P a g e | 5
INSTALLATION
C Series Autopsy Sink Installation Instructions
Locate package of smaller items (dissection board, tissue boxes, duct hose or filters)
The CA100 / CB100 / CC100 is utilized with Autopsy Carts that overhang the sink. It is imperative that the
mounting height is at 34” finished height and is exclusive of any slope in the floor of the building.
Placing a bubble level on the edge of the unit’s edge (do not place the level on the table’s surface) side to
side and front to back, assure the unit is completely level.
Mount the unit as required with an approved fastener. The fastener is suggested as being a minimum
dimension of 3/8” diameter and having a Tensile Pullout Load of at least 4,000 lbs. .
Once the unit is secured to the wall, the utilities can be hooked up. The cold water line is ¾” and the hot
water is ½” hard copper which will require customer supplied on/off valves. Mopec suggests using ball type
water valves to assure proper water flow for operating all table accessories. Drain lines are standard 1 1/2”
IPS and need to be vented to assure proper drainage.
Utility Connections
The only connections necessary are the Electric, Cold and Hot water supply, the drain connection and the
HVAC connections if in house ventilation. If the disposal option is ordered, the disposal will need to be
installed prior to the electrical hook up. The unit is already pre wired and plumbed for the options ordered.
Electrical Connection
All electrical, water and ventilation stubs should be prepared in accordance with our rough-in dimensions
shown on rough in diagram of this manual.
The electrical service provided for the MB600 must include:
1) A switch or circuit breaker for each circuit to which the wiring harnesses from the workstation will
be connected.
Each C Series autopsy sink has a three foot whip for attachment to the facility, unless a plug and cord are
requested. The whip leads are labeled. And depending on the options selected there are either 3 wires or 5
wires. L1 and C1 are for the unit electrical options such as lights, lifts and fans if a recirculation unit. L2 and
C2 are for the disposal.
115V Info
Without a disposal there are three wires:
Black Labeled as L1
White Labeled as C1
Green w Yellow Stripe Ground
With the Disposal there are five wires
Black Labeled as L1
White Labeled as C1
Green w Yellow Stripe Ground
Red Labeled L2
White Labeled C2
The flexible water tight conduit and connections is the preferred conduit and is not provided with the unit. This
is due to an unknown length needed and variances possible in the rough in process. If the disposal option is
purchased there MUST be two separate circuits for the two circuits on the autopsy sink. You must have
separate circuits for the disposal and unit.