Matsushima PFM-AD12S User manual

TPFM-013E
Rev. 0: Mar. 4, 2021
Contents
Safety precautions ........................... 1
1. Summary ................................... 2
2. Specifications ............................ 3
3. Outline ................................... 4
4. Mounting .................................. 5
5. Connecting ................................ 6-7
6. Part names and functions .................. 8
7. Start up .................................. 9-10
8. Set example of parameter .................. 11
9. Maintenance ............................... 12-15
10. Troubleshooting .......................... 16-17
11. Maintenance item ......................... 17
※An operator should read carefully this instruction manual and conduct correct handing.
INSTRUCTION MANUAL
FOR
Separate Type Air Dust Monitor
TYPE
PFM-AD12S
1-8-18 Norimatsu-Higashi,Yahatanishi-ku,Kitakyushu 807-0837 Japan
Phone No.(8193)691-3731 Fax No.(8193)691-3735
http://www.matsushima-m-tech.com
E-mail sales@matsushima-m-tech.com

- 1 -
Safety precautions
Be sure to thoroughly read the instruction manual before using the products.
Keep the instruction manual in a safe, convenient location for future reference.
All or part of the contents described in this manual may be changed without any notice.
Due to our constant striving for further improvement of products, parts or products that differ from those
described in this manual may be substituted.
WARNING (Failure to observe this WARNING may cause a fatal or serious injury.)
Be sure to confirm that any peripheral equipment does not move before installation work.
In addition, observe safety requirements for installation work where high-place work is expected.
Be sure to turn off the power source before wiring, mounting and transportation work. (Failure to observe
this WARNING may result in an electric shock/ injury or equipment damage due to short-circuit.)
Carry out wiring work correctly with reference to a proper drawing.
Never disassemble the equipment. (Failure to observe this WARNING may result in an electric shock.)
Do not open the cover under an explosive environmental condition when power is entered. (Failure to
observe this WARNING may result in an injury or equipment damage.)
Do not place or store the equipment in any hostile environmental place where it will be subjected to direct
sunlight, rain, water droplet, hazardous gas/water, etc.
CAUTION (Failure to observe this CAUTION may cause a moderate injury or equipment damage.)
Do not use the equipment for any purpose other than the original purpose of use.
Be sure to confirm the specification of equipment and use the equipment within the range of specification.
(Mounting conditions such as temperature, power source, frequency, etc.)
Make sure a correct wiring before applying power source.
Do not have a shock or strong impact to the equipment.
(Failure to observe this CAUTION may result in equipment damage.)
Be sure to connect necessary terminals (grounding, etc.).
Remove all wiring to the equipment before doing electrical welding work near the equipment.
Do not forcedly bend or pull the lead wire also do not use unnecessarily long wire.
Tighten the cover, lead outlet, etc. properly so that dust, rainwater, etc. do not enter inside the equipment.
Do not use the equipment under a corrosive condition (NH3, SO2, CI2, etc.).
Be sure to tighten the cable grand so that outer air does not enter inside the equipment.
When applying piping connection such as conduit, etc. instead of cable grand, apply putty or equivalents
on the cable entry so that outer air does not enter inside the equipment.
Do not directly touch the probe with hands regardless of power ON or OFF.
If you unavoidably have to touch the probe with hands when installation or maintenance process, be sure
to turn OFF the power and use non-conductive gloves such as rubber gloves during the process.
*Touching the probe with hand directly may cause a failure due to static electricity charge in a human body.
MAINTENANCE: When you clean the probe, be sure to turn OFF the power, put on non-conductive gloves
such as rubber gloves to clean the probe with moist cloth.*Touching the probe with hand directly may cause
a failure due to static electricity charge in a human body. Cleaning the probe with dry cloth may also cause
a failure due to friction charge generated between the cloth and probe.
IMPORTANT (indicates notes or information to help customers.)
Limitations of Warranty:
Warranty period shall be one year from the date of delivery (ex-factory).
Any damage of any other products that have occurred for use of the equipment is not covered by this
warranty. Also any loss induced by failure or malfunction of the equipment is not covered by this warranty.
Failure or malfunction caused by following are not covered by this warranty:
a. Modification or repair by a party other than MATSUSHIMA’s authorized personnel,
or replacement of parts not recommended by MATSUSHIMA.
b. Inadequate storage, installation, use, inspection or maintenance that does not comply
with specifications.
c. Cause for any peripheral equipment or device.
d. Accident beyond control and force majeure (fire, earthquake, flood, riots, etc.).
Lack of instructions to MATSUSHIMA for information or safety requirements that can be predicted only by
customers’ side.
This warranty conditions do not limit customers’ legal right.
Price for the equipment does not include any charge for services such as commissioning, supervising, etc.
!
!
!

- 2 -
1. Summary
Air Dust Monitor monitors the flying dust concentration for 24 hrs.
After automatically taking in the air including dusts from the air inlet, the dust solid
particles in the air make a transfer of electric charge to the electrode by touching with or
passing near the probe of Air Dust Monitor.
This transfer of electric charge is widely known as triboelectricity or contact charging.
This product will output a 4 to 20mA signal in proportion to the dust amount and some contact
signals, through filtering and amplifying the transfer amount of electric charge.
Important: Dust concentration measurement of triboelectric type Dust Monitor
The triboelectric type Dust Monitor will change an output value in its measurement principle,
corresponding to a variety of environmental factors such as “kind of particle”, “particle
diameter”, “specific gravity”and “flow velocity”.
However, obviously the linearity under a fixed environment will not change.
Regarding the concentration of the triboelectric type dust monitor, as mentioned above, it is
required to consider the effect of physical property change and it is not necessarily
corresponding to the specified concentration.
!

- 3 -
2. Specifications
1.
Power supply
AC100V -15%/+10% 50/60Hz (specified separately in case of AC 220V)
2.
Object to be measured
Floating dust
3.
Particulate material
All solid particles
4.
Particle size
≧0.3µm
5.
Particle concentration
0.1mg/m3to 1000mg/m3
6.
Measuring system
Triboelectric system
7.
Circumstances
7-1.
Temperature
-10℃to +50℃
* If using at Sensitivity +10℃to +40℃
7-2.
Humidity
Max. 95%RH
* If using at Sensitivity Max. 85%RH
8.
Output signal
8-1.
Analog output
DC4-20mA(Max.23.3mA/Isolated output/Load resistance: Max.500Ω)
(* Be sure to use a shielded cable)
8-2.
Contact output
Fault alarm:
x1 (AC250V, 2A/DC30V, 2A)
Upper limit:
x2 (AC250V, 2A/DC30V, 2A)
9.
Detection sensitivity level
1(Low sensitivity) to 9(High sensitivity) [9 ranges switching]
10.
Fan sound pressure
approx. 53dB
11.
Detection unit specifications
10-1.
Wiring distance
Exclusive use cable 10m Max. 20m
10-2.
Inhaling range of dust
Within approximate 10cm from the edge of the cover intake
10-3.
Dust conditions
≧0.3µm (* only case of airborne dust)
* Max. particle size dependent on specific gravity
≦40vol%
10-4.
Mounting
Pole mounting
U-Bolt mounting:(50A~90A)x 2
Wall mounting
Screw mounting: M8 screw x 4
12.
Converter unit specifications
12-1.
Connection method
Power line
Terminal block connection Screw size: M3.5
Signal line
Terminal block connection Screw size: M3
12-2.
Mounting
Wall mounting or put on the floor
12-3.
Protection
IP54

- 4 -
3. Outline (rough drawing)
+
*
*1) Make sure to connect the earth terminal.
*2) Make sure to tighten the lead outlet to prevent the outer air into Case.
For other connections like a conduit instead of our lead outlet plug, make sure to prevent
outer air from coming into the body by using putty etc. on the inlet lead wiring.
Fig. 1. Outline of Detection unit
Fig. 2. Outline of Converter unit

- 5 -
4. Mounting
This product is designed for wall-mounting or setting on floor.
4-1. Mounting Method
Please try to set pole-mounted with U-shape bolts sizing 50A to 90A, or wall-mounted with
the type M6 screws on the sensing station.
In case of wall-mounting converter component, use type M6 screws in 4 mounting holes. If
floor-setting, keep setting on a horizontal floor.
* Plastics attaching on the case door will be protective layers for sponge packing during
transportation.
Please remove them if using the product.
Warning: Not including U-shape bolts and screws for mounting.
Caution: Under any measuring environment and conditions outside the scope of specifications,
some malfunction may occur.
Important: In case of the sensing station under heavy rain more than 100mm, instrument
reading may be changed due to false detection by rain. If concerning false
detection by rain, set a roof to protect rain drops on the sensing station.
Caution: Keep the circuit protector to be OFF, if wiring.
!
!
!
!

- 6 -
5. Connecting
Detection unit internal device layout
Fig. 3. Detection unit internal device layout
Fig. 4. Converter unit internal device layout

- 7 -
Caution: Keep the circuit protector to be OFF, if wiring.
Important: Wiring power and signal individually.
Caution: Make sure to tighten the lead outlet to prevent outer air into the body.
For other connections like a conduit instead of our lead outlet plug, make sure to
prevent outer air from coming into the body by using putty etc. on the inlet of lead
wiring.
!
(1) Power source (2) Fault alarm output (SPDT)
AC110V -15% / +10% (3) Limit 1 output (SPDT)
or AC220V -15% / +10% (4) Limit 2 output (SPDT)
(For AC220V, specify separately) (5) Analog output (DC4 to 20mA output)
(6) Power supply for DC-Fan (DC24V output)
(7) Connection terminal for the sensor
!
*1 Make sure the Class D grounding.
Wire connecting terminal:
Max. 1.5mm2
*1
Fig. 5. Connection diagram
!

- 8 -
6. Part names and functions
(1) RANGE: measure range(1, 2, 3, 4, 5, 6, 7, 8, 9)set
(2) L1 ON: set ON point of limit 1 (set range:0-100%)
(3) L1 OFF: set OFF point of limit 1(set range:0-100%)
(4) L2 ON: set ON point of limit 2 (set range:0-100%)
(5) L2 OFF: set OFF point of limit 2(set range:0-100%)
(Minimum range of Hysteresis is still 1%.)
(6) INTEGRATION TIME: Integral time set (set range: 0-30 seconds)
(7) CORRECTION : Compensation value set (set range: 0.1-2.0)
【for indication】
POWER LED: Light on as power on
FAULT LED: CPU on when abnormal
L1 ON LED: Light on over limit 1set
L2 ON LED: Light on over limit 1set
4 beams indicate LED: detection density
will indicate per 0-100% complying with
setup.
【for button】
→: use when changing beam of
number/value.
↑: use when increasing number/value.
MOD: use when switching parameter set
mode and measure mode.
Pushing after set change will
memorize parameter and return to
measure mode.
SELECT:
Use when switching set items.
F-RESET:
Use when setting measure again.
Fig. 6. Front panel of Output unit
Matsushima Measure Tech Co., Ltd.
MADE IN JAPAN

- 9 -
7. Start up
7-1. Warm up operation
Dust Monitor requires warm up operation for about 1 hour after switch on.
7-2. Measure range set
Nine levels of measure range are provided, ranging from 1 (low sensitivity) to 9 (high
sensitivity).
If the display on the transducer is too close to the upper limit, increase the measure
range.
7-3. Integral set
Integration will even off measure signal per time fixed number up to maximum 30 seconds.
(optional set)
7-4. Limit set
Limit set can be done 2 different upper limits (alarm) in the range of 4-20mA (signal level
0-100%)
In case signal change gets severe however, chattering might rise often.
In such case you may be recommended to use integral function or delay function at/from
higher position (Customers expertise) respectively.
[Limit 1,2 and action chart of trouble alarm]
Limit output (upper limit alarm)
Power OFF to open COM-NO
Trouble alarm
Power OFF to close COM-NO
Trouble
Device
normal 機器
正常時
Device
normal
COM-NO close
LED On
COM-NO open
LED Off
COM-NO open
LED Off
COM-NO close
LED on
COM-NO open
LED off
COM-NO open
LED off
Fig. 7. Operation chart (upper limit alarm)
Fig. 8. Operation chart (Trouble alarm)
20mA
4mA
Time
Range to set
Current output
Limit set value(ON)
Limit set value(OFF)
Hysteresis

- 10 -
7-5. Compensation value set
Compensation value can output in a range of 0.1-2.0 (double) as compensation coefficient.
In such case as output is too high or too low for actual density when standard coefficient
is 1.0, this function can output nearest density to actual density by multiplying output.
Utilizing this function, output can be compensated as shown hereunder.
Fig. 9. Compensation value
This function is 0.1 in increments of input in the range of 0.1-2.0 (default is 1.0).
If management is only for trend of dust exhaust, you can do without using this function
(=set value 1.0). In case management requires the nearest value to actual density, you
seize actual density such as counter wash pulse time, etc. by manual analysis first and
then set counter wash pulse shown by device very much close to value of manual analysis.
Compensation value 1.0 trend
For actual density,
detect density outputs high.
Compensation value 0.5 trend
For actual density,
Detect density outputs almost
all alike and understood
compensation is effective.
Detection density
Real density
Detection density
Real density

- 11 -
8. Set example of parameter
8-1. Flow of parameter set
Parameter should set according to the flow as shown hereunder.
* In case setting up value over input range exceeding specification, setup value of
parameter will indicate [ERR] on indication LED, let you know setup value is
wrong and do not accept next operation.
* In this case, please change to correct set value again.
8-2. Example of parameter set
Following is a set example to set integral time at 15 second.
(Providing that set value was 0 second before change.)
(1) Push “MOD” button 1 time················ move to parameter window and LED (orange).
written Range on the left will light on.
(2) Push “SEL” button 5 time················· LED (orange) lit goes down in turn
and INTEGRATION TIME LED will light.
(3) By using both →button and ↑button,
set Indication LED at “15”.
(4) Push “MOD” button 1 time················· return to measure window and start
measure again at integral 15 second.
You operate other parameters with same procedure, please.
In case setting other parameters continuously, at (4) above instead of “MOD” you push
“SEL” and can perform both setting parameter and moving to plural items at the same time.
Reference: Parameter and light up of corresponding LED
RANGE : only RANGE LED light up
L1 ON : only L1 ON LED light up
L1 OF : only L1 OFF LED light up
L2 ON : only L2 ON LED light up
L2 OFF : only L2 OFF LED light up
INTEGRATION TIME : only INTEGRATION TIME LED light up
CORRECTION : both RANGE LED and CORRECTION LED light up same time
Measure
window
Parameter mode
Value set
Parameter select
MOD button
Press down
(to parameter mode)
SEL button
Push down
Respective
parameter select
→Move beam by
button
↑Each button
push down change
value
Push down MOD button
(after parameter set, to measure window)
SEL button push down
(after set parameter move
to other parameter item)

- 12 -
9. Maintenance
Updating for maintenance will be significantly changed because of environmental situations and
concentration of dust vacuuming.
Please try to decide the updating for maintenance at your side, after checking the points to be
maintained every day for the initial one month.
Keep the power to be OFF during maintenance.
Warning: Do not remove the earth wire during even maintenance because it may cause breakdown.
And please do not touch the probe on the sensor.
Important: If lacking any maintenance, low detection sensitivity or off the scale of
detection value may occur.
Surely perform a routine maintenance to keep normal detection sensitivity.
Please remove the butterfly bolt on the housing and take off the cover.
9-1.Maintenance of Probe
Release 2 patching locks fixing the sensor and take off the sensor in duct.
* Please do not remove the connecting wire between the sensor and the converter.
!
Fig. 11. Take off Sensor
Fig. 10. Take off Cover
!
Butterfly bolt
Cover
Take off Cover
Patching
locks
Sensor
Duct
Take off Sensor

- 13 -
Wipe the probe on the sensor with some wet waste cloth.
If oily fouling, use some liquid to remove oil.
9-2.Maintenance of Duct
Remove the wingnuts on the housing without the sensor and take off the duct.
* Please do not remove the connecting wire between the sensor and the converter.
Fig. 13. Take off Duct
Fig. 12. Sensor
Probe
Wingnuts
Duct
Plates

- 14 -
Next, please take off the air intake plate and the DC fan after removing the butterfly bolts.
Loosen the butterfly bolts (type M4) at 8 points to take off.
Wipe inside of the duct with air blower or waste cloth after taking off as below Fig. 15.
If unable to clean up, please try brushing.
After cleaning, please set on the sensor by inverse process.
Fig. 15. Individual Duct (after took out)
Fig. 14. Take off Duct
Butterfly bolts
Air intake
plate
DC fan
Exhaust plate
Finger guard
Butterfly bolts
Duct
Inside of Duct

- 15 -
9-3.Maintenance of DC Fan
Warning: The DC Fan is a suggested part of replacement by each 5 years to keep flow rate.
Warning: 5 years will be recommendable replacement term, not guarantee for operation.
Please take off the DC Fan according to Fig. 14 in 9-2.
Wipe inside of the DC fan removed from the duct with air blower of waste cloth.
If unable to clean up, please try brushing.
!
!
Fig. 16. Individual DC Fan (after took off)
Butterfly bolts
DC Fan

- 16 -
10. Troubleshooting
If the product can not recover by the below countermeasures, contact with us.
Table 1. Troubleshooting
Problem
Presumable cause
Confirmation
No light on
power LED
and
other LEDs
・Power not supplied
・Power specification is
changed from the
ordered specification.
・Circuit protector is
OFF.
・CPU abnormal.
・Confirm power supply voltage on Circuit protector terminal
base.
・Confirm the voltage on name plate of Transducer is the same
as the supplied power.
If it is different, return us Transducer, then the power
spec. will be changed at out site.
・Switch Circuit protector to ON.
・Power OFF, then ON.
Instable
output
indication
・Warm up operation after
setup is not complete.
・Unconnected between
Sensor and Converter
・The wiring between
Sensor and Transducer
is either off or wrong.
・Material sticking on
Probe electrode.
・Condensation on
electrode
・Micro amount solid
particles flying.
・Make warm up operation for 1 hour and check output again.
・Confirm cable connection between sensor part and converter.
If unconnected, please contact us.
・Confirm wiring of exclusive cable.
(See Page 5-6, 5.Connecting.)
If wiring is wrong, rewire correctly.
・Remove Sensor and check if any sticking on Probe electrode.
If yes, power off and wipe them off with moist cloth. (See
NOTE)
・Remove sensor and confirm electrode.
If condensation on electrode, wipe off with a wet rag after
power-OFF. (See NOTE)
・Change measuring sensitivity or integration setting.
Output
indication
off-scale
and no
return
・Duct or pipe is short-
circuited with Probe
electrode due to the
sticking material.
・Condensation on
electrode
・Device is in trouble.
・Too many solid
particles in gas
・CPU abnormal.
・Remove Sensor and check if any sticking on Probe electrode.
If yes, power off and wipe them off with moist cloth.(See
NOTE)
・Remove sensor and confirm electrode.
If condensation on electrode, wipe off with a wet rag after
power-OFF. (See NOTE)
・Consult with Sales Dept. to check the soundness of Device.
・Change measuring sensitivity.
・Power OFF, then ON.
No change
on
indication
・Power not supplied
・The wiring between
Sensor and Transducer
is either off or wrong.
・Material sticking on
Probe electrode.
・Device is in trouble.
・No intake/exhaust of
air
・CPU abnormal.
・Confirm power supply voltage on Circuit protector terminal
base.
・Confirm wiring of exclusive cable.
(See Page 5-6, 5.Connecting.)
・Remove Sensor and check if any sticking on Probe electrode.
If yes, power off and wipe them off with moist cloth. (See
NOTE)
・Consult with Sales Dept. to check the soundness of Device.
・Confirm that DC fan connector is connected. If not, connect
it.
.
・Power OFF, then ON.

- 17 -
Fault LED
is
lighted.
・Wiring is wrong.
・The supplied power
supply is out of
allowable range.
・CPU abnormal.
・Check wiring of power/output signal lines, exclusive cable.
(See Page 5-6, 5.Connecting.)
If wiring is wrong, rewire correctly.
・Check power supply with a Tester etc.
If it is out of the allowable range, apply the one within
the range.
・Power OFF, then ON.
Fan does
not
turn.
・No power supplied.
・Circuit protector is
OFF.
・Connector is
unconnected.
・Confirm power supply wire to connect to Terminal No.AC.
Confirm power supply voltage on Circuit protector terminal
base.
If power is not supplied, supply power.
・Switch Circuit protector to ON.
・Connect DC fan connector.
Note: After cleaning, make sure to tighten securely when mounting.
11. Maintenance item
The recommended items for periodical maintenance are in the below Table 2.
Table 2. Maintenance item list
Maintenance item
Type
Performance
Frequency
Duct unit
PFM-AD12-D
Replacement
Every 5 years
Packing set
PFM-AD12-P
Replacement
Every 3 years
Table of contents