Lauda Varioshake VS 60 OI User manual

Operating Instructions
Varioshake Shaking Incubator
VS 60 OI

2 / 26 Varioshake Shaking Incubator 09/2020

09/2020 Varioshake Shaking Incubator 3 / 26
The Shaking Incubator LAUDA Varioshake VS 60 OI is universally applicable for all tasks in standard and research laborato-
ries that require exactly reproducible orbital motions and temperatures and a clear view of the cultures, for example in incu-
bations, fermentations, homogenisations, chemical and biochemical reactions, enzyme and tissue studies.
Temperature and shaking frequency are microprocessor-controlled. The temperature can be set within a range of 20 °C
(from approx 8 K above ambient) to 70 °C. The shaking frequency can be set from 20 min-1 to max 250 min-1 and oper-
ates with a fixed amplitude of 30 mm. The micro-processor-controlled timer continuously displays the current remaining
incubation time and emits an acoustic signal at end of time.
Control panel with main switch as well as display and operating elements of the timer and the electronic regulators for interi-
or temperature and shaking frequency.
Before installation, please check whether contents of package are in good order and complete. Should you note any damag-
es or have any reasons for complaint, please contact your supplier or directly:
LAUDA-GFL Gesellschaft für Labortechnik mbH
Schulze-Delitzsch-Str. 4+5
30938 Burgwedel - Germany
Telefon: +49 (0)5139 9958 0
Fax: +49 (0)5139 9958 21
E-Mail: info@lauda-gfl.de
Internet: www.gfl.de

4 / 26 Varioshake Shaking Incubator 09/2020

09/2020 Varioshake Shaking Incubator 5 / 26
Content
Operating Instructions ...........................................................................................................................................................................1
1Use of the Shaking Incubator........................................................................................................................................................7
Intended Use..........................................................................................................................................................................7
Improper Use .........................................................................................................................................................................7
2Warranty .........................................................................................................................................................................................7
3Before installation ..........................................................................................................................................................................7
4Transport, Assembly and Location of the Shaking Incubator ....................................................................................................8
5Operating Voltage..........................................................................................................................................................................8
6Inbetriebnahme ..............................................................................................................................................................................8
Operation ...............................................................................................................................................................................9
Cooling Coil............................................................................................................................................................................9
Operation and display elements of the control panel........................................................................................................9
6.3.1
Incubation time .......................................................................................................................................................... 10
6.3.2
Incubation temperature............................................................................................................................................. 11
6.3.3
Shaking frequency...................................................................................................................................................... 11
RC Operation via PC......................................................................................................................................................... 12
Shutdown............................................................................................................................................................................. 12
7Error Codes and Temperature Cut-out .................................................................................................................................... 12
Error Codes......................................................................................................................................................................... 12
Temperature Cut-out ........................................................................................................................................................ 13
8Functional description ................................................................................................................................................................ 13
9Maintenance, Support and Repair............................................................................................................................................. 13
Cleaning and Disinfection.................................................................................................................................................. 13
Temperature Adjustment................................................................................................................................................... 14
Checking the Acrylic Glass Cover .................................................................................................................................... 14
Shutdown caused by overload or power failure ................................................................................................................ 14
Exchanging the fuses.......................................................................................................................................................... 14
Maintenance Procedures................................................................................................................................................... 14
10 Disposal of Old Units.................................................................................................................................................................. 15
11 Technical Data ............................................................................................................................................................................. 16
Varioshake Shaking Incubator VS 60 OI ......................................................................................................................... 16
12 Circuit Diagram........................................................................................................................................................................... 17
Varioshake Shaking Incubator VS 60 OI ......................................................................................................................... 18
13 Connection to the mains supply ................................................................................................................................................ 19
Electrical fuses .................................................................................................................................................................... 19
14 List of spare parts ........................................................................................................................................................................ 20
Varioshake Shaking Incubator VS 60 OI ......................................................................................................................... 20
15 Accessories................................................................................................................................................................................... 22

6 / 26 Varioshake Shaking Incubator 09/2020

09/2020 Varioshake Shaking Incubator 7 / 26
1Use of the Shaking Incubator
Intended Use
Die Informationen dieser Bedienungsanleitung müssen unbedingt gelesen und beachtet werden. Nur dann ist die ein-
wandfreie Arbeitsweise des Schüttelapparates gewährleistet. Nur Personen, die sich mit dieser Bedienungsanleitung ver-
traut gemacht haben, dürfen die Geräte installieren und bedienen. Die Frequenz der Schüttelbewegung ist elektronisch
geregelt einstellbar. Zu schüttelnde Laborgefäße sind durch den Einsatz entsprechender Zusatzeinrichtungen betriebssi-
cher zu fixieren. Die maximal nutzbare Bewegungsfrequenz wird durch Art und Gewicht der Nutzlast mitbestimmt. Zum
sicheren Abstellen eingesetzter Zusatzeinrichtungen ist ausreichend Arbeitsfläche im direkten Umfeld des Gerätes einzu-
planen.
Caution:
due to high risk of injuries, never reach into the unit while the shaking device is still in operation.
Improper Use
LAUDA Varioshake Shaking Incubators must not be used in potentially explosive environments. The temperature work and
the shaking motion must not create an explosive atmosphere in the vicinity of the unit.
LAUDA Varioshake Shaking Incubators, operated in a laboratory, are no Medical Devices. They fall neither under national
nor international Medical Device Directives and have to be used and applied accordingly.
2Warranty
For all laboratory apparatus and their accessories from LAUDA-GFL Gesellschaft für Labortechnik mbH, there is a war-
ranty claim, as well for spare parts, repairs and modifications, carried out by LAUDA-GFL. In order to identify defective
units, we require both model and serial number on the nameplate of the Shaking Incubator and, if applicable, a copy of the
invoice.
3Before installation
The information in these operating instructions must by all means be carefully read and observed. Only then a perfect
functioning of the Shaking Incubator can be guaranteed.
Vital information within the manual are emphasized in bold letters. Safety precautions are additionally marked with the
following symbols.
Read and observe the
operating instructions
Warning of hand injuries
Warning of hot surfaces
Warning of dangerous
electrical voltage
General warning
Before maintenance and repair
disconnect the unit all-pole from
the electrical mains (pull the plug
from the socket).

8 / 26 Varioshake Shaking Incubator 09/2020
4Transport, Assembly and Location of the Shaking Incubator
Provide suitable transport means to transport the unit to the location. Doors, lifts and hallways/ corridors to the location
must permit safe transport.
Caution:
the Shaking Incubator weighs 51 kg. Lift and carry the unit with at least two persons when transporting it
to the location.
Place only on firm, stable and level surfaces indoors. The location must offer sufficient space and must safely carry the
total weight of the unit (unit weight as per technical data, chapter 10 of these operating instructions, plus weight of the
load).
For safe handling of removed, warmed-up loads make sure to allow sufficient workspace in the direct vicinity of the unit.
The unit is not approved for operation in potentially explosive surroundings.
5Operating Voltage
The Shaking Incubator must only be connected to a properly installed power connection with earth conductor (PE), ac-
cording to the local regulations. The Shaking Incubator is an electric appliance, protection class I, therefore, a connection
to earth conductor (PE) must be provided. The value of the required mains fuse is stated in the technical data, chapter 11
of these instructions. The electrical connection must be configured so as to secure that power can be cut all-pole at any
time.
It must be secured that, even after opening the cover, the mains connection cable does not touch any warm parts of the
interior or of removed, warmed-up load. The main switch on the control panel of the Incubator must be off
(position O). The operating voltage stated on the nameplate (on the back of the unit) must be identical to
the mains voltage. If they are identical, connect to the mains.
6Inbetriebnahme
Caution:
due to high risk of injuries, never reach into the unit while the shaking device is still in operation.
Caution:
make sure that neither the shaking movement nor the heating process create an explosive atmosphere
near the Shaking Incubator during operation.
Caution:
hot surfaces in the interior when temperature is set to 50 °C or more. Danger of burns. Wearing suitable
safety gloves is recommended.

09/2020 Varioshake Shaking Incubator 9 / 26
Operation
After switching on the main switch the fan motor starts.
Incubation time, incubation temperature and shaking frequency must be set and activated individu-
ally. The shaking motion and the heating can only be initiated when the cover is closed and will be cut
off when the cover is opened. Please refer to the following pages for instructions on initiation of the
individual functions.
Cooling Coil
The serial cooling coil for connection to the domestic water supply increases the temperature range of the Shaking Incu-
bator (lowest operating temperature of the Incubator +20 °C). The connections of the cooling coil are situated at the back
of the unit. Cooling water is to be supplied to the Shaking Incubator through pressure-proof water hoses with an interior
diameter of approx. 13 mm. The connected hoses are to be secured with hose clamps (hoses and clamps are not part of
the standard scope of supply).
Operation and display elements of the control panel
1 Main switch
Operation panel incubation time
2 Display incubation time
3 Pilot lamp "time" in the display, hours
4 Pilot lamp "time" in the display, minutes
5 Pilot lamp "total run-time" in the display
6 Change-over key for total run-time and for entry of higher set points
7 Entry of lower set points
19 Key to switch off alarm signal after expiry of preset run-times
20 Change-over key for alteration of set points
21 Confirmation of new total run-times

10 / 26 Varioshake Shaking Incubator 09/2020
Operation panel temperature controller
8 Display for set and actual temperatures
9 Pilot lamp "actual temperature" in the display
10 Pilot lamp difference value of over-temperature alarm in the display
11 Pilot lamp set temperature in the display
12 Change-over key between display and entry of higher set points
13 Entry of lower set points
22 Key to switch off alarm signal
23 Pilot lamp difference value of under-temperature alarm in the display
24 Change-over key for alteration of set points
25 Confirmation of new set points 9 Kontrolllampe Ist-Temperatur im Display
Operation panel shaking frequency control
14 Display for actual and set shaking frequencies
15 Pilot lamp "actual shaking frequency" in the display
16 Pilot lamp "set shaking frequency" in the display
17 Change-over key between the displays and entry of higher set points
18 Entry of lower set points
26 Shaking movement on/off
27 Change-over key for alteration of set points
28 Confirmation of new set points
6.3.1 Incubation time
Press key 20 of the keypad to switch the timer to entry mode.
The display begins to flash and pilot lamp 4 glows (display in minutes). The required incubation time can now by set by
pressing keys 6 (+) and 7 (-) (max. 59 minutes). Press key 21 to confirm the set minutes. The entry mode then switches
to entry of set hours. Pilot lamp 3 glows. Use keys 6 (+) and 7 (-) to set the required incubation time in hours (max. 999
hours). Press key 21 to confirm the set hours. The timer is now activated, the set incubation time starts to run down.
After activation of the timer the set incubation time can only be interrupted or altered by switching the Shaking Incubator
off with the main switch.
The currently remaining run-time is shown in the display. If the remaining run-time is more than one hour, pilot lamp 3
glows. The time is displayed in hours. If the remaining run-time is less than one hour, pilot lamp 4 glows. The time is dis-
played in minutes. While the incubation time is running down, press key 6 to choose between display of the remaining run-
time and the set total run-time. When display of the total run-time is chosen, pilot lamp 5 glows. Approx. 30 seconds
after the last entry, the digital timer switches back to normal operation mode. Expiry of the set incubation time is signalled
by a continuous acoustic alarm. Press key 19 to switch off the alarm signal. The timer merely triggers the alarm signal; the
unit is not switched off automatically.

09/2020 Varioshake Shaking Incubator 11 / 26
6.3.2 Incubation temperature
The microprocessor-controlled PID-type regulation permits quick availability of the set temperature, at the same provid-
ing a high temporal temperature constancy of +/- 0.2 K. Display as well as setting the temperature is done digitally on an
LED display, in 0.1 K increments. The operating temperature is approx. 8 K above ambient temperature to 70 °C. The
serial cooling coil for connection to the domestic water supply or to a refrigerated circulator permits operation at set tem-
peratures as of 20 °C.
After switching the unit on, display 8 flashes and shows the last memorized set temperature, pilot lamp 11 glows. Use keys
12 (+) and 13 (-) to enter a new set temperature. Press key 25 to confirm the set temperature. The temperature con-
troller starts operation. Approx. 30 seconds after the last entry pilot lamp 11 goes out and pilot lamp 9 glows. The display
now switches from set temperature to actual temperature.
Three different temperatures can be selected and changed by pressing key 12:
1. Set temperature, pilot lamp 11 glows.
2. Difference point for over-temperature alarm, dependent on the set temperature (set temperature plus differ-
ence point = alarm point), pilot lamp 10 glows. Factory aligned at 04.0 = 4.0 K.
3. Difference point for under-temperature alarm, dependent on the set temperature (set temperature minus dif-
ference point = alarm point), pilot lamp 23 glows. Factory aligned at 09.0 = 9.0 K.
When key 24 is pressed, the displayed set point starts to flash and can be reset through keys 12 (+) and 13 (-). Press key
25 to confirm the new set point. Over-temperature and under-temperature alarms are signalled acoustically as well as
visually by error codes in the display. For further information on this subject, please see chapter 7.1 Error Codes of these
instructions. When opening the Incubator's cover, both heating and shaking drive will be switched off. After closing the
cover, the Incubator will restart with the set values of temperature and shaking frequency.
Caution,
the temperature in the Incubator's interior may rise up to 70 °C during operation. Danger of burns. Wea-
ring suitable safety gloves is recommended.
6.3.3 Shaking frequency
An overload-protected a. c. motor produces an orbital shaking motion with a shaking amplitude of 30 mm through the
robust, durable drive mechanism. The motor is controlled through a PID type microprocessor frequency regulation. The
shaking frequency can be set between 20 min-1 to 250 min-1, independent of the load. It is equipped with a gentle start-
up and can be set in increments of 1 min-1.
Press key 26 to switch the shaking movement on and off. After switching on the shaking device, the Incubator starts shak-
ing with the last set shaking frequency. Pilot lamp 15 glows, display 14 shows the actual shaking frequency. In order to
reset the shaking frequency, first press key 17 and then 27. Pilot lamp 16 glows, the display flashes and shows the set
shaking frequency. Use keys 17 (+) and 18 (-) to reset the displayed value. Press key 28 to confirm the new set value.
Even when the shaking device is switched off, the set shaking frequency can be read by pressing key 17 (pilot lamp 16
glows = set frequency). It can be reset as described above. Approx. 30 seconds after the last entry the Incubator auto-
matically switches back to operation mode (shaking device on or off). To switch off the shaking device, press key 26. The
shaking frequency is cut off slowly, the operation panel is switched off.

12 / 26 Varioshake Shaking Incubator 09/2020
When opening the Shaking Incubator's cover, a control switch will cut off both heating and shaking device. After closing
the cover, the Incubator will restart operation with the set values of temperature and shaking frequency.
Caution:
only after the shaking platform has come to a complete stop, the cover may be opened to reach into the
unit.
RC Operation via PC
An interface module with a terminal on the back of the Incubator produces the data transmission format RS 232. The
interface enables reading of current actual and set values at any time. In order to set values for the temperature and shak-
ing frequency in a defined time-span, the Incubator has to be switched to remote control mode via a PC signal. Being in
remote control mode, the incubator cannot be operated manually. An interface protocol is available for operation of the
Shaking Incubator's RS 232 interface. Please advise model and serial number of the incubator.
Caution:
before operating the unit in RC mode, both unit and interior assembly of the Incubator must be carefully
checked. The Incubator must be operated in direct sight. Take special care in the vicinity of the automati-
cally operated unit.
Shutdown
To turn off the unit for longer shutdown times, first switch the main switch to position O and disconnect the unit from the
mains. Empty the interior, if necessary, dry and clean the interior to prevent bacterial contamination.
7Error Codes and Temperature Cut-out
Error Codes
The alarm functions for over- and under-temperature are switched on once the set temperature
has been reached for the first time after the temperature regulation has been started. Errors that
are due to external influences or to an operational fault are shown acoustically and visually. The
display shows either E01 for over-temperature alarm, or E02 for under-temperature alarm. The
heating is switched off until key 22 is pressed to cut off the alarm signal.
In case of under-temperature the acoustic alarm is cut off immediately, the visual alarm is cut off
after exceeding the set alarm point. In case of over-temperature both acoustic and visual alarms
are switched off at the same time. If the alarm occurs recurrently, the Incubator should be
checked by a trained electrician.

09/2020 Varioshake Shaking Incubator 13 / 26
Temperature Cut-out
As an additional safety precaution to over-temperature and under-temperature alarms, a thermostatic temperature cut-
out protects the unit against overheating. If either temperature regulator or fan motor should fail, the thermostatic over-
temperature cut-out will prevent overheating of the incubator. Only after the unit has cooled down, the temperature cut-
out, which is situated at the back of the Incubator, can be deblocked. For deblocking, loosen the black cap nut on the back
of the unit. Within the thread, a white plastic pin will now be visible which must be gently pushed inside (e.g. with a pen)
until a soft clicking sound can be heard. After deblocking, watch the operation of the Shaking Incubator carefully. If the
fault re-occurs, the cause has to be removed by a trained electrician.
8Functional description
A cover made of 8 mm thick acrylic glass encompasses the cabinet interior. The interior is heated by a heating element in a
temperature range of 20 °C (from approx. 8 K above ambient) to 70 °C. The fan optimises the temperature distribution.
The built-in cooling coil permits works at set temperatures near to the ambient temperature. The temperature controller is
a PID controller and is equipped with a monitoring function that signals malfunctions and switches off the heating in case
of a malfunction.
The shaking device in the cabinet interior is driven by an a.c. motor, is electronically controlled and permits orbital motions
from 20 to 250 min-1 and a fixed shaking amplitude of 30 mm.
The timer allows times to be set between 1 minute and 999 hours and 59 minutes. The timer continuously shows the
current remaining run time and emits an acoustic expiration signal.
The electronic regulation is microprocessor-controlled.
The housing is made of lacquered, moulded ABS resin elements and of electrolytically galvanised and powder-coated
sheet steel. The shaking platform, made of anodised aluminium, is equipped with four plastic pins for secure attachment of
accessory equipment. The heating element and the cooling coil as well as the rear wall and bottom plate of the interior are
made of stainless steel.
An RS 232 interface, set into the rear side of the unit, not only serves to register set and actual values of temperature and
shaking frequency, but also for remote operation through PC.
9Maintenance, Support and Repair
Caution:
Before opening the Distillation Apparatus as well as before cleaning procedures, discon-
nect the unit from the mains! (Unplug the unit or switch off the main switch).
Danger of electrical shock!
Cleaning and Disinfection
The lacquered or powder-coated surfaces of the housing, the shaking platform made of anodised aluminium and the acryl-
ic glass cover may be cleaned with mild, non-abrasive and pH neutral detergents. Never use solvent-containing cleaning
agents. The stainless steel surfaces of the interior can be cleaned and the original shine be restored with commercial stain-
less steel polishing agents (e. g. “Helios Brillant“ from M/s Ecolab).
To disinfect the Shaking Incubator by a wipe process, approved disinfecting agents with proven compatibility with stainless
steel, lacquered or powder-coated steel and ABS resin elements, aluminium or acrylic glass can be used. The instructions
for use of the disinfectant must by all means be observed. If in doubt, test the disinfectant in a small local area. Should the
Incubator have been contaminated with dangerous substances, cleaning and decontamination measures must be carried

14 / 26 Varioshake Shaking Incubator 09/2020
out according to the safety data sheets of these substances. It may be necessary afterwards to have the unit tested for
function and safety by a trained electrician or by the service department in the manufacturing plant.
Temperature Adjustment
LAUDA Varioshake Shaking Incubators are set and adjusted at a temperature of 45 °C. Instructions for adjustment of the
temperature regulator are available on request for adjustment purposes during maintenance. Please state model and serial
numbers of the Shaking Incubator.
Checking the Acrylic Glass Cover
The acrylic glass cover is connected to the unit's housing by hinges. The closing motion of the cover is
slowed down by two gas pressure springs for safety reasons. The safe function of these gas pressure springs
must be checked in regular intervals when opening and closing the cover. Make sure that the cover does
not fall down undamped.
Shutdown caused by overload or power failure
The drive motor of the shaking device is protected against over-heating due to overload. The device will be switched off.
Caution, after the motor has cooled down, the shaking device restarts automatically (without gentle start-up control).
After a mains failure, the Incubator does not restart automatically, but has to be started as described in chapter 6.
Exchanging the fuses
The unit's two fuses are situated in a drawer below the unit's plug. In order to check or to exchange these
fuses, press the fixing clips on both sides of the drawer and pull.
The fuses may only be exchanged against 4 Amp fuses of the same type. Information on the fuse type can
be found on the nameplate next to the unit plug and in the spare parts list of these operating instructions. If
the unit is shut down due to an operation fault, always switch it off on the main switch before opening the cover and
touching parts in the unit's interior.
Maintenance Procedures
Please do not hesitate to contact your local dealer or us for technical support. Please contact your dealer or us before
returning a defective unit and do not return it before consulting us. A defective unit may only be returned thoroughly
cleaned, decontaminated, if necessary, and carefully packed, accompanied by a detailed description of the defect, to
LAUDA-GFL Gesellschaft für Labortechnik mbH
Schulze-Delitzsch-Straße 4+5
D-30938 Burgwedel
Servicings, repairs and modifications must be carried out according to the General Technical Rules and Regulations (§2,
chapter 2, BGV A3) by a trained electrician (§2, chapter 3, BGV A3). Only original spare parts may be used. Always
demand a confirmation (company, date, signature) on the kind and extent of the service(s) carried out.

09/2020 Varioshake Shaking Incubator 15 / 26
10 Disposal of Old Units
LAUDA-GFL will take responsibility, within the scope of the legal directives, for an environmentally sound handling and
disposal of all used LAUDA-GFL units as of the production year 1995 that are returned to us free of charge and will have
it materially recycled. Before the unit is returned, a legally binding declaration must be provided from the sender, confirm-
ing that the unit is free from harmful and/or hazardous contaminations as well as from hazardous substances caused by the
previous use of the unit. LAUDA-GFL laboratory apparatus are exclusively designed for industrial use and may not be
disposed of through public waste disposal authorities.
EAR Registration Number WEEE-ID.NO.DE 67770231

16 / 26 Varioshake Shaking Incubator 09/2020
11 Technical Data
Varioshake Shaking Incubator VS 60 OI
Exterior dimensions (W x D x H)
560 mm x 690 mm x 630 mm
Interior dimensions (W x D x H)
Volume / Capacity
450 mm x 450 mm x 360 mm
Approx. 60 l
accepts one shaking tray A000045
Maximum load
depending on weight of load and set shaking fre-
quency
up to 12 kg
Control
digital
Temperature range (Incubation)
Temperature constancy (temporal)
Over-temperature cut-out
Under-temperature cut-out
Temperature limiter
+20 °C (from approx 8 K above ambient) to +70 °C
serial built-in cooling coil
+/- 0,2 K
electronically settable up to 9.9 K above the set value, to protect the test substances
electronically settable up to 9.9 K below the set value, to protect the test substances
electromechanical temperature limiter with capillary tube sensor, cut-off temperature
135 °C / -10 K
Shaking motion type
Shaking amplitude
Shaking frequency
Incubation time
orbital, can be switched on and off
30 mm
20 - 250 min-1 depending on load
1 min –999:59 h
Electrical connection /
Mains connection
230 V, +/-10 %, 50 / 60 Hz
Mains fuse
On-site
Internal
10 A
4 A T
Power
650 W
Protection class / - type
I / IP20
Environmental conditions
only for indoors use (not in potentially explosive surroundings)
Ambient temperature
+ 10 °C to + 30 °C
Humidity
max 70 % relative humidity,
non-condensing
Emission sound pressure level
< 70 dB (A)
Weight
51 kg

09/2020 Varioshake Shaking Incubator 17 / 26
12 Circuit Diagram
A1 Electronic control power board
A2 Electronic control timer
A3 Electronic control temperature regulator
A4 Electronic control shaking frequency
A5 Pulse generator
A6 Interface inlet RS 232
A7 Line filter
B1 Temperature sensor PT100
C1 Motor condenser
E1 Tubular heating element
F1 Fuse 4 A T
F2 Fuse 4 A T
F3 Temperature limiter
M1 Fan motor
M2 Drive motor
S1 Main switch
S2 Control switch lid
S3 Change-over key for display of total run-time or entry of higher set points
S4 Entry of lower set points
S5 Switch off alarm signal
S6 Change-over key alteration of total run-time
S7 Confirmation of new total run-times
S8 Change-over key display and entry of higher set points
S9 Entry of lower set points
S10 Switch off alarm signal
S11 Change-over key alteration of set points
S12 Confirmation of new set points
S13 Change-over key display and entry of higher set points
S14 Entry of lower set points
S15 Shaking movement on / off
S16 Change-over key set points
S17 Confirmation of new set points
V1 Triac

18 / 26 Varioshake Shaking Incubator 09/2020
Varioshake Shaking Incubator VS 60 OI

09/2020 Varioshake Shaking Incubator 19 / 26
13 Connection to the mains supply
LAUDA Varioshake Shaking Incubators are supplied with a pre-assembled, cast-on shock-proof plug (CEE 7/7). Make
sure to connect to a protective conductor terminal.
Colour decoding
Stromnetz
ge/gr –yellow/green
PE (Protective earth)
bl –blue
N
sw –black
L1
Electrical fuses
Type
Power
Power consumption at mains voltage *
On-site mains fuses (F4, F5)
VS 60 OI
0.65 kW
3.0 A at 230 V
10 A (max. 16 A)
* see nameplate

20 / 26 Varioshake Shaking Incubator 09/2020
14 List of spare parts
Varioshake Shaking Incubator VS 60 OI
Pos. No. Part-No. Article
1 0012268 Heating element 230 V 500 W
2 0012430 Micro switch
3 0012843 Triac
4 A00 Gas pressure spring
5 0015533 Pivot shaft
6 0026283 Angular ball bearing
7 0015522 Ball bearing seat
8 A000097 Pulse generator
9 0014315 Stand
10 A000073 Motor
0015531 Belt pulley for motor
11 A000112 Driving belt
12 0026285 Flat spring (complete set)
13 0013111 Temperature sensor PT 100
14 A000092 Temperature limiter
15 0030010 Mains connection cable
16 0012044 Plug
0012045 Drawer
0012902 Fuse 4A T
17 0013757 Microprocessor temperature regulator
18 0013759 Microprocessor shaking frequency display
19 0013768 Microprocessor power board
20 0013758 Microprocessor timer
21 A000127 Main switch
22 0014413 Hinge
23 0012166 Radial fan
0012154 Protection shield for fan
This manual suits for next models
1
Table of contents
Other Lauda Accessories manuals
Popular Accessories manuals by other brands

Psooo
Psooo PS-400 user manual

Colorado time systems
Colorado time systems Dolphin Scoreboard Adapter quick start guide

PCB Piezotronics
PCB Piezotronics IMI SENSORS 641B10 Installation and operating manual

Antelope
Antelope GOLIATH HD owner's manual

Grizzly
Grizzly H8369 instructions

Niles
Niles CS120 Installation & operation guide

Protocol
Protocol ALL-IN-ONE BOXING SET instruction manual

PCB Piezotronics
PCB Piezotronics IMI SENSORS EX640B61 Installation and operating manual

Hioki
Hioki 9667 user manual

celesco
celesco MTA-3 installation guide

llaza
llaza STORBOX 250 installation manual

IFM Electronic
IFM Electronic Efector 200 OL Series operating instructions