Parker SciLog LabTec Series Manual

LabTec
®
Methadone Dispenser
Installation, Operation & Maintenance Instructions Manual

2
(This Page Intentionally Left Blank)
Parker Hannifin Corporation has a continuous policy of product development and although the Company
reserves the right to change specification, it attempts to keep customers informed of any alterations. This
publication is for general information only and
SciLog®, Tandem®and MabTec®are registered trademarks of Parker Hannifin Corporation.

3
Precautions
READ this manual BEFORE operating or servicing this equipment.
FOLLOW these instructions carefully.
SAVE this manual for future reference.
DO NOT allow untrained personnel to operate, clean, inspect, service or tamper
with this equipment.
ALWAYS DISCONNECT this equipment from the power source before cleaning
or performing maintenance.
CALL PARKER for parts, information and service.
WARNING
DISCONNECT ALL POWER TO THIS UNIT BEFORE
INSTALLING, SERVICING, CLEANING OR REMOVING THE
FUSE. FAILURE TO DO SO COULD RESULT IN BODILY
HARM AND/OR PROPERTY DAMAGE.
CAUTION
OBSERVE PRECAUTIONS FOR HANDLING
ELECTROSTATIC SENSITIVE DEVICES.
WARNING
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS
EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS,
TEST AND ADJUSTMENTS THAT MUST BE MADE WITH
POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS
CAN RESULT IN BODILY HARM.
WARNING
FOR CONTINUED PROTECTION AGAINST SHOCK HAZARD,
CONNECT TO PROPERLY GROUNDED OUTLET ONLY. DO
NOT REMOVE THE GROUND PRONG.
IF the power cord is lost or damaged, contact Customer Service to obtain a new
one. Do not replace it on your own

4
Précautions
LISEZ ce manual AVANT de faire fonctionner ou d’entretenir cet équipment.
SUIVEZ attentivement ces instructions.
CONSERVEZ ce manuel pour future référence.
NE LAISSEZ PAS du personnel non qualifié utiliser, nettoyer, inspecter,
entretenir, réparer ou manipuler cet équipement.
DÉBRANCHEZ TOUJOURS cet équipement de la source de courant avant de
nettoyer ou d’exécuter l’entretien.
APPELEZ PARKER pour pièces détachées, renseignements et entretien.
ATTENTION
DÉBRANCHEZ TOUT COURANT DE CETTE UNITÉ AVANT
DE FAIRE L’INSTALLATION, D’EFFECTUER L’ENTRETIEN, LE
NETTOYAGE OU AVANT DE RETIRER LE FUSIBLE. NE PAS
OBSERVER CES PRÉCAUTIONS RISQUERAIT DE CAUSER
DES BLESSURES CORPORELLES OU/ET D’ENDOMMAGER
L’ÉQUIPEMENT.
PRUDENCE
SOYEZ PRUDENT LORSQUE VOUS MANIPULEZ DES
APPAREILS SENSIBLES À L’ÉLECTROSTATIQUE.
ATTENTION
ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS
EQUIPMENT. EXERCISE CARE WHEN MAKING CHECKS,
TEST AND ADJUSTMENTS THAT MUST BE MADE WITH
POWER ON. FAILING TO OBSERVE THESE PRECAUTIONS
CAN RESULT IN BODILY HARM.
ATTENTION
POUR ASSURER UNE PROTECTION CONTINUE CONTRE
UNE DÉCHARGE ÉLECTRIQUE, BRANCHEZ UNIQUEMENT
SUR UNE PRISE CORRECTEMENT RELIÉE Á LA TERRE. NE
RETIREZ PAS LA FICHE DE TERRE.
SI le cordon d'alimentation est perdu ou endommagé, contactez le service
clientèle pour en obtenir un nouveau. No le remplacez pas par vous-même.

5
Table of Contents:
PRECAUTIONS.......................................................................................................................4
PRÉCAUTIONS.......................................................................................................................5
LABTEC METHADONE MAINTENANCE FORM ..................................................................9
SPECIFICATIONS.................................................................................................................11
INSTALLATION OF THE USB DRIVER................................................................................12
QUICK START: “DISPENSING METHADONE” .....................................................................14
LABTEC METHADONE DISPENSER: OVERVIEW .............................................................15
PART A: LABTEC™ METHADONE HARDWARE ................................................................16
1.0 Hardware Overview.....................................................................................................16
2.0 Pump Head/Motor:......................................................................................................16
3.0 Front Panel: Data Entry and Display .........................................................................16
4.0 Back Panel: Interface Options.....................................................................................18
5.0 LabTec Cleaning & Maintenance: RH1-CKC Pump Head..........................................20
6.0 LabTec Methadone Calibration: RH1-CKC Pump Head.............................................21
PART B: LABTEC™ METHADONE SOFTWARE.................................................................23
1.0 Software Overview: Main Menu .................................................................................23
2.0 Serial Mode.................................................................................................................25
3.0 Volume Mode..............................................................................................................27
4.1 Manual Mode ..............................................................................................................27
TROUBLESHOOTING ..........................................................................................................28

6
Standards:
EN 61326-1:2006, Class B
EN 6100-3-2:2006
EN 6100-3-3:1995 +A1:2001 +A2:2006
EN 61010-1 Issued: 2001/03/01
Conforms to UL STD 61010-1:2012 Ed.3+R:29Apr2016
Certified to: CAN/CSA-C22.2 No 61010-1-12:2012 Ed.3+U1:U2
Installation & Start-Up:
Installation of the LabTec™ System must be carried out only by trained personnel in accordance
with the relevant regulations and this operations manual.
Make sure that the technical specifications and input ratings of the LabTec™ are observed. See
“LabTec™ Specifications”.
The protection provided by this equipment may be impaired if the LabTec™ is used in a manner
inconsistent with this manual or for purposes not specified by the manufacturer.
Maintenance & Cleaning:
The LabTec™ Methadone requires routine daily maintenance. The pump head in particular requires
daily flushing with an approved cleaning solution and distilled water for trouble free operation. This
is especially true with flavored methadone solutions. Please read the section on Cleaning
Procedures.
To remove dust, dirt and stains, the outer surfaces of the LabTec™ may be wiped using a soft, non-
fluffing cloth moistened with water. If required, you may also use a mild detergent or 2-propanol.
Introduction:
You will find the LabTec™ Methadone System easy to use and easy to calibrate. The state- of-the-
art hardware and software design of the LabTec™ allows you to control measure and document
your dispensing processes. With proper maintenance, the LabTec™ System will provide many
years of excellent service and performance.

7
Please read the following instructions carefully!
Inspections: Remove the products carefully from the shipping container. Check the contents
against the purchase order to verify that all parts are included and undamaged.
Please do the inspection now, even if the products are not used immediately. Many carriers must
receive damage claims within seven days of delivery. Please retain all packing material so unit may
be shipped safely, if necessary.
Customer Service: Parker customer service: If assistance is required, please contact us at:
Parker Hannifin Manufacturing Ltd Parker Hannifin Corporation
Bioscience Division – EMEA Bioscience Division – N.A
Durham Road 2340 Eastman Avenue
Birtley, Co. Durham Oxnard, California, USA 93030
DH3 2SF, England toll free: 877 784 2234
phone +44 (0)191 4105121 phone: +1 805 604 3400
fax +44 (0)191 4105312 fax: +1 805 604 3401
email: [email protected] email: [email protected]
www.parker.com/bioprocessing www.parker.com/bioprocessing
Parker customer service personnel will be able to serve you more efficiently if you have the
following information:
Serial number and model name of the equipment
Installation procedure being used
Concise list of symptoms
List of operating procedures and conditions in use when problem arose
Warranty
Country specific information can be found at: www.parker.com/termsandconditions
LabTec Maintenance
Factory based preventative maintenance is recommended on an annual basis.
Contact your Territory Manager or Parker Technical support to obtain a TSP Number and pricing on
this procedure.

8
Specifications:
Mechanical:
Dimensions: Width: 5.75 in (14.6 cm); Height: 8.5 in (21.3 cm); Depth: 11in (27.9 cm)
Weight: 14 lbs (6.4kg)
Enclosure: Aluminum / Steel; Corrosion Resistant, Recessible Handle
Pump Head: RH1CKC Piston Head, max stroke of 0.10 ml, set at approx. 0.05 ml/stroke.
Flow Rate: 160 ml/min for US systems, 290 ml/min for overseas systems.
Motor: 3400 RPM, Optically encoded and servo controlled.
Electrical:
Power: 90 - 264 V , 47-63 Hz, 150 VA, listed Class 2 switching power supply; double
fused: 1A-T, 250V (CE: IR35A 250V ).
Battery: CR1220, used to support the internal clock only, not user serviceable.
Motor:3400 RPM at 24V , 3.8 Amperes, Variable Pump Speed optically encoded,
servo- controlled motor.
I/O Ports:
“Printer”, Female DB9 connector for data collection with Printer or PC.
“S1”, Male DB9 connector, Not utilized on LabTec Methadone. Do not remove the cover.
“S2”, Male DB9 connector, Not utilized on LabTec Methadone. Do not remove the cover.
“S3”, Male DB9 connector, Not utilized on LabTec Methadone. Do not remove the cover.
“External I/O”, Female DB37 connector, Not utilized on LabTec Methadone.
“V”, Female DB15 connector, Not utilized on LabTec Methadone.
“Temperature”, 2 pin Conxall connector, Not utilized on LabTec Methadone.
“P1, P2, P3”, RJ11 connectors, Not utilized on LabTec Methadone.
“USB”, USB-A connector, for RS-232 data collection with a PC. Driver on enclosed CD.
“Ethernet”, RJ-45 connector, Not utilized on LabTec Methadone.
Display: Two line LCD, 20 characters each, back-lit.
Data Entry: Membrane keyboard with auditory feedback.
Software:
Main Menu with the following operational modes:
Serial Mode: Normal Methadone dispensing mode, implemented upon power up.
Volume Mode: Not used for Methadone dispensing.
Weight Mode: Not used for Methadone dispensing.
Setup: Selection of user preferences and interface options.
Manual: Simple pump control, on/off and rate, no alarms.
Environmental:
Temperature range: 4 to 40C.
Altitude: up to 2000 Meters
Indoor, dry environments only, clean-up is wipe down only. (IP 20)
Relative humidity: 0-95%
Voltage fluctuations +/- 10%
Pollution degree: 2

9
Installation of the USB Driver:
Upon connecting the LabTec to the PC via a USB cable, the following “New Hardware Wizard”
window appears. Select ‘No, not at this time” and click “Next”. The second screen appears:
Insert the CD containing the LabTec Operating Manual into the PC, choose “Install the software
automatically” and click “Next”. The following screen appears:
Choose “Continue Anyway”, and the driver will finish loading, allowing you to communicate with the
LabTec via the assigned Com Port.
By opening Windows Device Manager and clicking on the + for Ports, you can determine the Com
Port assigned to the LabTec. It will be listed as “USB SciExpert”. (COM8 as shown)

10
Windows 7 and 8 users often will not have the “New Hardware Wizard” run properly, and the driver
will not be installed. When this occurs:
1. Open Device Manager and look under Other Devices for “Unknown Device” or similar and
double-click on that device.
2. Click on “Update Driver” on that screen or on the Driver Tab. Click on “Browse my computer”
on the next screen and browse to the CD drive that contains the LabTec Manual and click
on OK.
3. As mentioned earlier you will get a security warning about windows verification for the driver.
Click on Install anyway and the installation will finish properly.
Please review Device Manager at completion of this process to confirm the COM port
assigned to the LabTec, now listed under Ports as “USB SciExpert”.
Windows 10 will automatically create a “USB Serial Port” for the pump, no drivers are
needed!

11
Quick Start: “Dispensing Methadone”
Equipment: The following is needed to get started:
Parker P/N Description Quantity
200-METH-1342 LabTec Methadone Dispenser, FM-520 1
090-158 USB Cable, A/B Style 1
400-001M 316SS Sinker, small 1
400-216 #16 Tygon Tubing (one package) 50 ft
400-410 Stand and Clamps for Tubing 1
400-492 Dispensing Tip 1
Methadone container and cups
Setup:
A. Unpack all components, visually identify and inspect for damage.
B. At the Dispensing Window, place the Methadone container to the far left, the LabTec to the
right, and the Stand and Clamps furthest to the right.
C. Connect the LabTec to the PC using the USB cable (P/N 090-158) may be used. The driver
is found on the thumb drive as this manual for all except Windows 10 users. Windows 10 will
recognize this as a USB Serial Device automatically.
D. Plug in and power up the LabTec.
E. Remove the short piece of tubing installed in the pump for shipping, and retain it for use at
when sending the system in for maintenance, or for storage purposes.
F. Cut two pieces of the Tygon tubing. Make one long enough to go from the bottom of your
Methadone container to the input of the pump head, on the lower left. Insert the sinker in the
Methadone end of the tubing, and connect the other end to the head by loosening the
compression nut, inserting the tubing as far as it will go, and tightening the nut.
G. Cut the other piece long enough to go from the output of the pump head on the upper right
to the clamp at the dispensing location. Install one end in the pump head as before, and
insert a dispensing tip at the other end, and clamp it in place above the cup location.
H. The LabTec is ready for use with your Clinical Software. Use the software to Prime the
system, check for leaks, and check the Calibration.

12
LabTec Methadone Dispenser: Overview
The LabTec Methadone Dispenser provides high precision, high accuracy, and programmable
Methadone dispensing capability. The LabTec utilizes an optically encoded, servo-controlled motor,
assuring a highly reproducible pump performance. The LabTec system comes with an RH1CKC
piston pump head (Parker P/N: 080-402) which is powered by 3400 RPM motor (Parker P/N: 010-
017-2).
The LabTec has been used for many years on a nationwide and now international basis, and
provides long term accurate service in the dispensing of Methadone and the control of inventory in
the clinic when properly maintained and calibrated.
PumpSense™: Pump Overload Protection
When Methadone is allowed to dry inside the piston pump head, the LabTec software will recognize
this condition and go into a stand-by mode, the pump motor is turned off and the following message
is displayed:
Before continuing with your pumping application, immediately replace the methadone container with
your cleaning solution, press the A button twice to return to the normal screen and try dispensing
again. If this works, run the solution for at least 2 minutes before flushing and returning to normal
operations. If the situation continues, contact Parker Customer Service for assistance. The
PumpSense™feature has been implemented as a failsafe device to protect your pump head and
motor from permanent damage.
NOTE: Parker has implemented a LabTec preventative maintenance and loaner program.
Send in your LabTec unit to Parker for pump head cleaning and seal replacement. The
service also includes functional test, calibration as well as a written performance validation.
The typical turn-around time for LabTec repairs is 10 business days. You can rent a LabTec
loaner from Parker while your unit is being serviced.
Contact your Parker TM or Technical Support at 805-604-3400 To obtain the appropriate form and
TSP number prior to shipping your unit in for service.
CHECK PUMP HEAD
Press Any Key

13
Part A: LabTec™ Methadone Hardware:
1.0 Hardware Overview:
The LabTec has been designed to automate routine dispensing of Methadone from the 0.5 ml to 30
ml range, or the equivalent of 5 to 300 mg of Methadone as used in the United States. The LabTec
allows you to easily calibrate the unit at 200 mg (100 mg for the older powder blue units) and does
an excellent job of helping your clinical software maintain accurate inventories. Refer to the
calibration instructions later in this manual.
2.0 Pump Head/Motor:
The LabTec™Methadone Dispenser comes with a 3400 RPM, optically encoded, servo-
controlled motor and a RH1-CKC piston pump head. Materials of construction: the pump body is
made of Kynar, the piston and cylinders are made of ceramic. Both Kynar and ceramic are inert
materials and do not interact/react with Methadone. The head is equipped with a vernier style
adjustment that makes calibration easy.
NOTE: It is critical that the head be kept clean when not in use. This cannot be emphasized
enough to the clinical staff. Most Methadone in use is flavored, containing sugar and other
ingredients that will crystallize inside the head and cause it to freeze up, yielding the “Check Pump
Head” error mentioned above. Please refer to the cleaning procedures section of this manual.
Red ring;
Stationary
Set Screw;
Do not loosen!
Inlet Port
Outlet Port
Silver textured ring;
Used to make
adjustments during
Calibration

14
3.0 Front Panel: Data Entry and Display
The front panel consists of a user interface, which includes an alphanumeric display and a
membrane keypad to select operational modes and alarm settings. The display is a two line, 20
character each, liquid crystal display (LCD). The display is backlit to allow easy viewing over a wide
range of lighting conditions.
The lower line on the LCD is used to signify the function of the “soft keys” marked “A”, “B” and “C”.
The “soft key” current labels are displayed in the lower line of the LCD. If you press these keys,
then the function displayed above it will be performed.
The main keypad consists of eight “hard” keys whose function does not change. These keys are
used for basic control and programming of the LabTec.
The A B and C keys along with the Exit button are the only ones used in normal operation.
Pressing any key will take one out of Serial Mode, the mode used with this unit.
If the unit displaying something other than “Serial CW 160 ml/m” on the top line, press the Exit
button to reach “Mode Select Serial”, and then press C to select, and A to execute to return to the
proper screen.
Two LED’s are also on the front panel, just to the left of the main keypad. These indicate the current
pump status. A green light indicated the pump is in motion; the red light indicated that the pump has
stopped.

15
4.1 Back Panel: Interface Options
The LabTec back panel provides interfacing ports for:
Parker Printer (P/N: 080-095) or PC RS-232 connection: Female DB9, labeled “Printer”.
Electronic Scale: Male DB9, labeled “S1, S2 and S3” are not utilized in the LabTec
Methadone. Please do not remove the covers.
Foot Switch (P/N: 080-059): Male DB37, Labeled “External I/O”. Not utilized in the LabTec
Methadone unit.
SciPres Disposable Pressure Sensors: 3 RJ11 jacks, one for each pressure sensor.
Labeled “P1, P2, P3”. The LabTec Methadone does not use these sensors. (Do not
connect the LabTec to a telephone system.)
SciTemp Disposable Temperature Sensor: Conxall 2 pin connector. Labeled
“Temperature”. Also not used by the LabTec Methadone unit.

16
4.1 PRINTER PORT: The LabTec Methadone can be connected to a PC for control by the
clinical software in use. An RS-232 Serial Cable (P/N 080-073) is used for this connection.
Alternatively, a USB cable (090-158) may be used.
4.2 SCALE PORTS: Not normally used by the LabTec Methadone unit. The male DB9 ports
labeled “S1”, “S2” and “S3” are RS-232 ports for electronic scales. For the LabTec,
only S1 is used. (Please do not remove the covers on the unused ports.) This port
allows you to interface with a number of different electronic scales: i.e. Mettler, Ohaus,
and Sartorius top-loading scales. The following scale cables are required:
Mettler: PGS, PM, Viper Models: P/N: 080-067PGS
Ohaus: GT, “Precision Advanced”, “Explorer”, “Voyager” Models: P/N: 080-066
Ohaus: IP Series High Capacity: P/N: 080-067
Ohaus: Adventurer Pro Series: P/N: 080-067PGS
Sartorius: Most Series Scales: P/N: 080-068
In the LabTec Setup: Scale mode, select the scale manufacturer; the LabTec will
automatically implement the correct communications parameters. Check that the proper
communications parameters are also implemented in the scale being used.
4.3 PRESSURE SENSOR PORTS: The LabTec Methadone unit does not make use of these
jacks or Sensors.
4.4 TEMPERATURE PROBE PORT: The LabTec Methadone unit does not make use of these
jacks or Sensors.
4.5 VALVE V PORT: Not used in conjunction with the LabTec Methadone.
4.6 USB PORT: Used for connection to a PC, providing a Com Port. Can be used for data
communication as an alternative to the Printer port. The USB driver is included on the CD
that contains this manual.
4.7 ETHERNET PORT: Not utilized by the LabTec Methadone Unit.
4.8 EXTERNAL I/O CONNECTOR: Not utilized by the LabTec Methadone Unit.
Parts List: LabTec Methadone Dispenser Accessories:
400-001M 316SS Sinker: Prevents tubing from getting stuck to the inside of Methadone reservoir
400-216 Tygon Tubing: ¼ OD x 50 feet, tubing fits into RH1-CKC pump head.
400-410 Stand & Clamps: holds dispensing end of tubing.
400-492 Dispensing Tip (Fits Tygon tubing #400-216 above.)
Call Parker Sales at 805-604-3400 for a pricing on these items, or to re-order.

17
5.0 LabTec Cleaning & Maintenance: RH1-CKC Pump Head
NOTE: It is critical that the head be kept clean when not in use. This cannot be emphasized
enough to the clinical staff. Most Methadone in use is flavored, containing sugar and other
ingredients that will crystallize inside the head and cause it to freeze up if not removed well at the
end of the shift, yielding a “Check Pump Head” error and stopping the dispensing process.
Cleaning the Pump Head:
Parker strongly recommends the use of a 50:50 mixture of distilled water and isopropyl
alcohol (IPA) (final concentration of 35-40%) for cleaning purposes. Alternatively, an Alconox soap
solution may be used. (www.alconox.com) If Alconox is used, heat the solution when mixed to be
sure it is fully dissolved. It may also be heated prior to cleaning.
This solution must be pumped through the pump head for at least two (2) minutes or until the
cleaning solution appears clear at the discharge port of the pump. Stop the pump and leave the
cleaning solution in your pump head, or remove it and follow up with a rinse of distilled water. Do
not remove the tubing from the solution reservoir until you are ready to prime the pump with
Methadone the following day.
Do not leave Methadone in the pump head overnight. Preventive maintenance is very
valuable and ensures a long operational pump life.
NOTE: Never use tap water for flushing, the water “hardness” associated with most tap water
supplies will cause the pump head to “freeze”, i.e. the pump head becomes inoperable.
NOTE: If the LabTec Methadone unit is not used frequently or is to be stored for an extended
period of time, use the cleaning procedure outlined above then fill a 12” piece of tubing with
cleaning solution or distilled water, and connect the tubing between the pump inlet and outlet. Turn
the pump on and run in Manual Mode at 100% for a brief period to ensure the pump head is filled
with cleaning solution. If stored for more than 30 days, repeat this process once a month. This
simple procedure prevents the pump head from drying out and thus remains operable for a long
time.
These cleaning procedures work for all versions of the LabTec Methadone pumps.
The following Calibration Procedure on the next page
is used for the new (Dark Blue) units.
For the older (Powder Blue) units, the procedure is the
same except please calibrate only at 100 mg.
Call Parker Tech Support at 805-604-3400
if additional assistance is required.

18
6.1 LabTec Methadone Calibration: RH1-CKC Pump Head
Routine calibration checks and adjustments are commonplace and easy to perform. They keep your
system running at its best and your inventory accurate. If assistance is needed, call Parker Tech
Support at 805-604-3400.
Equipment needed:
A calibration tool to accurately measure volume or weight.
a. Good: a 25 ml graduated cylinder (glass, not plastic!);
b. Better: a 20 ml Volumetric flask (also glass);
c. Best: a balance with 0.1-gram resolution or better.
(All calibration will be done with water in this scenario.)
d. Larger graduated cylinders, syringes, or medicine cups
are not recommended due to inherent lack of accuracy.
If using a) or b) above, please read them level, and at the
bottom of the meniscus as shown:
Procedure:
A. Gather your calibration vessel. This can be a 20 ml volumetric flask,
a 25 ml graduated cylinder or a balance with a 0.1 gm resolution or better.
B. Enter the Calibration section of your clinical software and instruct the pump to dispense 200
mg. (This is equivalent to a 20 ml volume.) (Older, powder blue units, use 100 mg / 10 ml)
C. Measure the volume dispensed, and adjust the pump head accordingly:
a. Referring to the image below, if the dispense is over, turn the textured collar
clockwise (right) to reduce the volume.
b. Using the edge of the red label at the top as a reference point, turn it 2-4 of the small
marks on the Black ring at a time.
c. If the dispense is under, turn the collar counter-clockwise (left) to increase the
volume.
d. After adjustment, perform 2-3 more dispenses at 200 mg to confirm, re-adjust as
needed.
e. Once satisfied with the results, Calibration is complete, return to dispensing.
73
Red ring; stationary. Use
the edge of red label as a
reference point.
Set
Screw;
Do not loosen!
Black ring, has numbers on it,
turns with the silver ones.
Silver textured ring; Use to make
ad
j
ustments durin
g
Calibration
Outlet
Inlet Port

19
Tubing:
#16 (1/4” OD x 1/8” ID) Food Grade Tygon tubing is normally used in this system. It can be
purchased from Parker Customer Service in boxes of 50 ft. (PN 400-216)
When asked, we recommend replacing the entire tubing set once every 6 months, and when doing
so, cut both pieces a foot longer than needed. (This can be done more frequently if the clinic wishes
to.) This allows you to remove the tubing from the pump once a month and cutting of the last 2” of it
that has been trapped inside the head. The tubing does swell over time in these conditions and will
cause under-dispensing issues resulting in problems with the calibration and accuracy of the
dispensing at the end of the day.

20
Part B: LabTec™ Methadone Software
1.0 MAIN MENU
Table of contents
Popular Dispenser manuals by other brands

Edbak
Edbak PROCLEAN 1 installation manual

DEMA
DEMA Laundry Master 830 installation instructions

BOWMAN
BOWMAN CL007-0111 manual

Bunn
Bunn TCD-1 Installation & operating guide with illustrated parts

Gordon
Gordon 93140 Assembly and operating instructions

KWC
KWC EXOS EXOS618X Installation and operating instructions

AUTOMATED EQUIPMENT
AUTOMATED EQUIPMENT RAM 280-F Equipment manual

Kimberly-Clark PROFESSIONAL
Kimberly-Clark PROFESSIONAL ICON user guide

OK International
OK International TS350 user guide

Follett
Follett 15 Series Installation, operation and service manual

Kogan
Kogan KA25HWDSPNA user guide

HÖR
HÖR HÖR-DE-006A user manual