Major Science Winpact SI-200 User manual

2

3
Table of Contents
Table of Contents.........................................................................................................................3
Packing List ....................................................................................................................................5
WARNING .......................................................................................................................................6
I. Introduction .................................................................................................................... 11
II. Product Specification..................................................................................................... 12
III. Component Guide......................................................................................................... 13
IV. Installation Instruction .................................................................................................. 15
Installing Adjustable Angle Tube Rack (SI-200-02 / SI-200-03).........................................................15
Installing Universal Spring Rack (SI-200-04) ...........................................................................................19
Installing Sticky Pad Platform (SI-200-05)................................................................................................21
Installing Universal Cushioned Crossbar (SI-200-06)........................................................................23
Installing Microplate Holders (SI-200-07) ...............................................................................................26
Installing Flask Holder (SI-200-08~13).....................................................................................................28
Adjusting the Foot Stand ............................................................................................................................34
V. Operation Instruction................................................................................................... 36
Features of Control Panel...........................................................................................................................37
Program Mode Settings...............................................................................................................................39
Naming the Program Mode Operation ..........................................................................................40
Setting Up the Program Parameters ...............................................................................................41
Starting the Program Mode Operation ..........................................................................................44
Continuous Mode Settings .........................................................................................................................45
Setting Up the Continuous Mode....................................................................................................46
Starting the Continuous Mode Operation ....................................................................................47
System Setup .................................................................................................................................................49
Optional Device: PT-100 ....................................................................................................................50
Temperature Permissible ...................................................................................................................51
System Total Operation Time............................................................................................................53
System Software Version....................................................................................................................54
System Calibration Value....................................................................................................................55
Machine ID............................................................................................................................................56
Remote Control Operation.......................................................................................................................57
Installing SI-200 Software on User’s Personal Computer...........................................................57
Using SI-200 Software to Operate the Device.............................................................................62

4
VI. Troubleshooting ............................................................................................................. 88
Troubleshooting Guide................................................................................................................................88
VII. Calibration & Maintenance........................................................................................... 91
Temperature Calibration.............................................................................................................................91
Maintenance...................................................................................................................................................92
Allowance of Shaking/ Vibrating Settings .................................................................................................93
VIII. Ordering Information.................................................................................................... 96
Warranty.................................................................................................................................... 97

5
Packing List
Winpact Shaking Incubator × 1
Instruction Manual × 1
Power Cord x 1
Foot Stand (Pre-installed) x 4
Universal Platform x 1

6
WARNING
This equipment has been tested and verified to comply with safety limits. These limits are
designed to provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment may generate, use, and radiate radio
frequency energy, and if not installed and used in accordance with the instruction manual, may
cause harmful interference with radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the user will be required to
correct the interference at their expense. Changes or modifications not expressly approved by
the party responsible for compliance could void the user’s authority to operate the equipment.
It is strongly recommended that the user carefully read the following prints before operating
any equipment.
1. Read and follow the instruction manual thoroughly.
2. Do not modify the equipment. Failure to adhere to these directions could result in personal
and/or laboratory hazards, as well as invalidate the equipment warranty.
3. Use a properly grounded electrical outlet with correct voltage and current handling
capacity.
4. Disconnect the equipment from power supply before servicing and maintenance.
5. In the event of spilling solutions accidentally into the instrument, disconnect the grounded
plug and user must have an appropriate decontamination. And replace the damaged parts.
6. Do not use in the presence of flammable or combustible material, which may cause fire or
explosion. The instrument contains the components that can ignite such materials.
7. Refer maintenance and servicing to qualified personnel.
8. Ensure that the system is connected to electrical service according to local and national
electrical codes. Failure to properly connect may cause a fire or shock hazard.
9. Ensure the use of appropriate materials and correct operation to avoid possible hazards of
explosion, implosion or release of toxic or flammable gases arising from the heated
materials.
10. Handle the culture vessel with cares to prevent the glass vessel from any damages or
crashes, especially before and after autoclave / sterilization.
11. Always use the supplied handle or appropriate protection to handle the culture vessel to
prevent hands from heat burns, especially after autoclave/sterilization.

7
12. The instrument is intended for scientific research use only, and must be operated by
qualified personnel who realize the potential risks of the use of this instrument. Major
Science makes no claim that its instrument are designed or certified as medical device; no
representation, promises, express warranty, or implied warranty will be made concerning
the suitability of these instruments for any medical use. Major Science will not provide
customers any notice or certification concerning its products being compliant as medical
device.
Safety Information
The device may only be operated by trained lab personnel who have carefully read the operating
manual and are familiar with the device functions.
Before operating SI-200 incubator, make sure that anyone involved with its operation has been
instructed in both general safety practices for laboratories and safety practices for this apparatus.
The user is also responsible for following local guidelines for handling hazardous waste and
biohazardous materials that may be generated from the use of this equipment.
WARNING! Risk of explosion and injury or death!
Do not use equipment with flammable substances or organisms with flammable by-products.
WARNING! Heavy!
-Do not attempt to lift the SI-200 by yourself.
-Ask for assistance or use suitable equipment when raising or handling the device.
WARNING! Risk of electric shock and/or damage to unit!
-Check that the voltage and frequency of your unit are compatible with mains/power supply.
-Do not remove caution label from unit.
-Use a grounded power supply.
-Turn off the device and disconnect from power supply before replacing fuses.
WARNING! Damage to device!
Never run shaker without a platform.

8
CAUTION! Still Running!
The shaker still runs after the cover has been opened.Wait for it to stop running.
NOTICE! A torpedo level is needed before operation.
To keep device as balanced as possible, it is recommended to apply a torpedo level especially
necessary at the first time delivery, adjust the foot to ideal level (refer to p.33).
Auto Shuts Off Situation:
-Device Interrupt shuts off power to SI-200 when lid opens.Auto
resume when the lid closes.
-Deviations form temperature set point.
-Over temperature warning.
Use high levels of precaution when using any electrical device. Before connecting the electric supply,
check to see if the supply voltage is within the range stated on the rating label, and see to it that the
device be seated firmly. Place the unit in a safe and dry location; it MUST NOT touch any surrounding
objects. Follow the safety precautions for chemicals/dangerous materials, and hot surfaces (after
autoclave/sterilization). If needed, contact a qualified service representative or

9
Environmental Conditions
Ensure the instrument is installed and operated strictly under the following conditions:
1. Indoor use only
2. ≦95% RH
3. 75 kPa –106 kPa
4. Altitude must not exceed 2000 meters
5. Ambient to 40℃operating temperature
6. Pollution degree: 2
7. Mains supply voltage fluctuations up to ± 10% of the normal voltage
Electrical Shock Precaution
Follow the guidelines below to ensure safe operation of the unit.
The Winpact Shaking Incubator has been designed for use with shielded wires thus minimizing any
potential shock hazard to the user. Major Science recommends against the use of unshielded wires.
1. Dry out for a period of time and restore the instrument to NORMAL CONDITION before
operation.
2. NEVER connect or disconnect wires leading from the power jacks when the red indicator light
on the “ON/OFF”key is on or when the device is running.
3. Wait at least 5 seconds after stopping a run before handling output leads or connected
apparatus.
4. ALWAYS make sure that user’s hands, work area, and instruments are clean and dry before
connecting any cables or operating the device.
5. ONLY connect the power cord to a properly grounded AC outlet.
Preventive Measures of Damaging Instrument
1. Do not attempt to operate the device if damaged.
2. Protect this unit from physical damage, corrosive agents and extreme temperatures (direct
sunlight etc.).
3. For proper ventilation, leave at least 10 cm of space behind the instrument, and at least
5cm of space on each side.
4. Do not operate the instrument in high humidity environments (> 95%), or where
condensation may occur.
5. Before using any cleaning or decontamination method except those recommended by the
manufacturer, user should check with the manufacturer that the proposed method will not
damage the equipment.

10
Equipment Operation
1. Check that the device and all the accessories are assembled well and undamaged before
starting operating.
2. It is recommended that to have stopper for tubes user possesses before the operation to
prevent any splitting of solution.
3. NEVER access any HAZARDOUS LIVE parts that violate International Biosafety
Regulations.
Symbols
The symbols used on the device are explained below:
The symbol indicates an area where a potential shock hazard
may exist.
The symbol indicates a warning. Consult the manual to avoid
possible personal injuries or instrument damages.
ATTENTION: Hot Surface!
Two people are required to lift the
device. Ask for assistance or use
suitable equipment when raising or
handling the device

11
I. Introduction
Major Science is introducing our newly innovated Winpact Shaking Incubator.The ultimate
bench-top shaking incubator provides user the stability and durability for all user’s fermentation
and incubation needs.
Features
Automatic shut off over temperature protection.
Sensitivity adjustable G-sensor with warning embedded for imbalance weight detects.
Safety interlock stops platform motion.
External temperature probe available upon order for precision temperature
measurement.
Programmable or continuous mode for user application.
Wide shaking speed 20-500 rpm satisfies different types of culturing.
Brushless DC motor for long and quiet operation, durable and maintenance free
usage.
Various selection for accessories
Temperature calibration for different user needs
A spare air/gas inlet to create particular environment for specific kinds of
cell/microbial

12
II. Product Specification
Unit Dimension
23.3” ×32.3” × 20.9”
(Approx. 59 × 82 × 53 cm)
Platform Size
18.1” x 18.1” (46 × 46 cm)
Shaking Orbit
3/4” (19 mm)
Speed Range
20 rpm –500 rpm
Speed Increment
1 rpm
Communication Port
RS485
External Temp. Port
PT-100
Heating Temperature Range
Ambient +5℃to 65℃
Output
600 W
Temperature Accuracy & Uniformity
± 0.25℃at 37℃
Timer
999 (hr):59(min)/ Continuous operation
Display
3.5”Color TFT LCD screen
Power Rating
100~240V, 50/60 Hz, 6.3A
Weight
Approx. 70 kg
Certification
CE, FCC
Carrying Capacity
Approx. 20 kg
Material
Acrylic, SPCC
Auto Resume Function
Yes

13
III. Component Guide
Transparent Shield
Handle
Control Panel
Power Socket
Voltage Switch
110/220 V
Air/gas Inlet (optional)
Power Switch
Spare Installation Hole
Adjustable Foot Stand
RS-485

14
PT-100 Sensor Port
(External Temp. Port)
Spare Installation
Hole
Fan

15
IV. Installation Instruction
Winpact Shaking Incubator is a ready-to-use product when delivered to user. Before user start
the device, make sure it is placed on a level, steady surface and stored in safe and dry
environment.
To adapt different laboratory containers and experiment purposes, Major Science provides a
variety of optional holders and trays for user. Its easy installation feature allows user to assemble
the accessories on the platform by themselves.Also, user can install different accessories
simultaneously on the platform for a more flexible setup to meet user’s personal requirement.
Regarding the detailed instructions on the installation of incubator accessories, please refer to
the following sections.
Installing Adjustable Angle Tube Rack
(SI-200-02 / SI-200-03)
☝Things Required:
Adjustable Angle Tube Rack (SI-200-02 or SI-200-03)
M4 × 6mm Truss Head Screws (the numbers of screws depend on the numbers of racks.)
M3 x 6mm Slotted Head Screws (the numbers of screws depend on the numbers of racks.)
Phillip Screwdriver
Adjustable Angle Tube Rack (16 × 50ml)
Adjustable Angle Tube Rack (33 × 15ml)
Phillip Screwdriver
M4 × 6mm Truss
Head Screw and
M4 Washer
M3 × 6mm
Slotted Head
Screw
SI-200-02
SI-200-03

16
Loosen the flathead screws on
both ends of the rack.
Lift up the rack from the bottom plate.
Flathead Screw

17
Place the bottom plate to the desired position on the
platform.And use four truss-head screws and their
corresponding M4 washer to fasten the bottom plate
through the slots.
The design of slots on bottom plate allows user to
arrange the tube racks in various angles without fixed
angle of racks.
Tilting Angle
Remove the slotted head screw first then
fasten onto the right angle holes for user’s
requirement.
M3 × 6mm
Slotted Head
Screw

18
Place the tube rack back to the
bottom plate.Adjust the tilt angle of the
rack before user fixes it on the bottom
plate.
A maximum of 4 tube racks can be
installed on the platform.

19
Installing Universal Spring Rack (SI-200-04)
☝Things Required:
Universal Spring Rack (SI-200-04)
M5 × 15mm Flathead Screws
Phillip Screwdriver
Universal Spring Rack (SI-200-04)
Remove four countersunk head
screws at corners. (Put them away in
the case of further usage)
Phillip Screwdriver
User can arrange the springs in any
way for various requirements of
usage with maximum 152 springs
(or purchase more if it is needed).
M5 × 15mm
Flathead Screw

20
Note: Spring rack should be fixed on either
side on the platform, instead of on any place of
platform.
Place the spring rack on the platform,
and fix it onto platform with four flathead
screws.
Leave the screws of the rest of
corner untouched if only one spring rack is
needed.
Table of contents
Other Major Science Accessories manuals
Popular Accessories manuals by other brands

DH Instruments
DH Instruments RUBBER SHOE ACCESSORY - SCHEMA P-N 400203 manual

Interlogix
Interlogix TX-6010-03-1 Installation sheet

Topcom
Topcom BW-4753 user guide

Elsys
Elsys ERS operating manual

PCB Piezotronics
PCB Piezotronics IMISensors Echo 670A01 Installation and operating manual

Dometic
Dometic 8952 1.4 0 Series operating instructions

Knocks
Knocks KDDM-2 operating instructions

Denver Electronics
Denver Electronics PBA-16001MK2 instruction manual

Lithonia Lighting
Lithonia Lighting OFLR 9LN 120 MO BZ M2 operating instructions

Lika
Lika SMK Series user manual

Rukket
Rukket Spor Net Assembly & instruction manual

QOLSYS
QOLSYS IQ MINI DW S Series Quick install guide