OxyCheq Expedition-X User manual

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
OxyCheq Expedition-X Oxygen Analyzer
Operator’s Manual
OxyCheq
3528 Russell Road
Marianna, Florida 32446
USA
Tel: 850.482.0385
Email: oxy@oxycheq.com
Web: http://oxycheq.com
CONTENTS
1.0 Introduction ………………………………………….. 2
2.0 Controls ……………………………………………… 2
3.0 Air Calibration …………………….…………………. 3
4.0 Oxygen Calibration …………………………………. 3
5.0 Operation ………………...………………………….. 4
6.0 Troubleshooting ……………………………..……… 4
7.0 Maintenance …………………………..……………..4
8.0 Care of Your Expedition-X ………… ………..…… 5
9.0 Safety ………………...……………………………… 5
10.0 Specifications ……………………………………….. 6
11.0 Spare Parts …………………………………………. 6

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
1.0 Introduction
1.1 The Expedition-X Oxygen Analyzer is designed to measure oxygen levels in the 0.1-100% oxygen
range. It can be used to measure the oxygen content for mixes used in Nitrox, Trimix and Heliox.
1.2 The analyzer can be used for cylinder oxygen level verification or for monitoring a gas mixing panel.
If the analyzer is used with a gas mixing panel, an alternate analyzer should be used to verify the
cylinder.
1.3 The Expedition-X has a digital display, an internally mounted temperature compensated sensor with a
3 year life span. The Expedition-X is powered by three AAA 1.5 volt batteries providing an expected
life span of up to three years of operation before replacement is required.
1.4 The Expedition-X has an environmentally sealed case, potentiometer, on/off/on with back light switch,
sensor, and digital panel meter. Water resistance is achieved by means of a combination of o-rings
or gaskets. The cable must be plugged in to ensure water resistance.
1.5 Expedition-X is shipped ready to use after you connect the cable to the analyzer and sensor to the
cable.
2.0 Controls
2.1 The Expedition-X has a splashproof on/off/on with back light switch. When the switch is centered, it
is in the off position. When not in use, the switch should be in the off position to conserve battery
power. When flipped up, the switch is in the on position. When flipped down, the switch is in the on
position with the panel meter back light activated. The back light option should only be used in limited
light conditions to extend the life of the batteries. Excessive use of the backlight will result in a
significantly shorter battery life.
When switched on, the analyzer’s display will indicate an oxygen reading. Please note that the
analyzer must be calibrated prior to use (see section 3.0).
2.2 The low battery warning is a battery symbol in the left corner of the display. When present, change
the batteries prior to using the analyzer.
2.3 A waterproof potentiometer/calibration knob is located on the right side of the analyzer. Turn it fully
from left to right and then left again. The display reading should decrease in value when turned to the
left and increase when turned to the right. If the reading does not change, refer to section 8.0 on
Maintenance.
2.4 The display has a 0.1% resolution and a sealed gasket. It also has a low battery indicator and a
backlight (controlled via the on/off switch).

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
3.0 Air Calibration
3.1 Calibration is essential prior to every use. For oxygen mixtures above 50% always calibrate in 100%
oxygen before measurement for most accurate results.
3.2 Ensure that the sensor plug and the sensor cap are removed and the display reading has stabilized.
Expose the analyzer to clean air for a minute and adjust the calibration knob until the display reads
20.9%. If this is not possible refer to section 6.0 on Troubleshooting or to section 7.0 Maintenance.
3.3 The analyzer is now ready for oxygen measurement.
WARNING
The analyzer is sensitive to Oxygen partial pressure. Calibration must always be carried out at the
same pressure as the oxygen measurement.
4.0 Oxygen Calibration
4.1 Connect 100% oxygen from a certified distributor and measure gas with a 2 to 3 liter per minute flow
rate.
4.2 Allow the reading to stabilize.
Figure #3
(sensor cap/flow restrictor)
Jack
OxyCheq
20.9
on/off/on
switch
Calibration
knob
Sensor
Display
Expedition - X O2 Analyzer
Cal
Pwr
http://oxycheq.com
Figure #1
Tee
Diverter
Figure #2

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
5.0 Operation
5.1 The Expedition-X comes complete with a sensor flow cap/restrictor (figure #3) that allows you to read
directly from a cylinder valve.
a) Connect the cable to the sensor and analyzer.
b) Connect the sensor flow cap by screwing it over the sensor threads. It should be snug, but not
tight.
c) Slowly open the cylinder valve until you hear a slow, steady hiss. Place the sensor flow cap
against the valve outlet facing you.
d) Once the display reaches a consistent reading, close the cylinder valve.
e) Note that after a few seconds after the gas flow has stopped, the display reading will begin to drop
until it reaches the ambient oxygen percentage (20.9%).
f) If in doubt, repeat the procedures above.
g) An alternative method to sample gas would be to attach an optional flow restrictor to the arm on
the sensor flow cap/restrictor and connect it to your BC low pressure hose.
5.2 The Expedition-X also comes with a tee and diverter (figure #2) that allows you to sample gas from
the cylinder valve.
a) Connect the cable to the sensor and analyzer.
b) Screw the diverter onto the sensor threads.
c) Place the diverter into the bottom of the tee.
d) Slowly open the cylinder valve until you hear a slow, steady hiss. Place the tee against the valve
outlet facing you making sure you are not introducing air into the tee via Venturi effect.
e) Once the display reaches a consistent reading, close the cylinder valve.
f) Note that after a few seconds after the gas flow has stopped, the display reading will begin to drop
until it reaches the ambient oxygen percentage (20.9%).
g) If in doubt, repeat the procedures above.
h) An alternative method to sample gas would be to attach the tee to the optional DIN flow restrictor.
Doing so would eliminate any introduction of air via Venturi effect.

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
6.0 Troubleshooting
Problem Possible Solution
Cannot calibrate analyzer Replace sensor
No display once analyzer is turned on Replace batteries
Low battery indicator is displayed Replace batteries
7.0 Maintenance
7.1 Battery Replacement
a) Remove the 4 screws located on the back of the analyzer and carefully remove the back cover.
b) Slide the batteries out of their holder and replace them with new AAA batteries. Dispose of
batteries in accordance with local regulations.
c) Make sure no wire leads are outside of the enclosure, and the o-ring is in place. Carefully secure
the lid and replace the screws.
7.2 Sensor Replacement
a) Remove existing sensor from cable.
b) Properly dispose of the old sensor in accordance with local regulations.
c) Connect replacement sensor to cable.
8.0 Care of the Expedition-X
8.1 The Expedition-X is an electronics instrument and should be treated as such. When not in use, store
it in the carrying case to protect it from sunlight and physical damage.
8.2 The Expedition-X is a splahproof analyzer when the cable is connected. The enclosure was designed
to withstand normal use by divers in harsh environments. However, it should not be intentionally
dropped in water or physically abused. Treat it as any other precision electronics instrument and it
will provide you with a long life.
8.3 Use a damp cloth to clean the analyzer.
8.4 Avoid extreme temperatures long exposures to direct sunlight to prevent any damage to the LCD
screen.
8.5 The sensor in the Expedition contains caustic electrolyte. Prior to touching the sensor, always make
sure there are no electrolyte leaks. In the rare case a leak is present, avoid direct contact with skin of
clothing. If physical contact occurs, refer to Safety (9.0).

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
9.0 Safety
9.1 When the batteries or sensor require replacement, dispose of them in accordance with local
regulations.
9.2 The sensor contains a KOH Potassium Hydroxide solution that is hazardous and can result in the
following:
Eye Permanent eye damage is possible. Flush eye with water for 15 minutes and seek
medical attention.
Ingestion Can be fatal if swallowed. Drink a lot of fresh water and seek medical attention.
Skin Chemical burn is possible. Wash area of skin exposed and remove any clothing that
was contacted. Seek medical attention if needed.
9.3 Sensor Handling
Additional sensors are supplied in a sealed package. Look closely at the sensor once the package is
opened to make sure a leak has not occurred. If a leak is noticed, avoid physical contact with the
sensor. If clean up is required, use rubber gloves and safety goggles. Rinse any contaminated area
with water and properly dispose of any rags.
10.0 Technical Specifications
Range 0.1 – 100% oxygen
Display Accuracy 0.1%
Response time 90% in <5seconds
Sensor Type Galvanic
Sensor Life More than 36 months in air
24 month guarantee in air
Battery 3 AAA 1.5 volt batteries
Battery life Up to 36 months use
Operating temp. -5 to 50 C
Storage temp. -5 to 50 C
Pressure Sensitive to partial pressures of oxygen
11.0 Spares and Accessories
Your Expedition-X oxygen analyzer is supplied with an oxygen sensor, three AAA batteries, a high
pressure sensor flow cap, a straight-shielded cable, a diverter, a deluxe tee and a deluxe carrying case.
Additional accessories for the Expedition include a BC LP flow restrictor, DIN valve restrictor, 10’ coiled-

OxyCheq Expedition-X Portable Oxygen Analyzer
Operators Manual 0602
shielded cable and a sensor saver. Spares include an oxygen sensor and an electronics enclosure o-
ring. You can order all of these items by contacting your supplier or OxyCheq directly. Our website is
http://oxycheq.com.
Table of contents
Popular Analytical Instrument manuals by other brands

Philadelphia Scientific
Philadelphia Scientific iBOS Pro 6 Series installation manual

Olympus
Olympus IPLEX TX instructions

Gibertini
Gibertini densimat instruction manual

Parkside
Parkside PEK 2.7 C2 Assembly, operating and safety instructions

Leica Geosystems
Leica Geosystems DISTO D110 manual

Rosemount
Rosemount 755A instruction manual