MARcus TD 2014 Series User manual

User Manual
TD 2014 Series Weighing Controller
Applications: Electronic platform scale, weighing scale, etc.

1
1. Specifications
A/D Converter type
Sigma-Delta
Sample rate
10 times / second, 40 times / second, 250 times / second
Display resolution
3000
Internal resolution
1,000,000
Load cell excitation
DC 5V
Load cell
Up to 4 x 350 Ω or 8 x 700 Ω load cell
Display
6-digit 0.4 inch LED
Division
1/2/5/0.1/0.2/0.5/0.01/0.02/0.05/0.001/0.002/0.005/10/
20/50/100/200/500/0.10/0.20/0.50/0.010/0.020/0.050
Power
DC 9V 1A adapter or AC 110-220V 50 Hz
Operating temperature
0℃~40℃
Storage temperature
-25℃~55℃
Relative humidity
85% RH
Size
95 × 48 × 110 mm
Weight
0.5 kg
Signal output
4-20 mA analog output
Options
a) Opto-isolated RS232 or RS485
b) 6 relay outputs with 2 opto-isolated inputs that can be used
to control the output of relay #1 and #2
2. Model Number
TD 2014□-□□-□
0 - No interface
1 - Optical -isolated RS232C
2 - Optical -isolated RS485
AC- AC 110-220V 50 Hz available for XK315A2-6A series only
DC- DC 9V 1A adapter
A - Basic model with 4-20 mA analog output
B - Basic model plus 6 relay outputs and 2 opto-isolated inputs
3. Key Functions
【* 】Function Select key or Enter key
【→】Right key
【→T←】Tare key or Increment key
【→0←】Zero key or Decrement key
4. Status
〖→T←〗On when tare and the weight is net weight
〖→0←〗On when the weight is zero
〖◣◢〗 On when the weight is stable

2
5. Error Code
〖OUEr〗 - when the weight is > 100% FS + 9d
〖-OUEr〗 - when the weight is < - 20d
〖HHHHHH〗- when zero reference is > +2% FS
〖LLLLLL〗- when zero reference is < -2% FS
6. Connections
POWER SUPPLY
AC110-220V
(DC9V)
4-20mA CURRENT OUTPUT
RS232/485
LOAD CELL
-
DIGITAL INPUT COM
RELAY OUTPUT COM
7. Interface Operation
Serial format 8-N-1: 8 data bits, No parity bit, 1 stop bit.
1. Adr=00 - continuous ASCII reverse data output, if weight is -123.45, the output is 54.321-
2. Adr=99 - continuous ASCII regular data output, if weight is -123.45, the output is -123.45
3. Adr=01-98 - data output using ModBus protocol
8. Operation
1. Turn on the indicator, it displays the version number and checks the LED from “0” to “9”
2. Zero
When the weight is stable between ±2% FS,press【→0←】zero the display and〖→0←〗
is on
When the zero reference is > +2% FS, it displays 〖HHHHHH〗
When the zero reference is < -2% FS, it displays 〖LLLLLL〗
3. Tare
When the weight is stable,〖→T←〗is on
When the weight is greater than zero and stable,〖◣◢〗 is on
Press【→T←】zero display and〖→T←〗is on
4. Clear Tare
When the weight is stable and the weight is net weight,〖→T←〗is on
Press【→T←】clears the tare and〖→T←〗is off

3
9. Configurations
Operation
Display
Descriptions
Press【→】
〖000000〗
Go to Calibrations and Configurations menu and enter
password 002003
Press【→】
Or【→T←】
〖002003〗
To enter password 002003:
Press【→】to move blinking digit to the right
Press【→T←】to increment blinking digit
Press【→0←】to decrement blinking digit
Press【* 】
〖CAL SP〗
In Calibrations menu
Press【* 】
〖-SEt-〗
In Configurations menu
Press【→】
〖d 0.01〗
Set division
Press【→T←】
〖d 0.05〗
Press【→T←】or【→0←】to select d
Press【* 】
〖100.00〗
Set FS
Press【→】
〖000000〗
Left most digit is blinking
Press【→】
Or【→T←】
〖0300.00〗
Press a combinations of【→】,【→T←】and【→0←】to
select FS
Press【* 】
〖FLt 2〗
Set filter. FLt=0-2
The bigger filter will result in slower weight result
Press【→T←】
〖FLt 1〗
Press【→T←】or【→0←】to select filter
Press【* 】
〖AUtP00〗
Set zero trace range and zero at start range
AUtP=00 no zero set at start,zero trace range is 0.4d;
AUtP=10 zero at start range is 20% FS, zero trace range
is 0.4d;
AUtP=20 zero at start range is 20% FS, zero trace range
is 0.8d;
AUtP=30 zero at start range is 20% FS, zero trace range
is 1.2d;
AUtP=40 zero at start range is 20% FS, zero trace range
is 1.6d;
AUtP=50 zero at start range is 20% FS, zero trace range
is 2.0d;
AUtP=60 zero at start range is 20% FS, zero trace range
is 2.4d;
AUtP=70 zero at start range is 20% FS, zero trace range
is 2.8d;
AUtP=80 zero at start range is 20% FS,zero trace range
is 3.2d;
AUtP=90 zero at start range is 20% FS,zero trace is
3.6d;
Press【→】
Or【→T←】
〖AUtP10〗
Press a combinations of【→】,【→T←】and【→0←】to
select AUtP
Press【* 】
〖Adr 00〗
Enter indicator address
Adr=00 continuous reverse data output
Adr=99 continuous regular data output

4
Adr=01-98 data output using ModBus protocol
Press【→】
Or【→T←】
〖Adr 12〗
Press a combinations of【→】,【→T←】and【→0←】to
select Adr
Press【* 】
〖b 2400〗
Set Baud rate
b=1200,2400,4800,9600
Press【→T←】
〖b 9600〗
Press【→T←】or【→0←】to select baud rate
Press【* 】
〖L00791〗
Calibrate the 4 mA analog output
L means current output is 4 mA
Press【→】
Or【→T←】
〖L00791〗
Press a combinations of【→】,【→T←】and【→0←】to
calibrate the 4 mA analog output
Press【* 】
〖H03723〗
Calibrate the 20 mA analog output
H means current output is 20 mA
Press【→】
Or【→T←】
〖H03723〗
Press a combinations of【→】,【→T←】and【→0←】to
calibrate the 20 mA analog output
Press【* 】
〖gAIn 1〗
Set magnification of load cell input
0=low magnification, it is 64 times
1=high magnification, it is 128 times
Press【→】
Or【→T←】
〖gAIn 0〗
Press a combinations of【→】,【→T←】and【→0←】to
change the magnification
Press【* 】
〖rAtE 1〗
Set AD sample rate
0=10 times / second
1=40 times / second
2=250 times / second
Press【→】
Or【→T←】
〖rAtE 2〗
Press a combinations of【→】,【→T←】and【→0←】to
change the sample rate
Press【* 】
〖0.00〗
Save all parameters
10. Calibrations
Operation
Display
Descriptions
Press【→】
〖000000〗
Go to Calibrations and Configurations menu and enter
password 002003
Press【→】
Or【→T←】
〖002003〗
To enter password 002003:
Press【→】to move blinking digit to the right
Press【→T←】to increment blinking digit
Press【→0←】to decrement blinking digit
Press【* 】
〖CAL SP〗
In Calibrations menu
Press【→】
〖CAL 00〗
Make there is no weight on the scale
Press【* 】
〖------〗
Confirm zero calibration
〖100.00〗
Load calibration weight > 30% FS on scale.
For more accurate calibration, use calibration weight close to
FS
Press【→】
Or【→T←】
〖002000〗
Press a combinations of【→】,【→T←】and【→0←】to the
actual weight
Press【* 】
〖------〗
Confirm FS calibration
〖00.00〗
Save calibrations

5
11. Check ADC and Set To Factory Setting
Operation
Display
Descriptions
Press【→】
〖000000〗
Go to Calibrations and Configurations menu and enter
password 002003
Press【→】
Or【→T←】
〖002003〗
To enter password 002003:
Press【→】to move blinking digit to the right
Press【→T←】to increment blinking digit
Press【→0←】to decrement blinking digit
Press【* 】
〖CAL SP〗
In Calibrations menu
Press【* 】
〖-SEt- 〗
In Configurations menu
Press【* 】
〖-A-d- 〗
In Check ADC and Set To Factory menu
Press【→】
〖365235〗
ADC count
Press【* 】
〖0.00〗
Current weight
〖-A-d- 〗
Operate step 1 to step 5 again
Press【* 】
〖FACt 〗
Set to factory setting
Press【→】
〖0.00〗
Indicator is set to factory setting:
d=0.001 FS=3.000 FLt=2 AUtP=10 Adr=00
b=2400 L=00749 H=03723 gAIn=1 rAtE=1
12. Setting set point
Operation
Display
Descriptions
Press【→】
〖000000〗
Go to Set Point menu and enter password 110906
Press【→】
Or【→T←】
〖110906〗
To enter password 110906:
Press【→】to move blinking digit to the right
Press【→T←】to increment blinking digit
Press【→0←】to decrement blinking digit
Press【* 】
〖A1 〗
Set point #1 for relay #1
Press【→】
〖000150〗
Press【→】to access set point #1
Press【→】
Or【→T←】
〖000200〗
Press a combinations of【→】,【→T←】and【→0←】to
set point A1 to 000200
Press【* 】
〖A2 〗
Set point #2 for relay #2,
Follow step 4 and 5 to change set point
Press【* 】
〖A3 〗
Set point #3 for relay #3
Follow step 4 and 5 to change set point
Press【* 】
〖A4 〗
Set point #4 for relay #4
Follow step 4 and 5 to change set point
Press【* 】
〖A5 〗
Set point #5 for relay #5
Follow step 4 and 5 to change set point
Press【* 】
〖A6 〗
Set point #6 for relay #6
Follow step 4 and 5 to change set point
Press【* 】
〖OUt-d 〗
Set state of relay output
Press【→】
〖010001〗
The six bits in OUt-d from left to right represent the
open/close state of the relays from #6 to #1
Example 1:OUt-d=010001 ( relay #5 bit is 1),A5=005000

6
When the weight is <5000, relay #5 will be closed.
When the weight is ≥5000,relay #5 will be opened
Example 2:OUt-d=000001 ( relay #5 bit is 0),A5=005000
When the weight is ≥5000,relay #5 will be closed.
When the weight is <5000, relay #5 will be opened
Press【→】
Or【→T←】
〖100110〗
Press a combinations of【→】,【→T←】and【→0←】to
enter OUt-d
Press【* 】
〖En-o1 〗
Set the output of relays #1
If En-o1=0, enable the output of relay #1
If En-o1=1, use input #1 to enable the output of relay #1
If En-o1=2, use input #2 to enable the output of relay #1
If En-o1=3, disable the output of relay #1
Press【→】
〖000002〗
En-o1=2
Press【→T←】
〖000001〗
Set En-o1=1 to use input #1 to enable output of relay #1
Press【* 】
〖En-o2 〗
Set the output of #2 relay
Follow step 14 to 16 to change relay output setting
Press【* 】
〖En-o3 〗
Set the output of #3 relay
Follow step 14 to 16 to change relay output setting
Press【* 】
〖En-o4 〗
Set the output of #4 relay
Follow step 14 to 16 to change relay output setting
Press【* 】
〖En-o5 〗
Set the output of #5 relay
Follow step 14 to 16 to change relay output setting
Press【* 】
〖En-o6 〗
Set the output of #6 relay
Follow step 14 to 16 to change relay output setting
Press【* 】
〖0.00〗
Save set points
13. Software version
VER2.01 - packing scale
VER2.02 - 3-speed rating scale
VER2.03 - check weighing
VER2.04 - 2-materials, 2-speed batching scale
VER2.05 - 3-materials, 2-speed batching scale
VER2.06 - 4-materials batching scale
VER2.07 - 5-materials batching scale
14. Precautions
1. Do not place indicator in direct sunlight, dusty and vibration environment
2. The weight (including the tare) cannot be more than maximum rated weight
3. Do not use strong solvent (such as: benzene, nitric acid oil) to clean indicator housing
4. Never pour or splash water onto the indicator
5. If indicator fails during operation, turn it off immediately. If you are not scale manufacturers,
please send it back to Caisun for repair
15. Services
Warranty service is provided up to one year from the date of sale for normal operation failures.
Services include a full range of technical on site services and workshop repair, preventative

7
maintenance and calibration facilities.
16. Relay output
Table of contents
Popular Controllers manuals by other brands

InnoMedia
InnoMedia ESBC 9378-4B Quick install guide

WindowMaster
WindowMaster WMS 306-n Installation instructions manual

Corsair
Corsair iCUE COMMANDER CORE XT quick start guide

Lily Tech
Lily Tech ZL-6280A instruction manual

Alfalaval
Alfalaval ESE02126-EN instruction manual

digitalview
digitalview SVP-1280 instructions