Nuvoair AIR NEXT User manual

AIR NEXT USER MANUAL

2
WHAT’S IN THE BOX
Air Next v 1.0
EN

3
SETTING UP YOUR DEVICE
Download the NuvoAir app
from AppStore or Google
Play Store.
1
Create an account or
sign in to an existing
(Air MD only)
2
EN

1
4
PERFORM A LUNG FUNCTION TEST
Insert a new turbine
by sliding it through
the device.
Turn the turbine
clockwise until it
locks into place.
2 3
Tap the plus button
to perform a lung
function tests.
The spirometer lights
up when connected.
Enable Bluetooth on your mobile device. Allow the app
to access and use Bluetooth and follow the instruc-
tions on the screen.
Enable Bluetooth on your
mobile device. Allow the app
to access and use Bluetooth
and follow the instructions
on the screen.
EN

Test
2s
5
Take a slow deep breath - as
deep as you can. Close your lips around the
opening of the turbine. Blow as hard and as fast
as you can into the mouth-
piece.
4 5 6
PERFORM A LUNG FUNCTION TEST
Your exhalation should
exceed 6 seconds in order
to obtain a high quality test.
Enable Bluetooth on your mobile device. Allow the app
to access and use Bluetooth and follow the instruc-
tions on the screen.
EN

Test
6s
Results Results
Flow
Volume
Battery: 100 %
6
PERFORM A LUNG FUNCTION TEST BATTERY
CHECK
Keep going until you
have completely
emptied your lungs.
You should rest
15 to 30 seconds
before starting a
new test.
Perform a minimum
of 3 tests for high
quality results.
The app displays the
best result.
Open app settings to
check battery levels.
7 8 9
You can share
the test via e-mail.
EN

7
INDICATIONS FOR USE
Air Next is intended to measure
the forced expiratory volume in
1 sec (FEV1) and the forced vital
capacity (FVC) in a forced expiratory
maneuver. These measures can be
used for detection, assessment and
lung function.
Air Next is intended to be used by:
• Healthcare professionals trained
to perform spirometry tests on
• Adults trained by healthcare pro-
fessionals (”competent adults”).
Competent adult can assist a child
The actual diagnosis shall be done
by a healthcare professional so
home use is only for indicative
purpose.
GENERAL INFORMATION
Air Next works with iPhone and
Android devices and connects via
bluetooth. Air Next is powered by
2 x 1.5 V AAA alkaline batteries.
Dispose replaced batteries and
discarded device as electronic waste.
Air Next is designed to work with
disposable and single use NuvoAir
disposable turbines. When perform-
ing a spirometry test, the user
exhales into the turbine. The air-
motion. Air Next registers the speed
of the spinning rotor, converts it and
transfers the data via Bluetooth to
the smartphone app.
LIMITATIONS OF USE &
CONTRAINDICATIONS
A spirometry test should only be car-
ried out when the user is at rest (i.e.
does not experience shortness of
breath) and in good health, and thus
in a suitable condition for the test.
An analysis of the results of a
to make a diagnosis of a clinical
condition. Test interpretation and
suggested treatment must be pro-
vided by a Healthcare professional.
A correct spirometry test depends
on the user’s capability to inhale and
exhale all air completely and as fast
as possible. If these fundamental
conditions are not respected then
the results obtained during spirom-
etry testing will not be accurate, and
therefore the test results are “not
acceptable”.
The acceptability of a test is the
responsibility of the user. Special
attention should be given when
EN

8
testing the elderly, disabled and
children.
The device should never be used
when it is possible or probable that
the validity of the results may be
compromised due to any external
factors.
Some conditions may pose a rel-
the validity of spirometry perfor-
mance and results. These include,
but are not limited to the following:
unstable cardiovascular
status, unstable angina, recent
myocardial infarction (within one
month) or pulmonary embolism,
haemoptysis of unknown origin,
recent pneumothorax, thoracic,
abdominal or cerebral aneurysms,
recent thoracic, abdominal or eye
surgery, acute disorders such as
nausea or vomiting, severe respira-
tory distress, physical
limitations, cognitive impairment,
dementia.
IMPORTANT SAFETY WARNINGS
Air Next should never be used
with a charging phone. Make sure
the phone is unplugged before
conducting a spirometry test. Don’t
use the device with a connected AC/
DC adaptor.
Once removed from its packaging,
check that there is no visible dam-
age on the device. In the case of
damage do not use it and return it
to the seller.
Air Next has been examine by an
independent laboratory which has
to the European Safety Standards
EN 60601-1 and guarantees the EMC
Requirements within the limits set in
the European Standard
EN 60601-1-2 and ETSI’s EN 300 328.
It is a medical device class IIa (“two
a”) product. Air Next is constantly
controlled during its production,
therefore the product conforms to
the essential requirements set by
the Council Directive 93/42/EEC for
medical devices.
• Use only with the devices
(and/or our online support:
www.nuvoair.com/support).
• Use only in environments
• Do not expose to liquids.
• Air Smart is not made with
natural rubber latex.
• Should you lose the battery
hatch with required and device
guarantee will no longer be valid
and NuvoAir cannot provide
further support or gurantee the
condition or functionality of the
product.
EN

9
Explanation of the safety signs and
symbols marked on the device:
Manufacturer’s name and address.
CE mark: indicates that the device
requirements of the 93/42/EEC
medical device directive.
device is protected against solid
objects over 2.5 mm entering as well
as faling drops of water, if the case is
disposed up to 15° from vertical.
Waste electrical and electronic
equipment: Dispose accordingly.
Do not dispose as unsorted
municipal waste.
Manufacturer’s LOT batch/lot code.
Follow operating instructions:
Indicates the need for the user
to consult the instructions for
important information.
The symbol is used in accordance
with IEC 60601-1-2:2014 (ED. 4)
for products including radio trans-
mitters and in accordance with
EN 300 328.
Type BF applied part: Device that
has conductive contact or medium
to long term contact with the
intended use.
OPERATING ENVIRONMENT
Air Next has been designed for use
setting or at home.
Air Next is not intended for use
in an operating theatre nor in the
or detergents, nor in the presence
(oxygen or nitrogen). The device is
not designed to be used in direct
air currents (e.g. wind), sources of
heat or cold, direct sun rays or other
sources of light or energy, dust, sand
or any other chemical substances.
NUVOAIR
DISPOSABLE TURBINE
Air Next is designed to be used with
a NuvoAir disposable single use
turbine. The disposable turbine is in-
dicated for single use single patient
only. The integrity and functionality
is guaranteed by:
• Never holding the turbine under
a jet of water or air and never let
it come into contact with high
• Not allowing dust or foreign
bodies to enter the turbine
sensor, in order to avoid incorrect
functioning and possible damage.
The presence of any impurities
such as hair, sputum, threads etc.
within the body of the turbine
sensor may seriously compromise
the accuracy of the measurements.
• Avoiding environmental
contamination by cleaning waste
products, the user must adhere
to all relevant regulations.
0598
IP32
EN

10
CAUTION: RISK OF
CROSSCONTAMINATION
Each new individual that uses Air
Next shall use a new unused dispos-
able NuvoAir disposable turbine. If
the turbine is not replaced between
individuals there is a high risk of
infectious cross contamination.
AIR NEXT TESTS
PARAMETERS SUCH AS:
FEV1 (Forced Expiratory Volume
in 1 second) The volume of air (l)
after a full inspiration.
FVC (Forced Vital Capacity) The
total volume of air (l) exhaled during
after a full inspiration.
FEV1/FVC (Ratio) The calculated
result obtained by dividing FEV1
with FVC.
PEF (Peak Expiratory Flow) The
after a full inspiration.
-
plication are always the best values
obtained after performing
the desired number of tests includ-
ed in a session. The recommended
number of tests per session is 3,
however the user can perform up
to 8 tests.
Displayed test are compared to a
reference table (GLI) and displayed
as percent predictive value
(%pred.).
The reference values for predicted
normal lung function (FEV1, FVC,
FEV1/FVC, PEF) are based on ethnici-
ty, gender, age and height.
Interpretation of spirometric results
should be made by a physician or
allied health care
in the performance and interpreta-
tion of spirometry.
Note: Home use of Air Next is
intended only for indicative purpose.
Interpretation of the test results and
diagnosis shall be done by a
healthcare professional.
MAINTENANCE
Handle Air Next device carefully.
Store it in clean and moisture free
conditions. Before use, always check
that the device is free from dust,
contamination or any particles.
Air Next does not require profes-
sional maintenance. If, however,
the outside of the device needs to
be cleaned make sure to only use a
damp cloth without any detergents.
Air Next device uses replaceable bat-
teries. NuvoAir Disposable Turbines
are disposable.
EN

11
ACCESS TO BLUETOOTH
Enable Bluetooth on your mobile
device in order to pair with the
Air Next. You can enable Bluetooth
in your mobile device settings.
Check your device manual on how
to do this.
BLUETOOTH PROXIMITY
Make sure that the Air Next is within
the Bluetooth proximity of your
phone. If your phone is too far away
from the Air Next, the connecton
could be lost.
TROUBLESHOOTING
• Make sure that the NuvoAir
Disposable Turbine is correctly
inserted into Air Next.
• Make sure that the turbine is not
spinning before you do a test.
• Make sure that Air Next has
• Make sure that Bluetooth is
enabled on your mobile device.
• Make sure the app is allowed to
use Bluetooth.
environment if it is too bright.
Try dimming the light or move to
a darker location.
EN

12
TECHNICAL FEATURES
Flow/volume measurement system Bi-directional optical
Measurement method Infrared interruption
Power supply 2 x 1.5V AAA alkaline bateries
Dimensions 98 x 62 x 26 mm
Weight 75g
Flow range 0 – 14 L/s
Volume accuracy ± 3%
Flow accuracy ± 5%
Dynamic resistance at 12 L/s <0.5 cm H2O/L/s
Operating conditions T: min +10°C/max +40°C
RH: min 10 %/max 95%
ALT: max 2000 m
Conditions of storage
T: min -20 °C/max +60°C
RH: min 10 % /max 95%
P: min 500 mBar /
Life expectancy 10 years
Regulations applicable:
MDD 93/42/EEC
ISO 26782:200
Electrical Safety IEC 60601-1
Electro Magnetic
Compatibility IEC 60601-1-2
ETSI EN 300 328
EN

13
NuvoAir AB
Riddargatan 17D
SE-11457 Stockholm
Sweden
NuvoAir AB guarantees that the
product is free from material
and fabrication defect, under
normal use conditions and under
the warranty period. For more
information about the warranty,
for reclamation or to provide
feedback, please visit
www.nuvoair.com
EN

14
Lieferumfang
Air Next v 1.0
DE

15
EINRICHTEN DES GERÄTS
Laden Sie die NuvoAir-App im
AppStore oder im Google
Play Store herunter.
1
Erstellen Sie ein Benutzerkonto
oder melden Sie sich mit einem
bestehenden Benutzerkonto an.
(nur Air MD)
2
DE

1
16
DURCHFÜHRUNG EINES LUNGENFUNKTIONSTESTS
Schieben Sie eine neue
Turbine in das Gerät ein.
Drehen Sie die Turbine
im Uhrzeigersinn, bis sie
einrastet.
2 3
Tippen Sie auf die
Plus-Taste, um einen
Lungenfunktionstest
durchzuführen.
Das Spirometer leuchtet
auf, wenn die Verbindung
hergestellt wurde.
Aktivieren Sie Bluetooth
an Ihrem Mobilgerät.
und die Nutzung von
Bluetooth durch die App
zu und befolgen Sie die
Hinweise im Display
DE

Test
2s
17
Atmen Sie tief ein - so tief
wie möglich. Schließen Sie Ihre Lippen um
Pusten Sie so stark und so
schnell Sie können in das
Mundstück.
4 5 6
DURCHFÜHRUNG EINES LUNGENFUNKTIONSTESTS
Für einen aussagekräftigen
Test sollte Ihre Ausatmungszeit
länger als 6 Sekunden sein.
DE

Test
6s
Results Results
Flow
Volume
Batterie: 100 %
18
DURCHFÜHRUNG EINES
LUNGENFUNKTIONSTESTS BATTERIE-
PRÜFUNG
Atmen Sie aus, bis Ihre
Lunge vollkommen
leer ist.
Sie sollten 15 - 30
Sekunden ruhen, bevor
Sie einen neuen Test
beginnen.
Führen Sie für aussage-
kräftige Ergebnisse min-
destens 3 Tests durch.
Die App zeigt das beste
Ergebnis an.
-
stellungen, um den
Batteriefüllstand zu
überprüfen.
7 8 9
Sie können die Testergeb-
nisse per E-Mail teilen.
DE

19
INDIKATIONEN FÜR DIE NUTZUNG
Air Next ist für die Messung des forcierten
Ausatmungsvolumen innerhalb von einer Se-
kunde (FEV1) und der forcierten Vitalkapazität
(FVC) vorgesehen. Diese Messungen können
zur Erkennung, Beurteilung und Überwa-
chung von Erkrankungen genutzt werden, die
die Lungenfunktion beeinträchtigen.
Air Next ist für die Verwendung durch
folgende Personen vorgesehen:
-
ersonal, das für die Durchführung von
• Erwachsene, die von professionellen
sind(“kompetente Erwachsene“). Kom-
Die tatsächliche Diagnose muss durch
einen Gesundheitsexperten erfolgen.
Die Anwendung zuhause dient daher nur
indikativen Zwecken.
ALLGEMEINE INFORMATIONEN
Air Next kann mit iPhones und
Android-Geräten verwendet werden und
verbindet sich über Bluetooth. Das Air Next
wird mit 2 AAA 1,5-V-Alkaline-Batterien be-
trieben. Wenn die Batterien ersetzt oder das
Gerät entsorgt werden müssen, muss beides
gemäß den lokalen Vorschriften entsorgt
werden. Das Gerät muss als Elektroschrott
entsorgt werden.
Air Next ist für die Nutzung mit wegwerfbaren
Einmal-Turbinen von NuvoAir vorgesehen.
Bei der Durchführung eines Spirometrietests
atmet der Benutzer in die Turbine aus. Der
Luftstrom versetzt einen Rotor in Bewegung.
Air Next erfasst die Geschwindigkeit des
drehenden Rotors, wandelt sie um und
überträgt die Daten über Bluetooth an die
Smartphone-App.
LIMITATIONS OF USE &
CONTRAINDICATIONS
Ein Spirometrietest darf nur durchgeführt
werden, wenn der Benutzer sich in einem
Ruhezustand (keine Kurzatmigkeit) und guter
gesundheitlicher Verfassung und damit in
einem geeigneten Zustand zur Durchführung
Die Analyse der Ergebnisse eines
Spirometrietests allein reicht nicht aus, um
eine Diagnose eines klinischen Zustands
zu stellen. Die Interpretation des Tests
und die Behandlungsempfehlung müssen
durch einen Gesundheitsexperten erfolgen.
Die korrekte Durchführung eines Spirometrie-
tests hängt von der Fähigkeit des Nutzers
zum vollständigen und schnellstmöglichen
Ein- und Ausatmen der gesamten Luft ab.
Werden diese grundlegenden Bedingungen
nicht erfüllt, sind die bei dem Spirometrietest
erhaltenen Ergebnisse ungenau und daher
„nicht akzeptabel“.
Die Akzeptabilität eines Tests obliegt der
Verantwortung des Nutzers. Besondere
Aufmerksamkeit ist erforderlich, wenntesting
DE

20
Ältere oder behinderte Personen und
Kinder untersucht werden.
Das Gerät darf nie verwendet werden,
wenn es möglich oder wahrscheinlich
ist, dass die Gültigkeit der Ergebnisse
aufgrund externer Faktoren gefährdet
ist. Einige Umstände können für Patient-
en eine relative Gefahr darstellen oder
die Gültigkeit einer Spirometrie und
ihrer Ergebnisse beeinträchtigen. Dazu
gehören unter anderem: instabiler kar-
diovaskulärer Zustand, instabile Angina,
kürzlicher Herzinfarkt (innerhalb eines
Monats) oder Lungenembolie, Hämop-
tyse ungeklärter Ursache, kürzlicher
Pneumothorax, thorakale, abdominale
oder zerebrale Aneurismen, kürzliche
Operation an Thorax, Abdomen oder
Augen, akute Störungen wie Schwindel
oder Erbrechen, schwere Atemnot,
körperliche Einschränkungen, kognitive
Beeinträchtigungen, Demenz.
WICHTIGE SICHERHEITSHINWEISE
Air Next sollte nie mit einem Telefon
verwendet werden, das geladen wird.
Stellen Sie sicher, dass das Telefon ausges-
teckt ist, bevor Sie einen Spirometrietest
durchführen. Verwenden Sie das Gerät
nicht, wenn ein Wechselstromadapter
angeschlossen ist.
Überprüfen Sie das Gerät nach dem
Auspacken auf sichtbare Beschädigungen.
Sollten Beschädigungen vorhanden sein,
verwenden Sie das Gerät nicht und geben
Sie es beim Händler zurück.
Air Next wurde von einem unabhängigen
Labor überprüft, das die Konformität des
Gerätes mit den Europäischen Sicherheits-
die Einhaltung der EMV-Anforderungen
innerhalb der im Europäischen Standard
EN 60601-1-2 und ETSI EN 300 328
festgelegten Einschränkungen garantiert.
Es handelt sich um ein medizinisches
Gerät der Klasse IIa („Zwei a“). Air Next
wird während der Produktion ständig
kontrolliert, daher erfüllt das Produkt
die grundlegenden Anforderungen
der Richtlinie 93/42/EEC des Rates für
Medizinprodukte.
• Nur mit den in dieser Bedienungsan-
leitung (und/oder im Online-Support:
www.nuvoair.com/support) angege-
benen Geräten verwenden.
• Nur in einer in der Bedienungsanleitung
genannten Umgebung anwenden.
• Das Gerät nicht mit Flüssigkeit in
Kontakt bringen.
• Air Smart ist nicht mit Naturkautschuk
hergestellt.
• Sollten Sie die Batterieklappe mit den
Informationen verlieren, erlischt die
Garantie und NuvoAir kann keinen
weiteren Support bieten oder den
Zustand oder die Funktionalität des
Produkts garantieren.
DE