Orbit Controls OC 505 User manual

OC505_GBM_21507
CALIBRATOR
MULTIMETER
OC 505
Owner’s Manual
ORBIT CONTROLS AG
Zürcherstrasse 137
CH-8952 Schlieren/ZH
Tel: + 41 44 730 2753
Fax: + 41 44 730 2783
e-mail: info@orbitcontrols.ch
www.orbitcontrols.ch

2

3
Stellen Sie sicher, dass Ihre Sendung das richtige Gerät, Orbit Controls Modell OC 505, beinhaltet,
einschliesslich einer Betriebsanleitung OC 505.
Vor dem Einschalten des Gerätes überprüfen Sie die Anschlüsse und die Versorgungsspannung. Ein
falsch angeschlossenes Gerät kann beschädigt werden und damit auch die mitverbundene
Folgeelektronik. Für falsche Handhabung wird jede Haftung abgelehnt.
Remove the Packing List and verify that you have received all equipment, including the following:
Orbit Controls Model OC 505 Handheld Calibrator.
Operator's Manual OC 505.
If you have any questions about the shipment, please call the Orbit Controls Customer Service
Department.
Shipment
- Calibrator-Multimeter Model OC505
- Charger 12V DC, 600mA
- Set of Cables 30cm with 4mm Banana and Crocodile.
- „K“ and „U“ Thermocouple Connector
- Owner’s Manual and Calibration Certificate
Vor dem Einschalten
ZU BEACHTEN
Dieses Gerät wurde sorgfältig verpackt. Falls es bei Ihnen in beschädigtem Zustand eintrifft,
benachrichtigen Sie unverzüglich den Orbit Controls Kundendienst (Tel: +41 44 730 2753 oder
Fax: +41 44 730 2783) und nehmen Sie einen Schadenrapport auf, welchen Sie auch von der
Transportgesellschaft unterschreiben lassen. Bewahren Sie bitte das Verpackungsmaterial für
eventuelle Reklamationen auf.
NOTE
When you receive the shipment, inspect the container and equipment for signs of damage. Note
any evidence of rough handling in transit. Immediately report any damage to the Orbit Controls
customer service, Phone +41 44 730 2753 or Fax +41 44 730 2783 and to the shipping agent.
The carrier will not honour damage claims unless all shipping material is saved for inspection. After
examining and removing contents, save packing material and carton in event the reshipment is
necessary.
Unpacking Instructions

4
INHALT
CALIBRATOR –MULTIMETER OC505 Page 5
1 OUTPUTS, INPUTS and KEYBOARD 6
2 GRAFIC DISPLAY 7
3 SPECIFICATIONS 7
3.1 CALIBRATOR 7
3.2 MULTIMETER 9
3.3 DATALOGGER 9
4 SELECTION-CALIBRATOR 10
4.1 DIRECT VALUE ENTRY 11
4.2 RAMPS 11
4.2.1 Ramp selection in Menu 11
4.3 RAMP STEPS 11
4.3.1 Selection of steps in Menu 11
4.3.2 Selection of individual steps 11
5 GRAFICS 12
5.1 GRAPHICS 12
5.2 TRANSIENTS 12
6 MENU STEPS 13
7 ADDITIONAL FUNCTIONS in the MULTIMETER MODE 14
7.1 ADDITIONAL FUNCTIONS 14
8 SOFTWARE CALIBRATION 14
8.1 CALIBRATION - CALIBRATOR 15
8.1.1 Calibration of Currents 15
8.1.2 Calibration of Voltages 15
8.1.3 Calibration of Resistors 16
8.2 CALIBRATION –MULTIMETER 16
9 HARDWARE 17
10 DATALOGGER 18
11 SOFTMANAGER OC505W 19
CALIBRATION CERTIFICATE 22
WARRANTY 23

5
CALIBRATOR - MULTIMETER OC 505
Current Calibrator 0/4 - 22mA, Source/Sink
Voltage Calibrator 0-25V
mV Outputs 0-27mV and 0-540 mV
DIN Thermocouples J, K, N, R, S, T, B, E
RTD Simulator Pt and Ni
Resistance Simulation to 3kOhm
Multimeter ±2V to ±200V DC and 0-100mA
Calibrates and Measures Simultaneously
Eight Memory slots for fast Transients
Graphics of Measurements
Steps, Ramps, direct value settings
Datalogger Function
OC505 is a Calibrator-Multimeter for generation of Currents 0-22mA Sink
or Source und Voltages to 25VDC. External voltages ±2V, ±20V and
±200VDC (firm ranges or auto ranging) and Currents to 0-100mA can be
measured simultaneously and shown at the display.
Further Functions contain generation of calibration mV Signals, Thermo voltages of DIN Thermocouples
and RTD Resistors and simulation of true resistors.
mV voltages of 0-27mV or 0-540mV are mainly used for calibration of Strain Gauges, Transmitters and
small signal inputs with resolution of up to 0.001mV.
Thermocouples J, K, N, R, S, T, B, E are simulated. The temperature is entered with the keyboard and
shown at the display. The cold junction is compensated to the ambient temperature. It can also be
switched-off.
RTD Thermometers can be simulated within the DIN temperature range. The temperature is entered with
the keyboard and shown at the display.
Ohm Source Resistance values up to 3kOhm are simulated. The resistance value is entered with the
keyboard and shown at the display.
Graphics and Memorising of measured signals is a standard function. The signals are continuously
stored and shown at the display as graphics. In apart of this eight memory slots for storing of fast signals
-Transients- are available. They can individually be stored and selectively recalled at the display.
Datalogger is an Option. The calibrating signals and the multimeter input are shown at the display and
stored as tables with date and time from internal RTC. They can be downloaded to the PC and edited
under Windows and Excel. A software program is available for Windows.

6
1 INPUTS, OUTPUTS and KEYBOARD
CALIBRATOR
Voltage and Current Outputs,
Resistance and RTD: + OUTPUT -
Thermocouples: Cu Terminals at the front
MULTIMETER
Voltage and Current Inputs: + INPUT -
CONTROLLER
The Functions of the Calibrator and the Multimeter are selected with the keyboard and supported by a
microcontroller. The factory settings and the calibrated points are stored in a nopn-volatile memory and
remain stored also when the instrument is switched off. The software calibration of all ranges is
accessible in the calibration menu and protected with a password. All ranges can be recalibrated via the
keyboard if required.
The graphic display is divided into two windows. The upper window is shows the multimeter functions,
the lower window the calibrator. In programming mode the display shows the parameters. In graphics
mode of operation the display shows the stored input signals.
1 Thermocouples Connector
2 Output Calibrator
3 Input Multimeter
4 LCD-Display
5 Main Keys
6 Cursor Keys
7 Memory Key
8 Recall Graphics
9 Datalogger
10 Selection of Calibrator Output Mode
11 Function Keys
12 Charger Jack
13 USB

7
2 GRAPHICS DISPLAY
The LCD Display is divided into two parts. The upper part is for the multimeter functions, the lower part
for the calibrator. The bargraph in the middle is the analogical representation of the measured input
signals. The left side shows various symbols:
Symbols in the upper display part
AUTO Automatic Voltage Range or firm measuring range
200V Measure Ranges 2V, 20V, 200V, 100mA
INP The display shows the input signal.
Symbols in the lower display part
RMP Automatic Ramp of the calibrator signal
SUP The battery is being charged
60% Battery voltage in %. Do not run the instrument when the battery shows 0%. Charge only
with the enclosed original charger.
3 SPECIFICATIONS
3.1 CALIBRATOR Voltage and Current: + OUTPUT –
Thermocouples: Cu Connector
Conversion: 16 Bit
Current Source Range 0 ... 22mA. Maximum external Shunt 750 Ohm
Accuracy ± (0.05% from Value + 0.1% from Range)
Resolution 0.001 mA
Current Sink Range 0 ... 22mA at 24V maximum
Accuracy ± (0.05% from Value + 0.1% from Range)
Resolution 0.001 mA
Voltage Source Range 0 ... 25 V, 0 … 560mV, 0 … 28mV
Accuracy ± (0.05% from Value + 0.1% from Range)
Resolution: 0…25.000V, 0…560.00mV, 0…28.000mV
Load: 0 - 25V maximum 1mA
0 - 560mV, 0 - 28mV:maximum load 1kOhm
Thermocouples According to ITS 90:
J (1200 ºC), K (1370 ºC), N (1300 ºC), R (1760 ºC),
S (1760 ºC), T (400 ºC), B (1820 ºC), E (1000 ºC).
Accuracy ± 0.3 … 2.5 ºC
Resolution 0.1ºC
Load: maximum load 1kOhm
Could Junction compensation with SMT160 can be menu calibrated.
MULTIMETER measured input voltage in
automatic range.
CALIBRATOR generates 12.000mA. The Battery
has 70% of capacity.

8
Working Ranges and specified Accuracies for Thermocouples
The accuracies shown are maximum values. Typical accuracies are lower.
RTD Thermometer According to ITS 90:
Pt-100, 200, 500, 1000 -140 ºC to 850 ºC, max. Load 0.1W.
Ni - 1000 -60 ºC to 170 ºC
Accuracy ± 0.1% from Value ± 0.5 ºC
Resolution 0.5 ºC
Resistors 45 Ohms to 3000 Ohm, maximum load 0.1W
Accuracy ± 0.1% from Value ± 0.5 Ohm
Resolution 0.1 Ohm
Tempco 25ppm/K
Temperature Working: -10 ºC ... +35 ºC, Storing: 0 ... 60 ºC
Reference Temperature: 23 ºC ± 5 ºC
Terminals 4mm plugs, gold plated
Compensated Cu Terminals for thermocouples
Supply Two rechargeable Li-Ion Batteries 3.7V-2000mAh.
Battery operation approx. 8 h with backlight switched-off
Charger Mains Voltage100-240V, 48-60Hz / 12V-600mA DC.
Charging time 4 hours. The batteries are fully charged when the green LED
illuminates continuously.
Cabinet Dimensions 200 x 90 x 40mm (L x W x D), weight 320 g
R
Range [°C]
-50 - 0
0 - 400
400 - 1760
Accuracy [°C]
1.6
1.6
1.6
S
Range [°C]
-50 - 0
0 - 500
500 - 1760
Accuracy [°C]
2.4
1.9
1.5
B
Range [°C]
50 - 800
800 - 100
1000 - 1820
Accuracy [°C]
2.5
1.5
1.3
J
Range [°C]
-140 -0
0 - 700
700 - 1200
Accuracy [°C]
1.1
0.6
0.6
T
Range [°C]
-270 --100
-100 - 0
0 - 400
Accuracy [°C]
1.7
0.7
0.3
E
Range [°C]
-120 - 0
0 - 370
370 - 1000
Accuracy [°C]
1.1
0.3
0.7
K
Range [°C]
-230 - -100
-100 - 0
0 - 1370
Accuracy [°C]
1.8
0.8
0.9
N
Range [°C]
-270 - -100
-100 - 0
0 - 1300
Accuracy [°C]
2.1
1.4
1.2
If the resistance is measured with external Ohm-meter, the polarity of the Ohm-meter
input and the OC505 outputs has to be identical.
All specifications are valid after a warm-up time of 10 minutes at an ambient temperature of 23 ºC ± 5 ºC.

9
3.2 MULTIMETER Terminals + OUTPUT -
Voltage Input Ranges ± 2V (1.83 M)
± 20V (593 k)
± 200V(563 k)
Auto range 0 …. ± 200V DC
Accuracy ± 0.1% from Range ± 1 Digit
Current Input Range 0 - 100mA DC (10 )
Accuracy ± 0.1% from Range ± 1 Digit
Sampling Two samples per second
Tare The Tare can be activated in the Menu and relates to the input signal.
When activated, the symbol TARE is displayed.
Filter An averaging filter with steps 0 to 9 can be selected.
Tempco 25ppm/K
Terminals 4mm plugs, gold plated
All specifications are valid after a warm-up time of 10 minutes at an ambient temperature of 23 ºC ± 5 ºC.
3.3 DATALOGGER (Software Option)
The generated signal and the measured signal are stored in a table format. The date and the time are
added. The stored data can be downloaded via the USB to a PC. A Soft Manager supports the data
transfer to a Windows PC.
Two Set Points LEVEL LOW and LEVEL HIGH and two Intervals INTERVAL 1 and INTERVAL 2 will be
selected. If the measured signal is between the both Set Points LEVEL LOW and LEVEL HIGH, the data
will be stored with INTERVAL 1. The INTERVAL 2 stores the date when the input signal is outside the
Set Points. The both Intervals can be selected from 2 sec. to 24 hours.
With the key LOGG is the function Datalogger active. The key MENU and the two cursor keys permit the
parameter selection. With OK the setting will be stored.
The Multimeter is always active and can be used independently from the calibrator. The ranges can be selected
with the keyboard. The current range is firm. The voltage ranges are firm 2V, 20V or 200VDC or can be set as
automatic range.
Minus Terminals of the Calibrator and the Multimeter are common.
LOGGER OFF Datalogger OFF
LOGGER ON Data are stored
MEMORY ERASE Erase the memory
INTERVAL 1: < 2 Sek. … 24 Std.
INTERVAL 2: < 2 Sek. … 24 Std.
LEVEL LOW: free settable
LEVEL HIGH: free settable

10
4 FUNCTION SELECTION - CALIBRATOR
The calibrator function is selectable with the keys. The output values can be set in steps, ramps or free
values.
The output function will be selected with OUT. Vertical cursor keys select the required function:
Current Source 0 - 22mA
Current Sink 0 - 22mA
Voltage Output 0 - 25V
Voltage Output 0 - 560mV
Voltage Output 0 - 28mV
Resistance Output 45 - 3000 Ohm
RTD Outputs Pt-100, Pt-200, Pt-500,
Pt-1000, Ni 1000
Simulation of Thermocouples
B, J, K, N, R, S, T
OK confirms the selection. The selected function remains memorised also when the power is switched-off.
If the resistance is measured with external Ohm-meter, the polarity of the Ohm-meter
input and the OC505 outputs has to be identical.

11
4.1 DIRECT VALUE ENTRY
When the key NUM is pressed, the display shows the previous value with flashing digit. This digit can be
set from the keyboard. The flashing digit will be positioned automatically. It can also be positioned with
the horizontal cursors.
4.2 RAMPS
The key RAMP activates the Ramp generation. The display shows RMP and increments or decrements
in steps selected in the Menu. The Ramp can be stopped or run with the key RAMP. By keeping pressed
the key RAMP for several seconds, the Ramp will start from beginning.
Parameter RAMP PAUSE permits the time selection between the ramp steps from <1 sec to 1 minute.
4.2.1 Ramp Steps selection in MENU
Press MENU. The step can be selected from firmly pre-programmed values with the vertical cursors.
Confirm with OK.
4.3 RAMP STEPS
With keys RAMP or STEP the display will increment in steps selected in the MENU.
4.3.1 Selection of Steps in the MENU
Press the key MENU. Use the vertical cursors to select the Step from pre-programmed values. Confirm
with OK.
4.3.2 Selection of individual Steps
The range and the steps can be set in two memory slots FN.1 and FN.2.
Press MENU and select RAMP SELECT. With the key MENU and vertical cursors select the range and
the steps.
Confirm with OK.
Example: Current Output 0 - 22mA
DN STEP: Display decrementing
UP STEP: Display incrementing
Individual Ramp Step scan be selected und stored
in FN.1 and FN.2.
Example: Current Output 0 - 22mA
DN STEP: Display decrementing
UP STEP: Display incrementing

12
5 GRAPHICS
For fast diagnostics of external signals measured by the multimeter the graphics representation can be
displayed. Two modes of operation are available:
- GRAPHICS the running measurement is automatically stored in 128 points (FI-FO)
- TRANSIENTS Eight individual memory slots are assigned for fast signals –Transients.
The sampling time and the Trigger level are selectable.
The Memory module for the Graphics and the Transients are common. By recalling the Transients, the
Graphics will be erased.
5.1 GRAPHICS The key GRAPH opens the graphic presentation.
Graphics deactivated - GRAPH OFF. The graphics can be activated with vertical cursors - GRAPH ON.
The latest stored signal will be displayed when OK is pressed.
5.2 TRANSIENTS
Fast signals at the multimeter input can be stored. Eight individual memory slots TRANSIENT NO.1 …
TRANSIENT NO.8 are available. The sampling rate is 1ms. Each transient contains 256 Point and can
be time limited from 0.25 to 300 seconds. The Trigger level is programmable from <10% to >90%.
With the key MENU the Transients can be stored or recalled at the display. The selection is done with
the vertical cursors.
The parameters will be selected with the key MENU:
Stored Transient 1 is shown at the display
Transient 1 deleted
The key MENU and vertical cursors select
the time
The key MENU and vertical cursors select
the Trigger level
The graphics at the display
Start of Transient 1
memorizing
0.25 … 300 sec.
<10% … >90%

13
6 MENU STEPS
When the key MENU is used in the measuring mode, the display will change into program mode.
Ramp and step selection
Step selection
Individual step, minimum value, User 1
Individual step, maximum value, User 1
Individual step, differential value, User 1
Individual step, minimum value, User 2
Individual step, maximum value, User 2
Individual step, differential value, User 2
Junction compensation ON - OFF
Temperature units ºC or ºF
Tare of the Multimeter display ON - OFF
Filter constant of the Multimeter display
Setting of the Time and the Date
Software Upgrade OFF –ACTIVE
IMPORTANT! This parameter has always been set to
OFF. If set to ACTIVE, the ON-OFF key will be
disabled. It is important to set this menu step to OFF
and shortly interrupt the battery supply by means of a
reset button through the aperture at the side.
With ESC one return step in Menu is possible. By pressing ESC three times, the display returns
into the measuring mode.
Ramp break

14
7 ADDITIONAL FUNCTIONS in the MULTIMETER MODE
The key MENU opens the Functions TARA and FILTER for the Multimeter:
7.1 ADDITIONAL FUNCTIONS
The numeric keyboard permits selection of following functions:
HOLD Display Hold (with CLEAR return)
MAX Maximum display reading of the Multimeter (with CLEAR return)
MIN Minimum display reading of the Multimeter (with CLEAR return)
CLEAR Cancel the above functions and switches into the measuring mode
8 SOFTWARE - CALIBRATION
The voltage and the current ranges of the multimeter as well as the output signals of the calibrator can
be calibrated from the keyboard. The calibration menu is password protected. The calibration menu
opens with the key OK pressed for 10 seconds.
Full calibration of all ranges or partial calibration of some selected ranges can be performed. A precision
5 digits multimeter with accuracy of 0.005% is required.
Point 1: Calibrator Outputs
Points 2-3: Multimeter Inputs
Point 4: Calibration of the Resistor Simulator
Point 5: Hardware Configuration
Point 6: Selection of Logo for customized instrument
Quit the Menu with ESC.
To enter the Calibration Menu use the factory set Password “8952”. The Password can be
changed in step “6”. NOTE THE NEW PASSWORD! If lost, consult the manufacturer.
The TARE can be activated
The Filter can be set for the input signal.

15
8.1 CALIBRATION of CALIBRATOR OUTPUTS
Select the required range (1) or (4).
Currents, Voltage, Resistors and Thermocouples will be calibrated in the Range (1).
In the Range (2) the Resistors and the RTD will be calibrated.
8.1.1 Calibration of Currents
Connect a mA-Meter to the calibrator outputs. Required accuracy 0.01% @ 25mADC
8.1.2 Calibration of Voltages
Connect V-Meter to the calibrator outputs. Required accuracy 0.01% @ 25VDC
Set the value with the keyboard until the connected mA-Meter
shows 4.000 mA. SOURCE MODUS.
Set the value with the keyboard until the connected mA-Meter
shows 20.000 mA. SOURCE MODUS.
Set the value with the keyboard until the connected mA-Meter
shows 4.000 mA. SINK MODUS.
Set the value with the keyboard until the connected mA-Meter
shows 20.000 mA. SINK MODUS.
Set the value with the keyboard until the connected V-Meter
shows 0.000 V.
Set the value with the keyboard until the connected V-Meter
shows 24.000 V.
Set the value with the keyboard until the connected V-Meter
shows 0.000 mV.
Set the value with the keyboard until the connected V-Meter
shows 500.000 mV.
Set the value with the keyboard until the connected V-Meter
shows 0.000 mV.
Set the value with the keyboard until the connected V-Meter
shows 25.000 mV.

16
8.1.3 Calibration of Resistances
Connect Ohm-Meter to the calibrator outputs. The required accuracy is 0.05%.
8.2 CALIBRATION MULTIMETER
Select the required range (2) or (3).
The Range (2) is for calibration of the current, the Range (3) is for the voltage calibration.
The key ESC terminates the calibration and the display returns into the measuring mode.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Set with the keyboard the resistance value which is being
measured with the connected Ohm-Meter.
Supply from external current calibrator 0.000 mA and confirm with
OK. The point 0.00A is calibrated.
Supply from external current calibrator 100.000 mA and confirm
with OK. The point 0.1A is calibrated.
Supply from external voltage calibrator 0.000 V and confirm with
OK. The point 0V is calibrated.
Supply from external voltage calibrator 2.000 V and confirm with
OK. The point 2V is calibrated
Supply from external voltage calibrator 20.000 V and confirm with
OK. The point 20V is calibrated.
Supply from external voltage calibrator 0.000 V and confirm with
OK. The point 0V is calibrated.
Supply from external voltage calibrator 0.000 V and confirm with
OK. The point 0V is calibrated.
.
Supply from external voltage calibrator 200.000 V and confirm with
OK. The point 200V is calibrated.

17
9HARDWARE
In this Manu step the 100mA Output Calibration Current can be activated. The 100mA Output Current
is an Option and has to be ordered from Orbit Controls AG, Switzerland.
Further menu steps contain the correcting factor of the battery charge, temperature correction of the
junction for Thermocouples, storing of new calibration values and recall of the factory settings. The last
position will be used when during the attempt of calibration errors have been done or the calibration
incorrect terminated.
The menu steps contain
- Correction of the junction temperature
- Battery charge display correction
- Restore of factory calibration values
- Backup of new calibration values.
Standard instruments are 0 … 20mA. The output option 100.00mA
has to be ordered at the factory.
The temperature of the output terminals will be measured with
external thermometer. The correction will be done in Set to.
Terminate and store with OK.
Check the voltage with external V-meter when the battery is fully
charged and enter the value. Terminate and store with OK.
Restore: Recall of factory settings.
Backup: Store new calibration values.

18
10 DATALOGGER
When the function Datalogger is activated, the calibrator output and the multimeter input signals will be
stored in the internal memory. This permits fast checking of transmitters and their calibration in the field.
Example: A T/C transmitter with mV input signal generates 0-10V at the output. Both signals are stored
as a table with date and time added from internal RTC.
The Data can be downloaded to the PC and displayed as tables and graphics. They can be used for
Excel and handled under Windows.
Set Points LEVEL LOW and LEVEL HIGH and Intervals INTERVAL 1 and INTERVAL 2 will be
selected. If the measured signal is within the set points LEVEL LOW and LEVEL HIGH, the data will be
stored with INTERVAL 1. Outside the set points the Data are stored with INTERVAL 2. The Intervals are
programmable from 2 seconds to 24 hours.
DATA STORE
After LEVEL LOW and LEVEL HIGH and INTERVAL 1 and INTERVAL 2 are set, the data logging can be
initiated with LOGGER ON and OK.
In the menu step LOGGER OFF is the logging terminated.
Beginning of the storing is announced with a beep tone.
During the storing period the display shows STO changing with %.
The key MENU and the vertical cursors permit the selection of the parameters. They will be stored with
the key OK.
RECAL of stored DATA
The stored data can be uploaded to the PC via USB terminal. Enclosed Softmanager OC505W supports
the communication.
Apart of the stored data, also the Transients cane be uploaded to the PC (see 5.2) and displayed as
tables and graphics and handled under Windows and Excel.
LOGGER OFF Data logging disabled
LOGGER ON Data logging enabled
MEMORY ERASE Stored data will be erased
INTERVAL 1: < 2 sec … 24 h
INTERVAL 2: < 2 sec … 24 h
LEVEL LOW: free selectable
LEVEL HIGH: free selectable

19
11 SOFTMANAGER OC505W
The Softmanager OC505W permits an upload of the stored data and the transients to a PC, displaying
the results in tables and graphics and handling under Windows and Excel.
OC505 Calibrator will be connected via USB to a PC and switched-on.
The Softmanager OC505W will be installed at the PC and opened. The window shows the selection:
Select COM in the Menu
Window
COM The data port will be automatically found with Find COM.
The Lamp changes the colour to green when the communication is running.
Assure the speed of the PC COM to run at high Baud Rate.
Default Pass:Default Password is 8952 and will be transferred to OC505.
An individual password previously set will be overwritten.
PC speed: Recommended is the position 2 from left.

20
Logger Table The stored data in OC505 are transferred to the PC with Upload. They appear in
selected units in a table format with date and time added from internal RTC.
When the calibrator signal sent to the transmitter under test is e.g. 0-22mA and
the transmitter output is a current, the transmitter signal has to be measured at the
multimeter input 100mA.
Stop Communication stops.
Save The data are stored in a Text Format *.txt or Excel Format *.xls.
Copy Copy the data.
Transient Data Table
Number Eight memory slots for fast signals at the multimeter input. Each Transient contains
256 samples, selectable Sampling Time and selectable Trigger Level (see 5.2).
Upload The selected Transiente will be transferred to the PC with Upload.
Save The data are stored in a Text Format *.txt or Excel Format *.xls.
Copy Copy the data.
Table of contents
Other Orbit Controls Multimeter manuals