MIC TECHNOLOGY MIC-B30 Series User manual

MIC-B30 SERIALS BIOLOGICAL MICROSCOPE
OPERATION MANUAL
READ THIS MANUAL BEFORE USING THE MICROSCOPE
CHONGQING MIC TECHNOLOGY CO.,LTD

MIC-B30 SERIALS BIOLOGICAL MICROSCOPE
OPERATION MANUAL
Ⅰ.Application
MIC-B30 serials biological microscope is based on the XSP-C serials, and combine
advanced modular design, developed a new products.
The products have high degree of standardization, beautiful appearance, and also have
the feature about rational structure, easy to use, clarity of visual field and quality
stability. And these products also are the general Equipment for medical, research, and
teaching.
Ⅱ.Specification of Standard Accessories
1. 1. Eyepieces
Category
Magnification
Diameter of view-field(mm)
Remark
Wide field eyepiece
WF10X
ф18
WF16X
ф11
2. 2. Objectives
Category
Magnification
Numerical aperture
Working distance
Remark
Achromatic
objectives
4X
0.10
37.5
10X
0.25
7.613
40X(s)
0.65
0.632
spring
100X(s.oil)
1.25
0.198
Spring oil
3. Total magnification
Objectives
tube
coefficient
Total magnification
Eyepiece 10X
Eyepiece 16X
4X
1X
40X
64X
10X
100X
160X
40X
400X
640X
100X
1000X
1600X
4. Mechanical tube length: 160mm
5. ABBE condenser, NA=1.25
6. Filter diameter: 32mm
7. Coarse/fine focal range:20mm
8. Stage size:X-Y:120*125mm

Longitudinal movement range: 70mm
Horizontal movement range: 30mm
9. lamp:LED lamp
10. power:AC input 220V
Ⅲ.Structure
This series of microscopes are composed of the following main components.
1. Objective lens case and Nosepiece:
Microscope objective lenses are engraved with various tags: magnification,
Mechanical tube length/ Cover slip thickness. Such as: objective 40X: 40/0.65;
160/0.17 means that the magnification is 40X; Mechanical tube length is 160mm;
Cover slip thickness is 0.17mm, and about the objective 100X oil, it is engraved with
‘oil’.
Nosepiece guarantees smooth rotation by quadruple revolving nosepiece.
2. Eyepiece: Using WF10X wide field eyepiece.
This series of microscope’s binocular Tube spacing adjustable range is 55mm-75mm
3. Stage: it can be operated much easily by coaxial knob in low position.
4. Condenser: ABBE condenser N.A. =1.25 with iris diaphragm, kosher illumi
nation is optional.
Condenser can move up and down, and the condenser through strict assembly and
calibration, and Users should not arbitrarily disassemble.
5. Course Focusing Adjustment:
Coarse/fine focal range is 20mm. and it is on the bottom of left and right side.
Rotating rough hands wheel, the working stage rapid decrease or increase. About the
micro hands wheel, it can make the stage move up and down fine. There is an
indexing on the micro hands wheel. Every lattice is 0.002mm.
6. Base:
This is the basis of the product; there are four Rubber machine feet, to make it more
Stable.

Ⅳ.Operation Instruction
1. Unpacking and installation:
Microscope packaging is foam and carton; host packed in plastic bags, Objective are on
the transparent plastic box, Eyepiece with a plastic bag, it should be carefully when install.
2. Adjustment of the microscope:
1)Insert the eyepiece into the eyepiece tube, and screw objectives into the nos
epiece in sequence of different magnification from low to high. Then put the s
pecimen on the stage and secure it the position with tablet and move it to the
center of stage.
2) Turn on the power switch and adjust the brightness form dark to bright slo
wly. After working, you must adjust the brightness to a little dark before you
turn it off.
3)Observe the specimen from lower magnification objective firstly and move t
he specimen to the center of view field, then rotate higher magnification object
ive. You may use the fine focusing knob to obtain the clear image.When100X
(soil) objective is used; you should fill up with cedar wood oil (without bubb
le) between the front of objective and the specimen surface. After working, it
should be wiped with a few xylems immediately.
4)In order to obtain bright and clear image, the illumination must be adjusted.
When different objective is chosen, you should adjust the iris diaphragm o
f the condenser and different brightness of the light.
5)When the lamp needs to replace, you should shut off the power switch and
replace it after the lamp is cool. Note: The contact must be firmed, and t
he filament center should be adjusted.
Ⅴ.Maintenance
1. Open the package carefully to avoid the damage of the lenses or accessorie
s
2. Operate correctly, and put the dust cover on the microscope after work.
3. Don’t dismantle the instrument rashly besides the replaceable lest the correc
t position should be breached.
4. Keep the instrument in dry and cool place and away from the pollution an
d corrosion.
5. Please send the instrument to the special repair shop if it goes out of orde
r.
6. When the objectives and eyepieces won’t be used for a long time, please p
lace them into a dry box, and put the dust cover onto the microscope.

VI. Full set of the instrument
1
Mainframe of bio-microscope
1 set
2
Eyepiece: WF10X WF16X
2 pcs of each
3
Tube
1 pc
4
Achromatic Objective:4X,10X,40X,100X(oil)
1pc of each
5
Filter: blue
1 pc
6
Cedar wood oil
0.3ml/bottle
7
Operation Manual
1 copy
8
Dust cover
1 pc
Table of contents