Omega HH550 User manual




CONTENTS
Title page
1. SAFETY INFORMATION................................................................................................1
2. SPECIFICATION.............................................................................................................2
2-1 General Specification...........................................................................................2
2-2 Electrical Specification.......................................................................................... 5
3. PARTS & CONTROLS....................................................................................................6
3-1 The Name of Each Parts and Positions................................................................ 6
3-2 Power Status........................................................................................................9
3-3 Quick Manual for Easy Set-up..............................................................................9
3-4 ACV
、
DCV and DCA Test Setting
..........................................................................9
3-5 UNIT Setting Step .............................................................................................. 10
3-6 ALARM Setting Step........................................................................................... 10
3-7 Real Time Clock , Time/ Date Setting Step........................................................ 11
3-8 Record/Print Start Time, Stop Time, Interval Time Setting Step......................... 11
3-9 Print Mode Setting Step...................................................................................... 12
3-10 Clear Record Step............................................................................................ 12
3-11 Disable Start
、
Stop
、
Intv and Setting Step.................................................... 12
3-12 Decimal Point Setting Step...............................................................................12
3-13 Consecutive Record and Print Setting Step ..................................................... 12
3-14 Reset System Setting Step............................................................................... 12
3-15 Record / Print Start and Step Operation Method.............................................. 13
3-16 Memory Full Warning....................................................................................... 13
3-17 Alarm Print in Graphic Mode ............................................................................ 13
3-18 Scale Change in Graphic Mode........................................................................ 14

3-19 How to ConnectsTL-(A37) Analog Output Cable.............................................. 14
3-20 Setup Process Transducer (4-20mA) Mode..................................................... 15
4. BATTERY & FUSE REPLACEMENT...........................................................................16
4-1 Battery Check & Replacement ...........................................................................16
4-2 Fuse Replacement............................................................................................. 16
4-3 Note.................................................................................................................... 16
5. DATALOGGING............................................................................................................16
5-1 Software Installation........................................................................................... 17
5-2 Default Settings.................................................................................................. 21
5-3 Meter Stand alone Operation ............................................................................. 36
5-4 Download Datalogger (Graphic Recorder
è
PC)............................................... 38
5-5 Data Conversion................................................................................................. 41
5-6 On-Line Operation with PC.................................................................................45

1
1. SAFETY INFORMATION
q
Read the following safety information carefully before attempting to operate
or service the meter.
qThe meter analog input protection voltage range is 24V ,current
range is 250V/0.5AFuse(MAX) , maybedamage forever ifoverload.
Environment conditions
:
Altitude up to 2000 meters
Indoor use
Relatively humidity 10~80%RH
Operation temperature 0
〜
40
℃
(32~104
℉
)
Maintenance & Clearing
Repairs or servicing not covered in this manual should only be
performed by qualified personnel.
Periodically wipe the case with a dry cloth. Do not use abrasives or
solvents on this instruments.
Do not use abrasives or solvents on this instruments.
Safety symbols
:
Meteris protected throughout by double insulation or
reinforced insulation.
Comply with EN-61010-1

2
2. SPECIFICATION
2-1 General Specification
Display :3 2
1LCD
LCD symbols:
1342 2656724
18 19
11 28
17 20
2227259142310
21
13
12
15
16
8
(Fig-1)
1.
SET
:
In setting mode symbol.
2.
RS-232
:
In RS-232 transmission mode symbol.
DATE
3.
– – :
Show the date of real time clock.
Show mode : month– day– year.
TIME
4.
: : :
Show the time of real time clock.
Show modeis hour : minute : second.
START
5.
: :
:
Show the start time of record / print.
Show modeis hour : minute : second.

3
STOP
6.
: : :
Show the stop time of record / print.
INTV
7.
: : :
Show the interval time of record / print.
8. :DC test symbol.
9. :AC test symbol.
10.
:Low battery indicator.
11. mVA :Analog test indicate unit.
12. G:The printer working in the graphic mode.
13. T:The printer working inthe text mode.
14. =:Memory record indicator.
15. MAX :MAX Analog input indicator.
16. MIN :MIN Analog input indicator.
17. LUX :LUX unit indicator.
18. Fc :Feet-candle unit indicator.
19. dB :dB unit indicator.
20. %RH :Relative Humidity%RH unit indicator.
21.
:
Printer in using
indicator
.
22. .
:
7 segments to show the read value and decimal point of
LCD.
23.
:Negative polarity.

4
24.
HI ALARM
:
Hi alarm
indicator
.
25.
LO ALARM
:
Lo alarm
indicator
.
26.
HiHi ALARM
:
Hi Hi alarm
indicator
.
27.
LoLo ALARM
:
Lo Lo alarm
indicator
.
28.
℃,℉:
Temperature unit
indicator
.
29. Over Load Indicator :Show or - 。
Other parts:
Input Protection :Voltage range 24V ,current range 0.5A/250V
Fuse(MAX)
Operation Temperature:0℃〜40℃( 32℉〜104℉) & (10~ 80% RH)
and Humidity
ConserveTemperature:-10℃〜60℃(14℉〜140℉) & (10~70% RH)
and Humidity
Dimensions :About 250 (L) ×100 (W) ×45(H) mm
Weight :About 500g with battery
Printer :Interval print :
00hour:00minute:03second〜23hour :
59minute : 59sec
Trigger print : Printthreedata
(date / time / Reading)
Accessories :Thermo carrying case, Test leads with alligator
clip, 2 rolls of thermo paper, 110V/220V AC to 9V
DC power adapter/power cord, RS-232 cable, P.C.
software disk, 6pcs 1.5V AA batteries, TL-A37
analog output cable, RS-232 convert header,
Operation manual.

5
2-2 Electrical Specification
Analog test range :DCV/ACV:2V,200mV
DCA:20mA
Accuracy :(at 23℃±5℃,below 80%RH)
DCè(±1%rdg+3dgts)
ACè(±1%rdg+5dgts) at (50~450Hz)
Resolution :200mV/2V range is 0.1mV/1mV
20mA range is 0.01mA
A/ D Sample Rate :2.5 times/sec
Power Source :6 pcs 1.5V,# 3 battery or 9V, DC adaptor
(inside negative, outside positive)
Battery Life :Approx. 100 hours (without printing)
Printer :Thermal printer (EPL1902of Panasonic)
MEMORY :60000 records (up to 130000 records)

6
3. PARTS & CONTROLS
3-1 The Name of Each Parts and Positions:
PRINT
600V
60V
24V
MAX
COM V
mA
FUSED
20mA
MAX
FEED
LINE MIN
MAX
UNITRANGE
1,
20
200mV 2V 200mV 2V 20mA
C¢X FcLux dB ¢X
¢X
RH A A
DATE TIME START STOP TG
INTV
HiHi Hi LoLoLo
TRIGGER
3 CONSECUTIVE
,
3,
F
¢X
SETTING
13456789
CLOCK ALARM
RECORD
SETUP
(Fig-2)

7
1.
200mV
C / F
HiHi
DATA
0
Toggle button for ( 200mV, ℃/℉, HiHi, DATA)
and digit “ o ”.
2.
2V
Lux/Fc
Hi
TIME
1/2/3
Toggle button for ( 2V, Lux/ Fc, Hi,TIME) and
digit “ 1/2/3 ”.
3.
200mV
dB
LoLo
START
4/5/6
Toggle button for ( 200mV, dB, LoLo, START)
and digit “ 4/5/6 ”.
4.
2
V
%RH
Lo
STOP
7/8/9
Toggle button for ( 2V,%RH, Lo, STOP) and
digit “ 7/8/9 ”.
5.
20m
A
A/
G/T
INTV Toggle button for ( 20mA, A / A,
G / T, INTV ) button and “ ” right shift
function.
6. 3'SETTING
SETUP
Press and hold this button for 3 sec will into
setting mode, press again will end setting
mode.

8
7. RANGE
FEED
LINE
Feed paper in normalmode, range function in
setting mode.
8. UNIT
MAX
MIN
MAX/MIN functionin normal mode “ UNIT ”
function in setting mode.
9.
CLOCK
PRINT
In normal mode, press 1 sec is trigger print
function, press and hold for 3 sec is start/stop
consecutive print function, but in setting mode
will be “ clock ” function.
10.
ALARM
RECORD
=
In normal mode, press 1 sec is trigger record (
MAX 255 times), In setting mode is “ alarm ”
function, press and hold over 3 sec is start/stop
consecutive record function.
11. Power button.
12 “ COM ” Measuring
Socket To connect negative lead of voltage, current.
(black color)
13 “ V ” Measuring
Socket To connect positive lead for voltage.(red color)
14 “ mA ” Measuring
Socket To connect positive lead for current.
15
LCD
Show test value, unit symbol and range indicator.
16 Thermal Printer Printing data.
17 Acryl Putting thermal paper.
18 RS-232 Connector
RS-232 used.
19 FEED Button Feed paper into printer
20 DC 9V Socket To connect adaptor.

9
3-2 Power Status :
As battery power is not sufficient, the , will be shown on LCD
replace with six new 1.5V batteries is required.
3-3 Quick Manual for Easy Set-up
STEP1 STEP2 STEP3 STEP4 STEP5
Push this
button 3 second. Select one. Choose one from specified
function Bar. Reconfirm
input setting.
Press to
return
normal.
a. You could only set up one ofthe functions at each time.
b. To change position of point decimal, follow step1 & step2, then press button.
c. To input digit, follow step1 & step3 to set the number.
3-4 ACV
、
DCV and DCA Test Setting :
1. ACV 2V Test setting sequence:Press and hold 3'SETTING for 3 sec
àpress RANGE àpress 2V àpress RANGE àpress
3'SETTING , to end setting mode.
2. ACV 200mV Test setting sequence:Press and hold 3'SETTING for 3
sec àpress RANGE àpress 200mV
àpress RANGE à
press 3'SETTING , to end setting mode.
3. DCV 2V Test setting sequence:Press and hold 3'SETTING for 3 sec
àpress RANGE àpress 2V
àpress RANGE àpress
3'SETTING , to end setting mode.
RANGE
UNIT
CLOCK
ALARM
200mV 2V 200mV 2V 20mA
cFLUX / FC dB %RH A / A
DATE TIME START STOP INTV
HiHi Hi LoLo Lo G / T
SETUP
RANGE
UNIT
CLOCK
ALARM
SETUP

10
4. DCV 200mV Test setting sequence:Press and hold 3'SETTING for 3
sec àpress RANGE àpress 200mV àpress RANGE à
press 3'SETTING , to end setting mode.
5. DCA 20mA Test setting sequence:Press and hold 3'SETTING for 3
sec àpress RANGE àpress 20mA
àpress RANGE à
press 3'SETTING , to end setting mode.
3-5 UNIT Setting Step:
1. ℃/℉Setting step:press and hold 3'SETTING for 3 sec àpress
UNIT àpress C/ F change℃/℉àpress UNIT à
press 3'SETTING , to end setting mode.
2. Lux/Fc Setting step:press and hold 3'SETTING for 3 sec àpress
UNIT àpress Lux/Fc change Luc/Fc àpress UNIT à
press 3'SETTING , to end setting mode.
3. dB Setting step:press and hold 3'SETTING for 3 sec àpress
UNIT àpress dB
àpress UNIT àpress 3'SETTING ,
to end setting mode.
4. %RH Setting step:press and hold 3'SETTING for 3 sec àpress
UNIT àpress %RH
àpress UNIT àpress 3'SETTING ,
to end setting mode.
5. / A Setting step:press and hold 3'SETTING for 3 sec àpress
UNIT àpress change A/ A àpress UNIT àpress
3'SETTING , to end setting mode.
3-6 ALARM Setting Step:
1. HIHI ALARM Setting step:press and hold 3'SETTING for 3 sec à
press ALARM àpress HiHi àassort 0~9 number èright
shift button setting “ALARM" àpress ALARM àpress 3'SETTING ,
to end setting mode.
2. HI ALARM Setting step:press and hold 3'SETTING for 3 sec into setting
mode àpress ALARM àpress Hi àassort 0~9 number

11
èright shift button setting “ALARM" àpress ALARM àpress
3'SETTING ,toend setting mode.
3. LOLO ALARM Setting step:press and hold 3'SETTING for 3 sec into
setting mode àpress ALARM àpress LoLo àassort 0~9
number èright shift button setting “ALARM" àpress ALARM à
press 3'SETTING ,to end settingmode.
4. LO ALARM Setting step:press and hold 3'SETTING for 3 sec into
setting mode àpress ALARM àpress Lo àassort 0~9
number èright shift button setting “ALARM" àpress ALARM à
press 3'SETTING ,to end settingmode.
3-7 Real Time Clock , Time/ Date Setting Step:
1. Real time clock , date setting step:press and hold 3'SETTING for 3
sec into setting mode àpress CLOCK àpress DATE à
assort 0~9number èright shift button setting date àpress
CLOCK àpress 3'SETTING ,to end setting mode.
2. Real time clock, time setting step:press and hold 3'SETTING for 3
sec into setting mode àpress CLOCK àpress TIME à
assort 0~9number èright shift button setting time àpress
CLOCK àpress 3'SETTING ,to end setting mode.
3-8 Record/Print Start Time, Stop Time, Interval Time Setting Step:
1.
Start time setting step
:press and hold 3'SETTING for 3 sec into
setting mode
à
press
CLOCK
à
press
START
à
assort
0~9
number
è
right shift button setting start time
à
press
CLOCK
à
press
3'SETTING to end setting mode.
2.Stop
time setting step
:press and hold 3'SETTING for 3 sec into
setting mode
à
press
CLOCK
à
press
STOP
à
assort
0~9number
è
right shift button
setting stop time
à
press
CLOCK
à
press
3'SETTING to end setting mode.
3. Interval time setting step:press and hold 3'SETTING for 3 sec into
setting mode àpress CLOCK àpress INTV àassort 0~9
number èright shift button setting interval time àpress CLOCK à

12
press 3'SETTING to end setting mode.
3-9 Print Mode Setting Step:
Printer work mode setting sequence:press and hold 3'SETTING for 3
sec into setting mode àpress ALARM àpress G/T change
G/T MODE àpress 3'SETTING ,to endsetting mode.
3-10 Clear Record Step:
Press RECORD
=+Power ON in the same time
The LCD should display “dEL”
3-11 Disable Start、Stop、Intv and Setting Step:
1.Disable START、STOP、INTV sequence:press and hold
3'SETTING for 3 sec into setting mode àpress START disable
start, stop, intv àpress 3'SETTING to end setting mode.
2.Disable ALARM sequence:press and hold 3'SETTING for 3 sec into
setting mode àpress STOP disable alarm àpress 3'SETTING
to end setting mode.
3-12 Decimal Point Setting Step:
Shift point setting sequence:press and hold 3'SETTING for 3 sec to
enter setting mode àpress INTV to change decimal point
position àpress 3'SETTING to end setting mode.
3-13 Consecutive Record and Print Setting Step:
Setting sequence:press and hold 3'SETTING for 3 sec to entersetting
mode àpress TIME setting / =àpress 3'SETTING ,to
end setting mode.
※If already in setting MODE, press TIME button only.
3-14 Reset System Setting Step:
Press 3'SETTING + Power ON

13
3-15 Record / Print Start and Step Operation Method:
(1). Start way :
1. Trigger Record / Print : in general mode press RECORD
=or PRINT
onesecond.
2. Consecutive Record/Print :
Automatic mode :
a). Into setting mode, setting Start/Stop/Into/Time.
b). Chose automate consecutive Record/Print (by 3-13 setting
way)
c). Back to general mode, on LCD =or should be
repeat.
Hand setting mode :
a). Into setting mode, disable Start, Stop,Intv. (by 3-11 stop
way)
b). Into setting mode, setting Intv.
c). Back to generalmode.
d). Press RECORD
=PRINT button 3second, the LCD will be show
=or symbol, into consecutive Record / Print of hand
setting mode.
(2). End Record/Print :
1. Automatic mode : The function will be stopped automatically while
the STOP time is up.
2. Hand settingmode :To stop consecutive Record/Print by user press
and hold RECORD
=or PRINT
button for 3 seconds. The LCD
RECORD
=or
PRINT
symbolwillbe disappear.
3-16 Memory Full Warning :
The =symbol will be flash.
3-17 Alarm Print in Graphic Mode :
When printing inG mode, if you havesettingany alarmpoints
(HiHi, Hi, LoLo,Lo) when the read valueover the set points,the
printer the will print alarm value on the paper.

14
3-18 Scale Change in Graphic Mode
HiHi
ALARM Hi
ALARM Lo
ALARM LoLo
ALARM Scale change by alarm setting
üNo change
üNo change
üNo change
üNo change
ü ü Change scale by HiHi Hi ALARM
ü ü Change scale by HiHi Lo ALARM
ü ü Change scale by HiHi LoLo ALARM
ü ü Change scale by Hi Lo ALARM
ü ü Change scale by Hi LoLo ALARM
ü ü Change scale by Lo LoLo ALARM
ü ü ü Change scale by HiHi Lo ALARM
ü ü ü Change scale by HiHi LoLo ALARM
ü ü ü Change scale by HiHi LoLo ALARM
ü ü ü Change scale byHi LoLo ALARM
ü ü ü ü Change scale by HiHi Lo ALARM
No setting ALARM, no change
ü: had setting
3-19 How to ConnectsTL-(A37) Analog Output Cable

15
CONSECUTIVE
,
mA
,
3
PRINT
1
CLOCK
COM
MAX
FUSED
20mA 24V
600V
MAX
60V
RECORD
ALARM
TRIGGER
V
1
SETTING
Lux
3
SETUP
0
,
HiHi
DATE
¢X
200mV
¢X
CFLo
STOP
RH
2V
RANGE
LINE
FEED
3
2465 87
LoLo
START
200mV
TIME
Hi Fc
2V dB ¢X
¢X
UNIT
MAX
MIN
9
20mA
INTV
G
ATA
Model Connecting way (DC Input) Connecting way (AC Input) Remark
TES-1300 Red lead connect to (V),
Black lead connect to (COM) No (AC input)
TES-1302 Red lead connect to (V),
Black lead connect to (COM) No (AC input)
TES-1303 Red lead connect to (V),
Black lead connect to (COM) No (AC input)
TES-1332 Red lead connect to (V),
Black lead connect to (COM) No (AC input)
TES-1334 Red lead connect to (V),
Black lead connect to (COM) No (AC input)
TES-1350 Red lead connect to (V),
Black lead connect to (COM) White lead connect to (V),
Black lead connect to (COM) AC,DC output
terminal same
TES-1351 Red lead connect to (V),
Black lead connect to (COM) White lead connect to (V),
Black lead connect to (COM) AC,DC output
terminal same
TES-1352 White lead connect to (V),
Black lead connect to (COM) White lead connect to (V),
Black lead connect to (COM) AC,DC output
terminal difference
TES-1353 White lead connect to (V),
Black lead connect to (COM) White lead connect to (V),
Black lead connect to (COM) AC,DC output
terminal difference
TES-1360 Red lead connect to (V),
Black lead connect to (COM) White lead connect to (V),
Black lead connect to (COM)
3-20 Setup Process Transducer (4-20mA) Mode :