Safety
Meaning of symbol
CAUTION! Please consult the operating instructions before using the device. In these operating instructions, failure to
follow or carry out instructions preceded by this symbol may result in personal injury or damage to the device and the
installations.
Meaning of symbol
This appliance is protected by double insulation or reinforced insulation. It does not have to be connected to an earth
protection terminal for electrical safety.
Thank you for purchasing this MS301 meter.
To obtain the best service from your unit:
- Read these operating instructions carefully,
- Comply with the precautions for use.
PRECAUTIONS FOR USE
Never use on a voltage network over 600V with respect to the earth connection. This voltage surge category III
current meter complies with stringent reliability and availability requirements, corresponding to fixed industrial and
domestic installations (see IEC 664-1).
Do not use on alternative and continuous voltages > 600V.
Indoor use in environments with a maximum pollution level of 2 (EN 50419:2006) temperature of -10ºC to +50ºC
and relative humidity below 90%.
Respect the value and type of the fuses to avoid damaging the instrument and canceling the warranty.
Fuse 500V/3A
Fuse 500V/10A
Use accessories corresponding to safety standards (EN 61010-1:2001) with 600V minimum voltage and surge
category III.
Before any measurement, ensure correct positioning of the leads on the current meter and of the switch. When the
value range of a measurement is not known, place the switch at the highest caliber, and then gradually reduce it until
the appropriate caliber is achieved: the reading should preferably be in the upper 2/3 of the range.
During current intensity measurements (without a clip-on current meter), stop circuit power supply before
connecting or disconnecting the current meter or changing caliber.
The leads must be disconnected to open the lower half of the meter case
Never connect to the circuit to be measured if the casing is not properly closed.