IFOV =1.58 mRad (Theoretical with 320x240 Detector and standard 28.9˚x21.9˚lens) Detect 1in. target @52ft.
6. MANUAL ADJUST: (Level and Span) Press SELECT until the Sensitivity (S number at top of display) or Level
(center number next to color palette) is highlighted then press the + or – button to adjust manually the value.
7. PAUSE IMAGE: While in the RUN mode press and release the RUN button. After a moment the image will
freeze and display FRZ in the mode area. To return to the RUN mode press and release the RUN button once
again. Don’t forget there is a delay in this process.
8. SAVE IMAGE: In either RUN or FRZ mode press and release the SAVE button.
NOTE: we highly recommend that you freeze the image before saving it, this allows you to verify its proper
focus and perspective. If you freeze before saving, after saving you will be returned to the frozen image and will
need to press and release the Freeze Button to return to a live image.
9. RECALL IMAGE: Press the RUN button to enter the FRZ mode, remember to wait for FRZ appear. Press the +
or – buttons to scroll through the saved images. Press RUN to return to the RUN mode.
10. PALETTE: To change between grey scale to the selected color palette, press SELECT repeatedly until the
PALETTE BAR appears Highlighted then pres either + or –.
To select a different color palette, press SELECT repeatedly until the MENU option becomes highlighted, then
press RUN to enter the command menu.
Press + or – to select COLOR menu item, then press select to view your choices at the bottom of the
display.
Press SELECT and + and – to navigate to and chose the desired color palette. (for now stay in COLOR
POS 356 then chose A/Rainbow, B/Brightness, C/Hot Iron or D/Medical) Press RUN to return to RUN
Mode.
11. RANGE: A To select a different color palette; press SELECT repeatedly until the MENU option becomes
highlighted, then press RUN to enter the command menu. Use + or – to Highlight RANGE
Press SELECT then + or – to highlight the desired RANGE. Press RUN to return to RUN mode.
12. EMISSIVITY AND BACKGROUND: Both can be adjusted by going into the COMMAND MENU as above,
then highlighting and selecting either EMISS or B. GROUND COMPENSATION then changing their values.
Temp. Range (Hi/Low)
Emissivity
Run/Freeze Mode
Sensitivity
(Span) Image Mode
Gain Control Date/Time
Menu/Cursor/Memo/Standby
Range
Level
Center Point Temp
Temp. Span Setting
(Sensitivity)