LaMotte WaterLink Spin Touch User manual

QUICK
START GUIDE
BEFORE FIRST USE
¬Fully charge battery using
the Spin Touch USB cable and
adapter plug in AC outlet.
¬Ensure your Spin Touch has
the latest firmware. A USB
connection to a Windows® PC is
required:
1. Download and install the
WaterLink Connect 2
application for Windows
at http://softwarecenter.
lamotte.com/.
2. Plug meter into the computer
with the provided USB
cable and launch WaterLink
Connect 2 application from
the Start Menu.
3. A prompt will appear if
firmware updates are
available. Select Update.
Testing and data transfer will
not be possible until firmware
has been updated.
NOTE: For water samples over 100 °F (38 °C) subtract 0.3 from pH result or, for the most accurate result, wait until
water sample is below 90 °F (32 °C) to test.
Remove salt residue daily. Salt will damage the meter and cause inaccurate results. See CLEANING section in manual.
TESTING
1. Press and hold until the meter turns on.
2. Tap . Select a sanitizer type. Tap to confirm.
3. Tap . Select a disk series (found on disk packaging). Tap to confirm. NOTE: Disk Series
are limited by Sanitizer Type selection.
4. Remove disk from packaging.
5. Use syringe (1189) to fill disk with pool or spa water sample.
6. Insert disk. Cover disk with Universal Disk Cover (1719). Close lid.
7. Tap to start test. Tap to cancel the test. If the test is cancelled discard the disk.
8. Results will be displayed.
9. Select an option.
- Tap to add a tag.
- Tap the highlighted to save the test results to the test log if Auto Save is not enabled.
- Tap the hightlighted to send the results to the enabled Mobile Bluetooth Printer.
- Tap the highlighted to send the results to a Bluetooth enabled device.
- Tap to return to the Test Screen.
10. Press and hold for 2 seconds to turn meter off.
FILLING: Single-Use Disks
When the syringe is placed in the water
sample, and the plunger is pulled all the way
up, the syringe will hold more than enough
water sample to adequately fill the disk. Hold
the syringe vertically and insert the tip into the
fill hole in the disk. Press the plunger slowly
and smoothly to fill the disk.
1. Fill the disk using slow, even
pressure. The sample water will fill
the spaces between the baffles in a
counterclockwise order. Each space will
fill from the bottom to the top. Sample
water should be added until the sample
water in the fourth chamber fills to the
top of the chamber slightly past the
embossed fill line. It is OK to fill slightly
past the fill line.
Correct Fill
(Fill line)
2. Do not overfill the disk. If the disk is
overfilled, sample water will flow out of the
overflow hole in the center of the disk. The
disk is not leaking. Dry the disk and run the
test.
Overfilled
(Dry disk before placing in meter)
3. Do not under fill the disk. If the disk
is under filled, the reagent chambers
will not fill entirely and results will be
inaccurate.
Underfilled
4. Do not introduce air bubbles into the disk.
The reagent chambers will not fill entirely
and results will be inaccurate. As soon as
a bubble starts to form, pull back on the
plunger to draw the bubble out of the disk.
Begin the filling process again.
Air
Bubble
5. Wet disks should be dried thoroughly
with a lint free wipe. The disk should be
handled by the edges.
6. Disks should be filled and used within 10 minutes. They cannot be
filled ahead of time. For filling and troubleshooting tips go to:
www.lamotte.com/spin_support/
Compatible with iPhone 6 Plus, iPhone 6, iPhone 5s,
iPhone 5c, iPhone 5, iPad Air 2, iPad Air, iPad mini 3,
iPad mini 2, iPad mini, and iPad (4th generation).
Compatible with Android devices.
[email protected]
410-778-3100 | 800-344-3100 | www.lamotte.com
Disk US Patent No. 8,734,734
FCl US Patent No. 8,987,000
FCl EU Patent No. EP2784503 A1
TCl US Patent No. 8,993,337
3580-QG | 11.04.21
Certified to
NSF/ANSI/CAN 50
FILLING: Multiple-Use Disks
When the syringe is placed in the water
sample, and the plunger is pulled all the
way up, the syringe will hold more than
enough water sample to adequately fill
the disk. Hold the syringe vertically and
insert the tip into the fill hole in the disk.
Press the plunger slowly and smoothly to
fill the disk.
1. Fill any unused section of the disk
using slow, even pressure. Add
the sample water until the Blank
Chamber is full and the Over Flow
Chamber is partially filled.
Correct Fill
Blank
Chamber
Over Flow
Chamber
2. Do not overfill the disk. If the disk is
overfilled, sample water will flow out
of the vent hole in the center of the
disk. The disk is not leaking. Dry the
disk and run the test.
Overfilled
(Dry disk before placing in meter)
Blank
Chamber
Over Flow
Chamber
3. Do not under fill the disk. If the disk
is under filled, the blank chamber will
not be full. The reagent chambers
will not fill entirely and results will be
inaccurate.
Underfilled
Blank
Chamber
4. Do not introduce air bubbles into
the disk. The reagent chambers
will not fill entirely and results will
be inaccurate. As soon as a bubble
starts to form, pull back on the
plunger to draw the bubble out of the
disk. Begin the filling process again.
Air Bubble
5. Wet disks should be dried thoroughly
with a lint free wipe. Handle the disk
by the edges. Between tests store
the disk on a flat surface. Do not
agitate a partially filled disk to avoid
leaking and contamination. Partially
used disks should be discarded at
the end of the day.
6. Disks should be filled and used within 10 minutes. They cannot be filled
ahead of time.
For filling and troubleshooting tips go to: www.waterlinkspintouch.com
See the WaterLink Spin Touch manual for additional information.
DOS & DON’TS
¬Hold syringe vertically when filling disks.
¬Do not fill disk while in the meter. Fill disk on clean, dry surface.
¬The disk should not contain any large air bubbles. Air bubbles will
result in erroneous results.
¬Only the Universal Disk Cover (Code 1719) can be used with the
WaterLink Spin Touch®.
¬Empty syringe of previous sample before filling with next
sample.
¬Remove filled disk from meter after testing.
¬Store disks at 70°-80°F/21°-27°C.
¬Keep the chamber clean and dry. Gently swab LED and
photodiode lenses located around the hub with a cotton swab
dampened with streak-free window cleaner. Do not use alcohol.
It will leave a thin film over the lenses when dry.
CONNECTING VIA BLUETOOTH
The WaterLink Spin Touch is capable of connecting to a
Bluetooth enabled device, such as a phone or tablet. The Spin
Touch is also compatible with a BLE Mobile Printer (5-0067).
Other Bluetooth printers are not supported.
Transferring results over Bluetooth to a mobile device requires
that a mobile app is installed on the device and an active
account for the associated software program. For example,
with an active WaterLink Solutions™account and the WaterLink
Solutions mobile app on a phone or tablet, results can be
transferred from the meter into the application. Mobile apps for
LaMotte software products are available from iTunes (for iOS
devices) and Google Play (for Android devices).
For detailed information read the Spin Touch manual. Go to
http://softwarecenter.lamotte.com/ to learn more about
LaMotte software products.
LaMotte, WaterLink, Spin Touch are ® registered trademarks of LaMotte Company | © 2021 LaMotte Company, All Rights Reserved

Other manuals for WaterLink Spin Touch

4

Other LaMotte Measuring Instrument manuals

LaMotte DC1500 SULFATE User manual

LaMotte

LaMotte DC1500 SULFATE User manual

LaMotte DC1500 CHLORINE User manual

LaMotte

LaMotte DC1500 CHLORINE User manual

LaMotte SMART Colorimeter User manual

LaMotte

LaMotte SMART Colorimeter User manual

LaMotte ColorQ Pro-11 User manual

LaMotte

LaMotte ColorQ Pro-11 User manual

LaMotte COD3 Plus User manual

LaMotte

LaMotte COD3 Plus User manual

LaMotte Tracer 1741 User manual

LaMotte

LaMotte Tracer 1741 User manual

LaMotte DC1500 Copper User manual

LaMotte

LaMotte DC1500 Copper User manual

LaMotte WaterLink Spin Touch DW User manual

LaMotte

LaMotte WaterLink Spin Touch DW User manual

LaMotte DC1500 CHLORINE User manual

LaMotte

LaMotte DC1500 CHLORINE User manual

LaMotte SMART3 User manual

LaMotte

LaMotte SMART3 User manual

LaMotte WaterLink Spin Touch DW User manual

LaMotte

LaMotte WaterLink Spin Touch DW User manual

LaMotte DC1500 CHLORINE User manual

LaMotte

LaMotte DC1500 CHLORINE User manual

LaMotte 1705-EU User manual

LaMotte

LaMotte 1705-EU User manual

LaMotte Tracer 1749-KIT-01 User manual

LaMotte

LaMotte Tracer 1749-KIT-01 User manual

LaMotte DC1500 CHLORINE User manual

LaMotte

LaMotte DC1500 CHLORINE User manual

LaMotte pH 5 Plus User manual

LaMotte

LaMotte pH 5 Plus User manual

LaMotte CON 6 PLUS User manual

LaMotte

LaMotte CON 6 PLUS User manual

LaMotte WaterLink SpinTouch 3580-AU User manual

LaMotte

LaMotte WaterLink SpinTouch 3580-AU User manual

LaMotte DC1500 CHLORINE User manual

LaMotte

LaMotte DC1500 CHLORINE User manual

LaMotte PockeTester 1766 User manual

LaMotte

LaMotte PockeTester 1766 User manual

LaMotte 1200-MN User manual

LaMotte

LaMotte 1200-MN User manual

LaMotte DC1500 FLUORIDE User manual

LaMotte

LaMotte DC1500 FLUORIDE User manual

LaMotte WaterLink Spin User manual

LaMotte

LaMotte WaterLink Spin User manual

LaMotte TRACER POCKETESTER TDS/SALT/TEMP User manual

LaMotte

LaMotte TRACER POCKETESTER TDS/SALT/TEMP User manual

Popular Measuring Instrument manuals by other brands

WPI 900A instruction manual

WPI

WPI 900A instruction manual

Fluidwell F070 manual

Fluidwell

Fluidwell F070 manual

Nexas NexLink NL101 user manual

Nexas

Nexas NexLink NL101 user manual

Bender ISOMETER isoGEN523-S4-4 manual

Bender

Bender ISOMETER isoGEN523-S4-4 manual

Troxler Enhanced RoadReader Plus 3451 Manual of Operation and Instruction

Troxler

Troxler Enhanced RoadReader Plus 3451 Manual of Operation and Instruction

Desco SCS-73627 Installation, operation and maintenance

Desco

Desco SCS-73627 Installation, operation and maintenance

Hanna Instruments HI97769 instruction manual

Hanna Instruments

Hanna Instruments HI97769 instruction manual

Teledyne Lecroy Frontline Sodera LE quick start guide

Teledyne Lecroy

Teledyne Lecroy Frontline Sodera LE quick start guide

Keysight Technologies N9010B Installation note

Keysight Technologies

Keysight Technologies N9010B Installation note

Endress+Hauser Proline Promass S 500 operating instructions

Endress+Hauser

Endress+Hauser Proline Promass S 500 operating instructions

AFRISO Unitop AdBlue operating instructions

AFRISO

AFRISO Unitop AdBlue operating instructions

LINSHANG LS136 user manual

LINSHANG

LINSHANG LS136 user manual

Extech Instruments 341350A user guide

Extech Instruments

Extech Instruments 341350A user guide

YOKOGAWA UPM101 user manual

YOKOGAWA

YOKOGAWA UPM101 user manual

olympia electronics BS-634 user manual

olympia electronics

olympia electronics BS-634 user manual

Elenco Electronics K-38 Assembly and instruction manual

Elenco Electronics

Elenco Electronics K-38 Assembly and instruction manual

sauter CP P3 instruction manual

sauter

sauter CP P3 instruction manual

Datakom DKM-411 Applications manual

Datakom

Datakom DKM-411 Applications manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.