Initial Setup
Complete the following steps to set up the FieldDAQ device.
1. Install the software and NI-DAQmx driver as outlined in the FD-11634 Quick Start.
Note You must install NI-DAQmx 18.6 or later for device calibration support.
2. Set up the FieldDAQ device as outlined in the FD-11634 Quick Start.
3. Configure the FieldDAQ device in Measurement & Automation Explorer (NI MAX) as
outlined in the FD-11634 Quick Start.
4. If the FieldDAQ device is not reserved automatically, select the device and click the
Reserve Network Device button. Refer to Reserving the Device in MAX for more
information.
5. Self-test your device in MAX by expanding Devices and Interfaces »Network Devices ,
right-clicking your FieldDAQ device, and selecting Self-Test. Self-test performs a brief
test to determine successful device installation. When the self-test finishes, a message
indicates successful verification or if an error occurred. If an error occurs, refer to ni.com/
support/daqmx.
Reserving the Device in MAX
When the FieldDAQ device is connected to a network, multiple users can access the device.
To perform any DAQ functionality on the device, including reset and self-test, you must
reserve the device in MAX. In MAX, an unreserved device or device reserved by another host
appears with an X and a reserved device appears as dark grey. Only one user at a time can
reserve the FieldDAQ device.
If the device was not reserved automatically after it was added (Add Device), you can reserve
the device in MAX by expanding Devices and Interfaces»Network Devices, selecting the
device, and clicking the Reserve Network Device button. The Override Reservation dialog
box opens when you attempt to explicitly reserve a device. Agreeing to override the
reservation forces the FieldDAQ device to be reserved by the current user.
Verification
Complete the following procedure to determine the As-Found status of the FieldDAQ device.
1. Set the calibrator to Standby mode (STBY).
2. Connect the calibrator to Bank 1 of the FieldDAQ device, as shown in the following
figure. Refer to Signal Connector Pinout for signal connection information.
4| ni.com | FD-11634 Calibration Procedure