
8
• The device must not be used during magnetic resonance imaging
(MRI).
• The Monidrop® W infusion monitor is designed as a measuring de-
vice for the administration of basic fluids, nutritional solutions, blood
products or antibiotics. The device is not applicable for use together
with strong medicinal substances requiring high accuracy.
• Only use infusion set suitable for the device, with a drip size ALWAYS
of 20 drops (gtt)/ml and accuracy of at least ±10%. Before use,
check that the infusion set is attached correctly to the device.
• The Monidrop® W infusion monitor always uses ml/h (millilitres per
hour) as the measuring unit for the infusion rate.
• Make sure that fluids can flow freely by straightening all lines.
• The height of the IV bag should be suitable for gravity infusion. The
height should be about 60 ... 90 cm above the patient.
• Make sure that the level of the drip chamber does not reach the
measurement range. Fill the drip chamber halfway at the most.
• When replacing infusion sets, follow the infusion set manufacturer’s
recommendations (note country-specific hygiene regulations).
• The operating conditions of the device must fulfil proper requirements
(e.g. VDE 0100, VDE 0107 or IEC regulations). Note country-specif-
ic guidelines and differences.
• The device has also been approved for home care in accordance
with standard 60601-1-11.
• Do not use the device in a room containing explosive substances.
• Protect the charger against moisture.
• If the device drops or suffers a heavy impact, it must be sent to a
service point for inspection.
• Only use the charger delivered with the Monidrop W device.
• Do not tilt the device heavily when using it. If the device is tilted by
1.3 PATIENT SAFETY