
Operation Instruction
1= Cover plate
2= Prism
3= Calibration screw
4= Mirror tube
5= Eyepiece (focus adjustment)
It is important to regulate the temperature 
of the sample, the instrument and the 
room as much as possible. Changes in 
temperature can affect the final result.
Temperature
Specification
Measuring Range 0-40%
Resolution 0.2
Measurements 27mm x 40mm x 160mm
Weight 175g You may need to calibrate the instrument 
before use or after a rise or fall in 
temperature.
1. Open the cover plate (2).
2. Drop 1/2 drops of distilled water onto 
the prism (1).
3. Adjust the meter to zero (0).
4. Close the cover plate (2) and press 
gently.
5. Rotate and adjust the calibration screw 
(3) until the light blue boundary line 
meets the zero (0) line.
Calibration
1. Take a sample of the Adblue mixture
2. Place a droplet on the prism and cover
3. Look through the eye-piece
4. Adjust the light if necessary to focus the 
line between the dark and light areas.
5. Line this up with the marked 
measurement scale
6. Take the reading
Measurement
• Use only soft tissue or cloth to clean 
the prism as it is very easily scratched.
• Dab away the sample and then use 
distilled water or a glass-safe solvent 
such as alcohol to clean the prism.
Cleaning
• Adjustment for the distilled water 
and sample should be in the same 
temperature. 
• Once the temperature changes, the 
zero (0) point should be adjusted once 
per 30 seconds.
• Do not immerse in water
• This is a precision optical instrument 
and should be handles gently 
• Do not touch or scratch the optical 
surface. 
• Please keep it in a dry, clean and non-
corrosiveness environment
• Avoid dropping 
Attention and Maintenance
1
2
3
4
5
00
55
10 10
15 15
20 20
25 25
30 30
35 35
40 40
AUS 32 20°C
[68°F]
00
55
10 10
15 15
20 20
25 25
30 30
35 35
40 40
AUS 32 20°C
[68°F]
Calibration with distilled water
5015_RefractometerAdblueInst.indd 25015_RefractometerAdblueInst.indd 2 09/08/2010 09:5209/08/2010 09:52