Introduction
Read this user manual carefully before you use
the Philips BlueControl and save the user manual
for future reference.To achieve optimal treatment
results, use the device according to the treatment
schedule and the instructions in this user manual.
List of product codes
- Kit PSK0202/10
- Device PSD1211/10
- Fixation strap PSS1210/10
- Adapter with USB cable PR3089
Intended use
The Philips BlueControl is intended to provide
phototherapeutic light to the body.The Philips
BlueControl is generally indicated to treat dermatological
conditions.The Philips BlueControl is a wearable medical
device using blue LED light to treat mild to moderate
psoriasis vulgaris for adults only (i.e. people over the age
of 18) in the home environment.
Important
Contraindications
Do not use the Philips BlueControl:
- If your skin is burnt or sunburnt.
- If you have a fever.
- If you have a known sun allergy.
- If you are allergic to the surface
materials of the Philips BlueControl
device (polycarbonate or polyester/
elastane mix).
ENGLISH

- If you have any known idiopathic
dermatosis such as porphyria,
polymorphic light eruption, chronic
actinic dermatitis, actinic prurigo
or solar urticaria that causes
photosensitivity.
- If you have any known disease
related to low immunity.
- If you have active skin cancer, if you
have a history of skin cancer or any
other localised cancer, precancerous
lesions or large moles in the areas
to be treated.
- If you are taking photosensitising
agents or medications, check the
package insert of your medicine and
never use the device if it is stated that
it can cause photo-allergic reactions,
photo-toxic reactions or if you have to
avoid sun when taking this medicine.
- If you are on steroids, it makes the
skin more susceptible to irritation.
- If you have infections, unhealed
tattoos, burns, inammation of hair
follicles, open lacerations, abrasions,
herpes simplex, or wounds in the
areas to be treated.
ENGLISH 7

- Avoid treatment of skin areas with
tattoos, there is no information
available yet about possible reactions.
- If you have uncontrolled arterial
hypertension, a history of stroke,
myocardial infarction, angina pectoris,
arteriosclerotic vascular disease (ASVD).
- If you are pregnant or breastfeeding.
- If you have schizophrenia, borderline
syndrome or severe depression.
- If you have any sensory deprivation
or have been diagnosed with
shingles or post-herpetic neuralgia,
especially in the area to be treated.
- If you have any anatomical pathology
or other physical limitation which
would prevent you from placing the
Philips BlueControl successfully.
- If you have an active implanted
device, such as a cardiac pacemaker,
debrillator, neurostimulator,
cochlear implant, or an active drug
administration device.
- If you are in poor general health.
ENGLISH8

Important safety information
Danger
- Keep the adapter dry.
Warning
- Never immerse the Philips BlueControl
in water or any other liquid.
- Never wear the Philips BlueControl
while it is charging.
- Never look directly into the light
source during use.
- Do not open or modify the Philips
BlueControl during the lifetime of
the device.
- Do not use the Philips BlueControl
in wet and/or hot surroundings,
such as a sauna or bathroom.
- The Philips BlueControl is not intended
to be used by persons younger than
18 and persons with reduced physical,
sensory or mental capabilities.
- This is a medical device. Keep the
device out of the reach of children.
- Only charge the Philips BlueControl
with the USB cable and the adapter
supplied (PR3089).
- The Philips BlueControl is specically
designed to treat psoriasis vulgaris.
Do not use it for any other purpose.
ENGLISH 9

- Always check the Philips
BlueControl, the USB cable and the
adapter for damage before use.
- Do not use the device if it is closer than
1 metre to wireless communication
equipment, such as wireless home
network routers and mobile phones.
This wireless communication
equipment may cause electromagnetic
interference that prevents the device
from operating properly.
- If the USB cable, the adapter or
the Philips BlueControl is damaged,
stop using it and contact the Philips
Consumer Care Centre at this
telephone number 00800-7445 4775.
Caution
- Remove the Philips BlueControl
from the xation strap before you
wash the xation strap.
- Use, charge and store the Philips
BlueControl at a temperature between
10°C and 35°C and at a relative
humidity between 30% and 90%.
- Read section ‘Contraindications’
to nd out if any apply to you
before you start using the Philips
BlueControl.
ENGLISH10

-
To prevent damage to the drum of the
washing machine, to other laundry in
the drum or the fabric of the xation
strap, close theVelcro fasteners of the
slings and wash the xation strap and
the slings in a ne-mesh washing bag.
- Always use the Philips BlueControl
in the xation strap with one of the
slings provided, as the xation strap
and the slings have been specially
designed for this device.
- Philips BlueControl is intended for
daily use.The exact usage time is
stated in chapter ‘Using your Philips
BlueControl’, section ‘Treatment
schedule’.
Compliance
The Philips BlueControl complies with the standard
of basic safety for medical devices (IEC60601-1)
and with the relevant collateral and particular
standards for non-laser light source medical
equipment for home use: EN 60601-1[1],
EN 60601-1-11 [3], EN 60601-2-57 [2],
and EN 60601-1-2 [4]. Regarding electromagnetic
compatibility (EMC) and electromagnetic elds
(EMF), the device should be kept away from radio
frequency (RF) equipment to avoid interference.
ENGLISH 11

General description (Fig. 1)
1 Philips BlueControl
2 On/off button and LED indicator
3 Micro USB socket
4 Sensor
5 Fixation strap
6 Short sling for lower arm or ankle
7 Medium-long sling for upper arm or knee
8 Long sling for leg
9 Adapter housing
10 European plug
11 UK plug
12 USB cable
13 Carry case
Charging
Fully charge the Philips BlueControl before you use
it for the rst time. When fully charged,
the Philips BlueControl has enough power for at
least four treatments. It is recommended to charge
the Philips BlueControl every day.You can charge the
Philips BlueControl with the adapter and
USB cable supplied.The adapter and USB cable have
an expected life time of at least 3 years in normal
intended operation.The device has a useful life of at
least 3 years if used in accordance with its intended
use. When the BlueControl is connected to the
charger and in charging mode, the charger PR3089 is
to be considered as a part of the medical system.
Note:Treatment is not possible when the Philips
BlueControl is charging.
ENGLISH12

Charging with the adapter
It takes approximately 3 hours to fully charge the
Philips BlueControl.
1 Put the European or UK mains plug in the
adapter and lock it into place.
2 Insert the standard USB plug into the socket
on the adapter.
3 Insert the adapter into the wall socket
(100V-240V).
4 Insert the micro USB plug into the socket on
the Philips BlueControl.
,When the Philips BlueControl is connected to
the mains, it beeps once and the LED indicator
starts ashing green slowly to indicate that
charging has started.
5 When the Philips BlueControl is fully charged,
the LED indicator lights up solid green.
6 Remove the adapter from the wall socket and
pull the micro USB plug out of the micro
USB socket on the Philips BlueControl.
ENGLISH 13

Using your Philips BlueControl
Treatment schedule
Always stick to the recommended treatment
schedule. Do not treat each plaque more than
once a day.
The Philips BlueControl has been designed for daily
use on psoriasis vulgaris plaques. We recommend
treating single, non-overlapping psoriasis vulgaris
plaques only once a day.A single treatment takes
approximately 30 minutes, but can be extended to
40 minutes depending on your skin sensitivity and
possible treatment pauses during treatment.
For example, if the infrared sensor senses a too high
skin temperature, the treatment is briey paused for
safety reasons.Treatment resumes automatically when
the skin has cooled down. Following this treatment
advice ensures that you do not exceed the daily light
dose of 90 J/cm² per plaque. Blue light treatment can
be combined with vitamin D creams.Apply the cream
after the treatment. For other combination treatments
with blue light, please consult your doctor.
We do not recommend using topical medication,
creams or lotions immediately before and during
treatment with the Philips BlueControl. In case of
unpleasant effects such as excessive, non-resolving skin
redness, worsening of plaques or pain,
stop the therapy immediately and consult your doctor
if necessary. It is possible that you experience slight
tanning of the area treated.This should diminish within
a few days after the therapy has been stopped.
A temporary increased redness of the treated area
after treatment is normal but should also diminish.
ENGLISH14

Preparing the Philips BlueControl for use
To place the Philips BlueControl in the xation strap
and to position it correctly on your body, follow
the steps below. The xation strap has an expected
lifetime of at least 1 year if used in accordance with
its intended use.There are three slings: a short sling
for lower arm or ankle, a medium-long sling for
upper arm or knee, and a long sling for the leg.
1 Place the Philips BlueControl in the xation
strap.
Note: Make sure that the LED treatment area is not
covered by the xation strap material or any other
textile. The on/off button must be visible through the
hole on top of the xation strap
2 Select the suitable sling for the area to be
treated on the arm or the leg.
3 To attach the selected sling to the xation
strap, pull one end of the sling through one ring
of the xation strap (1) and fasten it with the
Velcro fastener (2).
Note: After fastening the Velcro fastener, two small grey
Velcro pads should be on the top of the sling.
4 Remove clothing from the area to be treated.
5 Position the xation strap on top of the
psoriasis plaque on your arm or leg (1).
Wrap the sling around your arm or leg (2).
1
2
1
2
ENGLISH 15

6 Pull the sling through the other ring (1).
Then wrap the sling around your arm or
leg once more and fasten it with theVelcro
fastener (2).
Note: The two small grey Velcro pads on top of the
other end of the sling enable the tightest adjustment.
Treatment procedure
1 Press the on/off button to switch on the Philips
BlueControl.
Caution: Never look directly into the light source
during use.
Do not intentionally touch the sensor in the centre
of the Philips BlueControl when it is switched on.
Do not cover the Philips BlueControl with other
textile materials than the xation strap when it is
switched on.
2 The Philips BlueControl ashes blue to indicate
that it is ready for treatment. The treatment
starts soon after the sensor detects the skin to
be treated.
,The LED indicator lights up solid blue and
a sequence of two tones indicates the treatment
is in progress.
3 The Philips BlueControl automatically switches
off after the treatment time
(approximately 30 minutes) has elapsed.
2
1
ENGLISH16

Interrupting treatment
1 If you want to interrupt the treatment briey,
simply remove the Philips BlueControl or press
the on/off button to switch it off.
Note: If you remove the Philips BlueControl or press the
on/off button to stop treatment before the treatment time
has elapsed and you do not resume the treatment within
2 hours, the timer automatically resets. If you resume the
treatment within 2 hours, the countdown of the treatment
time will continue where it left off. When a treatment is
resumed, there will be no sequence of 2 tones.
After treatment
Treatment ends automatically after the treatment
time has elapsed, unless you remove the Philips
BlueControl or press the on/off button to stop
treatment earlier.
Note: A sequence of two different tones indicates that
the Philips BlueControl switches off and treatment has
ended.
1 Open theVelcro fastener of the sling at one
end (1) and pull the strap out of the ring of the
xation strap (2).
2 Unwrap the sling (1) and remove the xation
strap from your psoriasis plaque (2).
3 Recharge the Philips BlueControl.You can take
the device out of the xation strap or charge
it while it stays in the xation strap.
1
2
2
1
ENGLISH 17

Light and sound signals during use
Overheating signal
Overheating of the device.
1 If the device measures a too high device
temperature, the Philips BlueControl switches
off for safety reasons.
,The LED indicator lights up solid orange and
a long 3-second tone indicates that the Philips
BlueControl switches off.
2 If you try to restart the Philips BlueControl
before it has cooled down, the LED indicator
ashes orange for 5 seconds. Cooling down
may take several minutes.
Overheating of the skin.
1 If the device measures a too high skin
temperature, the Philips BlueControl
temporarily switches off for safety reasons.
,The LED indicator ashes blue slowly until the
skin temperature has cooled down, the Philips
BlueControl switches on again automatically.
Battery level signals
When you press the on/off button and the LED
indicator lights up solid red and you hear 3 beeps,
the battery of the Philips BlueControl is empty.
1
You have to charge the battery before you can
start a new treatment. Charge for at least
45 minutes to be able to complete one full
treatment.
ENGLISH18

Cleaning
Do not immerse the Philips BlueControl in water
or rinse it under the tap.
Always remove the Philips BlueControl from the
xation strap before you put the xation strap in
the washing machine to clean it.
Caution: Clean the Philips BlueControl after each
treatment to ensure it continues to function properly.
1 Remove the Philips Blue Control from the
xation strap.
2 Clean the Philips BlueControl with a damp
cloth.
3 Clean the sensor with alcohol and a cotton
bud at an ambient temperature once a week.
4 The xation strap and the slings can be washed
in the washing machine.
Caution:To prevent damage to the drum of the
washing machine, to other laundry in the drum
or the fabric of the xation strap, close the Velcro
fasteners of the slings and wash the xation strap
and the slings in a ne-mesh washing bag.
Note: Use a gentle cycle and a temperature of 60°C
maximum. Dry the xation strap and the slings on the
air and not in a tumble dryer. Do not iron the xation
strap and the slings.
ENGLISH 19

Storage
The device comes with a carry case with storage
spaces for all parts.
1 Use the storage case to store the device and
its parts if you do not use it.
2 If you want to take the device with you to
work or while travelling, we advise you to put
it in the carry case for extra protection.
Troubleshooting
This chapter summarises the most common
problems you could encounter with the Philips
BlueControl. If you are unable to solve the problem
with the information below, contact the Philips
Consumer Care Centre.
Problem Possible cause Solution
I cannot switch
on the Philips
BlueControl even
though it is fully
charged.
The sensor is
dirty.
Clean the sensor with
alcohol and a cotton bud
(see chapter ‘Cleaning’).
ENGLISH20