NUI NCCD User manual

1
NCCD is a compact, lightweight chest
compression device to be used as a substute
to manual chest compression during CPR.
NCCD is powered by compressed gas, and
designed for use in hyperbaric environments.
NUI COMPACT CHEST COMPRESSION DEVICE (NCCD)
USER MANUAL
nui.no
January 2022

2
8. Hanson 600 series, 1/4” male
coupling
9. Chest belt
10. Velcro
11. Red markings on chest belt
12. Adjustable support strap
13. Extension hose (150 cm)
14. Scissors for cung fabric
NCCD COMPONENTS
1
23
4
6
5
7
8
9
10 11 10
13
Parts:
1-3. Compression unit
1. Piston house
2. Boom plate
3. Piston foot
4. Hand control unit
5. Compression release handle
6. Pressure hoses (100 cm)
7. Pressure hose (200 cm)
12 14

3
SAFETY
NCCD is only to be used by trained personnel with competence
in CPR and with experience in handling of pressurised gas.
Incorrect use may cause injury to the paent and user.
Use of NCCD can increase ppO2 and atmospheric pressure in
closed systems.
Do not use oxygen (>22%), or gas mixture with a O2 content
over 22% as supply gas.
Make sure not to stretch or pull hoses [6+7] to and from the
hand control unit [4].
Do not compress the hand control unit [4-5] while connecng
gas. The NCCD piston foot [3] can move and cause a pinching
hazard. Apply NCCD before gas is connected.
Incorrect positioning will reduce quality of compressions and
may cause injury to the patient.
MALFUNCTION
If proper applicaon and use of NCCD is not possible,
start manual CPR immediately.
Recommended checks and acons when possible:
1. Verify 10 bar supply pressure.
2. Check unit for damage or loose couplings.
3. Tell us about your experience: [email protected]
Please include NCCD serial number (SN08…..), found
on the back of the piston house [1], with correspondence.

4
USE OF NCCD
NCCD is only to be used when CPR is initiated on adult
patients, and only by personnel that have sufficient CPR
training.
NCCD is not to be used if trauma is suspected or during
pregnancy.
Your gas outlet connection needs to be Hanson 600 series, ¼“
female coupling. If your gas connection is of a different model,
please contact NUI.
Do not modify the NCCD in any way.
Do not use NCCD if the case seal is broken or missing. Does not
apply to the training units.
The NCCD will function while submerged in water.
INSTRUCTIONS
For further instrucons, images and videos, please visit:
nui.no/nccd
If proper application and use of NCCD is not possible,
start manual CPR immediately.

5
PRECAUTIONS
Maximum tension on pressurised hoses and ngs: 5 kg (49 N).
Max. distance [6] from paent to hand control unit [4-5]: 100 cm.
Maximum distance from paent to gas outlet [6+7] : 300 cm.
Maximum distance from paent to gas outlet using extension
hose [6+7+13]: 450 cm.
When connecng the NCCD QC [8] to gas outlet, make sure the
pressure release handle [5] is not compressed.
NCCD is to be used with pressurised HeliOx as drive gas at depth.
Compressed air can be used at surface. Use of air as drive gas can
signicantly limit the NCCD funconality in a hyperbaric
environment.
NCCD can be used along with debrillator pads from an AED.
Make sure to place the pads under the NCCD chest belt.
Aer use:
Aer use, make sure to disconnect the NCCD from the gas outlet,
and execute a couple of compressions to bleed o pressure from
the system. Be aware of pinching hazard.

6
A. Undress the paent’s upper body.
Cut clothes using included scissors [14] if needed.
B. Apply NCCD as instructed
on the following pages.
Image show NCCD applied.
APPLICATION OF NCCD

7
1. OPEN NCCD CASE
Break sealing. It requires 6 kg of force.
2. TAKE OUT NCCD
Remove the NCCD with chest belt [9] (black) from the
case. The belt is pre-aached to the right side of the
boom plate [2].
APPLICATION OF NCCD

8
3. BELT UNDER PATIENT CHEST
Place chest belt centered under
the paent’s back, posterior in
the middle of the back, between
the shoulder blades.
4. PLACE NCCD ON THE
PATIENT’S CHEST
Place the NCCD compression unit
1-3 on paent chest so that
the piston foot 3 is located
where your hands would have
been placed during manual CPR.
Make sure not to place the piston
foot to far down towards the
p of the sternum or too high up
on the chest. This could cause
injury to the paent or/and
cause insucient compressions.
APPLICATION OF NCCD
3)
4)

9
5. FASTEN BELT
a) Thread chest belt through slot
in the boom plate 2
b) Tighten the belt so the
compression unit is secured
on the chest.
Fasten the Velcro 10, and make
sure it’s rmly fastened on both
sides. Use red markings 11 to
center the belt.
c) Adjust both sides if necessary.
Conrm that the NCCD is rmly
applied to the paent’s chest.
The NCCD shall not compress
the chest by applicaon.
Conrm correct placement of the
piston foot [3] according to the
hand placement used when
you are preforming manual CPR.
APPLICATION OF NCCD
a)
c)
b)

10
NCCD can provide sucient
compressions when the
paent is in a vercal posion
when the NCCD is secured by the
adjustable support strap 2.
Remove clothes from the paent
upper body, and fasten the NCCD,
skin ght around the chest (do not
compress the chest with the belt).
a) Remove the adjustable support
strap 2 from the package, and
hang it around the paent neck.
b) Aach the carabiner hooks to
the slots in the NCCD plate 2 as
shown in image b) and c).
The strap length and tension needs
to be adjusted so that it supports
the NCCD, not causing an angle
(perpendicular to paent frontal
plane), or causing a rotaon of
the NCCD.
c) Verify that the adjustable strap
is aached as shown in the image.
a)
c)
b)
APPLICATION OF NCCD - VERTICAL PATIENT
12

11
6. CONFIRM PRESSURE and ppO2
NCCD runs on pressurised gas.
Conrm 10 bar “over ambient” gas drive pressure.
Conrm that gas supply has approx. the same ppO2 as
compartment atmosphere.
7. CONNECT TO GAS
Locate quick connector (QC) [8] for gas connecon. Insert male QC
on NCCD into gas outlet female QC (as you would a BIBS mask).
USE OF NCCD

12
8. START COMPRESSIONS
Locate NCCD hand control unit [4]. Hold it as shown in the
illustraon. You might nd it necessary to hold the control unit
with both hands.
Compress handle all the way in and release it all the way out.
Conrm that piston is compressing the chest according to CPR
requirements. Repeat and connue preforming chest
compressions in accordance with current CPR protocol.
Not fully compressing and releasing the handle on the control
unit will cause insuccient chest compressions. Strive to
complete full compressions og decompressions using as much
me to compress the handle
as to release the handle on
the control unit.
USE OF NCCD
Compr./Decompr.
Time

13
Clean NCCD aer use, before returning to NUI. Use a damp cloth
with mild detergent to remove visible contaminaon.
CLEANING / DISINFECTION
Biological maer
If the unit is contaminated with biological maer the unit shall
rst be cleaned with a cloth with detergent and warm water unl
visibly clean. Then use sucient disinfecon like alcohol +75% or
Virkon®.
NCCD chest belt [9]
The belt can be machine washed at 90°C (194°F) up to one hour
program. Shirking of the fabric when washed: <1 %.
If the belt is contaminated with biological maer, it should be
cleaned with soap and water, or machine washed as described
above. Then cleaned with a sucient disinfecon like alcohol 95%
or Virkon®.

14
TRAINING UNIT
If you are using a training unit, please make sure to repack
it according to the illustraon.
SERVICE INTERVALLS
Once a year, or when used, NCCD shall be returned to NUI
for maintenance. You will be contacted by NUI when it’s
me for service.
A new device will be sent to replace the unit that is
scheduled for maintenance.
RETURN AFTER USE
NCCD shall be returned to NUI aer being used on a
paent. Please return complete unit in transport case to:
NUI AS
Gravdalsveien 245
5165 Laksevåg
Norway

15
SPECIFICATIONS
NUI Compact Chest Compression Device (NCCD) is gas driven,
and has no electrical parts. It is designed to function under
hyperbaric conditions, and in heliox atmospheres.
NCCD will function while wet.
Specifications
Operang temperature: -5°C (41F) to 60°C (140F).
Sound level at operang distance: 80 dB.
Storage temperature: 5°C (41F) to 40°C (104F).
Maximum ambient pressure: 30 bar (435 psi/300 msw).
Requirements
Gas drive pressure: 10 bar (145 psi) over ambient.
Chest circumference: 90-130 cm (31-51 in).
Dimensions
Chest belt [10]: L150 cm, W11 cm.
NCCD [1-11] weight: 1950 g.
Case size: W265 x D240 x H174 mm.
Case empty weight: 1200 g.
Case with NCCD and accessories weight: 3420 g.

16
NCCD is delivered by NUI AS in Bergen, Norway
For more than 45 years, we have accuired unique competence,
experience and facilies within underwater technology.
NUI provides the subsea industry with a great diversity of
products and services; ranging from research, development
and advisory, to hyperbaric conngency, engineering,
pressure- and equipment tesng.
Our aim is to connuously contribute to improving equipment
safety, lowering health risks and raising awareness around
hyperbaric operaons.
www.nui.no
+47 55 94 28 00
N C C D
Other manuals for NCCD
1
Table of contents
Other NUI Medical Equipment manuals