POSEIDON Salinity Pen Plus+ User manual

User Guide

2
Poseidon Flow Technologies sincerely thanks you for
purchasing the Salinity Pen Plus+ (SP+). The SP+is a water
resistant, portable, easy to use instrument that uses electri-
cal conductivity to determine the Salinity/Specific Gravity of
natural and/or artificial sea water.
The unit rapidly displays Salinity (ppt), Specific Gravity (SG)
and temperature (Celsius/Fahrenheit), on an easy to read
LCD screen. The SP+ can also be easily calibrated in seconds
using the included Salinity Standard Solution (30ppt).
This device has been factory tested to provide trouble-free
performance and contains certified components to
ensure safe, reliable and accurate, lab grade results.
Please read the entire user guide before using
your Salinity Pen Plus+. The information provided
in this manual is subject to change.
For the most up to date information please visit:
www.poseidonflow.com/support
User Support ................................................................... Pg 3
Your Pen at a Glance ...................................................... Pg 4
Specifications................................................................... Pg 6
Pen Calibration................................................................ Pg 6
Pen Usage ........................................................................ Pg 7
Pen Care........................................................................... Pg 8

3
Customer Support
Many of the steps in this guide have corresponding how-to
videos. View them on the Poseidon Flow Technologies web-
site at poseidonflow.com/support-videos
United States and Canada
Contact Poseidon Flow Technologies Directly for all warranty
service and returns. Visit poseidonflow.com/support
International
Contact your local distributor directly for all warranty service
and returns. Visit poseidonflow.com/retailers
Customer Support
Should you require customer support for your Salinity
Pen Plus+, help is offered in the following ways:
Via Phone:
M-F 8:00AM to 5:00PM EST 1-800-202-FLOW
Via support@poseidonflow.com
http://poseidonflow.com/support
International Support:
Contact your local distributor directly for all warranty issues
or returns. Visit poseidonflow.com/retailers to find the
contact information for your Poseidon Flow Distributor.

4
Sensor – Platinum Plated
Sensor Connecting Cap
DO NOT TURN
Display Panel
Displays measurement values
Battery Cap – Unscrew for battery
access. Requires 2x CR1632 Button
Batteries - Included.
MODE Button – Navigate between
operational modes. Temperature
(Celsius/Fahrenheit), Salinity(PPT), and
Specific Gravity (SG).
CAL/HOLD Button
Allows measurements to be locked
on the display and is used for Salinity
Calibration.
Power Button – Depress to power the
device on/off.
Your Salinity Pen Plus+at a Glance
DO NOT SUBMERGE
PAST THIS POINT

5
Usage Notice
To ensure accurate measurement, clean the sensor tip prior to use,
and promptly after measurement.
Upon completion of testing, rinse the sensor with fresh water, and
ensure to fully dry the sensor using a clean, dry cloth or sponge.
Always keep the sensor dry when not in use.
Use caution when cleaning the platinum sensor. Clean only using a
damp cloth or sponge with a neutral detergent.
DO NOT use chemicals such as acetone or benzene to clean any part
of the instrument.
Avoid storing the instrument in places of high humidity, or where it will
be exposed to direct sunlight, high temperature, dust, or moisture.
Remove the batteries when being stored for extended periods (over
one month).
DO NOT submerge the instrument past the sensor connector (white
collar).
This device IS NOT waterproof. Water Damage IS NOT covered
under warranty.

6
The Salinity Pen Plus+ is now ready for use!
It is recommended to recalibrate the Salinity Pen
Plus+ before each use.
Calibration of the Salinity Pen Plus+
Calibration of testing equipment is essential for accurate
and consistent results, and your new Salinity Pen Plus+is no
exception. All instruments are tested before they leave the
factory however, it is good practice to calibrate your Salinity
Pen Plus+before use.
Calibration Procedure:
Step 1: Turn on the Salinity Pen Plus+by depressing the
POWER button.
Step 2: Press the Mode button until the display reads PPT.
The calibration procedure WILL NOT WORK if the pen is not
in PPT mode.
Step 3: Immerse the tip of the pen in the calibration solu-
tion. NOTE: When not in use, the calibration solution should be
stored at room temperature.
Step 4: Press and hold the CAL/HOLD button for 10 full sec-
onds until you see the CAL message display on the LCD.

7
Correct
Proper Salinity Pen Plus+Usage Notice
Measurement:
Keep the sensor at least 0.5” (1cm) away from all edges of
the container when measuring.
Incorrect

8
1.
Rinse the sensor with fresh water prior to measurement.
2. Press the Power Button to turn the device on.
a) The unit displays temperature measurement in real
time. To toggle between Celsius/Fahrenheit press and
hold the MODE button for 3 seconds until the unit
change is observed on the display.
b) Press the MODE button to toggle between testing
modes. Salinity (ppt), Specific Gravity (SG) and tem-
perature (Celsius/Fahrenheit),
3. Immerse the tip of the sensor into the aquarium water.
Shake the sensor back and forth 5 times to remove
any air bubbles on the surface of the sensor, which will
cause an error of measurement.
a. Read the value after the measurement be-
comes stable.

9
4. Press the CAL/HOLD button to lock the measured value
on the display.
5. After measurement rinse the sensor with fresh water,
ensure to fully dry the sensor using a clean, dry cloth or
sponge. Always keep the sensor dry when not in use.
**WARNING!**
Air bubbles stuck to the sensor after it is
immersed into the water will lead to
measurement errors.
Always shake the sensor back and fourth five times to
remove these bubbles.

10
Poseidon Flow Technologies warrants to the original purchaser only that the Salinity Pen Plus+will be free from
defects in material and workmanship under normal use for a period of (1) year from the date of original purchase if
the product used consistent with the User Manual.
The Salinity Pen Plus+consists of a Platinum Plated Sensor that will be immersed in the aquarium and the main
unit body that must not be exposed to water at any time. Poseidon Flow Technologies warranty shall not apply:
(i) to any damaged Salinity Pen Plus+that has been subjected to accident, misuse, neglect, alteration, acts of
God, improper handling, improper transport, improper storage, improper use or application, improper testing
or unauthorized repair, (ii) to any Salinity Pen Plus+where the main body has been exposed to water or (iii)
to cosmetic problems or defects that result from normal wear and tear under ordinary use, and do not affect
the performance or use of the product. Poseidon Flow Technologies warranty applies only to a product that is
manufactured by or for Poseidon Flow Technologies and identified by the Poseidon Flow Technologies trademark,
trade name or product identification logo affixed to the product. If you notice a problem with your Salinity Pen
Plus+, do not attempt to repair it yourself. Any attempts at self-repair shall void this limited warranty. Poseidon Flow
Technologies shall not be held liable for damage to any aquarium, aquarium life forms, or other personal objects
due to improper use of the device. If the product develops a covered defect within the one-year period, Poseidon
Flow Technologies will, at its option, either repair or replace the Salinity Pen Plus+without charge, provided that the
Salinity Pen Plus+is returned during the warranty period.
To obtain warranty service or technical support, please contact Poseidon Flow Technologies customer service
department at 1-888-202-FLOW. You will receive instructions on how to return the product for repair or
replacement. To take advantage of this Limited Warranty you will be required to supply an original point of purchase
receipt.
POSEIDON FLOW TECHNOLOGIES DISCLAIMS ALL OTHER WARRANTIES WHETHER ORAL OR WRITTEN, EXPRESS
OR IMPLIED, TO THE FULLEST EXTENT ALLOWED BY APPLICABLE LAW. NO IMPLIED WARRANTY OF FITNESS
FOR A PARTICULAR PURPOSE OR MERCHANTABILITY SHALL APPLY EXCEPT AS REQUIRED BY APPLICABLE LAW.
POSEIDON FLOW TECHNOLOGIES DOES NOT AUTHORIZE ANY PERSON OR BUSINESS ENTITY, INCLUDING
ITS AUTHORIZED DEALERS, TO CREATE FOR IT ANY OBLIGATIONS, LIABILITIES, OR OTHER WARRANTIES IN
CONNECTION WITH THIS PRODUCT.
ALL WARRANTIES ARE VOID IF THERE IS DAMAGE TO THE SALINITY PEN PLUS+ THAT IS NOT A RESULT OF A
DEFECT IN THE MATERIALS OR WORKMANSHIP. THE SOLE AND EXCLUSIVE REMEDY FOR ALL CLAIMS THAT
YOU MAY HAVE ARISING OUT OF OR IN CONNECTION WITH YOUR USE OF THIS PRODUCT, WHETHER MADE OR
SUFFERED BY YOU OR ANOTHER PERSON AND WHETHER BASED IN CONTRACT OR TORT, SHALL BE REPAIR OR
REPLACEMENT OF THE PRODUCT. IN NO EVENT WILL POSEIDON FLOW TECHNOLOGIES BE LIABLE TO YOU OR
ANY OTHER PARTY FOR ANY DIRECT, INDIRECT, GENERAL, SPECIAL, INCIDENTAL, CONSEQUENTIAL, EXEMPLARY
OR OTHER DAMAGES ARISING OUT OF THE USE OR INABILITY OF TO USE THE PRODUCT (INCLUDING, WITHOUT
LIMITATION, DAMAGES FOR LOSS OF BUSINESS PROFITS, BUSINESS INTERRUPTION, LOSS OF INFORMATION
OR ANY OTHER PECUNIARY LOSS, DAMAGE TO YOUR AQUARIUM OR ANY AQUARIUM INHABITANTS), OR FROM
ANY BREACH OF WARRANTY, EVEN IF POSEIDON FLOW TECHNOLOGIES HAS BEEN ADVISED OF THE POSSIBILITY
OF SUCH DAMAGES. IN NO CASE SHALL POSEIDON FLOW TECHNOLOGIES LIABILITY EXCEED THE AMOUNT YOU
PAID FOR THE PRODUCT. IF YOU DISAGREE WITH THESE TERMS AND CONDITIONS DO NOT USE THE SALINITY
PEN PLUS+.
No part of this publication may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including but not limited to photocopying, recording, information retrieval systems, or computer
network without the written permission of Poseidon Flow Technologies.
Poseidon Flow Technologies and all other Poseidon Flow Technologies product names and slogans are
trademarks or registered trademarks of Poseidon Flow Technologies.
©2018 by Poseidon Flow Technologies.
CE Declaration of Conformity: The equipment complies with the Directive 2004/108/EC relating to electromagnetic
compatibility. This equipment meets the following conformance standards: EN 61326-1:2006.

Table of contents
Other POSEIDON Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

PerkinElmer
PerkinElmer FL 6500 installation instructions

Challenger Optics
Challenger Optics CO-OPM-48 Interactive operating manual

Check-line
Check-line StroboVision SMS-400B Operation manual

Challenger Optics
Challenger Optics CO-OPM-48 operating manual

red lion
red lion CUB5V manual

Pro's Kit
Pro's Kit MT-7601 user guide