OPERATOR’S MANUAL HEATED GUTTA PERCHA SYSTEM
Page 3
SAFETY INFORMATION
1. Please read the operator’s manual in its entirety before using the Obtura III MAX.
2. To avoid explosion hazard, do not use the Obtura Ill MAX in the presence of ammable materials.
3. Do not allow heated parts to come in contact with soft tissue. All handpiece components attached to the heating chamber are hot and may
cause burns if used improperly. During normal use, the handpiece is not intended to contact the patient. Always use a thermal protector for
added protection in the event of accidental patient contact. In addition, the use of a rubber dam will increase protection.
4. Do not abuse the handpiece cable. Ensure that the handpiece cable and power supply cable are free from cuts, nicks or other damages.
5. Place the main control unit within sufcient proximity to the patient so as not to disrupt operation by reaching too far with the handpiece.
6. This Obtura System complies with current Electromagnetic Compatibility standards (IEC 60601-1-2). However, the user should ensure that any
possible electromagnetic interference does not present an additional risk of interfering with or damaging the device or any other electronic
devices in the vicinity (see pages 12-15).
7. The Obtura System does not use RF energy to function. Therefore, it is not likely to cause any interference in nearby electronic equipment.
(see pages 12-15).
8. The Obtura System should only be used with the provided cables and accessories. Use of other cables or accessories has the potential to
damage the device or affect compliance to EMC standards.
9. Cables and Accessories:
• Handpiece Cord - max length 8’, IEC 60601-1
• Power Cord - max length 8’, IEC 60601-1
10. Requirements of cables and accessories:
• 2-wires unshielded cable - 1.5m, AC/DC Adapter Type: FW7555M/12
• Obtura Handpiece shielded cable - 2m, Handpiece Attachment/Detachment
Disclaimer: Young Innovations, Inc. does not assume any responsibility or liability for incorrect diagnosis or Jailed procedures due to operator error or equipment malfunction.
Clinicians who are not familiar with the techniques or intended uses of this product should attend courses and receive training on the subject prior to use.
To ensure continued compliance with EC Directives and safety, use only the power supply included with this product or supplied by
Young Innovations, Inc., part number 823-711.
OVERVIEW
Congratulations on your purchase of the Obtura III Max! Please read the entire manual prior to using this unit. Here are some useful tips that
will aid your usage of the Obtura III Max:
When you are ready to practice with your Obtura III MAX, you should become familiar with its components. After attaching the needle and thermal
protector, turn the Obtura III MAX ON and set it to operating temperature. It takes the unit approximately 120 seconds to reach the highest operating
temperature of 200°C. The high temperature setting will improve gutta percha ow, especially when used in the cold practice block (sold separately).
Begin by lling the plastic practice blocks with gutta percha per the instructions included. Use condensers to condense the material void free.
We recommend the Obtura Spartan S-Kondenser (available in 3 sizes).
• The soft silver applicator needles are designed to conduct heat to insure even at the needle tip; however, as the needle tip cools
between uses, you should express 3 cm of gutta percha to pre-heat the tip just prior to use. The needles are bendable to t the tooth,
however, they may break if they are repeatedly bent or if they are bent at the needle attachment point. Always use the Multi-Tool to bend
the needle. Do not kink the needle, as any restriction will impede ow. After a few practice sessions, proper needle bending will be easily
accomplished. Obtura Applicator Needles are designed for one use only.
• Do not loosen or remove the blue plastic Heater Nut. The Heater Nut is fastened to the gun with a set-screw; removal of the heater nut will
cause damage. The Heater Nut must be kept tightly fastened in order to secure the heater to the handpiece when expressing gutta percha.
• The stand for the handpiece may be positioned by itself or it may be attached to the bottom of the control unit. The stand can be attached to
either side of the control unit with four phillips-head screws that are fastened to the bottom of the stand (see page 5 for further details).
• The cord to the handpiece can extend from the front or rear of the unit. If a front exit is desired, gently press the handpiece cable into the
tracks on the bottom of the unit, to exit either on the left-front or the right-front (see page 5).
• Cleaning the Obtura III MAX is simple but very important; just follow the directions in this manual (see page 7) or on the provided laminated
cleaning instruction card. If your Obtura III MAX is cleaned on a regular basis it will provide years of reliable service. If you do not clean your
unit, it may become clogged and difcult to move the plunger due to a buildup of old, hardened gutta percha.
Thank you for purchasing the Obtura III MAX. We believe it is a benecial instrument that will enhance your practice. We are continually striving
to improve our products and services. If you have any questions or suggestions, we would appreciate hearing from you. We are also interested in
seeing any X-rays you may have of interesting cases. Feel free to contact us at: 1-800-344-1321 (or 1-877-485-3556 in Canada); we welcome the
opportunity to speak with you. Also, visit our website at: www.obtura.com.