Oldelft Benelux Easy DR User manual

User Manual: Easy DR
ODB-100-5070ENG 1 Rev.:17-04-2020
User Manual
Easy DR
Rev.: 17-04-2020
0344

User Manual: Easy DR
ODB-100-5070ENG 2 Rev.:17-04-2020
Blank Page

User Manual: Easy DR
ODB-100-5070ENG 3 Rev.:17-04-2020
List of revisions
Equipment
Easy DR
Document
User Manual
Document number
ODB-100-5070ENG
revision
Date
Comment
01
09-05-2011
First setup F Vijn
02
14-07-2011
Added label position (1.3) and user warning (4.10)
03
31-10-2011
Changed Symbols (page 5 and 6)
Added EMC warning (page 7)
Added Warning Squeezing hazards chapter 3.1 and 4.10
Changed text chapter 4.2 Powering up.
Added separate chapter 5.1 Cleaning.
Changed chapter 5.2 Safety check.
Changed technical specifications (chapter 6)
04
16-10-2012
Changes regarding 60601- 3e edition and text updates
05
31-05-2016
Changed Sedecal interface to CPI interface
Added detailed power on/off instructions.
Updated collimator info.
06
29-09-2016
List with manuals delivered with the Easy DR
Add Harmony3-ST Docking
07
02-03-2017
Correction telephone- and faxnumber
08
07-09-2018
Add Variant 183cm arm/correction environment conditions
/correction WEEE symbol
09
17-04-2020
Add Lateral positioning arm (option)

User Manual: Easy DR
ODB-100-5070ENG 5 Rev.:17-04-2020
Preface
Identification
Manuals are identified by a part number.
The part number is printed at the bottom of the title page. Numbers printed on
other pages are for internal revision control and may differ.
CE Classification
Manuals made for products that have been CE (Conformité Européenne) marked
have the CE mark on the front page. The number behind the CE Mark defines the
responsible notified body.
Oldelft Benelux B.V. systems are classified according to IEC 60601-1 as Class I,
type B. Oldelft Benelux B.V. systems meet the relevant provisions of the European
Medical Device Directive 93/42/EEC; this is based on conformity of the products
and the quality system according to ISO 13485:2016.
Symbols
Symbol
Meaning
Remarks
Location: Near or on
the type identification
plate.
Symbol for separated collection of
electrical and electronic equipment
per Directive 2002/96 of the
European Parliament and the
Council of the European Union
(Directive on Waste of Electrical
and Electronic Equipment - WEEE).
If applicable to the type of device,
it indicates legally imposed
obligations within EU member
states, Iceland, Norway and
Switzerland when the equipment is
disposed of, at the end of its life-
time.
Owners of marked equipment
should contact the organization that
imported the equipment into their
country, when they want to dispose
of the equipment, at the end of its
lifetime. The Directive prioritizes re-
use of equipment over re-use of
components over re-use of
materials over disposal as waste.
Article 5 part 2d allows producers to
decline the return of any used
equipment that is or may be
biologically or radiologically
contaminated.
Legal manufacturer

User Manual: Easy DR
ODB-100-5070ENG 6 Rev.:17-04-2020
Date of manufacture
Serial number
This mark indicated that this is a
Type B Applied Part according to
EN60601-1. This unit can be
installed in the patient
environment.
Caution: Consult Instructions for
Use
Consulting the accompanying
documents is advised.
Consulting the accompanying
documents is a mandatory action.
General warning sign
Placed together with a
supplementary symbol or text. The
text associated shall be an
affirmative statement (i.e., a safety
notice) describing the principal
risk(s) foreseen (e.g. “Causes
burns”, “Risk of explosion”, etc.).
0344
The CE Mark is a declaration by the
manufacturer that the product
complies with the requirements of
the applicable European Union (EU)
medical device directive and that
the product has been subject to
conformity assessment procedures
as provided in that directive.
The number behind the CE Mark
defines the responsible notified
body.

User Manual: Easy DR
ODB-100-5070ENG 7 Rev.:17-04-2020
Conventions
The Easy DR has been designed to meet all safety requirements applicable to
medical equipment. However, anyone using the Easy DR must be fully aware of
potential safety hazards. The information contained in this chapter is provided to
help users to operate the Ease DR safely.
Throughout this manual certain conventions are used. These are Warnings,
Cautions and Notes. They provide a means of prioritizing information to be brought
to the attention of the user. They are given as follows:
WARNING:
Indication of an important warning that is to be obeyed to
eliminate the risk of personal injury or an incorrect clinical
diagnosis.
CAUTION:
Indication of important information to avoid serious system
equipment damage.
Note:
Additional information for reader understanding or
simplification of a task.
Precautions
The user is expected to use the product (soft and hardware) in accordance with
the instructions given in this manual, which must be read before the system is
used. Any unauthorized deviation from the procedures laid down in this manual
can affect the contractual obligations between purchaser and vendor.
WARNING:
This product was designed and manufactured to ensure
maximum safety of use. It should be installed and serviced
in strict compliance with safety instructions described in
this Document.
WARNING:
Never attempt a shortcut to procedures which may degrade
the safety considerations of the system.
WARNING:
Class 1 MD: To avoid the risk of electric shock, this
equipment must only be connected to a supply mains with
protective earth.
WARNING:
The user shall take care that potentional electromagnetic
or other interferences between the unit and other
equipment is avoided. If necessary certain actions must be
taken.
WARNING:
Only trained and qualified personal are allowed to operate
the Easy DR system.

User Manual: Easy DR
ODB-100-5070ENG 8 Rev.:17-04-2020
WARNING:
Do not modify any part of Oldelft Benelux supplied
hardware, software or software configuration. Any
modification, installation or running an application that has
not explicitly been approved by Oldelft Benelux may
compromise patient safety and will result in termination of
all warranty and service contracts.
a. If this equipment is modified, repaired or
maintained, appropriate inspection and testing must
be conducted to ensure continued safe use of the
equipment.
b. Do not leave problems unattended that may affect
the safety of the product. In case you have doubts
concerning the Easy DR‘s save operation, call the
service number of your vender for further
instructions.
c. The Easy DR should be cleaned regular to prevent
the accumulation of dust. The contact surface should
be cleaned after each patient for hygienic reasons.
WARNING:
The Oldelft Benelux products are designed for use and
maintenance using only parts available from Oldelft
Benelux. No liability for failure can be accepted by Oldelft
Benelux through the use of parts obtained from other
suppliers unless written permission is obtained from
Oldelft Benelux.
Copyright
All rights reserved. No part of this manual may be reproduced or
transmitted in any form whatsoever, including photocopying and
recording without written permission of the copyright holder:
Oldelft Benelux B.V., Wiltonstraat 41, 3961 KW VEENENDAAL, The
Netherlands.
Concerning this Publication
Information contained in this publication is subject to change
without notice. Whenever the equipment is delivered with a
documentation medium such as a CD or DVD, the user shall
always check the medium for latest information. This information
is part of Release Notes and Document Upgrade files.
Neither Oldelft Benelux B.V. and its subsidiaries nor anyone else
who has been involved with the development, production or
delivery of the documentation be liable for any special, incidental
or consequential damages, whether based upon breach of
warranty, breach of contract, negligence, strict liability in tort or
any other legal theory.
This manual was originally drafted in the English language.

User Manual: Easy DR
ODB-100-5070ENG 9 Rev.:17-04-2020
1Introduction..................................................................... 10
1.1 Variants ........................................................................... 10
1.2 About this manual.............................................................. 10
1.3 Classification..................................................................... 10
1.4 Installation ....................................................................... 11
1.5 Technical description .......................................................... 11
1.6 Label and symbols ............................................................. 12
1.7 Reference manuals ............................................................ 12
1.8 General description/indications for use .................................. 13
1.9 Physical description............................................................ 15
1.10 Functional description......................................................... 18
2Safety devices .................................................................. 21
2.1 General safety................................................................... 21
2.2 System safety ................................................................... 22
3Operating controls ........................................................... 23
3.1 Hand control ..................................................................... 23
3.2 X-ray generator control panel .............................................. 24
3.3 Collimator controls............................................................. 25
3.4 The user interface of control computer .................................. 26
4Operating instructions ..................................................... 27
4.1 Releasing the shipping locks ................................................ 27
4.2 Powering up...................................................................... 27
4.3 Switching off, applying the shipping locks .............................. 29
4.4 Entering patient and exposure related data............................ 31
4.5 Patient data entry .............................................................. 32
4.6 Protocol selection............................................................... 34
4.7 Changing exposure settings................................................. 34
4.8 Making an exposure ........................................................... 35
4.9 Making a pa thorax X-ray.................................................... 36
4.10 Reviewing images .............................................................. 38
5Maintenance..................................................................... 40
5.1 Cleaning........................................................................... 40
5.2 Safety switch check............................................................ 40
5.3 Troubleshooting................................................................. 40
5.4 Sending images afterwards ................................................. 42
5.5 Calibrate .......................................................................... 43
5.6 Calibration procedure ......................................................... 43
5.7 Troubleshooting after calibration procedure............................ 47
6Technical specifications ................................................... 48

User Manual: Easy DR
ODB-100-5070ENG 10 Rev.:17-04-2020
1 Introduction
1.1 Variants
The Easy DR is produced in two variants:
Easy DR
FDD
Rotation
Use
1
145cm
0⁰~ -90⁰
Chest screening
Multipurpose
2
183cm
Non-rotation
Chest screening
FDD = Focus Detector Distance (distance between Focal spot of X-ray tube to
the detector)
Note:
This manual is based on the variant 1.
When the Easy DR is Variant 2:
the 145cm have to be substituted by 183cm;
rotation of the arm is not possible and can be skipped.
Al other functions are similar as variant 1.
1.2 About this manual
This user manual is a guide for the operator of the Easy DR multifunctional X-
ray unit. In this manual it is assumed that the operator is a professional and
trained X-ray user, who has gained sufficient knowledge to operate Easy DR
system safely.
In particular this document describes the daily operation and interpretation of
the user interface for the Easy DR. It also contains:
•A brief description of the Easy DR system
•System safety
•Operating procedures
•User maintenance procedures
Please study this manual carefully before attempting to use the Easy DR.
Proper functioning of the apparatus can be guaranteed only if the instructions
in this manual are adhered to.
1.3 Classification
The system is classified according to Class 1 type B.
The system meets the relevant provisions of the European
Medical Device Directive 93/42/EEC.

User Manual: Easy DR
ODB-100-5070ENG 11 Rev.:17-04-2020
1.4 Installation
The installation procedure can be found in the:
•Service Manual / ODB-100-5080ENG
1.5 Technical description
The technical description can be found in the:
•Technical specifications Easy DR / ODB-100-5090ENG

User Manual: Easy DR
ODB-100-5070ENG 12 Rev.:17-04-2020
1.6 Label and symbols
Label
Location
This label is described in document: Labels / ODB-100-5100ENG
1.7 Reference manuals
Note:
This document complements the reference manuals
provided with this system and cannot replace these
manuals. The user should therefore take note of the
reference manuals, especially the safety rules before
using the system. When this document and the reference
manual differs, the reference manuals are considered as
leading.
List with manuals delivered with the Easy DR
ODB-100-5070ENG
User manual
Key-User Manual Easy DR
Operation manual HF Generator
Computer
Monitor
Detector, CXDI-401G/C Compact
Power box PB-6
CXDI Control Software Operation manual
CXDI Control Software Setup guide
Harmony3-ST Docking

User Manual: Easy DR
ODB-100-5070ENG 13 Rev.:17-04-2020
1.8 General description/indications for use
The Easy DR is a film-less X-ray system suitable for a wide range of
radiological exams. The design is based on the WHO recommend WIS-RAD
system and together with the Rogan ROX and VPX viewing station is very
suitable for (mass) screening on TBC.
X-ray photons arrive at the DR plate (sensor) and the scintillation cause a
reaction in which light is released. This light is absorbed and converted into a
current through a photoelectric converter, a SI-sensor.
Example of a DR sensor
A sensor contains lines and rows with pixels. These pixels are read out directly
after exposure. All the different currents from the pixels are converted into a
corresponding value so the computer can display an image. The captured X-
ray image is stored in digital memory. Digitized images are transferred as
DICOM 3.0 compatible files to a Picture Archive and Communication system
(PACS). The PACS system is used to distribute to diagnostic viewing station.

User Manual: Easy DR
ODB-100-5070ENG 14 Rev.:17-04-2020
The Easy DR is equipped with an elevator stand, which is firmly bolted to the
floor. It can lift the support frame to the desired height. The support frame
can also turn between o and -90 degree. This allows for a bed trolley to be
placed over the detector so bed patients can be done. The support frame
consists of the detector unit and the X-ray tube unit.
The X-ray generator may be installed outside the operator area. The X-ray
generator control panel (on/off) is located in the X-ray protected area. Both
the detector and X-ray unit are controlled by the control computer.
Easy DR system main components
Depending on the configuration the Easy DR System may contain the following
sub-systems:
•The Easy DR unit (Elevator stand and support frame)
•The Control computer (separate manuals)
•X-ray equipment (separate manuals)
•Optional PACS and Diagnostic Viewing Station (separate manuals)
•Optional printer (separate manuals)
•Optional UPS/battery system

User Manual: Easy DR
ODB-100-5070ENG 15 Rev.:17-04-2020
1.9 Physical description
In this paragraph only the Easy DR main hardware items are described. These
are:
•Elevator stand
•Support frame
•Control computer
•X-ray tube unit
•X-ray generator and control panel
•Collimator unit
1.9.1 Elevator stand
The support frame is mounted on the elevator stand. The detector and X-ray
tube can move vertically on the elevator stand to adapt the chin rest to the
patient height suitable for PA Thorax imaging. The support frame can also turn
from 0 to -90 degree. Rotation can only be done when the support frame is
between 850 and 1350 mm above the ground. This prevents the unit from
being damaged by the floor or ceiling.
When an obstacle touches the safety switch that is mounted at the bottom
side of the detector unit, the downward and/or rotation movement of the unit
immediately stops. The down and/or rotation control switch on the hand
control unit is overruled as long as the safety switch is activated.
Another way to stop the operation of the elevator stand and support frame is
pressing the emergency switch located next to the collimator unit.
1.9.2 Support frame
The support frames holds the detector and X-ray unit. The system has a fixed
focus detector distance (FFD) of 145cm. This allows for a wide range of
radiological exams. The support frame is designed in such a way that in all
positions patient placement is flexible and no cables or frames are in the way.

User Manual: Easy DR
ODB-100-5070ENG 16 Rev.:17-04-2020
1.9.3 Control computer
The control computer is used to operate the detector and X-ray unit. The
system comes with a wide range of preconfigured protocols for an optimal
image quality. By selection of a protocol the correct X-ray settings (kV, Ma and
Ms) are automatically set. The Automatic Exposure Control (AEC) will give the
most optimal image quality at the lowest possible patient dose. The Control
computer is also used to temporarily store the X-ray images. It is interfaced
via a network to the PACS and Diagnostic Viewing stations.
1.9.4 X-ray tube unit
The X-ray tube unit consists of a X-ray tube, a filter unit and a collimator unit.
The X-ray tube is mechanically aligned with the detector unit. The X-ray tube
focus is always directed to the detector unit.
The filter unit contains a rotating disc with four types of X-ray filters. The filter
unit is intended for maintenance purposes only; its setting should not be
changed by the operator.
1.9.5 X-ray generator and control panel
The X-ray generator has its own documentation.
Advice to consult the accompanying documentation:
•Operation manual HF generator.
A two-step radiation exposure switch is connected to the (mini) generator
interface panel. The following functions are available on the generator control
panel:
•Switching the X-ray system ON or OFF

User Manual: Easy DR
ODB-100-5070ENG 17 Rev.:17-04-2020
1.9.6 Collimator unit
The collimator unit comprises an adjustable diaphragm and a field illumination
indicator.
The adjustable diaphragm has two setting knobs: one to adjust two vertical
shutters for symmetrical lateral adjustment, the other to move the horizontal
shutters to adjust vertical field size.
The field illumination indicator delineates the total extent of the field to be
exposed. It can be switched on and off by a push button. If left on, the light
will turn off automatically.
In general, the lateral diaphragm is adjusted to maximum for Posterior-
Anterior (PA) exposures of normal patients, and to less than maximum for
lateral exposures or for exposures of small patients.
1.9.7 Lateral Positioning Arm.
As an option, for the Easy DR, a Lateral Positioning Arm can be mounted to the
detector frame. The mounting hardware is already set by the manufacturer
when the equipment is ordered with this option. It only requires to insert the
arm bar into its working position.
To do this:
•Insert the Lateral positioning Arm into the support block located on the
detector frame.
•Lock the Lateral Positioning Arm into the extension rod by using the control
knob.
WARNING:
Ensure the Lateral Positioning Arm is properly secured
before using the system with patients.
WARNING:
In cases where it is deemed necessary the use of the Lateral
Positioning Arm, take caution before making any movement
of support frame mounted on the elevator stand, in order to
make sure that there is no risk of collision.
WARNING:
Max Load 15kg

User Manual: Easy DR
ODB-100-5070ENG 18 Rev.:17-04-2020
1.10 Functional description
The Easy DR operates without X-ray film: the X-ray image is captured with the
X-ray detector and stored in digital memory. Digitized images are stored as
DICOM 3.0 compatible files a local hard disk (short term) and automatically via
a network link to the PACS and diagnostic viewer station.
After positioning the patient, the operator prompts the Easy DR to make an X-
ray exposure. Before the actual X-ray exposure can be made the patient's data
have to be entered on the control computer. After patient data is entered the
X-ray exposure is taken.
1.10.1 Exposure interval
The actual exposure starts when the two stage exposure switch is pressed.
The AEC automatically switch of the radiation as soon as the desired dose is
measured by the system.

User Manual: Easy DR
ODB-100-5070ENG 19 Rev.:17-04-2020
1.10.2 Elevator stand
The elevator stand moves the support frame up and down to accommodate
the patient height. This movement is controlled by means of the elevator
control switches on the hand control unit. When actuated, a safety switch
fitted to the bottom of the detector unit prevents the elevator from moving
down.
In case of emergency an emergency stop button, located just next to the
collimator unit, can be pushed. Pushing the emergency stop button will
immediately stop the movement of the scanner unit. After removing the
threat, and after releasing the emergency stop button, the unit is ready for
use again.
1.10.3 Support frame
The support frame holds the detector and X-ray unit. The X-ray tube unit is
placed on a fixed, 145 cm, distance from the detector unit. The support frame
can be turned from 0 to -90˚. In the -90˚ positions a bed can be placed over
the detector to accommodate other radiological procedures. When actuated, a
safety switch fitted to the bottom of the detector unit prevents the elevator
from moving down.
In case of emergency an emergency stop button, located just next to the
collimator unit, can be pushed. Pushing the emergency stop button will
immediately stop the movement of the scanner unit. After removing the
threat, and after releasing the emergency stop button, the unit is ready for
use again.

User Manual: Easy DR
ODB-100-5070ENG 20 Rev.:17-04-2020
1.10.4 Control computer
The control computer is the controlling device for the Easy DR rather than a
means for advanced diagnostics, which would require very high
resolution/brightness as in fast viewing stations. These are beyond the scope
of this manual.
The control computer offers a graphical user interface to perform the following
functions:
•Input of patient data, either directly (touch screen) or via import from
peripheral systems (network, barcode scanner)
•Control of X-ray detector (Data acquisition, post processing and display)
•Control of X-ray unit (kV, mA and mS settings)
•Database management for patient data and related X-ray images
•Export of digital X-ray images to peripheral systems (network)
Radiation on/off switching is controlled by the Control computer and X-ray
generator interface but always executed by the two stage exposure switch.
Generator technique settings are controlled by the control computer. However
Radiographers should follow well-known guidelines and procedures for
operating the X-ray generator and are supposed to have read and understood
the manufacturer’s instructions for that device.
Advice to consult the accompanying documentation:
•Operation manual HF generator.
Table of contents
Popular Medical Equipment manuals by other brands

Mitaka Kohki
Mitaka Kohki PSMS2 user manual

Wallach Surgical Devices
Wallach Surgical Devices SummitDoppler Vista ABI user manual

ACL
ACL OR-CONSOLE Series Instructions for use

Allied Healthcare
Allied Healthcare ResusciTIMER manual

Freedom Innovations
Freedom Innovations RS2 Instructions for use

Smiths Medical
Smiths Medical Pneupac babyPAC 100 user manual