Quick Start Guide
Fit the appropriate mains adaptor to
the charger (see ‘Baery Charging’
page 8).
Connect the battery charger to a
mains power socket and the Airflo 24
charging socket.
BEFORE USE
CAUTION: The Airflo 24 must be put on charge for 12 Hours before first use.
To calibrate the battery and battery indicator, it is essential before the first
use to charge for 12 hours and then disconnect from the charger and leave
for 90 minutes to allow the battery management software to correctly
calibrate. Failure to do this could lead to incorrect battery charging and
battery ladder indication.
1 x Airo 24 compressor (A) ed with a
24v - 2200mAh Ni-MH baery pack.
1 x Baery charger (with mains adaptors) (B).
1 x User Instrucons.
1 x Mains Power Adaptor (Oponal).
A
B
STANDARD PACKAGE CONTENTS

Healthcare solutions designed for life
3
Press the hand control
buons to raise the
ELK or Camel.
4
Connect the 4-Way hand
control inlet hose to the
Airo 24.
2
Connect the 4-Way hand
control outlet hoses to the
corresponding numbered
connectors on the ELK or
Camel.
1
Press the ‘POWER’ buon to
turn the Airo 24 ‘ON’.
3
NOTE: Disconnect the baery charger before using the Airo 24.
English
Quick Start Guide
CHARGE THE BATTERY FOR 12 HOURS BEFORE
USE. To calibrate the battery and battery indicator, it
is essential before the first use to charge for 12 hours
and then disconnect from the charger and leave for
90 minutes to allow the battery management
software to correctly calibrate.
For the first 3 hours, the bottom light will flash
green. After 3 hours the first light will become a
steady green and the second light will start
flashing. After 6 hours the bottom 2 lights will
remain green and the third light will flash. After
9 hours the bottom 3 will remain green and the top
will be flashing. Then after 12 hours the charge is
complete and the top light will remain flashing and
other 3 will be off.
First 3
Hours
12
Hours
Fully
Charged
GREEN
GREEN
GREEN
GREEN
GREEN
GREEN
GREEN
GREEN
NOTE: In the case of a power cut, switching the power socket off or disconnecting the
charging lead, the charging process will not be interrupted as the unit remembers the
previous charge state.
OPERATING

The Baery Status Indicator lights serve 2 purposes:
1 - Whilst charging, the lights indicate which part of the charging cycle the battery is at.
2 - When the Battery Status Indicator Button is pressed (with the Airflo 24 removed
from
4
Quick Start Guide
charge), the lights indicate how much charge remains in the battery.
Baery Status
Indicator Buon
UNDERSTANDING THE BATTERY STATUS INDICATOR LIGHTS
TO DETERMINE THE BATTERY CHARGE STATE, YOU MUST DISCONNECT THE
CHARGING LEAD OR REMOVE THE AIRFLO 24 FROM THE STOWAGE BOARD
AND THEN PRESS THE ‘BATTERY STATUS INDICATOR’ BUTTON TO ACTIVATE THE
BATTERY STATUS INDICATOR LIGHTS
English
4 GREEN
OK to Use
GREEN
GREEN
GREEN
GREEN
3 GREEN
OK to Use
GREEN
GREEN
GREEN
2 GREEN
OK to Use
GREEN
GREEN
1 GREEN
OK to Use
GREEN
RED
DO NOT USE
RECHARGE
RED

Healthcare solutions designed for life
5
Quick Start Guide
DURING A LIFT
AFTER COMPLETING A LIFT
Disconnect the inflatable unit and hand control and pack away. Turn off the compressor
and the bottom green light will go out. The product will also auto turn off after 2.5 minutes
of inactivity.
Press the Battery Status Indicator Button.
4 GREEN
OK to Use
GREEN
GREEN
GREEN
GREEN
3 GREEN
OK to Use
GREEN
GREEN
GREEN
2 GREEN
OK to Use
GREEN
GREEN
1 GREEN
OK to Use
GREEN
RED
DO NOT USE
RECHARGE
RED
Each time the hand control is operated the unit does a battery test and indicates the
battery state. The hand control button is then released when a compartment in the lifting
device is filled.
4 GREEN
OK to Use
GREEN
GREEN
GREEN
GREEN
3 GREEN
OK to Use
GREEN
GREEN
GREEN
2 GREEN
OK to connue
the li
GREEN
GREEN
1 GREEN
OK to connue
the li
GREEN
RED
OK to connue
the li - Recharge
when the li is
complete
RED
If you see a single solid red light (if this is within a lift - the lift can be completed), the Airflo
should be put on unbroken charge immediately for a minimum of 3 hours. This will provide
sufficient capacity to complete one more lift.
Once the red light shows to achieve full capacity a full 12 hour charge is required.
English

6
Intended Use
The Airo 24 by Mangar Health is a low pressure
air compressor, which has been designed to
inate the Mangar ‘Camel’ and ‘ELK’ pneumac
liing equipment through a 4-Way pneumac
hand control. The 4-Way hand control is
supplied with the Camel and ELK.
Applied Parts (Accessories) for the Airo 24:
Mangar ELK (Emergency liing cushion)
Mangar Camel (Emergency liing cushion)
Contents
Intended Use Pg 6
Warnings Pg 6
Baery Status Indicator Pg 7
Baery Charging Pg 8
Mains Power Adaptor (oponal) Pg 9
Operaon Pg 9
Cleaning & Storage Pg 10
Maintenance Pg 10
Trouble shoong guide Pg 11
Specicaons Pg 11
Symbols used on Airo 24 and Charger Pg 12
Symbols used on packaging Pg 12
Warranty Pg 13
Declaraon of Conformity Pg 51
WARNINGS
•Be sure that you have read and understood
these instructions and those provided with any
other Mangar Health product, before
operating the equipment.
•To avoid the risk of electric shock, NEVER
recharge the battery in the bathroom,
outdoors or in a damp environment.
•Always disconnect from the battery charger
before cleaning.
•Use only the supplied battery charger.
•Do not operate the Airflo 24 compressor
in the presence of flammable anaesthetic
mixtures with air, oxygen or nitrous oxide, as
these mixtures can create a combustible
environment leading to the potential for fire or
even an explosion in the presence of a spark.
•Ensure that when in use, the Airflo 24 and its
connected hoses are laid out in such a way as
to avoid any risk of tripping or strangulation.
•Ensure that children and pets do not interfere
with the safe operation of the Airflo 24 and
inflatable product.
•Keep the Airflo 24 and inflatable product away
from sources of heat.
•Disconnect the Airflo 24 compressor from the
inflatable product before transporting so as to
avoid accidental inflation.
• The Airflo 24 should only be used for its stated
purpose and should not be modified in any way.
•The optional mains power adaptor is specific
to the Airflo 24 and is the only approved unit
to be used for this purpose.
•Do not operate the compressor outside of the
specified duty cycle of 20% ‘Operating Time’
and 80% ‘Non-Operating Time’. Maximum
continuous operating time = 5 minutes.
•To calibrate the battery and battery indicator, it
is essential before the first use to charge for 12
hours and then disconnect from the charger
and leave for 90 minutes to allow the battery
management software to correctly calibrate.
Failure to do this could lead to incorrect
battery charging and battery ladder indication.
This product and associated packaging does not
contain latex.
The ELK requires assistance to help li a paent
from the oor to an elevated sing posion.
Once in a sing posion, the paent can be
transferred to a chair or helped to a standing
posion.
The Camel can be operated with or without
assistance and when inated, lis the paent
into a raised seated posion ready to stand.
No special skills or training are required to
operate the Airo 24 and its accessories.
Further guidance on operaon can be found in
the individual accessory User Instrucons.
The Airo compressor and applied parts are
intended for connual re-use as required and
no special acons or operaons are needed to
achieve this.
English
Airo 24

Healthcare solutions designed for life
7
WARNING: The battery must be put on charge immediately as soon as the red indicator light
appears. Running the Airflo24 on RED until the battery is exhausted will damage the battery.
When the compressor is running, the baery status indicator lights will only indicate the state of
baery charge as it was at the start of the li. When the compressor stops running, the baery
status indicator will show a bottom green light (this will be red if the battery charge has become
low).
IMPORTANT: To maximize the life and lifting performance of the battery, fully recharge the
battery after use.
Fully charge the battery for a minimum of 12 hours before using the Airflo 24 for the first time
and when a new battery has been fitted. Then disconnect from the charger and leave for 90
minutes to allow the battery management software to correctly calibrate. Failure to do this
could lead to incorrect battery charging and battery ladder indication.
Power Button
When the ‘Power’ buon is pressed, the compressor will momentarily run and
then stop. The boom light on the indicator will show GREEN.
RED
Recharge
Immediately
2 GREEN
OK to Use
3 GREEN
OK to Use
4 GREEN
OK to Use
1 GREEN
OK to Use
English
Airo 24
Battery Status Indicator
When the Battery State Indicator button is pressed, the lights indicate how much
charge remains in the battery.

8
1
B
A
First 3
Hours
12
Hours
English
Battery Charging - fig. 1
•Check the baery charger casing and lead
for signs of damage and if damaged DO NOT
USE.
•Fit the appropriate mains plug adaptor to
the baery charger (A).
•Plug the baery charger into a mains power
socket and connect the charger output lead
to the Airo 24 charging socket (B).
NOTE: When first fitting a battery they must
be charged for a minimum of 12 hours to
achieve full capacity and then disconnect from
the charger and leave for 90 minutes to allow
the battery management software to
correctly calibrate. On subsequent charges
the Airflo calculates the charging period
based on usage and will not necessarily start
on the bottom light flashing.
When the battery is charged:
•For the first 3 hours, the bottom light will
flash green. After 3 hours the first light will
become a steady green and the second light
will start flashing. After 6 hours the bottom 2
lights will remain green and the third light
will flash. After 9 hours the bottom 3 will
remain green and the top will be flashing.
Then after 12 hours all lights will briefly
become steady green.
•The compressor will move into an energy
saving mode with the top LED flashing every
6 seconds and the other 3 will be off.
NOTE: The compressor will not run whilst the charger is connected.
To maintain battery condition, it is recommended that the Airflo 24 is left continuously on charge
when not in use. The batteries cannot be overcharged using this charger.
WARNING: DO NOT recharge the batteries in a bathroom or in a damp environment.
Airo 24

Healthcare solutions designed for life
9
AB
A
3
•Connect the four hoses (A) from the
4-Way Hand Control to the corresponding
numbered and coloured connectors on the
Mangar Camel or ELK.
•Connect the hose (B) from the hand control
to the Airo 24 air outlet socket (C). The
connector will ‘click’ into place when rmly
pushed onto the socket.
•To switch on, press the ‘Power’ buon (D).
The compressor will run when a buon on
the hand control is pressed and stop when
the buon is released. An internal pressure
switch will stop the compressor running
when each corresponding secon of the
inatable product has reached full pressure.
C
D
Mains Power Adaptor (optional)
An optional mains power adaptor is available to
power the Airflo 24 directly from a mains power
socket. The adaptor is specific to the Airflo 24
and is the only approved unit to be used for
this purpose. Connect the adaptor between a
mains power socket and the adaptor inlet
socket on the end of the Airflo 24 (Fig. 2).
NOTE: This adaptor, has a non-standard output
connector and powers the Airflo 24 directly,
it will not charge the battery. The only way to
charge the battery is to use the charger which
was supplied with the Airflo 24. With the power
adaptor connected, the Airflo 24 behaves
in the same way as if it was running on the
battery, the only functional difference is
that the battery indicator will show bottom
LED orange whilst the Airflo is running. With
the adaptor connected, the Airflo 24 will
also run if there is no battery fitted.
Operation - fig.
Disconnect the Airo 24 from the baery
charger and place it in a safe and convenient
posion. NOTE: The Airo compressor should
not be placed in a posion where it is dicult
to operate the ‘Power’ buon in an emergency.
2
•Pressing any of the numbered hand control buttons will deliver air to the corresponding
numbered hose (follow the User Instructions for the Mangar Health product that is to be
inflated).
•To switch the Airflo 24 off after use, press the ‘Power’ button. The bottom indicator light will go
out. The pump will also automatically turn off after 2.5 minutes of inactivity.
English
Airo 24
3

10
Maintenance - g. 4
The Airo 24 has been designed to be maintenance
free and apart from replacing the baery, the Airo
contains no other user serviceable parts.
Baery replacement:
•Turn the Airo 24 onto its end containing the
air outlet socket, remove the screw (A) and
slide the end panel (B) away from the housing.
•Disconnect the wire from the battery pack
and remove the battery pack from the battery
compartment.
•Fit the replacement battery pack, ensuring that
the connector is secure, refit the end panel and
secure the screw (A).
•To calibrate the battery indicator, it is essential
before use to charge for 12 hours and then
disconnect from the charger and leave for 90
minutes to allow the battery management
software to correctly calibrate.
A
B
4
Disposal
When the Airo 24 has reached the end of it’s working life, remove the baery for safe disposal
in accordance with local regulaons, and dispose of the Airo 24 at your local household waste
recycling centre (civic amenity site).
Cleaning
Unless contaminaon is suspected, periodic cleaning is all that is recommended. The Airo 24
should be cleaned/disinfected as follows:
•Unplug the Airo 24 from the baery charger. Clean the external surfaces using a non abrasive
damp cloth with warm water. Wipe dry thoroughly with a dry towel before reconnecng to the
baery charger. Do not allow moisture to enter any part of the compressor.
•Do not use solvent-based products.
Disinfecng
•Pre-clean using a damp cloth and then disinfect using a hard surface alcohol impregnated wipe.
Storage
•When not in use, store in a dry place together with these instructions. To maintain the
batteries, when being stored for long periods, leave the Airflo 24 on charge.
When it becomes apparent that the Airo 24 is producing fewer than expected liing cycles, the
baery is probably reaching the end of its working life and should be replaced.
If a fault should develop or if you require assistance in seng up using or maintaining the equipment,
please contact your supplier.
English
Airo 24

Healthcare solutions designed for life
11
Specifications
Airflo 24
Weight:
Output pressure:
Batteries:
Battery Life:
4.1Kg
0.35 bar
1 x 24v - 2200mAh Ni-MH Airflo 24 Battery Pack (Mangar Prt No. CD0314)
Greater than 500 charging cycles (2 years depending on usage)
Environmental protection: IP41 (Protection from entry by solid objects with a diameter or thickness greater
than 1.0mm & dripping water)
Type BF equipment
This equipment has been tested and meets the requirements of current EMC legislation
Charger: Model number - HX2902SR
Input:
Output:
AC 100v - 240v, 50-60Hz, 0.5A Max DC
30v, 0.45A
Class II Double insulated equipment.
Trouble shoong guide
Problem Remedy
•Compressor does not run when the 4 way hand
control buttons are pressed.
•Ensure that the Airflo 24 has been switched on.
•Ensure the charger is disconnected from the Airflo
•Check that there are no kinks in the air hose.
•Compressor keeps stopping when trying to
inflate the lifting aid. •Check that there are no kinks in the air hose.
•Bottom Battery status indicator Flashing Red
every 2 seconds
•Compressor causes electrical interference with
another piece of electrical equipment.
•Move either the compressor or other item
further away from each other until the
interference stops.
English
Airo 24
•Bottom Battery status indicator solid Red Light •Low Battery Capacity, recharge.
•Bottom Battery status indicator flashing Red every
½ second
•Internal Pump Temperature too high – allow Airflo
to cool down 60-90mins before attempting a lift.
•No Battery status indicator Lights •Unit switched off -turn unit on.
•Thermal fuse on battery has tripped – Allow unit
to cool down 60-90minutess. A solid red LED will
display after temperature normalizes and fuse
resets. This will have reset the charge sequence.
To avoid damage to the Airflo after 12hour
charge, disconnect from charger and allow
another 60-90minutes for battery cool down.
•Battery Voltage low – Re charge. If occurs
repeatedly replace the battery.

12
Symbols used on Airo 24 and Charger labels
Alternang current (AC)
Fuse (with rang)
Class II Double Insulated equipment
Dangerous voltage
Direct current (DC)
Duty cycle
I = Operang me
O = Non operang me
(Max connuous operang me = 5
minutes)
For indoor use only. Do Not connect
to mains power in a bathroom or
outdoors.
Recycle
Do not dispose of in household waste
Type BF equipment (protecon
against electric shock)
Max/Min operang environment
temperature
Aenon consult instrucons
Warning
Power Buon
Date of manufacture
Baery Status Indicator Buon
English
Airo 24
Symbols used on packaging
Keep packaging dry
Handle with care
Max/Min storage temperature
Net & gross weights
Stack no more than ve high
Recycle packaging
Do not dispose of in household waste
Max/Min humidity when transporng
or in storage
Max/Min pressure when transporng or
in storage
Max/Min temperature when
transporng or in storage

Healthcare solutions designed for life
13
Airflo 24 serial number: .............................
Date of purchase: .............................
Invoice/Delivery note number: .............................
Warranty
The warranty is given against defects in material and workmanship and comprises piece, parts
and labour. The warranty is not valid in the case of fair wear and tear, wilful damage, common
neglect, misuse or unauthorised alteration or repair by unqualified persons.
The following statements do not affect the consumer’s statutory rights. This product is sold
on the understanding that in the event of any defect of manufacture or material appearing
within the warranty period, the defect will be rectified free of charge providing that:
(a) Reasonable evidence is provided (e.g. purchase invoice, delivery note) that the claim is
being made within the warranty period.
(b) The defect is not attributable to accidental damage (either in transit or otherwise),
misuse or unauthorised repair.
Mangar Internaonal, trading as Mangar Health, reserves the right to alter product specicaons and/or any of the informaon contained within this
document without noce.
English
Airo 24

14
Deutsch
Kurzanleitung
Diese Kurzanleitung ersetzt nicht die Bedienungsanleitung.
Bie lesen Sie vor der Benutzung des Airo 24 die beiliegende volle
Bedienungsanleitung.
Bringen Sie den passenden
Netzsteckeradapter am Ladegerät an
(siehe „Auaden des Akkus“, Seite 20).
Verbinden Sie das Ladegerät mit dem
Ladeanschluss auf der Rückseite des
Air lo 24.
VOR DER NUTZUNG
WARNUNG: Der Airflo 24 muss vor dem ersten Gebrauch 12 Stunden lang aufgeladen
werden. Um die Batterie und die Batterieanzeige zu kalibrieren, müssen Sie vor dem
ersten Gebrauch 12 Stunden lang aufladen, dann vom Ladegerät trennen und 90
Minuten warten, damit die Batteriemanagementsoftware korrekt kalibrieren kann.
Andernfalls kann es zu einer falschen Batterieladung und Anzeige der Batterieleiter
kommen.
1 x Airo 24 Kompressor (A) mit 24V - 2200
mAh NiMH-Akku.
1 x Ladegerät (mit Adaptern) (B).
1 x Bedienungsanleitung.
1 x Netzstromadapter (Oponal).
A
B
STANDARD-PACKUNGSINHALT

Healthcare solutions designed for life
15
Deutsch
Kurzanleitung
HINWEIS: Den Kompressor NICHT laufen lassen, während das Ladegerät angeschlossen ist.
BENUTZUNG
Schalten Sie den
Kompressor ein (Power-
Knopf).
3
Schließen Sie die
Schläuche zunächst an das
Hebekissen an (4-farbig
und nummeriert).
1
Schließen Sie den
fünen Schlauch
des Kissens an den
Kompressor an.
2
Drücken Sie die
Handbedienung
zum Aufpumpen des
Hebekissens.
4
VOR GEBRAUCH DEN AKKU 12 STUNDEN LANG
AUFLADEN. Um die Batterie und die Batterieanzeige
zu kalibrieren, müssen Sie vor dem ersten Gebrauch
12 Stunden lang aufladen, dann vom Ladegerät
trennen und 90 Minuten warten, damit die
Batteriemanagementsoftware korrekt kalibrieren
kann.
In den ersten 3 Stunden blinkt die
unterste Leuchte grün. Nach 3 Stunden hört die
erste Leuchte auf zu blinken und beginnt grün zu
leuchten und die zweite Leuchte beginnt zu
blinken. Nach 6 Stunden leuchten die
beiden untersten Leuchten grün und die dritte
Leuchte blinkt. Nach 9 Stunden leuchten die 3
unteren Leuchten grün und die oberste Leuchte
blinkt. Nach 12 Stunden ist der Ladevorgang
abgeschlossen. Die obere Leuchte blinkt weiter
und die anderen 3 gehen aus.
HINWEIS: Bei Stromausfall, Ausschalten der Netzsteckdose oder Trennen des Ladegeräts von der
Stromzufuhr wird der Ladevorgang nicht abgebrochen, da das Gerät sich den vorherigen Ladezustand
merkt.
12
Stunden
Vollständig
geladen
Erste 3
Stunden
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN

16
Deutsch
Die Akkuladestandsanzeige hat 2 Aufgaben:
1 - Während des Ladevorgangs zeigen die Leuchten an, wie weit der Ladezyklus des Akkus
bereits fortgeschrien ist.
2 - Wenn der Airo 24 vom Ladegerät getrennt ist, zeigt die Akkuladestandsanzeige nach Druck
auf die entsprechende Taste den aktuellen Ladestand des Akkus an.
ERKLÄRUNG DER AKKULADESTANDSANZEIGE
UM DEN AKKULADESTAND KORREKT ANGEZEIGT ZU BEKOMMEN,
MUSS DAS LADEKABEL ZUNÄCHST ABGEZOGEN WERDEN
BZW. DER AIRFLO 24 MUSS AUS SEINER WANDHALTERUNG
GENOMMEN WERDEN. BETÄTIGEN SIE ANSCHLIESSEND DIE TASTE
„AKKULADESTANDSANZEIGE“, UM DIE KONTROLLLEUCHTEN DER
AKKULADESTANDSANZEIGE ZU AKTIVIEREN
Wird die Taste „Akkuladestandsanzeige“ sofort nach Abziehen des
Ladegeräts gedrückt, leuchten möglicherweise die vier Leuchten grün.
Dies liegt daran, dass am Akku immer noch die Ladespannung vorliegt,
die höher als die Ruhespannung des Akkus ist.
Kurz darauf wechselt die Akkuspannung in die Ruhespannung. Zu diesem
Zeitpunkt leuchten bei einem Akkutest unter Umständen drei oder zwei
Leuchten grün. Dies ist abhängig vom jeweiligen Akkustand.
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN
GRÜN
Kurzanleitung
Taste
“Akkuladestandsanzeige”

Healthcare solutions designed for life
17
Deutsch
Trennen Sie das aufblasbare Produkt und die Handbedienung voneinander und verstauen Sie
sie. Schalten Sie den Kompressor aus – die unterste grüne Leuchte erlischt. Das Produkt wird
auch automatisch ausgeschaltet, wenn es länger als 2,5 Minuten inaktiv ist.
Betätigen Sie die Taste „Akkuladestandsanzeige“.
4 GRÜN
Kann verwendet
werden
3 GRÜN
Kann verwendet
werden
GRÜN
GRÜN
GRÜN
2 GRÜN
Kann verwendet
werden
GRÜN
GRÜN
1 GRÜN
Kann verwendet
werden
GRÜN
ROT
NICHT VERWENDEN
– LADEN
GRÜN
Bei jeder Betätigung der Handbedienung führt das Gerät einen Akkutest durch und zeigt
den Akkuladestand an. Die Taste an der Handbedienung wird losgelassen, sobald ein
Kissensegment der Hebevorrichtung gefüllt ist.
4 GRÜN
Kann verwendet
werden
GRÜN
GRÜN
GRÜN
GRÜN
3 GRÜN
Kann verwendet
werden
GRÜN
GRÜN
GRÜN
2 GRÜN
Anhebevorgang
kann fortgesetzt
werden
GRÜN
GRÜN
1 GRÜN
Anhebevorgang
kann fortgesetzt
werden.
GRÜN
ROT
Anhebevorgang
kann fortgesetzt
werden – Nach
Abschluss des
Vorgangs auaden.
ROT
Wenn während eines Anhebevorgangs eine einzelne Leuchte rot leuchtet, kann der Vorgang
noch beendet werden. Allerdings sollte der Airflo danach sofort mindestens 3 Stunden lang
ununterbrochen aufgeladen werden. Danach hat er wieder ausreichende Kapazitäten für
einen weiteren Anhebevorgang.
Sobald das rote Licht leuchtet, ist ein Aufladen über volle 12 Stunden erforderlich, um die
volle Leistung wiederherzustellen.
Kurzanleitung
GRÜN
GRÜN
GRÜN
GRÜN
WÄHREND EINES ANHEBEVORGANGS
NACH ABSCHLUSS DES ANHEBEVORGANGS

18
Deutsch
Verwendungszweck
Der Airflo 24 von Mangar Health ist ein
Niederdruck-Luftkompressor, der zum
Aufblasen der pneumatischen Hebehilfen
„Camel“ und „ELK“ von Mangar Health mittels
einer pneumatischen 4-Wege-Handbedienung
konzipiert wurde. Die 4-Wege-Handbedienung ist
im Lieferumfang von Camel bzw. ELK
enthalten.
Mit dem Airflo 24 verwendbare Produkte
(Zubehör):
Mangar ELK (Notfall-Hebekissen)
Mangar Camel (Notfall-Hebekissen)
Beim ELK wird ein Helfer benötigt, der
dem Patienten vom Boden in eine erhöhte
Sitzposition aufhilft. Aus dem Sitzen kann der
Patient dann auf einen Stuhl transferiert oder in
den Stand gebracht werden.
Das Camel kann mit und ohne Unterstützung
bedient werden. Während des Aufpumpvorgangs
wird der Patient in eine erhöhte Sitzposition
gebracht, aus der er aufstehen kann.
Zur Bedienung des Airflo 24 und seines
Zubehörs sind keine besonderen Kenntnisse
oder Schulungen erforderlich.
Weitere Orientierungshilfen zum Betrieb
sind der Bedienungsanleitung des jeweiligen
Zubehörs zu entnehmen.
Airflo 24 und Zubehör sind für den wiederholten
Gebrauch nach Bedarf vorgesehen. Dazu
sind keine besonderen Maßnahmen oder
Vorgehensweisen erforderlich.
Inhalt
Verwendungszweck S. 18
Warnhinweise S. 18
Kontrollleuchte für Akkuladestand S. 19
Auaden des Akkus S. 20
Netzstromadapter (oponal) S. 21
Betrieb S. 21
Reinigung und Auewahrung S. 22
Wartung S. 22
Fehlerbehebung S. 23
Technische Daten S. 23
Symbole auf Airo 24 und Ladegerät S. 24
Symbole auf der Verpackung S. 25
Garane S. 25
Konformitätserklärung S. 50
WARNHINWEISE
•Bitte vergewissern Sie sich, dass Sie
diese Bedienungsanleitung und die
Bedienungsanleitungen zu anderen verwendeten
Produkten von Mangar International vollständig
gelesen und verstanden haben, bevor Sie die
Hebehilfen in Betrieb nehmen.
•Um die Gefahr eines elektrischen Schlags
zu vermeiden, laden Sie den Akku NIE im
Badezimmer, im Freien oder in einer feuchten
Umgebung auf.
•Trennen Sie den Akku vor der Reinigung stets vom
Ladegerät.
•Verwenden Sie zum Aufladen ausschließlich das
mitgelieferte Ladegerät.
Verwenden Sie den Airflo 24 Kompressor nicht in
Gegenwart von brennbaren Anästhetikagemischen
mit Luft, Sauerstoff oder Distickstoffoxid, da diese
Gemische leicht entflammbar sind und zu
Brandgefahr oder auch Explosionsgefahr durch
Funkenflug führen können.
•Achten Sie darauf, dass bei Verwendung des Airflo
24 Kompressors der Kompressor selbst und der
angeschlossene Schlauch so platziert werden, dass
keine Stolper- oder Strangulationsgefahr besteht.
•Stellen Sie sicher, dass Kinder oder Haustiere nicht
den sicheren Betrieb des Airflo 24 und des
aufblasbaren Produkts behindern.
•Halten Sie den Airflo 24 und das aufblasbare
Produkt von Wärmequellen fern.
•Trennen Sie den Airflo 24 Kompressor vor dem
Transport vom aufblasbaren Produkt, um ein
versehentliches Aufpumpen des Produkts zu
vermeiden.
•Der Airflo 24 darf nur zum angegebenen
Verwendungszweck verwendet und in keiner Weise
modifiziert werden.
•Der optionale Netzstromadapter wurde speziell für
den Airflo 24 entwickelt und ist der einzige für diese
Anwendung zugelassene Adapter.
•Den Kompressor nicht außerhalb des angegebenen
Arbeitszyklus von 20 % „Betriebsdauer“ und 80 %
„Ruhezeit“ betreiben. Maximale ununterbrochene
Betriebsdauer = 5 Minuten.
•Um die Batterie und die Batterieanzeige zu
kalibrieren, müssen Sie vor dem ersten Gebrauch 12
Stunden lang aufladen, dann vom Ladegerät
trennen und 90 Minuten warten, damit die
Batteriemanagementsoftware korrekt kalibrieren
kann. Andernfalls kann es zu einer falschen
Batterieladung und Anzeige der Batterieleiter
kommen.
Dieses Produkt und seine Verpackung
enthalten kein Latex.
Airo 24

Healthcare solutions designed for life
19
Deutsch
Kontrollleuchte für Akkuladestand
Wenn die Ein/Aus-Taste gedrückt wird, läu der Kompressor für einen Moment
und stoppt dann. Die unterste Kontrollleuchte der Anzeige leuchtet GRÜN.
HINWEIS: Der Akku muss sofort aufgeladen werden, wenn die Kontrollleuchte rot aufleuchtet.
Der Betrieb des Airflo 24 bei ROTER Kontrollleuchte bis zur völligen Entladung des Akkus hat
Schäden am Akku zur Folge.
Wenn der Kompressor läu, zeigt d ie Akkuladestandsanzeige n ur d en Akkuladestand zu Beginn
des Hebevorgangs an. Wenn der Kompressor stoppt, leuchtet die unterste Kontrollleuchte für
den Akkuladestand grün (bzw. rot, wenn der Akkuladestand nun gering ist). Um den tatsächlichen
Akkuladestand zu ermieln, drücken S ie d ie Taste „Akkuladestandsanzeige“; d ann zeigen die
Kontrollleuchten den aktuellen Akkuladestand an.
WICHTIG: Zur Ausschöpfung der maximalen Nutzdauer und Hebeleistung des Akkus laden Sie
den Akku nach Gebrauch vollständig auf.
Laden Sie den Akku mindestens 12 Stunden lang vollständig auf, bevor Sie den Airflo 24 zum
ersten Mal verwenden und wenn ein neuer Akku eingelegt wurde. Trennen Sie dann das
Ladegerät vom Gerät und lassen Sie es 90 Minuten lang stehen, damit die
Batteriemanagementsoftware korrekt kalibrieren kann. Andernfalls kann es zu einer falschen
Batterieladung und Anzeige der Batterieleiter kommen.
ROT
Sofort
aufladen
3 GRÜN
Kann
verwendet
werden
4 GRÜN
Kann
verwendet
werden
2 GRÜN
Kann
verwendet
werden
1 GRÜN
Kann
verwendet
werden
Airo 24
Akkuladestandsanzeige
Wenn die Taste „Akkuladestandsanzeige“ gedrückt wird, zeigen die Leuchten an,
wie hoch der Ladestand des Akkus noch ist.

Aufladen des Akkus - Abb. 1
20
Deutsch
1
B
A
Erste 3
Stunden
12
Stunden
•Prüfen Sie Akkugehäuse und Kabel auf
Anzeichen von Schäden und VERWENDEN
SIE sie bei Beschädigung NICHT.
•Bringen Sie den passenden
Netzsteckeradapter am Ladegerät an (A).
•Verbinden Sie das Ladegerät mit einer
Netzsteckdose und stecken Sie den Stecker
des Ladegeräts in die Ladebuchse des Airflo 24
(B).
HINWEIS: Die Akkus müssen mindestens
12 Stunden lang aufgeladen werden, damit sie
ihre volle Kapazität erreichen. Bei allen weiteren
Ladevorgängen berechnet der Airflo die
erforderliche Ladedauer anhand der bisherigen
Nutzung und eine Leuchte blinkt (nicht
unbedingt die unterste).
Während des Auadens des Akkus:
Airo 24
•In den ersten 3 Stunden blinkt die unterste
Leuchte grün. Nach 3 Stunden hört die
erste Leuchte auf zu blinken und beginnt
grün zu leuchten und die zweite Leuchte
beginnt zu blinken. Nach 6 Stunden
leuchten die beiden untersten Leuchten
grün und die dritte Leuchte blinkt. Nach 9
Stunden leuchten die 3 unteren Leuchten
grün und die oberste Leuchte blinkt. Nach
12 Stunden leuchten schließlich alle
Leuchten grün.
•Der Kompressor schaltet in einen Energiesparmodus, wobei die oberste LED alle 6 Sekunden
blinkt, während die anderen drei LEDs nicht leuchten.
HINWEIS: Der Kompressor läuft nicht, solange das Ladegerät angeschlossen ist.
Um die Akkus in gutem Zustand zu halten, wird empfohlen, den Airflo 24 bei Nichtgebrauch
ständig am Ladegerät zu lassen. Die Akkus können mit diesem Ladegerät nicht überladen werden.
WARNUNG: Laden Sie die Akkus NICHT im Badezimmer oder in einer feuchten Umgebung auf.
Um das Ladegerät vom Netzstrom zu trennen, ziehen Sie das Ladegerätekabel aus der
Netzsteckdose.
Other manuals for Airflo 24
1
Table of contents
Languages:
Other Mangar Health Air Compressor manuals
Popular Air Compressor manuals by other brands

Powermate
Powermate 200-2235 Operator's manual

Nash Elmo
Nash Elmo G 400 Series operating instructions

Porter-Cable
Porter-Cable CPF23400P instruction manual

ULTIMATE SPEED
ULTIMATE SPEED UMK 10 A1 translation of original operation manual

Koolatron
Koolatron 40232 user manual

Panasonic
Panasonic DD73C14GBU6 Specification sheet