4. Cleaning and Disinfecting
It is recommended that only CE marked cleaning products are used in the cleaning of
this device. Cleaning and Disinfection should be carried out by hand only.
Clean the device with warm water and neutral detergent and dry the surfaces
thoroughly using a soft cloth. Suitable disinfectants are: quaternary ammonium
compounds, isopropyl alcohol & chlorine bleach up to 1% (10,000 ppm). Apply
disinfectant by cloth, spray or disinfectant wipe. Following disinfection, wash off the all
surfaces with clean warm water and dry thoroughly using a soft cloth.
The product will be adversely affected and its life expectancy reduced if the following
cautions are not observed;
CAUTION:
-Do not steam clean or jet wash any area of this device.
-Do not soak or immerse the device.
-Do not use concentrated bleaching disinfectant solutions, organic solvent or
abrasive powders in the cleaning or disinfection of this product.
-Dilute all disinfectants in accordance with the manufacturer’s guidelines.
-Disinfectant products are corrosive in nature; failure to properly wash and
dry the product surface could leave a corrosive residue which may cause
damage to the product.
5. Product Maintenance
The Gardiner hand surgery system does not require servicing. However, before use,
visually inspect the device for any loose or damaged parts.
If any part of the Gardiner hand surgery system is damaged or faulty, the
device must be taken out of use with immediate effect and the fault reported
to Anetic Aid, your authorised dealer or maintenance department. The
Gardiner hand surgery system must not be used until the damage or fault has
been repaired.
CAUTION: In line with the MHRA Device Bulletin DB2006(5), maintenance
work should only be conducted by suitably trained personnel following
manufacturer’s guidelines.
6. Product Warranty
The product, when new, is guaranteed to be free from defects in materials and
workmanship and to perform in accordance with the manufacturer’s specification for a
period of 12 months from the date of purchase from Anetic Aid Ltd or their approved
Distributor. Anetic Aid Ltd will repair or replace, at their discretion, any components
found to be defective or at variance with the manufacturer’s specification within this
time at no cost to the purchaser. The warranty will take effect from the date of
purchase, subject to the purchaser registering the product with Anetic Aid to confirm its
receipt, installation date and product details.
The warranty does not provide cover for breakage or failure due to tampering, misuse,
neglect, accidents, modifications or shipping. The warranty is also void if the product is
not used in accordance with the manufacturer’s instructions or is repaired during the
warranty period by any persons other than Anetic Aid or its appointed agent. No other
expressed or implied warranty is given.
For details of our extended warranty packages please contact Anetic Aid or your
authorised dealer.