Liko Traverse User manual

3EN120401 Rev. 7
ENGLISH
Traverse Switch
rail switch system
Service Manual
Product No. 3124604

Service Manual Traverse Switch • 3EN120401 Rev. 7 2 www.liko.com lwww.hill-rom.com
Spare part statement
Liko AB will as manufacturer of staonary & mobile lis keep spare parts available
for 10 years aer the last manufacturing date of the product.
Spare parts are available in the spare parts list or in the Service Manual for each model of li.
Components (manufacturing parts) are not available aer last manufacturing date.
Repairs and maintenance may only be performed by personnel authorized by Hill-Rom
and using original Liko™ spare parts.
For informaon regarding Service and Maintenance educaon, contact your Hill-Rom representave.
For contact informaon please visit our website: www.hill-rom.com lwww.liko.com
In this document, this warning symbol indicates that special care should be taken.
If instrucons are not followed there is a risk of serious injury to paent or caregiver.

Service Manual Traverse Switch • 3EN120401 Rev. 7 3 www.liko.com lwww.hill-rom.com
Page
1.0 Technical Data................................................................4
2.0 Troubleshoong............................................................. 5
3.0 SparePartsAssemblyinformaon ................................ 6
4.0 Spare Parts List ..............................................................7
5.0 PeriodicInspeconInformaon.................................... 8
6.0 Product Changes............................................................ 9
Table of Contents
Traverse Switch

Service Manual Traverse Switch • 3EN120401 Rev. 7 4 www.liko.com lwww.hill-rom.com
1.0 Technical Data, Traverse Switch
2010-04-21, llasaba 2 (10)
Technical Data
© Copyright Liko AB 2012-09
Inspection and Maintenance
When necessary, clean the product with a moist cloth using common surface cleaners or disinfectants. NOTE! Do not use
cleaning agents that contain phenol or chlorine, since these can damage aluminium and plastic materials.
Service
Likos transfer system should be periodically inspected at least once a year.
Periodic inspection, repair and maintenance should be performed only in accordance with the Liko Service Manual
and by personnel authorized by Liko/HillRom and using original Liko spare parts.
Service Agreement
Liko/HillRom offers the opportunity to enter into service contracts for the maintenance and periodic inspection of your Liko
products.
Expected Service Life
The product has an expected life time of 10 years when correctly handled, serviced and inspected in accordance with Liko’s
instructions.
Transport and Storage
The environment in which the product is transported and stored should have a temperature of 10–40 ˚C and a relative humidity
of 30–75 %. The air pressure should be 700–1060 hPa.
Recycling
For instructions on how to recycle your Liko product, please visit the Liko website: www.liko.com.
Product Changes
Liko’s products undergo continous development, which is why we reserve the right to make product changes without prior notice.
Contact your Liko/HillRom representative for advice and information about product upgrades.
Design and Quality by Liko in Sweden
Liko is quality-certied in accordance with ISO 9001 and its equivalent for the medical device industry, ISO 13485.
Liko is also certied in accordance with environmental standard ISO 14001.
The device is intended for indoor use.
Type B, in accordance with the electrical shock protection class.
Class II equipment.
Prod. No. 3124604 (only in combination with Liko H-70 rail)
Dimensions:
A: 490 mm / 19.3 inch.
B: 160 mm / 6.3 inch.
C: 92 mm / 3.6 inch.
Built-in dimension, H: 92 mm / 3.6 inch.
Weight: 3.5 kg / 7.7 Ibs.
Max. load: 250 kg / 550 Ibs.
Material: Aluminium, Steel and plastic.
Protection class: IP X1

Service Manual Traverse Switch • 3EN120401 Rev. 7 5 www.liko.com lwww.hill-rom.com
2.0 Troubleshoong, Traverse Switch
The switch doesn't work. 1. Check mains power supply.
2. Ensure that the cable is properly
connected to the power supply.
3. Check that hand control is rmly
connected.
4. If the switch sll does not work correctly
contact Hill-Rom.
The switch stops in close or 1. Check cables to power supply.
open posion. 2. Check mains power supply.
3. If the switch sll does not work correctly
contact Hill-Rom.
If the switch makes an unusual Contact Hill-Rom.
sound.

Service Manual Traverse Switch • 3EN120401 Rev. 7 6 www.liko.com lwww.hill-rom.com
3.0 Spare Parts Assembly Informaon
Assembly Instrucons to Liko™ spare parts are aached with the products.
The assembly instrucons are also available in Agile and on the Partner Web.
Document number to spare parts assembly instrucons can be found in the spare parts list.
Liko technical documentaon can be found on the Partnerweb:
hps://partner.liko.com “My informaon” - “Documents”
If you don’t have access, send a mail for registraon to: se.customer[email protected]
NOTE
- Always make sure the spare part is mounted according to the assembly instrucon.
- Check all funcons related to the changed spare part.
- Repairs and maintenance may only be performed by personnel authorized by Hill-Rom
and using original Liko™ spare parts.
Actuator LA12
Ställdon LA12
Assembly Instruction
Monteringsanvisning
Prod. No. 31790001
3SE690217-02
2012-10-01
Tools required
17 mm
6 mm
Contains: Qty:
Actuator LA12 1
Locking ring D8 2
Screw K6S M10x35 1
Nut M10 1
Fits:
3124604 Traverse Switch
NOTE
- Make sure the spare part is mounted according to the assembly instruction.
- Check all functions related to the changed spare part.
-
Repairs and maintenance may only be performed by personnel authorized by Hill-Rom/Liko and using original Liko spare parts.
www.liko.com
© Copyright Liko AB
4.497
1. Move the traverse switch to closed position, with
handcontrolorxedswitch.
2. Unplug the current from the actuator.
3. With the long nose plier remove the lower locking ring
from the locking pin.
4. Remove the locking pin.
5. Unscrew the screw and nut.
6. Replace the old actuator.
7. Reassemble in reverse order.
Connected to controlbox or multistation.
2
3
4
5
Document Number
INFORMATION

Service Manual Traverse Switch • 3EN120401 Rev. 7 7 www.liko.com lwww.hill-rom.com
4.0 Spare Parts List, Traverse Switch
3124604 Traverse Switch
497
Traverse Switch
20590010 449 Hand Control 2-buon M220 pce 3SE650285
31790001 497 Actuator LA12 pce 3SE690217
31790002 498 Rail Warning label, 4 pce set 3SE690218
31290125 515 Damper
Required on Carriage, installed in system
with Traverse Switch, according to Installa-
on Handbook.
pair 3SE640257
31290126 516 Damper QR
Required on QR Carriage 3126008, installed
in system with Traverse Switch, according to
Installaon Handbook.
pair 3SE690224
STANNA INTE HAR / DO NOT STOP HERE STANNA INTE HAR / DO NOT STOP HERE
498
Spare Parts Posions
PROD. NO. PRODUCT
Liko™ Spare Parts
Spare Parts List • Overhead Lifts 31 www.hill-rom.com l www.liko.com
PROD. NO. POS. NO. NAME UNIT ASSEMBLY INSTR.

Service Manual Traverse Switch • 3EN120401 Rev. 7 8 www.liko.com lwww.hill-rom.com
5.0 Periodic Inspecon Informaon, Traverse Switch
According to ISO 10535:2006 Annex B, Periodic Inspecon of medical device Liko™ overhead rail systems
should be performed at least once a year. Periodic Inspecon Liko™ overhead rail system (3EN111001) is
available in Agile and on the Partner Web.
Periodic Inspecon protocol
Liko™ overhead rail system
3EN111001 Rev. 13
1(12)
INSTALLATION
Acon required:
Approved Not
approved
1 Decal rail marking
2General inspecon
3* Rails
4End stops
5Ceiling/Wall aachments
System informaon Customer reference
Type of rail system: Agreement No:
S/N: Name:
Approved Max Load: Address:
Year of installaon: Post Code
ADDITIONAL COMPONENTS
N/A S/N:
Acon required:
Approved Not
approved
6* Rail Switches
7 Ultra Twist
8Transfer Motor Traverse
9Acvity Rail
10 Pool Side Lier
ENVIRONMENTAL IMPACT
11 Corrosive environments
LOAD TESTING
Max Load applied:
Approved Not
approved
12 Maximum load rail system Kg: Lbs:
REQUIRED MEASUREMENTS:
N/A
Dimension and unit
Approved Not
approved
3* Joint Gap mm: Inch:
6B* Rail Switch play
mm: Inch:
6C* Turn Table clearance
mm: Inch:
Inspecon sign o
Complete inspecon according to the instrucons (page 2-11), ll in this page and sign below.
Approval to use the overhead system Approved Not approved
If the overhead system has one or more inspecon points “Not approved”, the system must not be used.
Acon required: Acons according to the inspecon items "NOT APPROVED" should be performed immediately. Aer performed acons
sign below. If anything is unclear or if you have quesons, please contact Hill-Rom or your local Hill-Rom representave.
Contact informaon is to be found at www.hill-rom.com.
Inspecon performed by: Date:
Next inspecon:
Inspecon performed in accordance with ISO 10535:2006 Annex B- Periodic inspecon
Liko™ Overhead Rail System must be thoroughly inspected at least once per year. Inspecon and service must be carried out by Hill-Rom authorized personnel.
See "Installaon condion checklist" on next page, to determine how to perform inspecon points 2 and 5.
Periodic Inspecons are available for: Liko™ Overhead Lis (3EN191001), FreeSpan and FreeStand (3EN301001), Accessories (3EN601001) and Mobile Lis (3EN371001)
North America only: For non Liko™ Rail system, use “Non Liko® Rail Installaon Periodic Inspecon” (193725)
If printed: Make a colour print of this instrucon.
Liko technical documentaon can be found on the Partnerweb:
hps://partner.liko.com - “My informaon” - “Documents”
If you don’t have access, send a mail for registraon to: se.customer[email protected]
Periodic Inspecon may only be performed by personnel authorized by Hill-Rom and using
original Liko™ spare parts.
Available Liko Periodic Inspecons:
• 3EN371001PeriodicInspeconLiko™mobilelis
• 3EN111001PeriodicInspeconLiko™OHrailsystem
• 3EN191001PeriodicInspeconLiko™overheadLis
• 3EN301001PIFreespan™-Freestand™freestandinglisystems
• 3EN601001PeriodicInspeconLiko™accessories
Also available is Periodic Inspecon Slings. This inspecon should be performed by a person who is
suitable, has good experience and sucient knowledge about design, use and maintenance of the sling.
• 7EN160834GeneralPeriodicInspeconProtocolLikoSlingsandSabinaSupportVests
NOTE!
Periodic Inspecon shall be performed at least once per year.
INFORMATION

6.0 Product Changes, Traverse Switch
Prod. No. 3124604
2014
Changes Revision Project
To prevent noise from carriages in rail curves, new design of dampers are developed.
The new dampers 31290125 (Damper) and 31290126 (Damper QR) are now delivered with the
Traverse switch, and are replacing 31290096 (Damper) and 31290097 (Damper QR).
Rev. 01 SUS000583
2012
Changes Revision Project
TwonewdesignsofDampers,thatpreventstheCarriagestobeparkedinthewrong
area of the Travers switch are now delivered with the Traverse switch.
The new Dampers 31290096 (Damper) and 31290097 (Damper QR) are replacing
31290072 (End Stop)
Rev. 01 1778
2011
Changes S/N Project
Newaachmentforthetraverseswitchactuator.Theconguraonallowschangeof
posionoftheactuator,fromthexedrailtothesecondaryrailintheoverheadsystem.
TheslimlinependanthasexpiredsotheTraverseswitchstabilizaonbarhasbeenupdated
totthependantsthatareused.
6052700 1102
2006
Changes
New Prod. No 3124604 (old Prod. No 3124603), due to changed traverse plate.
Prod.No3124604hasTraverseplateprimaccordingtodrawingR3024023-3
Prod.No3124603hasTraverseplateprimaccordingtodrawingR3024023-2
Newdesignontraverseplateandactuator.Thenewactuatoropensandclosesintwoseconds(fourmesfaster).
The switch has new traverse plate that prevent stop of the carriages between the rails.
NB! Always use end stop adapters on the carriages!
SpecialInstallaoninstruconforTraverseSwitch.
Service Manual Traverse Switch • 3EN120401 Rev. 7 9 www.liko.com lwww.hill-rom.com

Liko AB
Nedre vägen 100
SE-975 92 Luleå , Sweden
www.hill-rom.com Iwww.liko.com
© Copyright Liko R&D AB 2017
This manual suits for next models
1
Table of contents