PME miniDOT Clear User manual

WWW.PME.COM
PG. 2
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
TABLE OF CONTENTS
WARRANTY..................................................................................................................................... 3
Limited Warranty........................................................................................................................................3
SAFETY INFORMATION ................................................................................................................. 5
Bursting Hazard .........................................................................................................................................5
CHAPTER 1: QUICK START ........................................................................................................... 6
1.1 The Quickest Start Possible ................................................................................................................6
1.2 A Few Details.......................................................................................................................................8
1.3 Overview & General Maintenance.......................................................................................................9
CHAPTER 2: SOFTWARE............................................................................................................. 15
2.1 Overview and Software Installation ...................................................................................................15
2.2 miniDOT Control................................................................................................................................16
2.3 miniDOT Plot......................................................................................................................................16
2.4 miniDOT Concatenate .......................................................................................................................19
CHAPTER 3: MINIDOT CLEAR LOGGER...................................................................................... 21
3.1 Overview............................................................................................................................................21
3.2 Opening and Closing the miniDOT Clear Logger..............................................................................21
3.3 Electrical Connections and Controls..................................................................................................22
3.4 Battery Replacement .........................................................................................................................23
3.5 Installing Copper Mesh or Plate.........................................................................................................24
3.6 Final Mounting Instructions................................................................................................................25

WWW.PME.COM
PG. 3
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
WARRANTY
Limited Warranty
Precision Measurement Engineering, Inc. (“PME”) warrants the following products to be, as of the
time of shipment, free from defects in materials or workmanship under normal use and conditions for
the period indicated below corresponding to the product. The warranty period commences on the
original date of purchase of the product.
Product
Warranty Period
Aquasend Beacon
1 year
miniDOT Logger
1 year
miniDOT Clear Logger
1 year
miniWIPER
1 year
miniPAR Logger (Logger only)
1 year
Cyclops-7 Logger (Logger only)
1 year
C-FLUOR Logger (Logger only)
1 year
T-Chain
1 year
MSCTI (excludes CT/C-sensors)
1 year
C-Sense Logger (Logger only)
1 year
For valid warranty claims made and covered defects existing during the applicable warranty period,
PME will, at PME’s option, repair, replace (with the same or then most similar product) or repurchase
(at purchaser’s original purchase price), the defective product. This warranty extends solely to the
original end-user purchaser of the product. PME’s entire liability and the sole and exclusive remedy
for product defects is limited to such repair, replacement or repurchase in accordance with this
warranty. This warranty is provided in lieu of all other warranties express or implied,
including, but not limited to warranties of fitness for a particular purpose and warranties of
merchantability. No agent, representative, or other third party has any authority to waive or alter
this warranty in any way on behalf of PME.
WARRANTY EXCLUSIONS
The warranty does not apply in any of the following circumstances:

WWW.PME.COM
PG. 4
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
I) The product has been altered or modified without PME’s written authorization,
II) the product has not been installed, operated, repaired, or maintained in accordance with PME’s
instructions, including, where applicable, use of proper grounding to an earth ground source,
III) the product has been subjected to abnormal physical, thermal, electrical, or other stress, internal liquid
contact, or misuse, neglect, or accident,
IV) the product failure occurs as a result of any cause not attributable to PME,
V) the product is installed with ancillary devices such as flow sensors, rain switches, or solar panels that
are not listed as compatible with the product,
VI) the product is installed in a non-PME specified enclosure or with other incompatible equipment,
VII) to address cosmetic issues such as scratches or surface discoloration,
VIII)operation of the product in conditions other than that for which the product was designed,
IX) the product has been damaged due to events or conditions such as caused by lightning strikes, power
surges, unconditioned power supplies, floods, earthquakes, hurricane, tornados, vermin such as ants
or slugs or intentional damage, or
X) products provided by PME, but manufactured by a third-party company, which products are subject to
the applicable warranty extended by their manufacturer, if any.
There are no warranties that extend beyond the above limited warranty. In no event is PME
responsible or liable to purchaser or otherwise for any indirect, incidental, special,
exemplary, or consequential damages, including, but not limited to, lost profits, loss of data,
loss of use, business interruption, loss of good will, or cost of procuring substitute products,
arising out of or in relation to the product, even if advised of the possibility of such damages
or losses. Some states do not allow the exclusion or limitation of incidental or consequential
damages, so the above limitation or exclusion may not apply. This warranty gives you
specific legal rights, and you may also have other rights which vary from state to state.
WARRANTY CLAIM PROCEDURES
A warranty claim must be initiated within the applicable warranty period by first contacting PME at
info@pme.com to obtain an RMA number. The purchaser is responsible for proper packaging and
return shipment of the product to PME (including shipping expense and any related duties or other
costs). The issued RMA number and purchaser’s contact information must be included with the
returned product. PME is NOT liable for loss or damage of the product in return transit and
recommends that the product be insured for its full replacement value.
All warranty claims are subject to PME’s testing and examination of the product to determine if the
warranty claim is valid. PME may also require additional documentation or information from the
purchaser to evaluate the warranty claim. Products repaired or replaced under a valid warranty claim
will be shipped back to the original purchaser (or its designated distributor) at PME’s expense. If the
warranty claim is found to be not valid for any reason, as determined by PME in its sole discretion,
PME will notify the purchaser at the contact information provided by the purchaser.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 5
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
SAFETY INFORMATION
Bursting Hazard
Should water enter the miniDOT Clear Logger and come into contact with the enclosed batteries,
then the batteries may generate gas, causing the internal pressure to increase. This gas will likely
exit via the same location where the water entered, but not necessarily. The miniDOT Clear Logger
is designed to release internal pressure, as the black end cap is unscrewed, prior to the
disengagement of the black end cap threads. If internal pressure is suspected, then treat the
miniDOT Clear Logger with extreme caution.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 6
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
CHAPTER 1: QUICK START
1.1 The Quickest Start Possible
Your miniDOT Clear Logger has arrived completely ready to go. It is set to measure and record time,
battery voltage, temperature, oxygen concentration, and measurement quality once every 10
minutes and write one file of measurements daily. Open the miniDOT Clear Logger and move the
Logger Control Switch to the “Record” position. In this condition, the miniDOT Clear Logger will
record measurements for a year before the internal batteries are expended. You must re-close the
miniDOT Clear Logger prior to deploying it.
At the end of the deployment period, open the miniDOT Clear Logger and connect it to a HOST
computer via the USB connection. The miniDOT Clear Logger will appear as a ‘thumb drive’. Your
temperature and oxygen concentration measurements, together with a time stamp indicating the
time the measurements were made, are recorded in text files in the folder having the serial number
of your miniDOT Clear Logger. These files can be copied onto any Windows or Mac HOST
computer.
This Manual and other software programs are also recorded on the miniDOT Clear Logger.
•MINIDOT CONTROL PROGRAM:Allows you to see the state of the miniDOT Clear Logger as
well as set the recording interval.
•MINIDOT PLOT PROGRAM:Allows you to see the plots of the recorded measurements.
•MINIDOT CONCATENATE PROGRAM:Gathers all the daily files into one CAT.txt file.
Your miniDOT Clear Logger will return to recording measurements after you disconnect the USB
connection. If you wish to stop recording, then move the Logger Control Switch to the “Halt” position.
You may move the Logger Control Switch at any time.
Follow these steps to start the deployment, logging DO & T once each 10 minutes:
1. Open the miniDOT Clear Logger by unscrewing the clear pressure housing from the black
end cap. It opens like a flashlight. Remove the clear pressure housing completely. Inside you
will see the circuit pictured below:

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 7
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
2. Move the Logger Control Switch to the “Record” position. The LED will flash green 5 times.
The miniDOT Clear Logger will now record a measurement of time, battery voltage,
temperature, and dissolved oxygen every 10 minutes (or at some other interval you may
have set using the miniDOT Control program).
3. Inspect the o-ring seal for debris.
4. Close the miniDOT Clear Logger by screwing the clear pressure housing back onto the black
end cap.
5. Deploy the miniDOT Clear Logger.
Follow these steps to end the deployment:
1. Recover the miniDOT Clear Logger
2. Clean and dry all accessible surfaces except the ‘sensing foil’.
3. Open the miniDOT Clear Logger by unscrewing the clear pressure housing from the black
end cap. Remove the clear pressure housing completely, taking care that water does not drip
onto interior surfaces of circuits or other items inside the miniDOT Clear Logger.
4. Connect to a Windows HOST computer via the USB connection. The miniDOT Clear Logger
will appear as a ‘thumb drive’.
5. Copy the folder having the same serial number as the miniDOT Clear Logger (example
7450-0001) to the HOST computer.
6. (Suggested, but optional) Delete the measurement folder, but NOT the miniDOT Control
program or the other .jar programs.
7. (Optional) Run the miniDOT Control program to see the state of the miniDOT Clear Logger
such as battery voltage or to select a different recording interval.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 8
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
8. (Optional) Run the miniDOT PLOT program to see a plot of measurements.
9. (Optional) Run the miniDOT Concatenate program to gather all the daily files of
measurements into one CAT.txt file.
10. If no more recording is desired, then move the Logger Control Switch to “Halt”, otherwise
leave it set to “Record” to continue recording measurements.
11. Disconnect the miniDOT Clear Logger from the USB connection.
12. Inspect the o-ring seal for debris.
13. Close the miniDOT Clear Logger by screwing the clear pressure housing back onto the black
end cap.
14. Remove the batteries if storing the miniDOT Clear Logger for extended periods.
1.2 A Few Details
The previous section gives instructions for sampling at 10-minute intervals. However, there are a few
additional details that will enhance use of the miniDOT Clear Logger.
RECORDING INTERVAL
The miniDOT Clear Logger measures and records time, battery voltage, temperature, dissolved
oxygen concentration and measurement quality at equal time intervals. The default time interval is
10 minutes. However, it is also possible to instruct the miniDOT Clear Logger to record at different
intervals. This is accomplished by running the miniDOTControl.jar program supplied with the
miniDOT Clear Logger. Recording intervals must be 1 or more minutes and must be less than or
equal to 60 minutes. Intervals outside this range will be rejected by the miniDOT Control program.
(Contact PME for other recording intervals.)
Please refer to Chapter 2 for instructions on operating the miniDOT Control program.
TIME
All miniDOT Clear Logger times are UTC (formerly known as Greenwich mean time (GMT)). The
miniDOT Clear Logger internal clock will drift in the <10 ppm range (< about 30 seconds/month) so
you should plan to connect it occasionally to a HOST computer having an Internet connection. The
miniDOT Control program will automatically set time based on an Internet time server. If the logger
has an issue correcting its time, please contact PME.
Please refer to Chapter 2 for instructions on operating the miniDOT Control program.
FILE INFORMATION

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 9
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
The miniDOT Clear Logger software creates 1 file daily on the miniDOT Clear Logger’s internal SD
card. The number of measurements in each file will depend upon the sample interval. Files are
named by the time of the first measurement within the file based on the miniDOT Clear Logger’s
internal clock and expressed in YYYY-MM-DD HHMMSSZ.txt format. For example, a file having the
first measurement on September 9, 2014 at 17:39:00 UTC will be named:
2014-09-09 173900Z.txt.
Files can be uploaded from the miniDOT Clear Logger by connecting it to a HOST computer. Use
the HOST computer’s copy/paste functions to move the files from the miniDOT Clear Logger to the
HOST computer.
Each measurement within the files has a time stamp. The time stamp format is Unix Epoch 1970, the
number of seconds that have passed since the first moment of 1970. This may be inconvenient in
some cases. If so, then the miniDOT Concatenate program not only concatenates all the
measurement files, but also adds more readable statements of the time stamp.
Please refer to Chapter 2 for instructions on operating the miniDOT Concatenate program.
The miniDOT Clear Logger requires time and battery energy to work through the file directory on the
SD card to allocate new file space. A few hundred files on the SD card is not a problem, but as the
number of files grows large into the thousands then the miniDOT Clear Logger may suffer decreased
battery life or other performance problems. Please, at the earliest convenient time, copy the
recorded files to a HOST computer and delete them from miniDOT Clear Logger’s SD card. Also, do
not use miniDOT Clear Logger to store files unrelated to the miniDOT Clear Logger’s operation.
1.3 Overview & General Maintenance
CLEANING THE SENSING FOIL
The sensing foil can be cleaned at regular intervals depending on the fouling condition at the site.
The cleaning procedure of the sensing foil should be done with caution so that the protective coating
is not removed. If the fouling is calcareous it can normally be dissolved with household vinegar.
If the marine growth remains, then use Q-tips to gently wipe the sensing foil after it has been
softened by soaking in vinegar or perhaps diluted HCl. After cleaning the sensing foil, then it should
be rinsed well in clean tap water before storing or reuse. Do not use other organic solvents such as
acetone, chloroform, and toluene since these and others will damage the sensing foil.
Never peel away untreated organisms from the oxygen sensing foil. Doing so will likely damage it.
The sensing foil can also be cleaned using a 3% H2O2 solution or rinsing it with ethanol.
The clear pressure housing and black end cap can be gently scrubbed however the clear pressure
housing can easily scratch if an abrasive material is used. Contact PME for replace clear pressure
housing.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 10
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
Keep the sensing foil out of direct sunlight for extended periods of time.
AA ALKALINE BATTERY LIFE
Alkaline batteries will give somewhat less performance than lithium, especially at low temperatures.
Alkaline batteries are superior to lithium in one way: you can determine how much battery life
remains by measuring the battery terminal voltage. For short deployments of a month or two, then
alkaline batteries will provide adequate performance. For longer deployments, or for deployments in
cold environments, then substitute lithium batteries.
AA LITHIUM BATTERY LIFE
The miniDOT Clear Logger consumes battery power mostly from the measurement of dissolved
oxygen, but also slightly from simply keeping track of time, writing files, sleeping, and other activities.
The following table shows the approximate endurance of the miniDOT Clear Logger when powered
by the Energizer L91 AA lithium / ferrous disulfide batteries:
Sample Interval
(Minutes)
Main AA Battery Life
(Months)
Number of Samples
1
12
500K
10
>12
>52,000
60
>12
>8,000
Keep a general record of the miniDOT Clear Logger’s number of samples. It is not possible to
accurately tell the charge state of a lithium battery by measuring its terminal voltage. If you have a
general idea of the number of samples already obtained on a battery, then you can make a guess as
to how many more samples remain.
The numbers in the table above are, at the time of this writing, based upon extrapolations of testing
of 500K samples acquired at a 5-second interval. The 1-year performance at 1 minute is very likely.
Performance at longer sample intervals will be much longer, but how long is difficult to predict. In any
event, these AA batteries are easily available and relatively inexpensive compared to the cost of the
miniDOT Clear Logger. PME suggests you replace the batteries often, especially before any long
(months) measurement deployments.
Monitor battery voltage in the miniDOT Control program. You cannot tell from terminal voltage of a
lithium battery how long the battery will last, but you can tell if it will die soon. The Low Drain
Performance plot below gives an estimate of terminal voltage for both lithium and alkaline batteries.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 11
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
You can operate batteries down to about 2.4 Volts (for two in series, 1.2 Volts on the graph below).
Remove the batteries and measure each of them. If your combined battery voltage is less than 2.4
Volts, replace the batteries.
You may also use alkaline AA batteries such as Duracell Coppertop. They will not last nearly as
long, especially at low temperatures, but will likely be adequate for several weeks at 10-minute
interval.
When replacing batteries use only fresh batteries. Do not mix battery types. If one battery differs in
type or charge level from the other and the miniDOT Clear Logger runs them to full discharge, then
one battery may leak. SEE SECTION 3.4 FOR CAUTION ON BATTERY PLACEMENT.
Error on the side of caution when planning your deployment.
The recommended battery is the Energizer L91 lithium battery. For more information including the
performance at low temperatures, then click the link: http://data.energizer.com/PDFs/l91.pdf
The figure to the left gives a general idea of
terminal voltage vs. lifetime. Service life in
hours is incorrect since miniDOT Clear
Logger draws far less than 50 mA
continuously, but the general shape of the
voltage vs. time gives an estimate of life
remaining. This plot is taken from the
manufacturer’s specification. The plot is for a
single battery. The miniDOT Clear Logger
halts operation at a total of 2.4 Volts.
COIN CELL BATTERY LIFE
The miniDOT Clear Logger uses a coin cell battery for backup of the clock when the power is
switched off. This coin cell battery will supply many years of clock operation. Should the coin cell
battery discharge, then it must be replaced by PME. Contact PME.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 12
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
RECALIBRATION
The miniDOT Clear Logger will maintain its calibration without the necessity of adjustment by the
user. The miniDOT Clear Logger should be returned to PME for re-calibration. We suggest that this
be done yearly.
O-RING AND SEAL
When the clear pressure housing is screwed on to the black end cap, then it passes along the o-ring
located in the black end cap several revolutions. Keep this o-ring lightly lubricated with silicone
grease or an oil compatible with buna-N o-ring material.
It is important to keep the o-ring free of debris. Failure to do so can lead to a breaching of the seal
and water entry into the logger housing. Wipe off debris with a lint free clean cloth. PME
recommends Kimtech Kimwipes for this application. Next, re-lubricate the o-ring.
When the miniDOT Clear Logger is opened after deployment, then a small number of water drops
are deposited on the inner surface of the o-ring. When the clear pressure housing is screwed back
on to the black end cap, then these drops can become trapped inside the miniDOT Clear Logger. Be
sure to carefully dry the o-ring and adjacent surfaces (especially underneath) prior to closing the
miniDOT Clear Logger. Re-lubricate the o-ring at this time.
LED INDICATIONS
The miniDOT Clear Logger indicates its operation with its LED. The table below presents LED
indications:
LED
Reason
1 Green Flash
Normal. Presented immediately after new batteries are installed. Indicates that
the CPU has started its program.
1 Green Flash
Occurs at the time of sampling for sample intervals of 1 minute or less.
5 Green Flashes
Normal. Indicates that miniDOT Logger is beginning to record measurements.
This indication appears in response to switching the Logger Control Switch to
“Record.”
5 Red Flashes
Normal. Indicates the miniDOT Logger is ending recording of measurements.
This indication appears in response to switching the Logger Control Switch to
“Halt.”
Continuously Green
Normal. Indicates the miniDOT Logger is connected to a HOST computer via the
USB connection.
Continuously Flashing Red
SD card write error. Try removing/reinstalling batteries. Contact PME.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 13
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
VERIFYING CALIBRATION
You may from time to time want to verify the calibration of your miniDOT Clear Logger. Do this by
placing the miniDOT Clear Logger in a black 5-gallon bucket containing 4 gallons of fresh water.
(The picture below shows a white bucket so that the miniDOT Clear Loggers are more easily seen.)
The miniDOT Clear Logger’s black end cap is heavy and the miniDOT Clear Logger will tend to flip
so that this end is down. Prevent this somehow. The miniDOT Clear Logger must be placed in the
bucket with the black end cap upwards. Otherwise bubbles will accumulate in the black end cap area
and miniDOT Clear Logger will not sense the DO in the water correctly. Use an aquarium pump and
air stone in the water to provide a bubble stream. Cover the bucket with a black lid. The idea is to
prevent light from enabling algal growth.
Record measurements for several hours or a day, but
in any event long enough for the miniDOT Clear
Logger’s temperature to come to equilibrium with the
water. During the experiment, find the local air
pressure, either from measurements or from a local
weather station. Watch out... weather stations often
report barometric pressure referenced to sea level.
You must determine the absolute barometric pressure
at your elevation.
A more comprehensive experiment is to additionally
place ice in the bucket and mix until the water
temperature is close to zero degrees. Next, remove
the ice. Place the bucket on a towel or piece of cardboard and cover the top of the bucket with a
towel. Record for 24 hours as the bucket temperature gradually returns to room temperature.
After recording the bubbled water, you may also remove the air stone and gently mix a packet of
baker’s yeast into the bucket together with a tablespoon of sugar. The water must be only slightly
warm to the touch but not more than 30 deg C. These organisms will deplete all the dissolved
oxygen in the water. Cut a disc of thin plastic film just large enough to lay on top of the water. Place
this on top of the water. Do not stir or bubble after placing the film. Record measurements for at least
an hour or more.
Use the miniDOT Clear Logger’s miniDOT Plot program to examine the measurements. Saturation
values should be very close to 100%, depending upon the accuracy that you have determined
barometric pressure. If you placed ice in the bucket, then saturation values will still be 100%. You
will see the DO concentration and temperature change greatly as the bucket warms.
The recorded data, when using yeast should show 0% saturation and 0 mg/l dissolved oxygen
concentration. In practice the miniDOT Clear Logger often reports slightly positive values of about
0.1 mg/l, but within the accuracy of the miniDOT Clear Logger.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 14
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
CLOSING AND OPENING
Close and open the miniDOT Clear Logger like you would a flashlight; open by unscrewing the clear
pressure housing from the black end cap. Close by screwing the clear pressuring housing on to the
black end cap. When closing, do not tighten the clear pressure housing. Just screw it on until it
contacts the black end cap. See Chapter 3 for more instructions.
Caution: DO NOT remove the stainless-steel screws in the black end cap. There are no user serviceable parts here.
If the screws are removed, then you will damage the miniDOT Clear Logger and it will have to be returned for repair.
STORAGE WHEN NOT IN USE
Remove the batteries. Keep the black end covered with the cap supplied by PME. If the cap is lost,
then cover the black end cap with aluminum foil. There may be a calibration effect of ambient lighting
so attempt to keep ambient light from reaching the sensing foil as much as possible.
JAVA
miniDOT Clear programs depend on Java and require Java 1.7 or higher. Update Java at
https://java.com/en/.
ENVIRONMENTAL USE AND STORAGE CONDITIONS
miniDOT Clear is useful over the range of 0 to 150% saturation of dissolved oxygen, over the range
of 0 to 35 deg C temperature and may be continuously immersed in fresh or salt water to a
maximum depth of 100 meters. miniDOT Clear may be stored in environments ranging from 0 to
100% humidity and temperatures ranging from -20 deg C to +40 deg C.
ELECTRICAL POWER SPECIFICATIONS
miniDOT Clear is battery powered and requires 2 AA size expendable or rechargeable batteries.
Voltage requirement is 3.6 VDC. Maximum current demand is 30 mA.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 15
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
CHAPTER 2: SOFTWARE
2.1 Overview and Software Installation
The miniDOT Clear Logger arrives with these files on the SD card:
•miniDOTControl.jar program allows you to see the state of the miniDOT Clear Logger as well
as set the recording interval.
•miniDOTPlot.jar program allows you to see the plots of the recorded measurements.
•miniDOTConcatenate.jar program gathers all the daily files into one CAT.txt file.
•Manual.pdf is the manual.
These files are located on the root directory of the miniDOT Clear Logger.
PME suggests you leave these programs where they are on the miniDOT Clear Logger, but you may
copy them to any folder on your HOST computer’s hard drive.
miniDOT Control, miniDOT Plot, and miniDOT Concatenate programs are Java language programs
that require the HOST computer to have the Java Runtime Engine V1.7 (JRE) or later versions
installed. This engine is commonly required for Internet applications and will likely already be
installed on the HOST computer. You can test this by running the miniDOT Plot program. If this
program displays its graphical user interface, then the JRE is installed. If not, then the JRE can be
downloaded via the Internet from http://www.java.com/en/.
At this time, the miniDOT Clear Logger is supported on the Windows operating systems but may
also operate on Macintosh and perhaps Linux.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 16
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
2.2 miniDOT Control
Begin program operation by clicking
“miniDOTControl.jar”. The program presents the
screen shown below:
The miniDOT Clear Logger must be connected to
the HOST computer via the USB connection at this
time. When correctly connected, the miniDOT Clear
Logger’s LED will display a constant green light.
Click the “Connect” button. The program will
contact the miniDOT Clear Logger. If the
connection is successful, then the button will turn
green and display “Connected”. The Serial Number
and other parameters will be filled in from the
information taken from the miniDOT Clear Logger.
If the HOST computer is connected to the Internet,
then the current difference between an Internet time server’s time and the miniDOT Clear Logger’s
internal clock will be displayed. If more than a week has passed since time was last set, then the
miniDOT Clear Logger’s clock will be set and the check mark icon will appear. If the HOST computer
is not connected to the Internet, then no time services will occur. If the miniDOT Clear is unable to
automatically set time and there is a large time error at connect, please contact PME about
correcting this.
The current miniDOT Clear Logger’s sample interval will be displayed next to the “Set Sample
Interval” button.
To set the interval, enter an interval not less than 1 minute and not greater than 60 minutes. Click
the “Set Sample Interval” button. Shorter and faster intervals are available. Contact PME.
If this interval is acceptable, then the interval need not be set.
End the miniDOT Control program by closing the window. Disconnect the miniDOT Clear Logger’s
USB cable.
Upon disconnection of the USB cable, the miniDOT Clear Logger will begin logging or remain halted
as indicated by the position of the Logger Control Switch.
2.3 miniDOT Plot
Begin the program operation by clicking “miniDOTPlot.jar”. The program presents the screen shown
below.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 17
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
The miniDOT Plot program plots the files recorded by
the miniDOT Clear Logger. The program reads all the
miniDOT Clear Logger files in a folder, except the
CAT.txt file. The program will also compute air
saturation from the dissolved oxygen measurements.
To do this, the program must know the air pressure and
salinity. It calculates air pressure based on elevation of
the water surface above sea level or uses the
barometric pressure you enter if Barometric Pressure is selected. If the Surface Elevation is entered,
then no compensation for weather-induced barometric pressure variation is made. Enter elevation or
barometric pressure. Enter water salinity.
Select the folder that contains the files recorded by the miniDOT Clear Logger. If the miniDOT Plot
program is run directly from the miniDOT Clear Logger, then the program will suggest the folder that
is located on the miniDOT Clear Logger’s SD card. You may accept this by clicking “Plot”, or you
may click “Select DATA Folder” to browse to your HOST computer’s hard drive. If the number of
measurements recorded is small, for example a few thousand, then these can conveniently be
plotted directly from the miniDOT Clear Logger’s storage. However, it is best to copy large
measurement sets to the HOST computer and select them there. The file access to the miniDOT
Clear Logger is slow.
The miniDOT Clear Logger’s measurement folders must NOT contain any files besides those the
miniDOT Clear Logger recorded and the CAT.txt file.
Click “Plot” to begin plotting.
The program reads all the miniDOT Clear Logger’s data files in the selected folder. It concatenates
these and presents the plot shown below.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 18
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
You may zoom this plot by drawing a square from upper left to lower right (click and hold left mouse
button) that defines the zoom region. To zoom completely out, attempt to draw a square from lower
right to upper left. Right click on the plot for options such as copy and print. The plot can be scrolled
with the mouse while the Control key is held depressed. Copies of the plot can be obtained by right
clicking on the plot and selecting Copy from the pop-up menu.
Different DATA Folders can be selected during one session of the program. In this case the software
produces multiple plots. Unfortunately, the plots are presented exactly on top of one other and so
when a new plot appears it is not obvious that the old plot is still there. It is. Just move the new plot
to see previous plots.
The program can be re-run at any time. If an already processed DATA Folder is selected, then the
progrm simply reads the miniDOT Clear Logger’s measurement files again.
End the miniDOT Plot program by closing the window.
Special note: plotting of sample sets of more than 200K samples may consume all the memory
available to the JRE. The miniDOT Plot program will present a partial plot and freeze in this case. A
simple solution is to separate the files into multiple folders and plot each folder individually. A special
miniDOT Plot that sub-samples can be provided by PME. Please contact PME in this case.

WWW.PME.COM
TECHNICAL SUPPORT: INFO@PME.COM | (760) 727-0300
PG. 19
THIS DOCUMENT IS PROPRIETARY AND CONFIDENTIAL. © 2021 PRECISION MEASUREMENT ENGINEERING, INC. ALL RIGHTS RESERVED.
2.4 miniDOT Concatenate
Begin the program operation by clicking “miniDOTConcatenate.jar”. The program presents the
screen shown below.
The miniDOT Concatenate program reads and concatenates the files recorded by the miniDOT
Clear Logger. This program produces a CAT.txt file in the same folder as selected for the data. The
CAT.txt file contains all the original measurements and contains two additional statements of time
and air saturation. To compute saturation, the program must know the air pressure and salinity. It
calculates air pressure based on elevation of the water surface above sea level or uses the
barometric pressure you entered if the Barometric Pressure was selected. If Surface Elevation is
entered, then no compensation for weather-induced barometric pressure variation is made. Enter
elevation or barometric pressure. Enter water salinity.
Select the folder that contains the files recorded by the miniDOT Clear Logger. If the miniDOT Plot
program is run directly from the miniDOT Clear
Logger, then the program will suggest the folder
located on the miniDOT Clear Logger. You may
accept this by clicking “Concatenate”, or you
may click “Select DATA Folder” to browse your
HOST computer’s hard drive. If the number of
measurements recorded is small, for example a
few thousand, then these can conveniently be
plotted directly from the miniDOT Clear Logger’s
storage. However, it is best to copy large
measurement sets to the HOST computer and
select them there. The file access to the miniDOT Clear Logger is slow.
The miniDOT Clear Logger measurement folders must NOT contain any files besides those the
miniDOT Clear Logger recorded and the CAT.txt file.
Click “Concatenate” to begin concatenating files and create the CAT.txt file.
Table of contents
Other PME Measuring Instrument manuals