Mazet Sante BIOMOOV User manual

Page 1 on 20 DATE 13/04/2022
A0 version
BIOMOOV Manual
Proprioception sensor

Page 2 on 20 DATE 13/04/2022
A0 version
Instructions for use
&
Technical description
Please read these instructions carefully before using your new device!
This manual is an integral part of the device and must be kept until it
is destroyed.
This equipment has been designed and manufactured for use in a
therapeutic setting.
Use is reserved for professionals who have undergone the
appropriate training.
If you have any problems or do not understand this manual, please contact
your distributor (see stamp on the last page) or contact Électronique du Mazet
at :
Tel : (33) 4 71 65 02 16 - Fax : (33) 4 71 65 06 55
Please return the warranty certificate located on the last page of this manual
within 15 days of installation or receipt.

Page 3 on 20 DATE 13/04/2022
A0 version
1Presentation of the device
BioMoov is a proprioception sensor. The sensor has no therapeutic claims.
Equipped with a software, the BioMoov allows, through a Bluetooth connection to the Biostim,
to transcribe the patient's activities on a proprioception board.
The BioMoov is provided in a pouch, with a USB cord to recharge it, and a scratch to fix it on a
proprioception tray.

Page 4 on 20 DATE 13/04/2022
A0 version
Table of Contents :
1Presentation of the device ..................................................................................................3
2Description and technical information...............................................................................5
2.1 Symbols used...............................................................................................................6
2.2 Technical data..............................................................................................................7
2.2.1 General characteristics.................................................................................................................... 7
2.2.2 Technical characteristics of BioMoov.............................................................................................. 7
2.2.3 Accessories...................................................................................................................................... 8
2.3 Nameplate label...........................................................................................................8
2.3.1 BioMoov identification label ........................................................................................................... 8
3Warnings ............................................................................................................................9
4Precautions .......................................................................................................................10
4.1 Residual risks.............................................................................................................10
5Installation of the device..................................................................................................10
6User's Manual................................................................................................................... 11
6.1 Getting started with the device..................................................................................11
6.1.1 Powering up / starting up.............................................................................................................. 11
6.1.2 Using Bluetooth............................................................................................................................. 11
6.1.3 Warning screen ............................................................................................................................. 11
6.2 Choosing a treatment.................................................................................................12
6.3 Setting the parameters ...............................................................................................12
6.4 Run a treatment..........................................................................................................13
6.4.1 Start the treatment ....................................................................................................................... 13
6.4.2 During treatment .......................................................................................................................... 14
6.4.3 End of treatment........................................................................................................................... 14
6.5 Saving a treatment.....................................................................................................14
6.6 Technical Information, Configuration & Settings.....................................................14
7Clinical Guide ..................................................................................................................15
7.1 Target population.......................................................................................................15
7.2 Side effects ................................................................................................................15
8Maintenance, upkeep........................................................................................................15
8.1 Housing......................................................................................................................15
8.2 Sterilization :..............................................................................................................16
9Malfunction......................................................................................................................16
10 After-sales service and warranty......................................................................................17
11 Disposal............................................................................................................................17
11.1 Accessories ............................................................................................................17
11.2 Electronics..............................................................................................................17
12 Transport and storage.......................................................................................................18
13 CE declaration..................................................................................................................18
14 Manufacturer....................................................................................................................18
15 Certificate of guarantee....................................................................................................20

Page 5 on 20 DATE 13/04/2022
A0 version
2Description and technical information
-This user and maintenance manual is published to facilitate the handling of your BioMoov from
the initial reception phase, then the commissioning to the successive stages of use and
maintenance.
If you have any difficulty in understanding this manual, contact the manufacturer, Électronique
du Mazet, your dealer or distributor.
-This document must be kept in a safe place, protected from atmospheric agents, where it
cannot be damaged.
This document guarantees that the devices and their documentation are technically up-to-date
at the time of marketing. However, we reserve the right to make changes to the device and its
documentation without any obligation to update these documents.
-In the case of transfer of the device to a third party, it is mandatory to inform Électronique du
Mazet of the details of the new owner of the device. It is imperative to provide the new owner
with all documents, accessories and packaging related to the device.
-Only personnel who have been informed of the contents of this document may operate the
device. Failure to comply with any of the instructions contained in this document will relieve
Électronique du Mazet and its authorized distributors of the consequences of accidents or
damage to personnel or third parties (including patients).

Page 6 on 20 DATE 13/04/2022
A0 version
2.1 Symbols used
Warning: this logo draws your attention to a specific point
Operating Instructions: This logo informs you that the operating
instructions must be read for safe use of the device
Recycling: This device should be disposed of at an appropriate collection
and recycling facility. Consult the manufacturer.
Manufacturer
Date of manufacture
Country of manufacture
Lot number
Direct current
Non-ionizing electromagnetic radiation
Medical device
UDI
Deposit
Operating instructions
CE marking (European Conformity)

Page 7 on 20 DATE 13/04/2022
A0 version
2.2 Technical specifications
2.2.1 General characteristics
•Operating temperature: 0°C to 40°C.
•Storage temperature: -40°C to 70°C.
•Operating relative humidity: 30% to 75%.
•Operating altitude: < 2000 meters
2.2.2 Technical characteristics of BioMoov
•Housing dimensions: 5.8 x 5.8 x 1.9 mm
•Case weight: 159,425
•Case color: black
•Medical Class I equipment.
•Applied part type BF

Page 8 on 20 DATE 13/04/2022
A0 version
2.2.3 Accessories
This device is delivered with the following accessories as standard:
- USB cable
- Scratch
Accessories are for personal use only.
The BioMoov is not in contact with the patient. Indeed, it is intended to be fixed only on a
proprioception tray.
The use of accessories not recommended by the manufacturer does not engage his
responsibility
2.3 Identification label
The information and characteristics are given on the back of each device on a label.
2.3.1 BioMoov's identification label

Page 9 on 20 DATE 13/04/2022
A0 version
3Warnings
CAUTION: Install the unit on a flat, stable surface . Do not block the rear
ventilation openings (no objects closer than 4 cm).
CAUTION: The appliance must be plugged into an outlet with a grounding
terminal (Class I electrical appliance).
CAUTION: In case of emergency, disconnect the power cable directly from the
unit.
CAUTION: No modifications to the device are permitted. It is strictly forbidden
to open the housing of the device.
CAUTION: This device complies with applicable electromagnetic compatibility
standards. If you experience interference or other problems with another
device, contact Électronique du Mazet or the distributor for advice on how to
avoid or minimize the problem.
CAUTION: Operating altitude below 2000m.
The performance of the device decreases with altitude.
CAUTION: The device must be used with the accessories supplied by the
manufacturer

Page 10 on 20 DATE 13/04/2022
A0 version
4Precautions
4.1 Risks residual
If water enters the unit, it may not function properly. In this case, unplug the device
and disconnect the cables. In any case, avoid the presence of water in the environment
near the device.
5Installation of the device
Open the package, remove the accessories and the BioMoov device.
Check the contents against the packing list included with the documentation.
Check that the contents of the box are not damaged; if you have any doubts about the integrity
of the device or its accessories and that the proper functioning of the device could be
questioned, contact Électronique du Mazet
If the unit was stored in a cold place and there was a risk of condensation, let the unit rest for
at least 2 hours at room temperature before turning it on.
Before the first use, a cleaning of the device and its accessories is recommended see §
maintenance

Page 11 on 20 DATE 13/04/2022
A0 version
6User's manual
6.1 Handling the device
6.1.1 Powering up / starting up
Turn on the device with the on/off button located on the device.
The BioMoov turns on and a LED flashes.
6.1.2 Use of Bluetooth
See setting specification.
6.1.3 Warning screen
When the BioMoov device is connected to the Biostim, a logo is displayed on the Biostim
screen.

Page 12 on 20 DATE 13/04/2022
A0 version
6.2 Choice of treatment
Refer to the Biostim user manual.
6.3 Setting the parameters
Turn on the device and click on the Bluetooth logo (under the house). When the accessory is
connected, the Bluetooth logo turns green, and the logo of the connected accessory appears
on the right with a bar graph indicating the available battery level
The accessory is then used like the other probes, by choosing the channel to be used that
corresponds to the probe in the BFB menu.
If the Biostim is not connected to the PC, the duration of the programs is limited to 1 minute.
When accessing the treatment, and before launching it, it is necessary to set up the tray:
Clicking on the tray logo to the right of the Bluetooth symbol opens the configuration page.
This page allows you to configure the tray
type:
2 axis tray (ball type)
Or 1 axis (support on a bar)
Then, the calibration is done by putting the
device in the right stop, then before, and by
clicking on the adjustment logo for each
step.
Note: During adjustment, the movement of the dot on the target may be reversed. After
adjustment, the movement of the stage should cause the dot on the target to move in a
consistent manner.

Page 13 on 20 DATE 13/04/2022
A0 version
6.4 Run a treatment
6.4.1 Start the treatment
Use the BioFeedBack exercise module.
Reminder: Taring Biofeedback
The taring works automatically. However, it is possible to adjust the parameters calculated by
the machine with the buttons and .
Procedure for taring
•Place the probe or electrodes
•Start taring
•Ask the patient to make a sustained contraction (the gauges change automatically),
then release the effort for a few seconds.
•The BioStim automatically sets the operating range
•Go to the next page by pressing the arrow (or wait for the end of the taring time)
•During the exercise, it will always be possible to manually adjust the level of the tare
with the + and - buttons at the top right of the page
It is possible for each channel to activate or not the negative biofeedback, either in the
configuration page or during the taring.
If the option is enabled, you can see the work below the base level. This is particularly useful if
you are working on relaxation.
In the + version, it is possible to adjust the level of negative BFB to be displayed during the
exercise, using the slider in the right-hand panel:
Zero Patient: the bottom of the screen corresponds to the minimum reached by the patient
during the taring
This setting allows you to erase the basic tone.
Negative biofeedback
zone
Max Level
Basic Toning
Zero Patients
Zero Biostim
Max Level
Zero Patients
Zero Biostim

Page 14 on 20 DATE 13/04/2022
A0 version
•Zero Biostim : minimum measurable by the device : to work in negative BFB
•It is also possible to choose intermediate values:
6.4.2 During treatment
All parameters can be changed during processing by selection on the screen or by pressing the
buttons on the remote control (optional).
The treatment can be stopped or restarted by selection on the touch screen or by pressing the
keys on the remote control.
6.4.3 End of treatment
Refer to the Biostim user manual
6.5 Saving a treatment
Refer to the Biostim user manual
6.6 Technical Information, Configuration & Settings
Refer to the Biostim user manual

Page 15 on 20 DATE 13/04/2022
A0 version
7Clinical Guide
7.1 Target population
BioMoov is intended for anyone over 5 years of age, regardless of gender, in combination
with the BioStim device, it cannot work alone. The clinical part is detailed in the Biostim user
manual
7.2 Side effects
As the device has no therapeutic claims, there are no side effects.
8Maintenance, upkeep
The BioMoov device is designed to last 5 years.
To guarantee the preservation of the device's performance throughout its life, it is necessary to have
the device checked by Electronique du Mazet's technicians every 2 years.
8.1 Housing
The case only requires normal and periodic cleaning of its external surface which could be dirty.
The same applies to the USB cable, ....
Clean only with a dry or very slightly damp cloth.
Be sure to unplug the power cord before cleaning.

Page 16 on 20 DATE 13/04/2022
A0 version
8.2 Sterilization:
This device is not sterile,
Accessories are not sterile
9Malfunction
If you notice a malfunction that is not commented on in the documents accompanying the
device, please inform your distributor or the manufacturer.
In the case of a shipment of the device, please observe the following instructions:
▪Decontaminate and clean the unit and its accessories.
▪Use the original packaging, including the retaining flanges.
▪Attach all accessories to the unit.
▪Set up the different elements.
▪Make sure the package is properly closed.
Shipping address:
Electronique du Mazet
ZA Route de Tence
43520 Le Mazet St Voy
Tel : (33) 4 71 65 02 16
Fax: (33) 4 71 65 06 55
Possible operating anomalies:
Description of the anomaly
Possible causes
Actions
No connection to Biostim
Bluetooth connection problem
-Check that the Biomoov is charged
-Switching the Biomoov off and on
-Click on the Bluetooth logo (under the house).
No charging of the device
USB cord problem
Make sure the cord is not damaged
Flat trace in biofeedback
windows
- loss of communication with the
module.
- no sensor on the considered
input
- exit the current treatment and return to the
main office.
- check the path used
If the device is dropped or if water penetrates, it is imperative to have the device checked by
Électronique du Mazet to exclude any risk (patient and user) related to the use of the device.

Page 17 on 20 DATE 13/04/2022
A0 version
10 After-sales service and warranty
This device is warranted by your supplier under the conditions specified in this document,
provided that:
◼Only accessories supplied by Électronique du Mazet or its distributors should be used.
◼Any modification, repair, extension, adaptation and adjustment of the device must be carried
out by Électronique du Mazet or its authorized distributors for these operations.
◼The work environment meets all regulatory and legal requirements.
◼The device may only be used by competent and qualified personnel. Use must be in accordance
with the instructions in this user's manual.
◼Treatments should be used only for the applications for which they are intended and which are
described in this manual.
◼The device must be regularly maintained according to the manufacturer's instructions.
◼All legal requirements for the use of this device are met.
◼The device uses only the consumables or semi-consumables supplied or specified by the
manufacturer.
◼Machine parts and spare parts must not be replaced by the user.
Inappropriate use of this device or neglect of maintenance relieves Électronique du Mazet and its
authorized distributors of all responsibility for defects, malfunctions, damage, injuries and the like.
The warranty is void if the operating instructions in this manual are not strictly followed.
The warranty is 24 months from the date of delivery of the device.
The accessories are guaranteed for 6 months from the date of delivery of the device.
Consumables and semi-consumables are not guaranteed.
Shipping and handling costs are not included in the warranty.
Électronique du Mazet, or its distributor, agrees to provide the drawings, parts list,
instructions and tools necessary to repair the unit on the sole condition that qualified
technical personnel have been trained on this specific product.
11 Disposal
11.1 Accessories
As soon as any deterioration of an accessory is noted, the product must be cleaned with a broad
spectrum disinfectant and returned to the manufacturer.
11.2 Electronics
If the BioMoov device should fail or become unusable, it is requested to return it to the
manufacturer or to deposit it in a Recylum collection point.
Indeed, as part of its commitment to the environment, Électronique du Mazet finances the
Recylum recycling network dedicated to WEEE Pro, which takes back free of charge electrical
lighting equipment, control and monitoring equipment, and used medical devices (More
information on www.recylum. com ).

Page 18 on 20 DATE 13/04/2022
A0 version
12 Transport and storage
The device must be transported and stored in its original packaging or in packaging that
protects it from external damage.
Store in a clean, dry place at room temperature
13 CE declaration
ÉLECTRONIQUE DU MAZET makes available on request the CE declaration of this device .
The first affixing of the medical CE on this device took place on 18/05/2022.
14 Manufacturer
Électronique du Mazet is a company located in the heart of the Massif Central. Originally
a simple manufacturer of electronic cards, over the years it has developed its own brand of
medical equipment, mainly for physiotherapy.
Today, EDM studies, develops, manufactures and markets pressotherapy,
depressotherapy and electrotherapy devices (uro rehabilitation).
For any additional information, please contact us.
SAS Électronique du Mazet
ZA Route de Tence
43520 Le Mazet St Voy
Tel : +33 (0)4 71 65 02 16
Fax : +33 (0)4 71 65 06 55
facebook.com/electroniquedumazet www.electroniquedumazet.com

Page 19 on 20 DATE 13/04/2022
A0 version
Your dealer / distributor :
ELECTRONIQUEDU MAZET
ZA ROUTE DE TENCE
43520 LE MAZET SAINT VOY
Tél : +33 4 71 65 02 16
Mail : sav@electroniquedumazet.com

Page 20 on 20 DATE 13/04/2022
A0 version
15 Certificate of guarantee
Certificate of Guarantee
This form must be returned to Electronique du Mazet within 15 days of installation or receipt
of equipment.
I, the undersigned, .......................................
Organization: ........................................
Address : ........................................
.........................................
.........................................
Declares to have received the Biomoov device n° ......................... in working order.
I have received all the necessary instructions for its use, maintenance, care, etc...
I have read the instruction manual and I have noted the warranty and service conditions.
In the event that Electronique du Mazet or its distributors do not receive this form duly
completed and signed within one month of delivery, Electronique du Mazet will be relieved of
all responsibility with regard to the warranty and after-sales service, or any other consequence
due to misuse of the device.
Done at .............................. on .............
Signature
User:
To be returned to :
Electronics of the Mazet
Z.A. Route de Tence
43520 Le Mazet St Voy
Your distributor :