
-40 to +70 °Ctd
Instrument
accuracy
± 0.2 °C (-20 to +70° C) ± 1 digit
± 0.4 °C (rem. measurement range) ± 1 digit
The accuracy for humidity measurements
results from the accuracy of the connected
sensor.
Resolution 0.1 °C, 0.1 %rF
Operating
temperature
-20 ... +70 °C
Storage
temperature
-40 ... +85 °C
Battery type 1x lithium (TL-5903)
Life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP65
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight
approx. 220 g
Measuring cycle 1s – 24h (freely selectable, for online
measurement 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
EC Directive
2014/30/EC
Pos: 24 /TD/L ei st ung s bes chr eib ung /T ec hni sc h e Dat en/t es t o 17 6 H 2 @ 6\mod_1278341606049_79.docx@ 63457@ 5 @ 1
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
40
testo 176 H2 (0572 1766)
Feature
Values
Measurement
parameter
Temperature (°C/°F) and humidity (%rF,
%RH, °Ctd, g/m³, WB)
Sensor type 2 capacitive humidity sensors external
Measurement range
0 to 100 %rF (non-dewing)
-20 to +70°C
-40 to +70°Ctd
Instrument
accuracy
± 0.2°C (-20 bis +70°C) ± 1 digit
± 0.4°C (remaining measurement range) ± 1
digit
The accuracy for humidity measurements
results from the accuracy of the connected
sensor.
Resolution 0.1 °C, 0.1 %rF
Operating
temperature
-20 ... +70°C
Storage
temperature
-40 ... +85°C
Battery type 1x lithium (TL-5903)
Service life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP65
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight
approx. 430 g
Measuring cycle 1s – 24h (freely selectable, for online
measurement 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
EC Directive
2014/30/EC
Pos: 25 /TD/L ei st ung s bes chr eib ung /T ec hni sc h e Dat en/t es t o 17 6 P 1 @ 6 \mod_1278341642475_79.docx@ 63489 @ 5@ 1
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
41
testo 176 P1 (0572 1767)
Feature
Values
Measurement
parameter
Temperature (°C/°F), humidity (%rF, %RH,
°Ctd, g/m³), barometric pressure (mbar, hPa,
psi, inH2O)
Sensor type 2 NTC temperature sensors external or
2 capacitive humidity sensors external
1 absolute pressure sensor internal
Measurement range
600 mbar to 1100 mbar
-20 to +70 °C
-40 to +70 °Ctd
0 to 100 %rF (non-dewing)
Accuracy ± 0.2 °C (-20 bis +70 °C) ± 1 digit
± 0.4 °C (remaining measurement range)
± 1 digit
± 3 mbar (0 to 50 °C) ± 1 digit
The accuracy for humidity measurements
results from the accuracy of the connected
sensor.
Resolution 0.1 °C / 0.1 %rF / 0.1 mbar
Operating
temperature
-20 ... +70 °C
Storage
temperature
-40 ... +85 °C
Battery type 1x lithium (TL-5903)
Service life 8 years (15 min. measuring cycle, +25 °C)
Degree of
protection
IP54
Dimensions in mm
(LxWxH)
103 x 63 x 33 mm
Weight approx. 230 g
Measuring cycle 1s – 24h (freely selectable, for online
measuremant 2s – 24h))
Interface Mini-USB, SD card slot
Memory capacity 2 million readings
EC Directive
2014/30/EC
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3 Specifications
42
Battery life
The programming windows of the software provide you with typical
guide values for the expected lifetime of the battery. This lifetime is
calculated on the basis of the following factors:
• Measuring cycle
• Number of connected sensors
Since the battery life depends on quite a few other factors, the
calculated data can only serve as guide values.
The following factors have a negative effect on the battery life:
• longer flashing of the LEDs
• frequent reading out (several times per day) via the SD-card
• extreme fluctuations in operating temperature
The following factors have a positive effect on the battery life:
• display switched off
The battery capacity reading in the display of the data logger is
based on the calculated values. However, the data logger is
switched off when a critical voltage level has been reached. It may
therefore happen that:
• readings are still recorded, even though the battery capacity
reading says "empty".
• the measurement program is stopped, even though the battery
capacity reading just before indicated a still remaining battery
capacity.
In case of an empty battery or a battery change saved readings will
not be lost.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

4 First steps
43
4First steps
4.1. Unlock the data logger
1. Open the lock with the key (1).
2. Remove the lock (2) from the locking pin.
3. Pull the locking pin (3) out of the holes in the wall bracket.
4. Slide the data logger out of the wall bracket (4).
The data logger is delivered with the battery (type TL-
5903) inserted. The display of the data logger shows
rSt
.
4.2. Connecting the data logger to PC
For testo Comfort Software Basic 5:
The software is available in the Internet as a free download
requiring registration:
The instructions for the installation and operation of the
software can be found in the
testo Comfort Software Basic
5 instruction manual, which can be downloaded together
with the software.
For testo Comfort Software Professional und testo Comfort
Software CFR:
1. Install the software testo Comfort Software.
2. Connect the USB cable to a free USB port on the PC.
3. Loosen the screw on the right side of the data logger.
4. Open the cover.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5 Display and control elements
44
5. Plug the USB cable into the Mini USB port (1).
6. Configure the data logger, see separate operating instructions
testo Comfort Software.
5Display and control elements
5.1. Display
The data loggers testo 176 T1, testo 176 T3, testo 176 H2
have no display.
The display function can be switched on/off via the testo
Comfort Software.
Depending on the operating status, various information may
be shown in the display. A detailed representation of the
information that can be called up can be found under Menu
overview.
For technical reasons the display speed of liquid crystal
displays becomes slower at temperatures below 0 °C
(approx. 2 seconds at -10 °C, approx. 6 seconds at -20
°C).
This has no influence on the measuring accuracy.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5 Display and control elements
45
testo 176 T2, testo 176 T4, testo 176 H1, testo 176 P1
1 Measurement value channel 1, 2, 3 (depending on number of
channels and view)
2 Units channel 1, 2, 3 (depending on number of channels and
view)
3 Measurement value channel 2, 3, 4 (depending on number of
channels and view)
4 Units channel 2, 3, 4 (depending on number of channels and
view)
5 Number of saved measurement values with alarm value
violation (Log)
6 Current intermediate measurement value, appears in the
display, but is not saved (Act)
7 End of measuring program (End), measuring program is
running (Rec)
8 Waiting for start of measuring program (Wait), start criterion
Formula programmed (F)
9. Start criterion Date/ Time programmed
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5 Display and control elements
46
10 Battery capacity
Icon
Capacity
> 151 days
< 150 days
< 90 days
< 60 days
< 30 days
> Read out data and change battery (see
Replacing the battery).
11 Lower alarm value displayed channel (2, 3, 4):
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were fallen short of
12 Channel number (2, 3, 4)
13 Upper alarm value displayed channel (2, 3, 4):
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were exceeded
14 Identification of measurement value:
Max: highest saved measurement value
Min: lowest saved measurement value
15 Lower alarm value displayed channel (1, 2, 3):
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were fallen short of
16 Channel number (1, 2, 3)
17 Upper alarm value displayed channel (1, 2, 3):
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were exceeded
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5 Display and control elements
47
5.2. LED
Representation Explanation
Red LED flashes once
every 10 seconds
Remaining battery capacity has dropped
below 30 days
Red LED flashes twice
every 10 seconds
Remaining battery capacity has dropped
below 10 days
Red LED flashes three
times every 10 seconds
Battery is empty:
Red LED flashes three
times when pressing
the button
Limiting value exceeded/fallen short of.
Yellow LED flashes
three times
Instrument changes from Wait-mode to
Rec-mode.
Yellow LED flashes
three times when
pressing the button
Instrument is in Rec-mode.
Green and yellow LED
flash three times when
pressing the button.
Instrument is in End-mode.
Green LED flashes
three times when
pressing the button
Instrument is in Wait-mode.
Red, yellow and green
LEDs flash one after
the other
Battery was inserted, capacitor in charge
state.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6 Using the product
48
5.3. Key functions
A detailed representation of the screen displays can be found
under Menu overview.
✓Instrument in operating status Wait and start criterion Button
start programmed.
> Press [GO] for approx. 3 seconds to start the measurement
program.
- The measurement program starts and Rec appears in the
display.
✓Instrument is in operating status Wait:
> Press [GO], to toggle between the displays of upper alarm
value, lower alarm value, battery lifetime and last measurement
value.
The displays appear in the specified sequence.
✓Instrument is in operating status Rec or End:
> Press [GO], to toggle between the displays for highest saved
measurement value, lowest saved measurement value, upper
alarm value, lower alarm value, number of times the upper
alarm value has been exceeded, number of times the lower
alarm value has been fallen short of, battery lifetime and last
measurement value.
The displays appear in the specified sequence.
Show current measurement value
✓The last 10 seconds [GO] has not be pressed.
> Press [GO].
- The current measurement value is determined and appears in
the display.
- The display shows Act.
If [GO] is pressed again within the next 10 seconds, the
current measurement value for the next channel will be
determined and displayed.
6Using the product
6.1. Connecting a sensor
Observe the following points when connecting sensors to data
logger and measuring points.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6 Using the product
49
> Ensure correct polarity of the plugs.
> Press the plugs firmly into the ports to ensure leak tightness.
However, do not apply force!
> Make sure that the plugs are firmly connected to the data logger
or that the connections are closed with blanking plugs.
> Ensure correct positioning of the sensor to avoid disturbing
influences affecting the measurement.
> testo 176 T2, testo 176 T3, testo 176 T4, testo 176 H1, testo
176 H2, testo 176 P1: Make sure that you connect the
configured (with the testo Comfort Software) sensors to the
individual connections. The numbers of the connections are
printed on the housing.
6.2. Programming data logger
In order to adapt the programming of your data logger to your
individual requirements, you require the testo Comfort Software
Basic 5 software. It is available in the Internet as a free download
requiring registration ,
Service&Support | Download Center.
The instructions for the installation and operation of the
software can be found in the testo Comfort Software Basic
5 instruction manual that is downloaded together with the
software.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
Other manuals for 176
1
This manual suits for next models
1
Table of contents
Other TESTO Data Logger manuals