Porkka ProMed MC 70 User manual

SCIENTIFIC
Medical & Research
Medical and Laboratory
Refrigeration Cabinets
ENG Installation and operation manual
SWE Installations- och bruksanvisning
NOR Installasjons- og bruksanvisning
FIN Asennus- ja käyttöohjeet
FINLAND
Mukkulankatu 19,
FI-15210 Lahti, Finland
Tel.+358 20 5555 12
www.porkka.fi
NORWAY
Ringeriksveien 16
3414 Lierstranda, Norway
Tel. +47 66 98 77 77
www.porkka.no
SWEDEN
Industrigatan 21,
S-619 33 Trosa, Sweden
Tel. +46 15 63 48 40
www.porkka.se
UNITED KINGDOM
29B Greenhill Crescent,Watford
Hertfordshire, WD18 8YB, UK
Tel. +44 1923 233675
www.porkka.co.uk
ProMed MC 70, 180, 150/150 • Future MC 520, MC 720, MC 722,
MC 722GD, MC 1422GD, MC/MF 722, MF 520, MF 720, MF 722

REV. 032017
2
Installation and operation manual
Contents
Purpose of use, delivery and installation ...................................3
Introduction .............................................................................4
S1. Home screen ......................................................................5
S1.1 USB .................................................................................5
S2. Setting up .........................................................................6
S2.1 Start-up............................................................................6
S2.2 How to set language ........................................................6
S2.3 Time and date ..................................................................7
S2.4 Battery control..................................................................8
S3. Daily use ............................................................................9
S3.1 Temperature settings .......................................................9
S3.2 Temperature set value ......................................................10
S3.3 Temperature set value control ...........................................10
S3.4 Temperature alarm limits ..................................................10
S3.5 Temperature alarm limits, set-up .......................................11
S3.6 Temperature data recording intervals set-up......................13
Service menu ...........................................................................14
Alarms ....................................................................................14
Alarm history menu .................................................................15
Alarm signals ...........................................................................15
Device functions in the event of a power failure
(backup battery must be connected) .........................................16
Temperature data menu ...........................................................17
Export menu (USB)...................................................................18
Reset data and alarm menu......................................................19
Defrost information menu.........................................................20
Information menu ....................................................................20
Cleaning ..................................................................................21
Operation failures.....................................................................23
Disposal of the equipment........................................................23
Warranty..................................................................................24
Dear Customer. Thank You for purchasing Porkka Product.
This manual includes information for following Porkka models:
ProMed MC 70, 180 and 150/150 and Future MC 520, MC 720,
MC 722, MC 722GD, MC 1422GD, MC/MF 722, MF 520,
MF 720, MF 722.

REV. 032017
3
Installation and operation manual
Contents
Purpose of use, delivery and installation ...................................3
Introduction .............................................................................4
S1. Home screen ......................................................................5
S1.1 USB .................................................................................5
S2. Setting up .........................................................................6
S2.1 Start-up............................................................................6
S2.2 How to set language ........................................................6
S2.3 Time and date ..................................................................7
S2.4 Battery control..................................................................8
S3. Daily use ............................................................................9
S3.1 Temperature settings .......................................................9
S3.2 Temperature set value ......................................................10
S3.3 Temperature set value control ...........................................10
S3.4 Temperature alarm limits ..................................................10
S3.5 Temperature alarm limits, set-up .......................................11
S3.6 Temperature data recording intervals set-up......................13
Service menu ...........................................................................14
Alarms ....................................................................................14
Alarm history menu .................................................................15
Alarm signals ...........................................................................15
Device functions in the event of a power failure
(backup battery must be connected) .........................................16
Temperature data menu ...........................................................17
Export menu (USB)...................................................................18
Reset data and alarm menu......................................................19
Defrost information menu.........................................................20
Information menu ....................................................................20
Cleaning ..................................................................................21
Operation failures.....................................................................23
Disposal of the equipment........................................................23
Warranty..................................................................................24
The ProMed and Future medical and laboratory cold cabinets, freezers and combination
cabinets are specifically designed for storage of temperature sensitive pharmaceuticals and
samples.
On receipt of your product from the carrier, examine it carefully.Any signs of damage must
be indicated clearly on the delivery document and reported to both the carrier and the
retailer. Photographs of any damage to the packaging and product should be emailed to
the retailer.
NOTE!
The manufacturer warranty does not cover transportation damages, therefore claims for
damage in transit cannot be made unless they have been validated as detailed above
Examine the product carefully as soon as you receive it. If you notice that it has suffered
from damages during transportation, you must indicate this in the delivery document and
report to the carrier and the retailer.
It is very important to read these installation and operating instructions carefully before
you start using the equipment for the first time. Please keep these instructions in a safe
place for future reference or use by another operator.
By following these instruction, you can improve the product performance and reduce un-
necessary repair costs. Please note that it is important to have your equipment regularly
maintained by a qualified engineer!
The ambient temperature must not exceed +32°C or fall under +10°C. Air circulation is
required above the equipment. Avoid placing the cabinet close to a heat source..
NOTE! The equipment is for indoor use only.
NOTE! In order to comply with electrical safety regulations do not store liquids above the
equipment
The rear and sides of the cabinet can located against a wall, but when placing refrigerated
cabinets next to each other or against an exterior wall, ensure there is a minimum 10mm
ventilation gap between the cabinets.
Adjust the feet to ensure the cabinet is horizontal in all directions, checking with a spirit
level if necessary. Always ensure that all feet are in contact with the ground.
During cold weather deliveries, transporting the cabinet from the carrier into a warm envi-
ronment may cause a temporary build-up of condensation. In these circumstances DO NOT
connect to a power supply until the unit is dry and at room temperature.
When safe to do so connect the power cable from the cabinet to an earthed 230 V / 50 Hz
wall socket with its own 10 A safety fuse. Check the voltage and frequency of the cabinet
from the rating plate. The power socket must be positioned so that it is possible to discon-
nect the socket without having to move the unit.
NOTE! A damaged power supply cord must only be changed by a professional electrician,
authorized by the manufacturer
Purpose of use
Delivery
Installation
WARNING: Keep clear of obstruc-
tion all ventilation openings in
the appliance enclosure or in the
structure for building-in.
WARNING: Do not use mechanical
devices or other means to accelerate
the defrosting process.
WARNING: Do not damage the
refrigerant circuit.
ENGLISH

REV. 032017
4
Before using it, clean the cabinet carefully with mild cleaning agent, e.g. liquid food safe
detergent. Rinse and wipe dry, leave the door open for airing for a while.
ProMed MC 70, 180 and 150/150 models
Place the shelves on the brackets and position the door compartments in their correct
locations. Locate the condensation water container and remember to empty it regularly
Future MC, MF 520, 720, 722 and 1422 models
Place the shelf brackets at the required height and locate the shelves on the brackets.The
cabinet is equipped with anti-tild sliding shelves.The maximum load of the cabinet is 75
kg/shelf.
Pull-out shelves or drawers with bearers are not to be used with glass doors or doors with
glass windows.
Ensure cabinets has reached the required holding temperature prior to loading with
product.
NOTE!
In order to avoid evaporator corrosion always store materials with high acid content in air
tight containers.
NOTE!
In order to ensure good air circulation
within the cabinet DO NOT exceed the load
limit (red mark inside the cabinet on the
back wall) and DO NOT place products on
the floor inside the cabinet. Wire baskets
are available as an optional accessory for
products unsuited for shelves. Please con-
tact your retailer if required.
NOTE!
The freezer cabinet has been designed to hold temperature of item being stored and will
not chill or freeze large amounts of products being placed in the cabinet!
Introduction
Condensation water container
at the bottom of the cabinet

REV. 032017
5
Symbol legend:
(1) Probe displayed (P1, P2, P3 or P4)
(2) Probe temperature measurement
(3) Frame information
(4) Defrost icon appears when defrost is running
(5) Light icon appears when light output is activated
(6) The alarm icon will flash when an alarm is present
(7) Time (HH:MM) and date DD/MM/YYYY (format depending on EU
parameter)
(8) Percentage of memory used and data log: both percentage and icon
will be present only if the data log function is activated.
Keys:
HOME Return to the main menu from a submenu
PROBES To change the displayed probe.
NOTE: it browses only between the present probes.
SET Set point: temperature control menu
LIGHT To switch the light on or off (option)
ALARM Alarm: to enter the alarm menu
SERVICE Additional functions and service menu
BACK/EXIT/UPL BACK/EXIT returns one level up from a submenu
UPL backup copy in service menu
S1. Home screen
S1.1 USB
USB connector is below the battery backup switch.
Lorem ipsum
(1)
(2)
(3) (4) (5) (6) (7)
(8)
ENGLISH

REV. 032017
6
S2.1 Start-up
Connect the cabinet to the power supply, The start-up screed shown below will appear on
the controller.
Press the ENTER button below the text and the controller is ready for use.
S2.2 How to set the language
The default factory setting is English. It can be changed to Finnish, Swedish or German
as follows:
1. Go to SERVICE MENU with SERVICE button.
2. Select LANGUAGE submenu with arrow keys and press ENTER
3. Select the required language by pressing ENTER again (language starts to flash) and
with the arrow UP or DOWN.
4. Save the selection by pressing ENTER and return to the initial state with HOME key.
S2. Setting up
Lorem ipsum
Lorem ipsum
Lorem ipsum
Lorem ipsum

REV. 032017
7
S2.3 How to set time and date
Set the real-time clock at the correct time as follows:
1. Go to the SERVICE with SERVICE button.
2. Use arrow keys to go to the row CLOCK SETTINGS and press ENTER
3. Choose the day or time field you want to edit the UP and DOWN arrow keys and press
ENTER on the information to be changed when it begins to flash.
4. Edit the information with UP and DOWN arrow keys and press ENTER to save the
information.
5. Go back by using the HOME button.
Lorem ipsum
Lorem ipsum
ENGLISH

REV. 032017
8
S2.4. Battery backup (additional accessory in Future models)
The controllers of Porkka Future cabinets can be equipped with a 12h battery backup in
compliance with DIN 58345
1. Connect the cabinet to the mains and turn the key (K) 90° to the right to switch battery
backup on
NOTE!
Do not leave the key in the lock, but store the key in a safe accessible location.
2. Battery backup can be turned off during transportation or movement of the unit.Turn
the key 90° to the left to switch battery backup off.

REV. 032017
9
In the normal state, the display shows The internal air temperature (P1) with a +/- 2 °C
variance in normal use. The temperature is displayed with the accuracy of 1/10 degrees and
the set value is the coldest indoor air temperature. The variation goes towards warmer from
this set value.
Internal cabinet temperature displayed is affected by opening the door and changing
products, when the air temperature momentarily rises. The refrigeration is designed so that
the temperature variation will have no effect on the temperature of the stored product.
Additionally, the equipment has a separate control temperature probe (P4), which corre-
sponds with the stored product temperature.
S3.1 Temperature control (NORMAL STATE)
• Press the PROBES button and select the temperature displayed.
• P1 is indoor temperature that changes e.g. when the door is opened.
• P4 is product probe.This is equivalent to the product temperature stored inside
the cabinet.
S3. Daily use
Lorem ipsum
Temperature
Set value
ENGLISH

REV. 032017
10
S3.2 Temperature set point check (SET POINT display)
1. Press the SET button and release it immediately. The display shows the temperature set
point.
2. Press the HOME button and release it immediately or wait for 60 seconds, when the
display returns to the main menu.
S3.3 Temperature set point control
1. Press the SET button and release it immediately.
2. Press the ENTER button until the temperature set point starts flashing.
3. Change the value by pressing the UP or DOWN arrow key within 60 seconds.
4. Save the new set point by pressing the ENTER button for AT LEAST 2 SECONDS or wait
for 60 seconds.
5. Return to the main menu by pressing the HOME button.
S3.4 Temperature alarm limits
The inside temperature of the cabinet is controlled based on the set temperature value
taking into consideration the running and rest periods of the machinery. To ensure safe
storage of the products, the cabinet is equipped with temperature alarm function. It alarms
automatically if the internal temperature falls below the alarm limit (+ 2°C) or exceeds the
alarm limit (+ 8°C) staying outside these limits continuously for the time set to tempera-
ture alarm delay (15 min). This delay has been chosen by the manufacturer. It is so short
that the change of air temperature does not cause such a change of the product tem-
perature that would risk its safety. These alarm limits and delays can be changed in the
settings according to the instructions of the help menu.
S3. Daily use
Temperature
Set Value
Alarm Low
Alarm Up
Lorem ipsum

REV. 032017
11
S3.5 Setting of the temperature alarm limits
1. Go to the SERVICE menu by pressing the SERVICE button.
2. Select the PARAMETERS submenu and press ENTER.
3. Select the PARAMETER LEVEL 1 submenu and press ENTER.
4. Select the CONTROL OF TEMPERATURE ALARMS OF PROBE P1 with arrow keys and
press ENTER.
S3. Daily use
Lorem ipsum
Lorem ipsum
Lorem ipsum
ENGLISH

REV. 032017
12
5. Adjustable parameters are shown in two menus which can be browsed with arrow
keys. Select the required parameter with arrow keys and press ENTER
6. Parameters and their use are as follows:
• A1C alarm: Ab absolute i.e. actual measured temperature according to A1U and A1L
• Rel relative i.e. deviation from the temperature set point value according to A1U and
A1L
• A1U High temperature alarm level, either actual temperature or size of deviation
• A1L Low temperature alarm level, either actual temperature or size of deviation
• A1H Alarm deviation range specifies how many degrees the temperature must alter
before the alarm is automatically switched off (alter remains in the report).
• A1D general alarm delay
• d1o additional delay in start-up, when a warm cabinet is taken in use the alarms are
not connected before the cabinet has reached the required storage temperature.
When adjusting limits and delays, limits must be set outside the temperature
deviation range and the delay such that it will not cause unnecessary alarms
due to normal use of the cabinet, set such as opening the door and defrost
periods.
S3. Daily use
Lorem ipsum
Lorem ipsum

REV. 032017
13
S3.6 Temperature recording intervals setting
1. Go to SERVICE menu by pressing the SERVICE button.
2. Go to the PARAMETERS menu and press ENTER
3. Select Parameter level 1 and press ENTER
4. Select the CLOCK AND RECORDING SETTING and press ENTER
5. Set the required recording interval ItP (min), default 15 min.
Press ENTER, set with arrow keys and confirm by pressing ENTER.
NOTE! The recording interval affects the length of the displayed temperature diagram,
being always the last 100 recordings:
Interval Diagram
15 min 25 h
30 min 2 days 2 h
45 min 3 days 3 h
60 min 4 days 4 h
6. Go back to main menu by pressing the HOME.
All other service menu settings must only be altered by an authorised service agent.
S3. Daily use
Service menu
Lorem ipsum
Lorem ipsum
ENGLISH

REV. 032017
14
The controller can store in its internal memory the last 20 alarms with their start and finish
times. If any alarm is present, the HOME screen will display the relative icon.
Menu of recorded alarms
There are two ways to access the ACTIVE ALARM menu:
• By using the ALARM key from the HOME screen.
• By selecting the ALARM submenu from the SERVICE screen.
After accessing the menu ALARMS, all the active alarms will be displayed and it will be
possible to select each of them to have more information. If there are no active alarms, the
display will be empty.
View the active alarm details by pressing LOG. The first column shows the alarm code and
the second column gives the description of the alarm.
Pressing the LOG button again takes you to the alarm history menu.The 20 latest alarms
are stored in the history.
ACTIVE ALARMS menu
This menu contains all the information concerning the active alarms. The alarm / total
number of active alarms are displayed in the top right hand corner of the display (12)
Browse active alarms using the UP and DOWN arrows.
When an active alarm has been removed, the alarm information is transferred to the alarm
history menu (LOG).
Alarms
Lorem ipsum
Lorem ipsum
(12)

REV. 032017
15
Press the LOG key in the ACTIVE ALARMS screen to enter the ALARM LOG menu. In this
menu all the stored alarms can be analysed (one for each screen).
(13) = Alarm selected / total number of logged alarms
For each alarm the starting time and date and the finish time and date are recorded.
Press the EXIT key to return to the ACTIVE ALARMS menu.
The alarm code is displayed until the temperature returns to within the set parameters. All
the alarm codes alternate with the internal cabinet temperature, except for the parameter
P1 which is flashing.
Alarm history menu
(ALARM LOG)
Alarm signals
Code Cause Output
”P1” Regulating probe TR3 failure Alarm output ON; Compressor output accord-
ing to parameters Con and CoF.
”P2” EVP probe failure Alarm output ON; Other outputs unchanged
”P3” Logging probe TL1 failure Alarm output ON; Other outputs unchanged
”P4” Logging probe TL2 failure Alarm output ON; Other outputs unchanged
”HA1” High temperature alarm, probe P1 Alarm output ON; Other outputs unchanged
”LA1” Low temperature alarm, probe P1 Alarm output ON; Other outputs unchanged
”HA3” High temperature alarm, probe P3 Alarm output ON; Other outputs unchanged
”LA3” Low temperature alarm, probe P3 Alarm output ON; Other outputs unchanged
”HA4” High temperature alarm, probe P4 Alarm output ON; Other outputs unchanged
”LA4” Low temperature alarm, probe P4 Alarm output ON; Other outputs unchanged
”dA” Door open alarm Compressor and fans depend on “rrd”
”EA” External alarm Output unchanged.
”CA” Serious alarm All outputs OFF.
Lorem ipsum
(13)
ENGLISH

REV. 032017
16
The alarm code is displayed until the temperature returns to within the pre-set parameters.
All the alarm codes are shown alternating with the room temperature, except for the
parameter P1 which is flashing.
Silencing buzzer
Once the alarm signal is detected, the buzzer can be silenced by pressing any key.
Unless the cause of the alarm is corrected, the buzzer signals again after 30
min.
EE alarm
The control unit is provided with an internal check for the data integrity. The “EE” alarm
flashes when a failure in the memory data occurs. In such cases the alarm output is ena-
bled.
Acknowledgement of alarm and need for service
Faulty sensor alarms: P1, P2, P3 and P4. The device operates on backup system.
Acknowledge the alarm and contact an authorized service company.
Temperature alarms HA1, LA1, HA3, LA3, HA4 and LA4 are acknowledged automatically
when the temperature returns to normal.
Alarms EA and CA. Acknowledge the alarm and contact an authorized service company.
Power failure without temperature alarm
If the controller is connected to battery, during a power failure:
1. Alarm light (6) comes on.
2. You will hear an alarm signal.
Press any button to silence the alarm signal. If the power failure will go on unusually
long the signal becomes activated again after a delay.
Device functions
in the event of a
power failure, if the
backup battery is
connected
Lorem ipsum

REV. 032017
17
Power failure and temperature alarms
If the temperature alarm is given during a power failure:
1. Alarm light (6) comes on.
2. The alarm signal sounds continuously.
The alarm sound is muted by pressing any button and starts sounding again after a
delay if the power failure continues for an unusually long time.Temperature data and
alarms are stored during the power failure if the battery backup is switched on.
Press Data button in the main menu to go to DATA LOG menu
• DATA LOG contains all temperature data the controller has saved.
• DOWNLOAD exports the recorded data to a USB stick.
Data log menu
This menu contains all the information the controller has recorded.
1. Go to the data by pressing ENTER when the temperature history will appear on the
display. Wait until the hourglass on the left corner of the display disappears.
Temperature
data menu
Lorem ipsum
Lorem ipsum
ENGLISH

REV. 032017
18
This menu shows all the data belonging to the selected RECORD.
Scroll through all the recorded data by using UP and DOWN keys. The HOURGLASS
icon will appear on the left side of the screen while the display is collecting the data to
show.
RM Temperature measured by probe P1
EVP Evaporator temperature (for service)
P3 Condenser temperature (for service)
P4 Temperature probe for products
In case of probe failure the label ERR will be displayed. In case of probe absence the label
NP will be displayed.
ST: Status of the controller/loads:
“-” Working with no load activated
OFF Working in standby mode
DDefrost running
CCompressor working
A backup copy can be made to an external USB device of the data stored in the controller.
1. Press DATA button
2. Select DOWNLOAD
3. Insert the USB stick into the USB port (max. 4 GB)
4. Press ALARM or DATA to export the required data.
5. Display indicates: ”Copy in progress. Please wait”. During this notification, the
controller sends data to the USB device.
6. When the display shows “copy completed“remove the USB device from the controller.
Export menu (USB)
Lorem ipsum

REV. 032017
19
IMPORTANT!
USB DEVICE MUST NOT BE REMOVED DURING THE LOADING PHASE. IT MAY
DESTROY THE DATABASE AND THE USB RECORDER.
5. Disconnect the USB device.
6. Go back to normal mode using the HOME or BACK button.
NOTE!
An ERROR message will be displayed in case of a failure.This message shows for 60 sec-
onds. Press the HOME or BACK button.After that the DATA MENU screen will be restored.
Remove the USB device or repeat the download operation.
From the SERVICE menu, it is possible to select the RESET menu and delete all recorded
values from the internal database.
1. Enter the SERVICE menu.
2. Select RESET DATA AND ALARM submenu and press ENTER.
3. Select the item to reset by using ALARM or DATA button.
5. Wait for the process to finish.
6. At the end of the process the V icon will appear.
7. Exit the menu by pressing the HOME or BACK button.
Reset data and
alarm menu
Lorem ipsum
Lorem ipsum
ENGLISH

REV. 032017
20
You will see from the DEFROST INFORMATION menu when the previous defrosting has
occurred and when to start the next defrost cycle.
1. Enter the SERVICE menu.
2. Select DEFROST INFORMATION submenu and press ENTER.
3. The information related to the last and the next defrost will be displayed.
4. Exit the menu by pressing HOME or BACK.
The INFORMATION menu shows the following controller related information:
• Release firmware: XW777K / product family / release EEPROM.
• Release date in the format: dd/mm/yy
• EEPROM release
1. Enter the SERVICE menu.
2. Select INFORMATION submenu and press ENTER.
3. Exit the menu by pressing either the HOME or BACK button.
Defrost information
menu
Information menu
Lorem ipsum
Lorem ipsum
This manual suits for next models
11
Table of contents
Languages:
Other Porkka Refrigerator manuals