5Instructions for Use 2019-03-20
• Accurate canal measurement is not always possible depending on the shape and condition of the tooth as well as a
decline in the equipment’s performance.
• Do not use damaged file holders; an accurate measurement cannot be made with a damaged file holder.
• When a continuous tone is heard while the Power Switch is on and without any operation, some electrical part may
be malfunctioning. Do not use the device and send it to Poldent Co. Ltd. for repairing.
• A rubber dam should be used when performing endodontic treatment.
• Some care must be taken concerning electromagnetic compatibility (EMC) when using the device. Refer to the user’s
manual and other attached documents for EMC information regarding installation and operation.
• Both portable and movable radio frequency transmitters may have some effect on the device.
• Using replacement parts or accessories not supplied by Poldent Co. Ltd. could adversely affect the EMC performance
of the device.
• As far as possible, do not use the Endostar Navigator near or simultaneously with other devices. If this cannot be
avoided, observe carefully and make sure both the Endostar Navigator and the other device operate normally.
• No modification of the device is allowed.
* Poldent Co. Ltd. is not responsible for any accidents or other types of trouble that are caused by not following
the precautions noted below.
• Do not use the device in conjunction with an electric scalpel or on patients who have a pacemaker.
• Blocked canals cannot be accurately measured.
• Except for ways described in this manual, the device must not be connected to or used in combination with any other
apparatus or system. It must not be used as an integral component of any other apparatus or system. Poldent Co.
Ltd. will not be responsible for accidents, equipment damage, bodily injury or any other trouble which results from
ignoring this prohibition.
• Illumination devices such as fluorescent lights and film viewers which use an inverter can cause the Endostar Naviga-
tor to operate erratically. Do not use the Endostar Navigator near devices such as these.
• Electromagnetic wave interference could cause the Endostar Navigator to operate in an abnormal, random and pos-
sibly dangerous manner. Cellular phone, transceivers, remote controls and all other devices which transmit electro-
magnetic waves located inside the building should be turned off.
• Do not perform maintenance while using the device for treatment.
Precautions
Indications for Use
: This indicates when not to use the equipment.
The Endostar Navigator is a dental device, Apex Locator.
It can be used to detect the apex of root canal.