Pfeiffer Vacuum Tele TC cable 400 User manual

Translation of the original instructions
Operating Instructions
EN
PT 0310 BEN/B (1104)
Tele TC cable 400

2
Table of contents
Table of contents
1 About this manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1 Validity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Proper use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.3 Improper use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4.1 Preparatory work. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4.2 Connection to the turbopump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4.3 Connection to the electronic drive unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
5 Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
5.1 Installing an air cooling unit to the electronic drive unit . . . . . . . . . . . . . . . 6
6 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

3
About this manual
1 About this manual
1.1 Validity
This installation manual describes the functioning of the designated product and
provides the most important information for the safe installation and use of the
unit. The description follows applicable EU guidelines. Up-to-date operating in-
structions can also be downloaded from www.pfeiffer-vacuum.net.
2Safety
2.1 Safety precautions
• Do not loosen any plug connection during operations.
• Keep leads and cables well away from hot surfaces (> 70 °C).
• The unit has been accredited with protection class IP 54. When installing into
ambient conditions, which afford other protection classes, the necessary mea-
sures must be taken.
2.2 Proper use
The Tele TC cable 400 serves to the connection between the turbopump HiPace and
the respective electronic drive unit TC 400 for separate operation.
2.3 Improper use
Improper use will cause all claims for liability and warranties to be forfeited. Im-
proper use is deemed to be all use for purposes deviating from those mentioned
above, especially:
• Connection to pumps or units which are not suitable for this purpose according
to their operating instructions.
• Connection to units which have exposed voltage-carrying parts.
NOTE
CE conformity
The manufacturer's declaration becomes invalid if the operator modifies the original
product or installs additional components!
Following installation into a plant and before commissioning, the operator must
check the entire system for compliance with the valid EU directives and reassess it
accordingly.
Installation location weather protected (indoors)
Protection category IP 54
Protection class III
Temperature • +5 °C to +25 °C, max. drive power = 175 W
• +5 °C to +35 °C, max. drive power = 125 W
• +5 °C to +35 °C, max. drive power = 250 W (only with air
cooling unit for the electronic drive unit)
Relative humidity 5-85 %, non dewing
Atmospheric pressure: 75 kPa - 106 kPa
Installation altitude 2000 m max.
Degree of pollution 2
Overvoltage category II

4
Product description
3 Product description
To correctly identify the product when communicating with Pfeiffer Vacuum, al-
ways have the information from the rating plate available.
3.1 Scope of delivery
• Tele TC cable 400, length and cable departure according to the ordering
• Allen head screws (2x)
• Operating instructions
4 Installation
4.1 Preparatory work
Turn off the vacuum pump, vent to atmospheric pressure and allow to cool, if
necessary.
Remove the vacuum pump from the system, if necessary.
Dismount the electronic drive unit of the turbopump according to the operating
instructions of the pump.
Only separate the pump and the electronic drive unit from each other after dis-
connecting the supply voltage and the complete standstill of the pump.
Do not exercise any mechanical load on the electronic drive unit.
Preserve the fixing screws of the electronic drive unit.
Fig. 1: Overview of Tele TC cable 400, connectivity
A-C Configuration cable departure (C=standard)
801 Connecting cable (various lengths on request)
802 Adapter, TC side
803 Adapter, pump side
802
801 803
C
AB
CAUTION
Damages to the pump and drive
Even after the mains power is switched off, the subsequently running pump delivers
electric power to the electronic drive unit. There is a danger of electric body contact by
premature separating the pump from the electronic drive unit.
Never separate the electronic drive unit from the pump when the mains power is
connected or the rotor is running.

5
Accessories
4.2 Connection to the turbopump
Attach and fix the pump-sided adapter of the Tele TC cable to the connection of
the pump lower part.
4.3 Connection to the electronic drive unit
Fix the electronic drive unit to the TC-sided adapter of the Tele TC cable using
two allen head screws 57.
– Preserve the third screw of the TC fastening, if necessary for air cooling.
Use free boreholes (M4) of the TC-sided adapter for fixation on external con-
structions (refer to "Dimensions").
5 Accessories
When using a Tele TC cable, the connectivity of accessory is nearly the same as for
the standard configuration of a turbopump and its electronic drive unit. Exceptions
are:
Fig. 2: Tele TC connection to the turbopump
3 Pump lower part 803 Adapter, pump side 813 Allen head screw (enclosure)
803
3
813
Fig. 3: Tele TC connection to the electronic drive unit
8 Electronic drive unit TC
802 Adapter, TC side
57 Allen head screw (TC)
57c Allen head screw, spare
8
802
57
57c
NOTE
Installation and operation of accessories
Pfeiffer Vacuum pumps can be equipped with a series of adapted accessories. The in-
stallation, operation and maintenance of connected devices are described in detail in
the operating instructions of the individual components.
For information on order numbers of components, see "Accessories".
Use original accessory parts only.

6
Product description
Install connecting cables between the accessories on the pump side and the
electronic drive unit.
– Order connecting cables in similar lengths of the Tele TC cable separately.
5.1 Installing an air cooling unit to the electronic drive unit
The electronic drive unit must be cooled additionally, when the respective ambient
temperatures are attained.
Fix the air cooling to the free borehole of the electronic drive unit, using the in-
stallation material (enclosure) and the remaining third allen head screw (57c).
Plug in and fix the accessory’s control lead to the corresponding accessory con-
nection port on the TC 400.
Designation Order number
Air cooling with cable for TC 400, M12 plug for Tele TC cable 400 PM Z01 304
Fig. 4: Mounting an air cooling unit to the electronic drive unit
57c Allen head screw
802 Adapter, TC side
850 Air cooling unit
855 Hexagon nut
856 Spring washer
802
57c 850
855
856
NOTE
Air cooling units on turbopump and electronic drive unit
Where the respective ambient temperatures are attained, Pfeiffer Vacuum recom-
mends the additional use of an air cooling unit for the turbopump with connection to the
TC.
Connect the air cooling unit using a Y-connector of the accessories programme.
Fig. 5: Connection example of Pfeiffer Vacuum accessories to a turbopump with Tele TC cable
400

7
Dimensions
6 Dimensions
Fig. 6: Tele TC cable 400, pump side
Fig. 7: Tele TC cable 400, TC side
B
AC
130
94 25.5
23.7
12
45
20.5
23.7 52.5 130
115
5
100 M4 - 9/10 deep
M4 - 9/10 deep
76.2
35
130.5
45
26
13

You are looking for a
perfect vacuum solution?
Please contact us:
Germany
Pfeiffer Vacuum GmbH
Headquarters
Tel.: +49 (0) 6441 802-0
info@pfeiffer-vacuum.de
Benelux
Pfeiffer Vacuum GmbH
Sales & Service Benelux
Tel.: +800-pfeiffer
benelux@pfeiffer-vacuum.de
China
Pfeiffer Vacuum
(Shanghai) Co., Ltd.
Tel.: +86 21 3393 3940
info@pfeiffer-vacuum.cn
France
Pfeiffer Vacuum France SAS
Tel.: +33 169 30 92 82
info@pfeiffer-vacuum.fr
Great Britain
Pfeiffer Vacuum Ltd.
Tel.: +44 1908 500600
sales@pfeiffer-vacuum.co.uk
India
Pfeiffer Vacuum India Ltd.
Tel.: +91 40 2775 0014
pfeiffer@vsnl.net
Italy
Pfeiffer Vacuum Italia S.p.A.
Tel.: +39 02 93 99 05 1
contact@pfeiffer-vacuum.it
Korea
Pfeiffer Vacuum Korea Ltd.
Tel.: +82 31 266 0741
sales@pfeiffer-vacuum.co.kr
Austria
Pfeiffer Vacuum Austria GmbH
Tel.: +43 1 894 17 04
office@pfeiffer-vacuum.at
Sweden
Pfeiffer Vacuum Scandinavia AB
Tel.: +46 8 590 748 10
sales@pfeiffer-vacuum.se
Switzerland
Pfeiffer Vacuum (Schweiz) AG
Tel.: +41 44 444 22 55
info@pfeiffer-vacuum.ch
United States
Pfeiffer Vacuum Inc.
Tel.: +1 603 578 6500
contact@pfeiffer-vacuum.com
www.pfeiffer-vacuum.net
Pfeiffer Vacuum stands for innovative and custom
vacuum solutions worldwide. For German engineering art,
competent advice and reliable services.
Ever since the invention of the turbopump, we´ve
been setting standards in our industry. And this claim
to leadership will continue to drive us in the future.
Leading. Dependable.
Customer Friendly
Table of contents
Other Pfeiffer Vacuum Cables And Connectors manuals