
OPERATION OPERATION
The power meter can now be used for measuring RF power. For full instructions and
uncertainty calculations, refer to the power meter Operating Manual, Chapter 3.
Linearity Factor Correction for 6920 Series Sensors
At -20 dBm, corrected reading =Displayed reading +Correction Factor
where Correction Factor =Sensor Lin Factor - 8%
For each 3 dB decrease in power level, this correction figure should be halved. At
-30 dBm or lower the error will be negligible and the above correction is not necessary.
The linearity factor of 6920 Series sensors can vary from unit to unit. Since the 6950 has
two preset linearity factor correction values, measurement errors will result when using 6920
Series sensors with a linearity factor that is significantly different from this value, and when the
power level is above -30 dBm. If necessary, the displayed reading can be corrected to give a
more accurate power measurement figure, as follows:
Adjust the ZERO control for zero reading on the meter, using the special tool
provided with the power meter.
With no power applied to the sensor, select the most sensitive range by turning the
RANGE switch fully counter-clockwise.
Connect the sensor to the SENSOR INPUT socket of the power meter using the
sensor cable supplied with the power meter.
(2)
(3)
It may be helpful in setting zero to adjust the RESPONSE TIME control on the rear panel
to reduce noise. You may also find it easier to set zero first on a less sensitive range, as
slight adjustments of the ZERO control have considerable effect on the most sensitive
range.
Hint...
WITH 6950 POWER METER
Zeroing
(1)
With no power applied to the sensor, press [AUTO ZERO]. Five dashes appear
on the display, representing the power meter's five ranges. When the last of these
disappears, all five ranges have been zeroed. This takes approximately 25
seconds.
Connect the sensor to the SENSOR INPUT of the power meter, using the sensor
cable supplied with the pow.ermeter.
(2)
WITH 6960 SERIES POWER METER
Zeroing
(1)
(2) Attach the appropriate magnetic range scale to the skirt of the RANGE switch if
required and set the switch to 0 dBm.
Use the N-type to SMA adapter for 6913/6914/6934
Use the 75 Qto 50 Qadapter for 6919
Use the 30 dB pad for 6920
Use the 30 dB pad and N-type to SMA adapter for 6923/6924
Calibration
(1) Connect the sensor to the POWER REFERENCE output of the power meter.
(5) Adjust the GAIN control for full-scale meter reading.
(3) Set the CAL FACTOR control to the value of the reference calibration factor
given on the sensor label. Some sensors, e.g. the 6910, do not have this
information on the label; in these cases the default value of 100% is assumed.
(4) On the rear panel, switch POWER REF to ON and the
UN
F : 6%/8% switch to
the setting which is nearest to the value of the linearity factor shown on the
sensor label or its calibration data chart. (If the
UN
F switch positions on your
6950 are designated 75 Qand 50 Q, these should be interpreted as 6% (75 Q) and
8% (50
Q)
and the switch set accordingly as above.)
(6)
(7)
Switch POWER REF off and disconnect the sensor from the POWER
REFERENCE socket.
Determine the calibration
factor
for the measurement frequency, whether from
the graph on the sensor label or from its calibration data chart. Set the CAL
FACTOR control to the same value.
Calibration
(1) Determine the linearity factor, either from the label on the sensor or from its
calibration data chart, Press [UNEARITYFACTOR] and enter this value in the
power meter.
(2) Connect the sensor to the POWER REFERENCE output of the power meter.
Use the N-type to SMA adapter for 6913/6914/6934
Use the 75 Qto 50 Qadapter for 6919
Use the 30 dB pad for 6920
Use the 30 dB pad and N-type to SMA adapter for 6923/6924
(3) Press [CAL FACTOR] and enter the value of the reference calibration factor
given on the sensor label. Some sensors, e.g. the 6910, do not have this
information on the label; in these cases the default value of 100% is assumed,
3-2 46882-150K
Feb. 93
46882-150K
~~
N