Olympus IX2-UCB User manual

INSTRUCTIONS
IX2-UCB
U-HSTR2
CONTROL BOX FOR IX2
HAND SWITCH
A X 6 4 4 4
This instruction manual is for the Olympus Control Box model IX2-UCB and Hand Switch model
U-HSTR2, both for use with the IX2/GX motorized microscope. To ensure safety, optimum perfor-
mance and familiarize yourself fully with the use of the motorized microscope, we recommend that
you study this manual thoroughly before operating the system. Retain this instruction manual in an
easily accessible place near the work desk for future reference.
Optical Microscope Accessory
Downloaded from ManualsNet.com search engine

In accordance with European Directive on Waste Electrical and Electronic Equipment, this
symbol indicates that the product must not be disposed of as unsorted municipal waste, but
should be collected separately.
Refer to your local Olympus distributor in EU for return and/or collection systems available in
your country.
For Korea only
ࣾ匏ࣜ匶匶ࣜࣜࣤ儆洛殯ࣜ愯暧皻柦匶沖沲ࣥ
決ࣜ匶匶垚ࣜ儆洛殯ࣤࣾ匏ࣥࣜ洊沖砒洇穯匶匶嵢昢ࣜ渂嵢ࣜ儆洛櫖昢ࣜ斲殯穞垚ࣜ冉汊
徯洇求嵢ࣜ穞彶ࣨࣜ微姦ࣜ滆櫳櫖昢ࣜ斲殯穦ࣜ朞ࣜ沎枻城埪࣪
This product is applied with the requirements of standard IEC/EN61326-1 concerning
electromagnetic compatibility.
- Immunity Applied to industrial and basic environment requirements.
NOTE: This product has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference when the product is operated in a commercial environment. This product
generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance
with the instruction manual, may cause harmful interference to radio communications.
Operation of this product in a residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
FCC WARNING: Changes or modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the product.
Downloaded from ManualsNet.com search engine

IX2-UCB/U-HSTR2
CONTENTS
IMPORTANT — Be sure to read this section for safe use of the equipment. — 1-2
1 NOMENCLATURE
2 OPERATION
2-1 Control Box IX2-UCB-2............................................................................................................................................................. 5
1Turning Power On 2Functions of Indicator LEDs
2-2 Hand Switch U-HSTR2 .............................................................................................................................................................. 5
1Attaching Indication Stickers 2Grouping Panel Sheet
2-3 Operation Selection of the DIP Switches ........................................................................................................ 6
PROPER SELECTION OF THE POWER SUPPLY CORD.......................................................... 17- 18
3-4
5-6
IX2 SERIES MOTORIZED SYSTEM
þNote on operation in stand-alone mode ý
When the IX81 motorized revolving nosepiece is rotated in stand-alone mode (without a PC), automatic
escape movement of an objective (prevention of interference between the objective and stage) is not
performed. Therefore, when using an objective with a short working distance, let the motorized
revolving nosepiece escape once before pressing the objective switching button.
3-10
3 SPECIFICATIONS
4 TROUBLESHOOTING GUIDE
5 ASSEMBLY - If you intend to assemble the unit by yourself, read this section first. -
GX SERIES MOTORIZED SYSTEM
1 MOTORIZED SYSTEM DIAGRAM
2 NOMENCLATURE
7
8
9-10
11- 16
12-13
14-16
Downloaded from ManualsNet.com search engine

1
IMPORTANT
The IX2-UCB (Type 2) control box is the basic module for controlling the drive functions of the IX2
microscope with motorized specification. It also incorporates the power supply for the microscope. It
can be combined with one of the Type 2 microscope frames: IX81S1F-2, IX81S8F-2 and IX81F-2.
Motorized operation from the U-HSTR2 hand switch or a PC is possible. Besides, the functions of the
buttons can be set as desired by using the IX2-BSW software (Ver. 01.03 or later).
The IX2-BSW software (Ver. 01.03) is compatible with Windows 2000 and Windows Me. For details, see the
IX2-BSW's help file.
When your GX Series motorized system uses the IX2-UCB (Type 2) control box and U-HSTR2 hand switch,
read this “Important" section and “GX Series Motorized System" section on pages 11 to 16.
Please do not use this control unit with BX2 series components.
SAFETY PRECAUTIONS
1. Before connecting cables, be sure to set the main switch of the IX2-UCB control box to “ ” (OFF)”.
2. Make sure to ground the equipment for safety and for maintaining the electrical safety performance.
3. When installing the control box, leave spaces of more than 10 cm around it. (Note that the control box also has a
ventilation air inlet on the front panel).
4. Never insert a metallic object, etc. into the air vent of the control box. Otherwise, electric shock or malfunction may result.
Also be careful not to allow the provided small articles, including the push ring driver and Allen wrench, enter the air vents
of the associated module used together with the control box.
5. Distribute the power cord and other cables away from the lamp housing and its surroundings. Otherwise, the cord or
cable coating may melt by heat and cause an electric shock hazard.
The following symbols are found on the unit. Study the meaning of the symbols and always use the equipment in the
safest possible manner.
Safety Symbols
Symbol Explanation
Before use, carefully read the instruction manual. Improper use could result in personal injury
to the user and/or damage to the equipment.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
l
Indicates the illumination, which is limited only to a 12 V, 100 W halogen bulb with this system.
Bottom surface is hot, and should not be touched with bare hands.
Downloaded from ManualsNet.com search engine

2
IX2-UCB/U-HSTR2
2Caution
If the equipment is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the
equipment may also be damaged. Always use the equipment as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
: Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment).
# : Indicates that failure to follow the instructions could result in damage to equipment.
} : Indicates commentary (for ease of operation and maintenance).
1Getting Ready
1. The control box and hand switch are precision equipment. Handle them carefully by avoiding any shock or impact, and
also connect each cable gently.
2. Do not use the equipment under a direct sunlight, in a place under high temperature and humidity or in a place subject
to vibrations. (For the operating environmental condition, see chapter 3, “SPECIFICATIONS” on page 7.)
3. While the main switch of the IX2-UCB control box is set to “ I” (ON), do not replace any module, plug or unplug any cable
or switch the light path manually to prevent malfunction.
4. Never disassemble any part of the unit as this could result in mafunctions or reduced performance.
3Intended use
This instrument has been designed to be used to observe magnified images of specimens in routine and research
applications.
Do not use this instrument for any purpose other than its intended use.
Downloaded from ManualsNet.com search engine

3
NOMENCLATURE
Make sure to connect the Olympus-specified module to each connector.
The PC in use should meet the IEC60950 requirements.
If any non-specified equipment is used, Olympus cannot guarantee any performance of the system.
Control Box IX2-UCB-2
Indicator LEDs
· RMT: Lights at the time of remote control (in orange).
· ERR: Blinks in case of an error (in red).
Lights during attachment (in green)/blinks during operation.
· NP: 6-position motorized revolving nosepiece
· PRC: Motorized light path selector
· MU: Either the IX2-RFACA or U-FWR
· CDT: Either the IX2-LWUCDA or U-FWR
· FW: Either the U-FWO or U-FWR
· BP: IX2-TVRAC
· Z: U-ZPCB
·
· Spare
·
Main switch ( I: ON, : OFF)
RS232C connector (9-pin male)
PC connector
DIP switches
Used for selection of operation
settings. (p.6)
U-HSTR2 (Hand Switch) connector
RFACA/FW1 connector
MAIN 1 connector
MAIN 2 connector
MAIN 3 connector
100 W halogen lamp housing connector
(Note) The connector on the right cannot be used.
Option slot
Z board can be installed.
Option slots (2 ch)
Power cord connector
LWUCDA/FW2 connector
FW3 connector
IX2 SERIES MOTORIZED SYSTEM
Downloaded from ManualsNet.com search engine

4
IX2-UCB/U-HSTR2
Hand Switch U-HSTR2
Grouping Panel Sheets (3 types) Indication Stickers
TSHT (transmitted light shutter) button
RSHT (reflected light shutter) button
Projections for blind-
touch operations
Connector
Auxiliary button
Objective switching buttons
}The button functions shown below are the functions in stand-alone mode or the initial functions set at the moment the
system is started after having installed the IX2-BSW software (Ver. 01.03 or later) in the PC.
The button functions can be modified as desired by using the software.
}Above each button, attach the indication sticker corresponding to the function set for the button. When you use the buttons
with their initial settings, simply attach the indication stickers as shown in the following figure.
}The current position can be identified by the lighting of the positions where the indication stickers are attached (objective
in use/TSHT/RSHT/BP).
for direct selection of 1 to
6 when the IX81 is used
Light shielding
stickers
Indication
stickers
Blank stickers
Meanings of the and buttons
Each press of the or button changes the position number in the decremental direction (1
-
>6
-
>5
-
>4
-
>3
-
>2
-
>1) or in the incremental direction (1
-
>2
-
>3
-
>4
-
>5
-
>6
-
>1) respectively.
CDT– (condenser switching) button
CDT+ (condenser switching) button
( )
MU+ (mirror unit switching) button
MU– (mirror unit switching) button
FW+ (filter wheel switching) button
FW– (filter wheel switching) button
BP (bottom port) observation button
Downloaded from ManualsNet.com search engine

5
OPERATION
2-1 Control Box IX2-UCB-2
1Turning Power On (Fig. 1)
Fig. 1
Ensure that the modules to be used are connected properly. (P. 9)
1. Set the main switch @ to “ I” (ON).
2. Ensure that the LED indicators ² corresponding to the connected mod-
ules are lit.
2Functions of Indicator LEDs (Fig. 1)
1. RMT: Lights only at the time of remote control.
2. ERR: Blinks in case of an error. At this time, the associated indicators
blink as described below.
3. NP to Z: Each indicator lights when the corresponding module is installed.
2-2 Hand Switch U-HSTR2
Fig. 2
1Attaching Indication Stickers (Fig. 2)
1. Attach each piece of the provided function indication stickers onto the
dented area @ above the button where the corresponding function is
set.
2. The indication stickers are given weak adhesive force intentionally so
that they can be removed and re-attached easily.
3. The indication stickers include two types of stickers carrying no indication
on them.
· Light shield sticker: Attach to the dented area above a button with no
function set.
· Blank sticker: Create a custom indication sticker by writing the function
name with oily ink and attach to the dented area above
the required button.
Also usable for substitution of light shield stickers when
they have run out.
Fig. 3
2Grouping Panel Sheet (Fig. 3)
Two sheets showing the function groups of buttons with enclosing lines
@ and a blank sheet ² are provided. Select and use the sheet that is
most convenient.
The blank sheet can be used by drawing desired grouping lines with oily
ink pen.
}Sheet 1(the topmost sheet) has been designed for use in stand-alone
mode or when you use the buttons with their initial settings.
@
²
@
²
@
Downloaded from ManualsNet.com search engine

6
IX2-UCB/U-HSTR2
2-3 Operation Selection of the DIP Switches
}The allocated functions of the DIP switches are shown in the table below.
# Make sure the main switch is set to “ ” (OFF) before setting the DIP switches. The unit detects the new settings
only when the power is switched on, making those settings effective.
}The shaded sections show factory settings (all set at OFF).
Regarding the RS232C Cable
# Be sure to use a commercially available RS232C straight cable. (Use of other cables may cause malfunction.)
Use a D-Sub 9P (female)–D-Sub 9P (female) connector. Be sure to set the main switches of the control box and PC
to “ ” (OFF) before connecting the RS232C cable.
Setting of U-ZPCB etc.
# When combining the U-ZPCB or other focus control boards, set the following dip switches on the board to ON and
insert to the control box;
· S3-No.4
· S4-No.5
Be sure to set the main switches of the control box to “ ” (OFF) before inserting the board.
OFF
12345678
OFF
ON
OFF
OFF
OFF
OFF
OFF
ON
OFF OFF OFF OFF OFF OFF OFF OFF
DIP Switch
Switch Position Function Detail
SW1 (upper row)
Buzzer prohibited Activates buzzer.
Does not activate buzzer.
Reserved for manufacturer
Initialization prohibited
when power is switched on
All fixed at OFF.
Reserved for manufacturer
SW2 (lower row)
Does not initialize.
Initializes.
OFF
Reserved for manufacturer
Fixed at OFF.
Fixed at OFF.
Downloaded from ManualsNet.com search engine

7
SPECIFICATIONS
Item Specification
Control Box IX2-UCB-2
Power supply rating Input rating: 100 to 120/220 to 240 V , 50/60 Hz, 3.5/1.5 A
LED indicators · RMT (Remote) LED · ERR (Error) LED
· Module installation check LED x 10, (Spare x 3)
Option slots Power capacity (single slot)
Per board Total of 3 slots
+5 V 1 A max. 2 A max.
+15 V 1 A 1 A (normal) + 1 A (motor load 20% duty)
+24 V 1 A max 2 A max.
Dimensions & weight 125(W) x 216(H) x 310(D) mm, approx. 5 kg (11 lb.)
Hand Switch U-HSTR2
Button functions Connects to the IX2-UCB-2 for use in the control of following operations.
Also operable from a PC (which meets the IEC60950).
Stand-alone mode or initial settings at the start of PC
· FW–/FW+/CDT–/CDT+/MU–/MU+ buttons
· RSHT/TSHT/BP buttons
· Objective direct selection buttons (1 to 6)
Dimensions & weight 147 (W) x 32 (H) x 108 (D) mm., approx. 0.37 kg (0.81 lb.)
Operating environment · Indoor use.
· Altitude: Max. 2000 m.
· Ambient temperature: 10 to 35°C.
· Maximum relative humidity: 80% (up to 31°C). The maximum humidity
decreases linearly at above 31°C, through 70% (at 34°C), 60% (at 37°C)
to 50% (at 40°C).
· Supply voltage fluctuation: ±10%.
· Pollution degree: 2 (in accordance with IEC60664).
· Overvoltage category: II (in accordance with IEC60664).
Downloaded from ManualsNet.com search engine

8
IX2-UCB/U-HSTR2
TROUBLESHOOTING GUIDE
Under certain conditions, performance of the microscope may be adversely affected by factors other than defects. If a
problem occurs, please review the following list and take remedial action as appropriate. If you cannot solve the problem
after checking the entire list, please contact your local Olympus representative for assistance.
Problem Cause Remedy Page
a) ERR (Error) indicator blinks. Module corresponding to the indicator
blinking simultaneously with ERR is not
connected properly.
Connect the motorized module of the
simultaneously blinking indicator prop-
erly.
b) Power cannot be turned on by set-
ting the main switch to “ I” (ON).
Power cord is unplugged. Connect the power cord properly.
c) Communication through RS232C
is impossible.
The RS232C cable in use is wrong. Use the specified RS232C cable.
The RS232C cable is not connected
properly.
Connect it properly.
d) The Hand Switch buttons do not
work or the indicators do not light.
The hand switch is not connected prop-
erly.
Connect it properly.
e) A motorized module does not work
or its indicator does not light.
The motorized module is not connected
properly.
Connect it properly.
f ) The lamp will not light. The lamp housing connector is not con-
nected properly.
Connect it properly.
The lamp on-off switch of microscope
is set to OFF.
Set it to ON.
The lamp is blown. Replace the lamp.
9
10
9
10
9
9
–
–
6
The IX2-BSW (Ver. 01.03 or later) is not
logged in.
Log it in. –
The lamp housing is plugged into the
right-side connector. 9
–
g) The objectives do not escape
when they are switched.
h) The rotation direction of the focus
adjustment knob and the move-
ment of the objectives do not agree
with each other.
The on-board DIP switches on the
U-ZPCB Z-board are not set correctly.
Set them correctly by referring to the
instruction manual for the IX81.
Plug it into the left-side connector.
The on-board DIP switches on the
U-ZPCB Z-board are not set correctly.
Set them correctly by referring to the
instruction manual for the IX81.
They do not escape in stand-alone
mode.
Install IX2-BSW (Ver. 01. 03 or later) and
start it up.
–
–
Downloaded from ManualsNet.com search engine

9
ASSEMBLY
5-1 Assembly Diagram
Motorized shutter*
IX2-SHA
Motorized fluorescence
mirror unit cassette
IX2-RFACA
Motorized long-WD universal
condenser
IX2-LWUCDA
Filter wheel
U-FWO
U-FWR
Microscope
frame
IX81S1F-2
IX81S8F-2
IX81F-2
100 W halogen bulb lamp housing
U-LH100L-3
IX-HLSH100 + U-RMT (extension cord)
IX-LH100-3 + U-RMT (extension cord)
Rear panel
Hand Switch
U-HSTR2
Power cord
RS232C cable
Control Box
IX2-UCB-2
Be sure to set the main switch to “ ” (OFF)
before connecting cables.
Focusing knob unit
U-FH
Focusing
interface
U-IFFH
PC
*Plug the IX2-SHA into the SHA1 (top) connector when it has been mounted in the transmitted illumination light path,
and plug it into the SHA2 (bottom) connector when it has been mounted in the reflected illumination light path.
Downloaded from ManualsNet.com search engine

10
IX2-UCB/U-HSTR2
5-2 Detailed Assembly Procedures
Fig. 4
Make sure that the main switch @ of the control box to “ ” (OFF)
before connecting the cable of the Hand Switch (and other modules)
and the power cord. (Fig. 4)
The power cord and connection cables are sensitive to bending or
twisting. Do not apply excessive force to them.
1Connecting the Hand Switch Cable (Fig. 4)
Align the connector of the Hand Switch with the HS connector ² on the
front panel of the Control Box and plug in firmly.
2Connecting the Power Cord (Figs. 5 & 6)
Always use the power cord provided by Olympus. If no power cord
is provided, please select the proper power cord by referring to the
section “PROPER SELECTION OF THE POWER SUPPLY CORD” at
the end of this instruction manual. If the proper power cord is not
used, product’s safety performance cannot be guaranteed.
1. Insert the connector | of the power cord into the power cord connector
³.
Connect the other end of the power cord to a power outlet with 3
conductors including the grounding line. If the power outlet is not
grounded properly, the electrical safety performance intended by
Olympus cannot be guaranteed.
2. Insert the plug ƒ on the other end of the power cord into the power
outlet ….
Distribute the power cord at a distance from the lamp housing. If the
power cord comes in contact with a hot part around the lamp hous-
ing, the cord may be melted, causing an electric shock hazard.
3. Bundle the connection cables using the Velcro (5 pieces) provided with
the microscope frame.
Fig. 5
Fig. 6
@
²
ƒ…
|
³
Downloaded from ManualsNet.com search engine

11
GX SERIES MOTORIZED SYSTEM
For the motorized modules that can be mounted on the system, see Chapter 1, "MOTORIZED SYSTEM DIAGRAM" on
page 12 Using an unsuitable module hinders the system from manifesting its full performance.
Combination of the IX2-UCB-2 control box and U-HSTR2 hand switch are required to control the motorized modules.
Configuration of Instruction Manuals
The instruction manuals for the motorized modules are provided separately as shown in the following table. This
instruction manual from this page on pertains to the instructions of the GX-RTUA motorized mirror unit turret and the
features specific to the system using it.
Manual Name Main Contents
IX2-UCB/U-HSTR2
Control Box/Hand Switch
Functions of the control box (including power supply) and hand switch in
combination with the IX2 Series are described in the first half. Operations
with the original assembly of the GX71/GX51 motorized system are
described in the second half.
U-FWT/FWR/FWO
Motorized Filter Wheels
Functions of the motorized filter wheels (only the U-FWR is usable with
this system) are described.
If your GX71 or GX51 is provided with the manual entitled “GX71/GX51 Motorized System", please
discard it because it has errors.
NOTE
Downloaded from ManualsNet.com search engine

12
IX2-UCB/U-HSTR2
MOTORIZED SYSTEM DIAGRAM
Be sure to connect the Olympus-specified module to each
connector. If other module than specified is connected, Olympus
can no longer warrant the performance of the system.
Be sure to distribute the cables away from the lamp housing
and the surroundings (particularly, take special care to the
filter wheel). If a cable comes in contact with them, it may melt
and electric shock may result.
Motorized Revolving Nosepiece
U-D6REM
U-D5BDREM
(Note 1)
To prevent disconnection of the cables
of the motorized revolving nosepiece and
mirror unit turret, mount the provided
cable cover (for how to mount and how
to use, see next page).
Filter Wheel
U-FWR
Inverted Metallurgical
System Microscope
GX71
GX51
Motorized Mirror Unit Turret
GX-RTUA
<Mountable only on the GX71>
Connection Cable
U-REMMT
(Note 1)
Rear
Control Box
IX2-UCB-2
Set the main switch to “ ” (OFF) before connection.
Power cord
Hand Switch
U-HSTR2
(Note 1)
Downloaded from ManualsNet.com search engine

13
Fig. 7
Attaching and Using the Cable Cover (Figs. 7 to 9)
1. Connect the connector @ of the motorized revolving nosepiece connec-
tion cable.
@
Fig. 8
Fig. 9
2. Attach the cord clamp ³, provided with the GX71/GX51, to the bottom
of the cable cover ² by adhesion.
3. Pass the motorized revolving nosepiece cable through the slot | on
the cable cover and fit the cable cover around the revolving nosepiece
connector.
|
²
³
³
…
ƒ
|4. Clamp the cable cover by lightly tightening the two clamping knobs on
the bottom and left side panel.
# Do not tighten the knobs too much, for this will damage the revolving
nosepiece connector.
5. Pass the revolving nosepiece cable ƒ and mirror unit turret cable …
through the cord clamp ³ and clamp the cables by reserving certain
slacks on them.
Downloaded from ManualsNet.com search engine

14
IX2-UCB/U-HSTR2
NOMENCLATURE
}The functions of the IX2-UCB-2 control box and U-HSTR2 hand switch described in the following pertain only to
those available when this system configuration is used. The GX71/GX51 motorized system does not use the
controls that are not described below.
Control Box IX2-UCB-2
Indicator LEDs
· RMT: Lights in remote mode. (Orange)
· ERR: Blinks in case of an error. (Red)
Each of the following indicators lights (in Green)
when the corresponding module is connected
and blinks when it is functioning.
· NP: Motorized revolving nosepiece
· MU: Motorized mirror turret unit GX-RTUA
· FW: Filter wheel U-FWR
Main switch ( : ON. : OFF)
HS (Hand Switch) connector
RFACA/FW1 connector
(Connector for GX-RTUA)
MAIN3 connector
(Connector for U-REMMT)
Power cord connector
CAUTION
FW3 connector
(Connector for U-FWR)
Downloaded from ManualsNet.com search engine

15
Hand Switch U-HSTR2
}As this system configuration does not use a PC, the functions of the hand switch buttons become as shown below when
the cable is connected. (Stand-alone mode)
}See page 5 for how to attach the indicator stickers and how to use the grouping panel sheets.
Not used
Not used
When the U-D5BDREM is used
Connector
MU– (mirror unit switching) button
For GX-RTUA
MU+ (mirror unit switching) button
For GX-RTUA
Objective switching buttons
For direct selection of 1 to 5 or 1 to 6
(with motorized revolving nosepiece)
Meanings of the and buttons
Each press of the or button changes the position number in the decremental direction (1
-
>6
-
>5
-
>4
-
>3
-
>2
-
>1) or in the incremental direction (1
-
>2
-
>3
-
>4
-
>5
-
>6
-
>1) respectively.
FW+ (filter wheel switching) button
For U-FWR
FW– (filter wheel switching) button
For U-FWR
Downloaded from ManualsNet.com search engine

16
IX2-UCB/U-HSTR2
Motorized Mirror Unit Turret
GX-RTUA
(Note 1) The applicable microscope is only the GX71.
(Note 2) Replace the manual mirror unit turret, provided as standard with the GX71, with
the motorized mirror unit turret.
(Note 3) Remove the mirror units from the manual mirror unit turret and put them in the
motorized mirror unit turret.
Motorized control cable
Mirror unit replacement cover clamping screw
Detachable using the Allen screwdriver provided
with the microscope.
Cable connector
Mirror unit replacement cover
*Mirror unit indicator numbers
Indications of 1, 2, 3, 4, 5 and 6
* Place the indicator plate of the mounted mirror unit's number in the indicator plate
pocket with the corresponding number.
Mirror unit indicator plate pocket*
Indicator plate
Downloaded from ManualsNet.com search engine

17
PROPER SELECTION OF THE POWER SUPPLY CORD
If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to “ Specifications ” and
“ Certified Cord ” below:
CAUTION: In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant the
electrical safety of the equipment.
Specifications
Voltage Rating
Current Rating
Temperature Rating
Length
Fittings Configuration
125V AC (for 100-120V AC area) or, 250V AC (for 220-240V AC area)
6A minimum
60°C minimum
3.05 m maximum
Grounding type attachment plug cap. Opposite terminates in molded-on IEC con-
figuration appliance coupling.
Table 1 Certified Cord
A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked with an
agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of agencies listed in
Table 1. In case you are unable to buy locally in your country the power supply cord which is approved by one of the
agencies mentioned in Table 1, please use replacements approved by any other equivalent and authorized agencies in
your country.
Country Agency Certification
Mark Country Agency Certification
Mark
Argentina
Australia
Austria
Belgium
Canada
Denmark
Finland
France
Germany
Ireland
IRAM
SAA
ÖVE
CEBEC
CSA
DEMKO
FEI
UTE
VDE
NSAI
Italy
Japan
Netherlands
Norway
Spain
Sweden
Switzerland
United
Kingdom
U.S.A.
IMQ
KEMA
NEMKO
AEE
SEMKO
SEV
ASTA
BSI
UL
JET, JQA, TÜV,
UL-APEX / MITI
Downloaded from ManualsNet.com search engine
This manual suits for next models
1
Table of contents
Other Olympus Controllers manuals