Labomed Magna User manual

MAGNA
User Manual
Operating Microscope Dentistry
Part No.: 6129000-795
Issue 1.7
Printed on May,2020
To ensure proper use of this instrument as well as to avoid injury while operating instrument,
Understanding this manual completely before use is highly recommended.

MAGNA
6129000-795Issue 1.7
Printed on May,2020
Magna
MAGNA is a Trade name for LABOMED endodontics microscope.
LABOMED is a registered trademark of Labo America, Inc.
All other trademarks are the property of their respective owners.
The information contained in this document was accurate at the time of publication. Specifications are subject to change
without prior notice. LABOMED reserve the right to make changes to the product described in this user manual without
notice and without incorporating those changes in any products already sold.
ISO 13485 Certified - LABOMED products are designed and manufactured under quality processes that meet ISO
13485 requirements.
No part of this publication may be reproduced, stored in a retrieval system, or transmitted in any form or by any
means: electronic, mechanical, recording, or otherwise, without the prior written permission from LABOMED.

Magna
6129000-795Issue 1.7
Printed on May,2020
MAGNA
LIST OF CONTENTS
1.
2.
3.
4.
5.
5A.
5B.
5C.
6.
7.
8.
8A.
8B.
9.
10.
11.
12.
13.
14.
14.1.
14.2.
14.3.
14.4.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
1
2
3
4 - 9
10 - 11
12
13 - 14
15 - 19
20 - 22
23 - 24
25 - 26
28
29 - 30
31
32
33
34
35 - 36
37 - 38
39
40 - 41
42 - 43
44 - 45
46
46
47
48- 49
50
50
51
52
52
53
54 - 55
56 - 59
60
61
62 - 63
Introduction, Intended Use and Contraindications
Product Description, Floor Stand
Product Description, Ceiling Mount
Installation References (Ceiling Mount)
Ceiling Mount Installation
Drop Ceiling Applications
Leveling The Ceiling Mount System
Installing Ceiling Mount On Concrete Ceilings
Product Description, Wall Mount
Wall Mount Installation
Wall Mount Installation Procedure
Wall Mount Installation on Wooden Wall
Wall Mount Installation on Masonry Wall
Construction Requirements (Ceiling / Wall Mount)
Warning and Cautions
Explanation of Symbols
Standards and Directives
Unpacking
Installation of Cross Base (Mobile Stand)
Installation of Microscope (Mobile Stand)
Electrical Connections
Wiring Coding Diagram
Control Elements
Using the Microscope
Use of Accessories
Thermal Cut-off
Tension Adjustment
Moving & Storage the Instrument
Care and Maintenance
Cleaning and Disinfection
Autoclaving
Schedule of Autoclavable Caps
Ambient Requirement
Disposal
Technical Specification
Troubleshooting Table
Guidance Tables
Dimensions and Weight
Glossary
Warranty

Magna
MAGNA
1. INTRODUCTION AND INTENDED USE
Issue 1.7
Printed on May,2020
1
6129000-795
LABOMED Magna is a surgical and diagnostic microscope that is adaptable for different surgical needs for
consistent visualization during all intra operative phases of endodontics by providing a magnified view of the
surgical field without compromising performance.
The microscope provides extremely high optical image quality, good depth of focus, and a wide field of view for
precise surgery. Illumination control and high eye point of the observation head helps reduce the user’s work
fatigue and allows comfortable use over a long period.
Salient features of this microscopes are:
1. The observation head can easily be positioned with the help of a suspension arm.
2. An advanced Galilean Zoom 1:6, also convertible to 8 step magnichanger allows optimal magnification
for various needs at different magnification levels.
3. Cold light illumination with a high intensity 50W LED lamp is provided using a fiber optic guide for
proper illumination.
4. An effortless articulation system ABA (Automatic Balancing Arm) features an electromagnetic brake system
that enables the ultimate flexibility in maneuvering the microscope’s viewing angle and position.
5. The Labomed NuVar CMO with variable working distance objective comes standard, and provides user
greater convenience in achieving a comfortable working distance.
6. The rigid Cross base with caster wheels provides the instrument with stability and mobility.
7. When the microscope is not in use, the suspension arm can be folded over the main body for compact storage.
INTENDED USE
The Magna microscope is an AC powered device intended for use during diagnosis and surgery to provide a
magnified view of the region of interest.
Note:
Use Magna as specified in the intended use.
CONTRAINDICATIONS - NONE
CONFIGURATIONS
Microscope Catalogue No.
Magna Floor Stand (std arm)
Magna Ceiling Mount (std arm)
Magna Wall Mount (std arm)
Magna Ceiling Mount long arm
Magna Wall Mount long arm
Magna Floor Stand with motorised filter (std arm)
Magna Ceiling Mount with motorised filter (long arm )
Magna Wall Mount with motorised filter (long arm )
Magna Ceiling Mount (std arm)with motorised filter
Magna Wall Mount (std arm) with motorised filter
Ceiling Mount Options (Heavy Type / 170 mm Dia)
Ceiling Height
Column Length (Microscope)
10 Ft (305 cm) 2.4 Ft (740 mm)
12 Ft (366 cm) 4.4 Ft (1350 mm)
6129005
6129006
6129007
6129008
6129009
6129010
6129011
6129012
6129013
6129014

Magna
MAGNA
2. PRODUCT DESCRIPTION (FLOOR STAND)
6129000-795
1
2
3
4
5
6
7
22 8
9
10
11
12
15
16 17
20
21
23
24
25
19
18
14
13
2
Issue 1.7
Printed on May,2020
16. Suspension Arm Hydraulic Movement Lock
17. Suspension Arm
18. Suspension Arm Spring Tension Adjustment
19. Suspension Arm Movement Locking Knob
20. Swivel Arm
21. 7 pin connector for ABA Arm
22. Left Hand Release (Electromagnetic Clutch)
23. Wire Mesh
24. Fibre Light Guide
25. Filter Change knob
1. Caster wheel with Brakes
2. Cross Base Top
3. Cross Base Bottom
4. Column
5. Hand Rest
6. Swivel Arm Locking Knob
7. Common Main Objective with Fine Focusing
8. Right hand Release (Electromagnetic Clutch)
9. Eyepieces
10. Ergo Head 0 - 210° adjustable
11. Zoom Magnification Knob
12. 2 pin connector for Clutch Buttons
13. Illumination Control Knob
14. ABA (Auto Balancing Arm) Electromagnetic Clutch
15. Auto Balancing Arm Locking Knob
Figure-1

1. Column
2. Cover Mount (to be used in false ceiling only)
3. Swivel Arm
4. Suspension Arm
5. Suspension Arm Hydraulic Movement Lock Knob
6. Auto Balancing Arm Lock Knob
7. ABA (Auto Balancing Arm) Magnetic Clutch
8. Illumination Control knob
9. Ergo Head 0 - 210˚ adjustable
10. Eyepieces
11. Right Hand Release (Electromagnetic Clutch)
12.
Zoom Magnification Knob
13. Common Main Objective with Fine Focusing
14. Left Hand Release (Electromagnetic Clutch)
15. Master Plate
MAGNA
3. PRODUCT DESCRIPTION (CEILING MOUNT)
6129000-795Magna
3
Issue 1.7
Printed on May,2020
1
15
2
3
5
6
7
8
910
11
12
14 13
4
Figure-2

MAGNA
6129000-795Magna
4
Issue 1.7
Printed on May,2020
4. INSTALLATION REFERENCES (CEILING MOUNT)
CEILING MOUNT SYSTEM FOR 12 FEET CEILING HEIGHT
1064
500 1000 150
2309
3660 (12')
1351 (4.43')
360°
310°
360°
220°
170mm
Figure-3

MAGNA
6129000-795Magna
5
Issue 1.7
Printed on May,2020
INSTALLATION REFERENCES (CEILING MOUNT)
CEILING MOUNT SYSTEM FOR 11 FEET CEILING HEIGHT
1064
500 1000 150
2309
3352 (11')
1043 (3.42')
360°
310°
360°
220°
170mm
Figure-4

MAGNA
6129000-795Magna
6
Issue 1.7
Printed on May,2020
INSTALLATION REFERENCES (CEILING MOUNT)
CEILING MOUNT SYSTEM FOR 10 FEET CEILING HEIGHT
1064
500 1000 150
2309
3050 (10')
741 (2.4')
360°
310°
360°
220°
170mm
Figure-5

MAGNA
6129000-795
7
Marking References
Issue 1.7
Printed on May,2020
Magna
INSTALLATION REFERENCES (CEILING MOUNT)
Master Plate
Spacer
Ceiling Mount Column
Ø168MM
175mm175mm
175mm175mm
175mm175mm
175mm175mm
400mm
400mm
Figure-7
MASTER PLATE
Figure-6

MAGNA
6129000-795Magna
8
Issue 1.7
Printed on May,2020
INSTALLATION REFERENCES (CEILING MOUNT)
Operator Position.
Patient is right below the
microscope optical head.
R1800mm
1385.3mm
694mm
1199mm
Figure-8

MAGNA
6129000-795Magna
9
Issue 1.7
Printed on May,2020
INSTALLATION REFERENCES (CEILING MOUNT)
Operator Position.
Patient is right below the
microscope optical head.
694mm
1198mm
Figure-9

Magna
MAGNA
6129000-795
10
Issue 1.7
Printed on May,2020
5A. DROP CEILING APPLICATIONS
Many newer constructions use drop ceilings. Whenever this type of application is used, all supporting
structure and electrical service (115/230 VAC, 15A ) is the customer's responsibility and must be completed
before the installation.
WARNING: THE SUPPORTING STRUCTURE FOR THE CEILING MOUNT AND WALL MOUNT INSTALLATION
MUST BE CAPABLE OF SUPPORTING A MINIMUM WEIGHT OF 600 LBS (272 kg) AND BE STABLE
LATERALLY AND VERTICALLY, AND FREE FROM VIBRATION.
WARNING:
• CEILING MOUNT PLATE AND COLUMN ASSEMBLY TOGETHER CAN WEIGHTS APPROXIMATELY UPTO
66LBS (30KG) WITH MICROSCOPE.
•TO AVOID PERSONAL INJURY TWO (2) PEOPLE ARE REQUIRED FOR PROPER INSTALLATION.
PROCEDURE:
1. Ensure that there is a 350mm diameter hole in the drop ceiling to mount the master Ceiling plate.
2. Labomed Ceiling Mount system for wooden ceiling can be installed by fixing a flat wooden membrane
at least 90mm thick to the joist and then mount the Master Ceiling Plate to the wooden membrane as
shown in Fig.10 and Fig.-11.
3.Using the Paper template provided, mark the positions of the six (6) mounting holes on wooden
membrane refering figure.
4. Drill six (6) 1/4”(6mm) pilot holes either direct into Joist or the additional wooden membrane as per the
roof available.
5. Place a 10mm flat washer over each of the six M10 X 100mm hex head lag screws.
6. Keeping welded bosses clown side, line up the holes in the plate with the pilot holes in the mounting
surface.
7. Insert the hex head lag screws into the holes and tighten the master Plate with a 17mm Hex Socket
Spanner.
8. Position the ceiling Mount Provided against the master plate and insert 6 socket head bolts M12 X 40
after aligning the 6 mounting holes.
9. Tighten all the bolts using Allen key 10mm slightly.
10. Use a level to ensure the column assembly is Levelled vertically. Refer Section “Leveling the ceiling
Mount system” to correct any leveling concerns.
11. All the jobs from Part 1 to 10 here are the responsibility of the customer to provide. Labomed represen
tative will complete rest of the microscope Installation after this.
12. Plug the support system power cord into the hospital grade earth-ground AC outlet, and verity that
there is power to the system.
13. Put the drop ceiling tile back into place.
5. CEILING MOUNT INSTALLATION

Magna
MAGNA
6129000-795
11
Issue 1.7
Printed on May,2020
Wooden membrane
Heavy Duty vibration free Ceiling Mount
Joist
Master Plate
Shim
Column
Level
Joist
Master Plate
Shim
Column
Level
Normal Ceiling Mount
Wooden membrane
Figure-10
Figure-11

Magna
MAGNA
6129000-795
12
Issue 1.7
Printed on May,2020
5B. LEVELING THE CEILING MOUNT SYSTEM
It is important to ensure the column for the ceiling mount is Plumb after installation. The column must be
perfectly vertical to prevent the microscope system from drifting from side to side.Use following procedure:
1. Use the 10mm Allen key to slightly loosen all of the 6M10 bolts.
2. Place a level across the bottom of ceiling mount as shown in Figure and to check the Vertical leveling.
Use Shims as required and adjust the RAWL Bolts/ Lag bolts until the column is Vertical.
3. Tighten all screws securely.
NOTICE:Shims and Shim materials are to be provided by the customer and are not furnished by
Labomed. There are numerous shims and shim materials commercially available and no specific type or
brand name is required.
Heavy Duty vibration free Ceiling Mount Normal Ceiling Mount
Figure-13 Figure-14
Joist
Master Plate
Shim
Column
Level
Wooden membrane
Wooden membrane
Joist
Master Plate
Shim
Column
Level

Magna
MAGNA
6129000-795
13
Issue 1.7
Printed on May,2020
5C. INSTALLING CEILING MOUNT ON CONCRETE CEILINGS
Fixing of Master Plate:
1. Use the Paper template provided to mark the position for the anchors of Master Plate.
2. Drill the Six (6) mounting holes 14mm diameter and 100mm deep.
3. Place a 10mm flat washer and a split washer over each of the six M10 X 75 hex head RAWL Plug.
4. Keeping welded bosses downside, line up the holes of Master plate with the pilot holes in the ceiling.
5. Insert the hex head RAWL screws into the holes and tighten the Master Plate with a 17mm hex socket
spanner.
Installation of Ceiling Mount:
6. Position the ceiling Mount Provided against the master plate and insert 6 socket head bolts M12 X 40
after aligning the 6 mounting holes.
7. Tighten all the bolts using Allen key 10mm slightly.
8. Use a level to ensure the column assembly is Leveled vertically. Refer Section “Leveling the ceiling
Mount system” to correct any leveling concerns.
9. All the jobs from Part 1 to 8 here are the responsibility of the customer to provide. Labomed representa-
tive will complete rest of the microscope Installation after this.
10. Plug the support system power cord into the hospital grade earth-ground AC outlet, and verity that
there is power to the system.
11. Put the drop ceiling tile back into place.
Concrete Ceiling
Master Plate
Column
Level
Normal Ceiling MountHeavy Duty vibration free Ceiling Mount
Concrete Ceiling
Master Plate
Shim
Column
Level
Shim
Figure-15 Figure-16

MAGNA
6129000-795Magna
14
Issue 1.7
Printed on May,2020
INSTALLATION REFERENCES (CONCRETE CEILING MOUNT)
LIST OF CONTENTS:
•Rawl Plug with Cap (Ceiling- 6pcs./ Wall- 6pcs) -
•Washers (Ceiling - 6pcs / Wall - 6pcs) -
•Acrylic Plate -
•Hole Mounting Template (Ceiling) -
•Ceiling Mount Cover -
•Column -
•Hole Mounting Template (Wall) -
•M4x12 Pan Head Screws for Acrylic Plate- 10 pcs -
•Threaded Spacer - 12 pcs -
•M10x40 Socket Head Cap Screw- 4pcs -
•M6x10 Socket Set Screw- 3pcs -
RAVAL PLUG M10x75mm
SS-854 (8-NOS.)
TO BE FIXED IN CONCRETE CEILING
BOLT M1Ox75mm
SS-854 (8-NOS.)
FIXED IN RAVAL PLUG
SS-666
M4x10 PH. PAN HEAD SCREW
(10-NOS.)
BOLT M12x50mm
SS-994 (6-NOS.)
SPRING LOCK WASHER 12MM
SS-995 (6-NOS.)
18.2700
Figure-17
Part No. SS- 854
Part No. 6134010 - 451
Part No. 6129001 - 043
Part No. 6129001 - 404
Part No. 6137200 - 014
Part No. 6129001 - 117
Part No. 6129002 - 403
Part No. SS - 666
Part No. 6129001 - 122
Part No. SS - 707
Part No. SS - 174

Magna
MAGNA
6. PRODUCT DESCRIPTION (WALL MOUNT)
6129000-795
13
14
12
15
11
10
9
8
7
6
5
4
3
2
1
Issue 1.7
Printed on May,2020
1. Wall Mount Bracket
2. Mount Shaft
3. Swivel Arm
4. Suspension Arm
5. Suspension Arm Hydraulic Movement Lock Knob
6. Auto Balancing Arm Lock Knob
7. ABA (Auto Balancing Arm) Magnetic Clutch
8. Illumination Control knob
9. Ergo Head 0 - 210˚ adjustable
10. Eyepieces
11. Right Hand Release (Electromagnetic Clutch)
12. Zoom Magnification Knob
13.
Common Main Objective with Fine Focusing
14. Left Hand Release (Electromagnetic Clutch)
Figure-18

MAGNA
INSTALLATION REFERENCES (WALL MOUNT)
Operator Position.
Patient is right below the
microscope optical head.
Magna
6129000-795
16
Issue 1.7
Printed on May,2020
Figure-19

Magna
6129000-795
17
Issue 1.7
Printed on May,2020
MAGNA
INSTALLATION REFERENCES (WALL MOUNT)
Rotation of Suspension Arm about
the Horizontal Arm on both the
Ceiling Mount and the floor stand is
360°.
Rotation of Horizontal Arm about the
Column on the Wall Mount is limited
with the wall acting as a stop.
Figure-20
Other manuals for Magna
1
This manual suits for next models
1
Table of contents
Popular Dental Equipment manuals by other brands

W&H
W&H 945 Instructions for use

Nevadent
Nevadent SMD 24 A1 Operating instructions and safety instructions

Monitex
Monitex BlueLex LD-105 user manual

Bego
Bego Otoflash operating instructions

ROLENCE ENTERPRISE
ROLENCE ENTERPRISE Ultra-Lite 1000E DUAL user manual

Bien Air
Bien Air Lubricare 2 HIM-2 Instructions for use